Transcript
1" Bladder Pump Operating Instructions Model 407 SS 1" Dia. Operating Principles When the Solinst Bladder Pump is placed in a well or borehole, water rises inside the bladder and sample tubing to static level. Compressed nitrogen or air is supplied to the pump via the drive tubing using a control unit. Applying pressure causes the bladder to compress and closes the bottom check valve, forcing water from the bladder into the sample tubing. During a vent cycle the pressure is released from the drive tubing. The bladder returns to its initial state as water re-enters the pump, while the top check valve prevents water already in the sample tubing from falling back into the bladder. Cycling the drive and vent provides water flow, the rate of which can be adjusted for purging or sampling.
Notes: 1. The maximum depth for 1" Bladder Pump operation is 500 ft. (150 m) below grade. 2. DO NOT exceed an operating pressure of 250 psi.
3. The pump has been decontaminated before leaving Solinst, however, you may wish to decontaminate your pump before use. The pump should be decontaminated between wells.
Portable Sampling Setup Sample Line
Tubing Reel
Pump Assembly Portable: The Bladder Pump can be connected to skip-bonded, dual 1/4" OD tubing, mounted on a reel. a) Push the tubing onto the drive and sample stems of the Bladder Pump and secure using dual-ear Oetiker clamps. b) Lower the assembled Bladder Pump into the well, using a stainless steel safety line connected to the eye safety stem on the pump. The Solinst Model 103 Tag Line can be used as a safety line if required. c) Connect the supply line with the in-line dryer from the compressed gas source to the Control Unit. The drive line connects from the Control Unit to the reel (drive and supply lines come with the Model 464 Control Unit). d) Attach a short (3 ft. or 1 m) length of 1/4" OD sample line to the sample connector on the reel. Dedicated: The Bladder Pump is assembled by connecting to a Dedicated Wellhead with 1/4" OD sample and drive line tubing. a) Cut the tubing to desired length. Push the tubing onto the drive and sample stems of the Bladder Pump and secure using dual-ear Oetiker Clamps. b) Attach the sample line and drive line to the appropriate push fitting on the underside of the Wellhead (see diagram at right for use of push fittings and adaptors). c) Lower the Bladder Pump into the well, using a stainless steel safety line connected to the suspension hook on the Wellhead, if desired. Push the Wellhead down firmly onto the riser casing. d) Attach a short (3 ft. or 1 m) length of sample line to the sample fitting on the Wellhead (see diagram at right for use of push fittings and adaptors). e) Connect the supply line with the in-line dryer from the compressed gas supply to the Control Unit. The drive line connects from the Control Unit to the top of the Wellhead (drive and supply lines come with the Model 464 Control Unit). For detailed pumping instructions, please see the Solinst Model 464 Control Unit Operating Instructions.
Drive Line
Supply Line
Air Compressor or Cylinder Electronic Control Unit
Bladder Pump in Screened Well
Dedicated Wellhead Setup Drive Line 3/8" (9.5 mm) Push Fitting
Sample Line 3/8" (9.5 mm) Push Fitting
1/4" (6 mm) Push Fit Adaptor Drive Line Quick Connect
Sample Line 3/8" (9.5 mm) Push Fitting
7/8" (22 mm) Access Hole
Suspension Hook
1/4" (6 mm) Push Fit Adaptors
Wellheads come with three 1/4" push fit adaptors so that either 1/4" or 3/8" tubing can be used, as preferred. To attach tubing just push into push fitting. To release tubing push down on both sides of the top ring and pull tubing out.
®Solinst is a registered trademark of Solinst Canada Ltd.
(Page 1 of 2)
High Quality Groundwater and Surface Water Monitoring Instrumentation
Disassembly 1. Remove the Filter Retainer, Filter Screen and Top and Bottom Platens, being careful not to lose the Check Balls. Remove the Pump Body. 2. Remove the Top and Bottom Retainers, then remove the Centre Rod from the Bladder Sleeve by pushing one end of the Centre Rod onto a solid surface, while gripping the Bladder Retainer. It may be necessary to use one of the Retainers to push the Centre Rod out. 3. Remove the Bladder from the Bladder Sleeve using a pair of needle nose pliers.
10. Drop a Check Ball into the Top Platen and screw the Pump Body onto it. 11. Shake the assembled pump to hear if the Check Balls rattle. If not, repeat steps 8 to 10, to ensure that the Check Balls are positioned correctly. 12. Fit the Filter Screen onto the Bottom Platen and screw the Filter Retainer into place.
Drive Stem
Decontamination
Safety Line Attachment 1/4" Oetiker Clamp 100164
Sample Stem Top Platen
Notes: 1 . Always follow your local guidelines and standard protocols.
2. Do not use acetone on the O-rings.
107285
1. Completely disassemble the Pump. 2. Wash all pump components with phosphate-free soap or a detergent. 3. Rinse all components thoroughly with deionized water. 4. Replace any worn O-rings and Bladder if necessary,and reassemble. 5. The Pump’s interior can be cleaned without disassembling by flushing your cleaning solution up and through the Pump’s intake. With the Pump’s Filter removed, use a peristaltic pump connected to the Bottom Platen, to push the cleaning solution through the Pump. Then flush the system with deionized water.
Bladder Sleeve 104892
V-116 O-ring 103493
Check Ball 5/16"OD 100732
V-010 O-ring 100257
V-111 O-ring 104679
Top Retainer 104894
Centre Rod (Riser) 104855
Reassembly 1. Manually slide the new Bladder into the Bladder Sleeve. It should extend about 1/4" beyond the Bladder Sleeve, at both ends. 2. Using your finger, flare both ends of the Bladder. 3. If necessary install new O-rings on both Top and Bottom Retainers and both Platens (8 in total). 4. Hold both ends of the Replacement Bladder to minimize slip within the Sleeve. Liberally lubricate a Retainer with water and insert it first at a 45° angle and rotating gently back and forth until the Bladder has been pushed about 1/4" past the second O-ring on the Retainer. 5. Insert the Centre Rod into the Sleeve until it rests in the Retainer. 6. Lubricate the second Retainer with water, align it onto the Centre Rod and working gently to avoid twisting or crumpling of the Bladder, push the Retainer into the Bladder until it is about 1/4" past the second O-ring. Trim off excess as necessary. 7. Take the Bottom Platen and drop the 5/16" OD Teflon Check Ball into it. 8. Fit the end of the assembly firmly into the Bottom Platen. 9. Slip the Pump Body over the assembly and screw it onto the Bottom Platen. If tight, use a single wrap of Teflon Tape.
Bottom Retainer 104893
V-111 O-ring 104679
V-012 O-ring 100291
Teflon® Bladder Pump Body 104850
104891 (or)
LDPE Bladder 108521
50 Mesh Filter Screen 103115
Check Ball 5/16"OD 100732
V-116 O-ring 103493
Bottom Platen 104348
Filter Retainer 103113
1" Stainless Steel Bladder Pump
Printed in Canada December 13, 2013 (#108369) (Page 2 of 2)
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
1.66" Bladder Pump Operating Instructions Model 407 PVC 1.66" Dia. Operating Principles When the Solinst Bladder Pump is placed in a well or borehole, water rises inside the bladder and sample tubing to static level. Compressed nitrogen or air is supplied to the pump via the drive tubing using a Control Unit. Applying pressure causes the bladder to compress and closes the bottom check valve, forcing water from the bladder up into the sample tubing. During a vent cycle the pressure is released from the drive tubing. The bladder returns to its initial state as water re-enters the pump, while the top check valve prevents water already in the sample tubing from falling back into the bladder. Cycling the drive and vent provides water flow, the rate of which can be adjusted for purging or sampling
Notes: 1. The maximum depth for PVC Bladder Pump operation is 500 ft. (150 m) below grade. 2. DO NOT exceed an operating pressure of 50 psi.
3. The pump has been decontaminated before leaving Solinst, however, you may wish to decontaminate your pump before use. The pump should be decontaminated between wells.
4. Tube fittings are based on use of 1/4" drive line and 3/8" sample line. A 3/8" to 1/4" adaptor is also supplied, if dual 1/4" tubing is preferred.
5. Refer to the Model 464 operating instructions for detailed pumping instructions.
Portable Sampling Setup Sample Line
Tubing Reel
Pump Operation Portable: The Bladder Pump will be attached to skip-bonded, dual 1/4" OD tubing, mounted on a reel. a) Push the tubing into the drive and sample compression fittings of the Bladder Pump. The compression fitting nuts may have to be loosened before the tubing is inserted. Tighten the nut 1-1/4 turns past finger tight for a proper seal (see overleaf). b) Lower the assembled Bladder Pump into the well. The Solinst Model 103 Tag Line can be used for this purpose. c) Connect the supply line with the in-line dryer from the compressed gas source to the Control Unit. The drive line connects from the Control Unit to the reel. d) Attach a short (3 ft. or 1 m) length of 3/8" OD sample line to the sample connector on the reel. Dedicated: The Pump will come with a roll(s) of tubing to be cut to length, as required, for attachment to a Wellhead. a) Cut the tubing to desired length. See step a) above for tubing connection instructions to the Pump. b) Attach the sample line and drive line to the appropriate Push Fitting on the underside of the Wellhead (see diagram at right for use of Push Fittings). c) Lower the Bladder Pump into the well, using a stainless steel safety line if desired, and push the Wellhead down firmly onto the riser casing. d) Connect the supply line with the in-line dryer from the compressed gas source to the Control Unit. The drive line connects from the Control Unit to the Wellhead. e) Attach a short (3 ft. or 1 m) length of 3/8" OD sample line to the sample connector on the Wellhead. f) For detailed pumping instructions, please see the Solinst Model 464 Control Unit Operating Instructions.
Drive Line
Supply Line
Air Compressor or Cylinder Electronic Control Unit
Bladder Pump in Screened Well
Dedicated Wellhead Setup Sample Line 3/8" (9.5 mm) Push Fit Connector 1/4" (6 mm) Drive Line Quick Connect Suspension Hook
Sample Line 1/4" (6 mm) Push Fit Adaptor Sample Line 3/8” (9.5 mm) Push Fit Connector
7/8" (22 mm) Access Hole Sample Line 1/4" (6 mm) Push Fit Adaptor
Wellheads come with two Push Fit Adaptors so that either 1/4" or 3/8" tubing can be used, as preferred. To attach tubing just push into Push Fitting. To release tubing push down on both sides of the Top Ring and pull tubing out. ®Solinst is a registered trademark of Solinst Canada Ltd.
(Page 1 of 2)
High Quality Groundwater and Surface Water Monitoring Instrumentation
Disassembly 1. Unscrew and remove the Filter Retainer, Vyon Filter and Top and Bottom Platens, being careful not to lose the Check Balls. Remove the Bladder Cartridge from the Pump Body. 2. Use Oetiker pliers to remove the Oetiker clamps from the Top and Bottom Retainers of the Bladder Cartridge Assembly. 3. Remove the Top and Bottom Retainers from the Bladder. Remove the Riser from inside the Bladder.
Model 407 2 ft. x 1.66" dia. PVC Bladder Pump 1/4"Drive Fitting 100232
3/8"Sample Fitting 102045
Top Platen
100232
103147
V-124 O-ring Oetiker Clamp
Decontamination
103150
Notes: 1 . Always follow your local guidelines and standard protocols.
1/4"Eye Bolt 101717
Includes Extra 1/4" x 1/4" Compression Fitting
101133
Check Ball 3/8" OD 101399
V-012 O-ring
2. Do not use acetone on the O-rings.
100291
V-117 O-ring
1. Completely disassemble the pump. 2. Wash all pump components with phosphate-free soap or a detergent. 3. Rinse all components thoroughly with deionized water. 4. Replace any worn O-rings and reassemble.
103144
Top Retainer 103143
Riser
103149
Reassembly Assembling the Bladder Cartridge 1. Insert the Riser into the Bottom Retainer until the O-ring properly seats.
Teflon® Bladder 103602 or
LDPE Bladder 108524
Note: The Bottom Retainer is identified by the four holes at the bottom of its stem.
2. Slide the Bladder over the Riser until the Bladder overlaps the O-ring on the Bottom Retainer. 3 Using an Oetiker clamp, fasten the Bladder to the Bottom Retainer. 4. Fit the Top Retainer onto the top of the Riser until the O-ring properly seats. 5. Slide the top of the Bladder over the Top Retainer until the Bladder overlaps the O-ring. Use an Oetiker clamp to fasten the Bladder to the Top Retainer. Assembling the Pump 1. Slide the Vyon Filter over the Filter Retainer until firmly seated in the groove at the bottom of the Retainer. 2. Thread the Bottom Platen onto the Filter Retainer until the O-ring seats and parts are finger-tight. Insert the Check Ball into the top of the Bottom Platen. 3. Insert the bottom end of the assembled Bladder Cartridge into the Bottom Platen until the O-ring firmly seats. 4. Slide the Pump Body over the Bladder Cartridge. Thread the Pump Body onto the Bottom Platen until the O-ring seats and parts are finger tight. 5. Carefully place the Check Ball in the top of the Top Retainer. 6. Holding the Bottom Platen with one hand, take the Top Platen and thread into the top of the Pump Body until the O-ring seats and the parts are finger tight. Note: With PVC Pumps there is the option of using a stainless steel weight threaded into a tapped Filter Retainer to reduce buoyancy during operation. An optional Weight and Filter Retainer Assembly (part #105881) is available from Solinst
Bottom Retainer 103145
V-117 O-rings 103144
V-012 O-ring 100291
Check Ball 3/8" OD 101399
Oetiker Pump Body Clamp 103522
103150
V-124 O-ring 101133
Bottom Platen 103146
V-116 O-ring
Vyon Filter
103493
103603
Filter Retainer 103803
Tubing Eye Bolt Compression Fitting Nut
Nylon Compression Fitting
Top Platen ®Teflon is a registered trademark of DuPont Corp.
Printed in Canada July 29, 2011 (#109172) (Page 2 of 2)
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
Drop Tube Assembly Instructions Model 407 PVC 1.66" Dia.
Tools and Materials Needed 1. 407 1.66" PVC Drop Tube Assembly (#105225) Includes: • Top Drop Tube Extension (#104647) • Bottom Drop Tube Extension (#104870) • 2 x 1/2" Nylon Compression Fittings (#104360) • 2 x 3/8" OD Teflon® Check Balls (spare) (#101399) • Filter Retainer for Weight (#103803) • 2 x #V116 O-rings (one as spare) (#103493) 2. 1/2" OD x 3/8" ID LDPE tubing
Notes: 1 . See separate operating instructions for the 1.66" OD PVC Model 407 Bladder Pumps. The Pump Body must remain submerged during operation.
2. Refer to the Solinst Model 464 operating instructions for detailed pumping instructions.
3. This Drop Tube Assembly is also suitable for use with the 1.66" OD Model 408 PVC Double Valve Pumps. (See separate operating instructions.)
3. Tubing Cutters PVC Pump Body
4. Weight (optional)
103142
Operating Principles
Bottom Platen
A Drop Tube Assembly allows samples to be obtained from greater depths then the standard maximum 100 ft (30 m) below grade for Solinst PVC Bladder Pumps. The Drop Tube places the screened filter intake of the Pump as deep as necessary to access the sample zone desired.
3/8"Tefon Check Ball
103146
#V116 O-ring 103493
Top Drop Tube Extension 104647
The Drop Tube Assembly is easily installed on existing Pumps, even in the field. Pumps are operated in the same manner as those without a Drop Tube, which maintains high sample integrity.
Installation 1. Select and cut the appropriate length of 1/2" OD x 3/8" ID tubing. The most common type of tubing used is low density polyethylene (LDPE).
1/2" Compression Fitting 104360
1/2" OD x 3/8"ID LDPE Tubing
Note: Other sizes and types of tubing may also be used, however, the end user will need to supply their own tubing and tubing adaptors.
2. Unscrew and remove the Filter Retainer (with the white Vyon filter) from the Bottom Platen of the Pump. Keep the Check Ball. 3. Remove the white Vyon filter from the Filter Retainer and slide it over the new Filter Retainer with the 7/8" thread. 4. Insert the Check Ball into the stem of the Top Drop Tube Extension. 5. Ensure there is a #V116 O-ring over the groove at the shoulder of the Top Drop Tube Extension, and tighten by hand into the Bottom Platen of the pump until the O-ring properly seats. 6. Push one end of the 1/2" OD x 3/8" ID tubing into the compression fitting on the bottom of the Top Drop Tube Extension. Note: To use the compression fittings, loosen the compression fitting nut, and then insert the tubing all the way into the fitting. Then re-tighten the nut snug by hand, then 1-1/4 turns more past finger-tight.
7. Take the Bottom Drop Tube Extension and connect it to the other end of the tubing using the compression fitting. 8. Thread the Filter Retainer (with the white Vyon filter) into the bottom of the Bottom Drop Tube Extension, ensuring the O-ring on the Filter Retainer properly seats.
1/2" Compression Fitting 104360
#V116 O-ring 103493
White Vyon Filter 103515
(#109022)
Bottom Drop Tube Extension
104870
Filter Retainer with 7/8" thread 103803
Weight (optional)
104670 (Sold separately)
Model 407 PVC 1.66" Diameter Drop Tube Assembly 9. Thread a weight, if required, into the thread on the bottom of the Filter Retainer to complete the installation. 10. The Bladder Pump is operated in the same manner as those without a Drop Tube. See separate operating and pumping instructions. The entire Pump Body must be submerged when operating using a Drop Tube.
®Solinst is a registered trademark of Solinst Canada Ltd.
Printed in Canada July 29, 2011
101399
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
1.66" Bladder Pump Operating Instructions Model 407 SS 1.66" Dia. Operating Principles When the Solinst Bladder Pump is placed in a well or borehole, water rises inside the bladder and sample tubing to static level. Compressed nitrogen or air is supplied to the pump via the drive tubing using a Control Unit. Applying pressure causes the bladder to compress and closes the bottom check valve, forcing water from the bladder into the sample tubing. During a vent cycle the pressure is released from the drive tubing. The bladder returns to its initial state as water re-enters the pump, while the top check valve prevents water already in the sample tubing from falling back into the bladder. Cycling the drive and vent provides water flow, the rate of which can be adjusted for purging or sampling.
Notes: 1. The maximum depth for stainless steel Bladder Pump operation is 500 ft. (150 m) below grade. 2. DO NOT exceed an operating pressure of 250 psi.
3. The pump has been decontaminated before leaving Solinst, however, you may wish to decontaminate your pump before use. The pump should be decontaminated between wells.
4. Tube fittings are based on use of 1/4" drive line and 3/8" sample line. A 3/8" to 1/4" adaptor is also supplied, if dual 1/4" tubing is preferred.
5. Refer to the Solinst Model 464 operating instructions for detailed pumping instructions.
Portable Sampling Setup Sample Line
Tubing Reel
Pump Operation Portable: The Pump will be attached to skip-bonded, dual 1/4" OD tubing, mounted on a reel. a) Push the Tubing Inserts into the tubing and into the drive and sample compression fittings of the Bladder Pump. The compression fitting nuts may have to be loosened before the tubing is inserted. Tighten the nut 1-1/4 turns past finger tight for a proper seal (see overleaf). b) Lower the assembled Bladder Pump into the well. The Solinst Model 103 Tag Line can be used for this purpose. c) Connect the supply line with the in-line dryer from the compressed gas source to the Control Unit. The drive line connects from the Control Unit to the reel. d) Attach a short (3 ft. or 1 m) length of 1/4" OD sample line to the sample connector on the reel. Dedicated: The Pump will come with a roll of tubing to be cut to length, as required, for attachment to a Wellhead. a) Cut the tubing to desired length. See step a) above for tubing connection instructions to the Pump. b) Attach the sample line and drive line to the appropriate Push Fitting on the underside of the Wellhead (see diagram at right for use of Push Fittings). c) Lower the Bladder Pump into the well, using a stainless steel safety line if desired, and push the Wellhead down firmly onto the riser casing. d) Connect the supply line with the in-line dryer from the compressed gas source to the Control Unit. The drive line connects from the Control Unit to the Wellhead. e) Attach a short (3 ft. or 1 m) length of 3/8" OD sample line to the sample connector on the Wellhead. f) For detailed pumping instructions, please see the Solinst Model 464 Control Unit Operating Instructions.
Drive Line
Supply Line
Air Compressor or Cylinder Electronic Control Unit
Bladder Pump in Screened Well
Dedicated Wellhead Setup Sample Line 3/8" (9.5 mm) Push Fit Connector 1/4" (6 mm) Drive Line Quick Connect Suspension Hook
Sample Line 1/4" (6 mm) Push Fit Adaptor Sample Line 3/8” (9.5 mm) Push Fit Connector
7/8" (22 mm) Access Hole Sample Line 1/4" (6 mm) Push Fit Adaptor
Wellheads come with two Push Fit Adaptors so that either 1/4" or 3/8" tubing can be used, as preferred. To attach tubing just push into Push Fitting. To release tubing push down on both sides of the Top Ring and pull tubing out. ®Solinst is a registered trademark of Solinst Canada Ltd.
(Page 1 of 2)
High Quality Groundwater and Surface Water Monitoring Instrumentation
Model 407 2 ft. x 1.66" dia. SS Bladder Pump
Disassembly 1. Unscrew and Remove the Filter Retainer, Filter Screen and Top and Bottom Platens, being careful not to lose the Check Balls. Remove the Pump Body. 2. Remove the Top and Bottom Retainers, then remove the Centre Rod from the Bladder Sleeve by pushing one end of the Centre Rod onto a solid surface, while gripping the Bladder Sleeve. It may be necessary to use one of the Retainers to push the Centre Rod out. 3. Remove the Bladder from the Bladder Sleeve using a pair of needle nose pliers.
1/4"Drive Fitting 1/4"x 3/8" Compression Adaptor (108520)
1/4"Eye Bolt 101717
3/8"Sample Fitting Top Platen 103526
V-124 O-ring
Tubing Inserts for 1/4" ID, 0.17" ID
101133
Check Ball 3/8" OD 101399
V-012 O-ring 100291
Decontamination
V-115 O-ring 100457
Notes: 1 . Always follow your local guidelines and standard protocols.
2. Do not use acetone on the O-rings.
Top Retainer 106131
1. Completely disassemble the Pump. 2. Wash all pump components with phosphate-free soap or a detergent. 3. Rinse all components thoroughly with deionized water. 4. Replace any worn O-rings and Bladder if necessary,and reassemble.
Centre Rod 106376
Teflon® Bladder 106364 or
LDPE Bladder
Reassembly 1. Manually slide the new Bladder into the Bladder Sleeve. It should extend about 1/4" beyond the Bladder Sleeve at both ends. 2. Using your finger, slightly flare both ends of the Bladder. 3. If necessary install new O-rings on both Top and Bottom Retainers and both Platens. (Total of 8.) 4. Hold both ends of the Replacement Bladder to minimize slip within the Sleeve. Liberally lubricate a Retainer with water, and insert it first at a 45° angle and rotate gently back and forth until the Bladder has been pushed about 1/4" past the second O-ring on the Retainer. 5. Insert the Centre Rod into the Sleeve until it rests in the Retainer. 6. Lubricate the second Retainer with water, align it onto the Centre Rod and working gently to avoid twisting or crumpling of the Bladder, push the Retainer into the Bladder until it is about 1/4" past the second O-ring. 7 Take the Bottom Platen and drop the 3/8" OD Teflon Check Ball into it. Do not force the ball in as the correct ball should drop in easily. 8. Fit the Bottom Retainer of the assembly firmly into the Bottom Platen. 9. Slip the Pump Body over the assembly and screw it onto the Bottom Platen. If tight, use a single wrap of Teflon Tape. 10. Drop a 3/8" OD Teflon Check Ball into the Top Platen and screw the Pump Body onto it. 11. Shake the assembled pump to hear if the Check Balls rattle. If not, repeat steps 8 to 10 to ensure that the Check Balls are positioned correctly. 12. Fit the Filter Screen onto the Bottom Platen and screw the Filter Retainer into place.
Printed in Canada July 29, 2011 (#103181) (Page 2 of 2)
108524
Bottom Retainer 106130
V-115 O-rings 100457
V-014 O-ring 103492
Bladder Sleeve
Check Ball 3/8" OD 101399
106185
V-124 O-ring 101133
Bottom Platen
Pump Body
106140
103522
V-116 O-ring 103493
50 Mesh Filter Screen
Filter Retainer
103603
103581
Nut
Note:
Ferrule Tubing Eye Bolt Compression Fitting Nut (Ferrule inside)
Top Platen
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
Tubing Inserts Compresion Fitting
Be careful not to lose the two pieces of the ferrule if loosening or removing the nut from the compression fitting.
Drop Tube Assembly Instructions Model 407 SS 1" Dia.
Tools and Materials Needed Pump Body
1. 407 1" Drop Tube Assembly (#107902) Includes:
104850
• 2 x 14.0 Single Ear Oetiker Clamps (#100165) • 1" dia. x 6" SS Weight (#107903)
Bottom Platen 104348
2. 1/2" OD x 3/8" ID LDPE tubing 3. Single Ear Oetiker Clamp Pliers • Solinst (#100144) or equivalent
1/2" OD x 3/8" ID LDPE Tubing
4. Tubing Cutters
Single Ear Oetiker Clamps 100165
Drop Tube Assembly A Drop Tube Assembly allows samples to be obtained from greater depths then the standard 500 ft. (150 m) below grade for stainless steel Solinst Bladder Pumps. The Drop Tube places the screened filter intake of the Pump as deep as necessary to access the sample zone desired. The Drop Tube Assembly is easily installed on existing Pumps, even in the field. Pumps are operated in the same manner as those without a Drop Tube, which maintains high sample integrity.
1" dia. x 6" SS Weight 107903
Installation 1. Select and cut the appropriate length of 1/2" OD x 3/8" ID tubing. The most common type of tubing used is low density polyethylene (LDPE). Note: Other sizes and types of tubing may also be used, however, the end user will need to supply their own tubing and tubing adaptors.
2. Unscrew and remove the Filter Assembly (Filter Retainer and Filter Screen) from the Bottom Platen of the Pump. Set aside, as it will be used to complete the Drop Tube Assembly. 3. Slide an Oetiker Clamp over one end of the tubing. Insert the stem of the Bottom Platen into this end of the tubing. 4. Center the Oetiker Clamp over the groove in the stem, and use the pliers to crimp the Oetiker Clamp to secure the tubing to the Pump. 5. Slide the second Oetiker Clamp over the other end of the tubing. Insert the unthreaded stem of the 1" dia. x 6" SS Weight into this end of the tubing. 6. Center the Oetiker Clamp over the groove in the stem of the SS Weight, and use the pliers to crimp the Oetiker Clamp to secure the tubing to the SS Weight. 7. Thread the Filter Assembly onto the bottom stem of the SS Weight to complete the installation. 8. The Bladder Pump is operated in the same manner as those without a Drop Tube. See separate operating and pumping instructions.
50 Mesh Filter Screen 103115
(#109023)
103113
Model 407 1" Diameter Drop Tube Assembly
Notes: 1. See separate operating instructions for the 1" Model 407 Bladder Pump
2. Refer to the Solinst Model 464 operating instructions for detailed pumping instructions.
3. One end of the 1" dia. x 6" SS Weight is threaded (NPTF) to receive the Filter Assembly. The opposite end of the Weight is not threaded, and will be fitted with the 1/2" OD x 3/8" ID tubing and clamped with one Oetiker Clamp.
®Solinst is a registered trademark of Solinst Canada Ltd.
Printed in Canada July 29, 2011
Filter Retainer
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
Drop Tube Assembly Instructions Model 407 SS 1.66" Dia.
Tools and Materials Needed 1. 407 1.66" SS Drop Tube Assembly (#103231) Includes: • Top Drop Tube Extension (#103232) • Bottom Drop Tube Extension (#103233) • 2 x 3/8" NPTM x 3/8" Tube Adaptors (#105884) • 2 x 3/8" OD Teflon® Check Balls (spare) (#101399) • 4 x single ear 17.0 Oetiker clamps (2 spare) (#100166) • 2 x #V116 O-rings (one as spare) (#103493) 2. 1/2" OD x 3/8" ID LDPE tubing (#101083) (length as required)
Notes: 1. See separate operating instructions for the 1.66" OD SS Model 407 Bladder Pumps. The Pump Body must remain submerged during operation.
2. Refer to the Model 464 Pump Controller operating instructions, for detailed pumping instructions.
3. This Drop Tube Assembly is also suitable for use with the 1.66" OD Model 408 SS Double Valve Pumps. (See separate operating instructions.)
3. Tubing Cutters 4. Oetiker Pliers (#100144) SS Pump Body 103522
Operating Principles A Drop Tube Assembly allows samples to be obtained from greater depths then the standard maximum 500 ft. (150 m) below grade for stainless steel Integra Bladder Pumps. The Drop Tube places the screened filter intake of the Pump as deep as necessary to access the sample zone desired. The Drop Tube Assembly is easily installed on existing Pumps, even in the field. Pumps are operated in the same manner as those without a Drop Tube, which maintains high sample integrity.
106140
#V116 O-ring 103493
1. Select and cut the appropriate length of 1/2" OD x 3/8" ID tubing. The most common type of tubing used is low density polyethylene (LDPE). Note: Other sizes and types of tubing may also be used, however, the end user will need to supply their own tubing and tubing adaptors.
2. Being careful to retain the check balls, unscrew and remove the Filter Retainer (with the filter screen) from the Bottom Platen of the Pump. Keep this for attachment later. 3. Insert the Check Ball into the stem of the Top Drop Tube Extension. 4. Ensure there is a #V116 O-ring over the stem of the Top Drop Tube Extension, and tighten by hand into the Bottom Platen of the pump. 5. Place one Oetiker clamp over the Tube Adaptor on the bottom of the Top Drop Tube Extension. 6. Push one end of the 1/2" OD x 3/8" ID tubing over the Tube Adaptor on the bottom of the Top Drop Tube Extension. 7. Use the Oetiker pliers to crimp the clamp and secure the tubing to the Tube Adaptor. 8. Take the Bottom Drop Tube Extension and place an Oetiker clamp over its Tube Adaptor. 9. Use the Oetiker pliers to crimp the clamp and secure the other end of the tubing to this Tube Adaptor.
Oetiker Clamp
(#109207)
101399
Top Drop Tube Extension Tube Adaptor 105884
1/2" OD x 3/8"ID LDPE Tubing 101083
Oetiker Clamp
Tube Adaptor 105884
100166
Bottom Drop Tube Extension 103233
#V116 O-ring 103493
Filter Screen 103603
Filter Retainer 103581
Model 407 SS 1.66" Diameter Drop Tube Assembly
10. Thread the Filter Retainer (with the filter screen) into the bottom of the Bottom Drop Tube Extension, ensuring the O-ring on the Filter Retainer properly seats. 11. The Bladder Pump is operated in the same manner as those without a Drop Tube. See separate operating and pumping instructions. The entire Pump Body must be submerged when operating with a Drop Tube Assembly.
®Solinst is a registered trademark of Solinst Canada Ltd.
Printed in Canada April 7, 2011
3/8"Teflon Check Ball
103232
100166
Installation
Bottom Platen
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
Bladder Pump to DVP Conversion Instructions Model 407 to 408 1" x 2 ft SS
Tools and Materials Needed 1. Model 407 1" x 2 ft SS Bladder Pump (#108207) 2. 1" dia. x 2 ft 407 to 408 Conversion Kit (#106925) Includes:
• 2 x V-010 O-ring (#100257) (1 as spare) • 3 x V-012 O-ring (#100291) (1 as spare) • 2 x V-116 O-ring (#103493) • 2 x 3/8" OD Teflon® Check Ball (#101399) • Valve Body for 1" dia. Double Valve Pump (#103529) • Riser for 1" x 2 ft Double Valve Pump (#106928)
Teflon Bladder 104891
Top Platen 107285 O-ring V-010
100257
O-ring V-016 103493
5/16" OD Check Ball 100732
Note: These instructions assume that the Model 407 Solinst Bladder Pump has been completely decontaminated and is fully assembled before beginning the conversion.
Top Retainer
Top Platen
O-ring V-011
O-ring V-010
104894
107285
Disassembling the Bladder Pump
Bladder Sleeve
1. With each hand, hold the Top and Bottom Platens and rotate each counter clockwise, opposite to each other.
Centre Rod 104855
3. Slide the Pump Body off the Bladder Sleeve. Pull the Bladder Sleeve out of the Bottom Platen. Set the Bladder Sleeve aside for future use; it contains the Top and Bottom Retainers, Centre Rod, and Bladder.
104893
O-ring V-011 107285
O-ring V-012 100291
Assembling the Double Valve Pump
5/16" OD Check Ball
1. Take the 3/8" OD Check Ball from the Conversion Kit and drop it into the Bottom Platen.
100732
Pump Body 104850
Note: See separate instructions for Double Valve Pump operation.
103493
(#106926)
104850
O-ring V-012 3/8" OD Check Ball 101399
Valve Body 103529
O-ring V-012 100291
3/8" OD Check Ball Bottom Platen 104348
104348
Model 407 1" dia. x 2 ft Bladder Pump Assembly
Model 408 1" dia. x 2 ft Double Valve Pump Assembly
Stainless Steel Bladder Pump Conversion to a Double Valve Pump
®Solinst is a registered trademark of Solinst Canada Ltd.
Printed in Canada July 6, 2011
Pump Body
101399
O-ring V-016 Bottom Platen
4. Insert the bottom end of the riser into the Valve Body. (The bottom end of the riser has holes in the side of it.)
6. Take the Top Platen and while lining up the top of the Riser into the centre of the Top Platen, thread the Pump Body to the Top Platen.
106928
100291
5. Check all O-rings for wear or damage, and replace as necessary.
5. Slide the Pump Body over the Riser and thread onto the Bottom Platen.
Riser
Bottom Retainer
4. Remove the 5/16" OD Check Ball from the Bottom Platen. Be careful not to lose either of the Check Balls, as they will be required for future use.
3. Insert the other 3/8" OD Check Ball into the top of the Valve Body.
100257
104892
2. Once loosened, unscrew the Top Platen from the Pump Body. Be careful not to lose the 5/16" OD Check Ball. Unscrew the Pump Body from the Bottom Platen.
2. Take the Valve Body and insert into the Bottom Platen until the O-ring firmly seats.
107285
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
Bladder Pump to DVP Conversion Instructions Model 407 to 408 1.66" x 2 ft SS
Tools and Materials Needed 1. Model 407 1.66" x 2 ft SS Bladder Pump (#106138) 2. 1.66" dia. x 2 ft 407 to 408 Conversion Kit (#106896), Includes: • 408 1.66" DVP Replacements Part Kit (#103062) • 3 x V-012 O-rings (#100291) • 2 x V-124 O-rings (#101133) • 1 x V-116 O-ring (103493) • 2 x 3/8" OD Check Balls (#101399) • 1.66" DVP Bottom Platen (#103525) • 1.66" x 2 ft DVP Riser (#103528) • 1.66" DVP Valve Body (#103529)
3/8"Sample Fitting 1/4" Drive Fitting
1/4"Sample Fitting
Bladder
Teflon-106364 (or) Single Use LDPE-108523
Top Platen
O-ring
V-012 100291
Bladder Sleeve
Note: These instructions assume that the Model 407 Solinst Bladder Pump has been completely decontaminated and is fully assembled before beginning the conversion.
106185
3/8" Check Ball
103526
O-ring
101399
V-012 100291
Top Retainer 106131
O-rings
1. With each hand, hold the Top and Bottom Platens and rotate each counter clockwise, opposite to each other.
103528
106376
3. Slide the Pump Body off the Bladder Sleeve. Pull the Bladder Sleeve out of the Bottom Platen. Set the Bladder Sleeve aside for future use; it contains the Top and Bottom Retainers, Centre Rod, and Bladder.
O-rings V-015 100457
O-ring
4. Remove the 3/8" OD Check Ball from the Bottom Platen. Be careful not to lose either of the Check Balls, as they may be required for future use.
1. Thread the Filter Retainer with Filter Screen into the Bottom Platen. 2. Take the 3/8" OD Check Ball from the Conversion Kit and drop it into the DVP Bottom Platen. 3. Ensure there is an O-ring on the DVP Valve Body, and push fit, with the holes pointing downward into the DVP Bottom Platen until the O-ring firmly seats. 4. Insert the other 3/8" OD Check Ball into the top of the Valve Body. 5. Insert the bottom end of the DVP Riser into the Valve Body. (The bottom end of the riser has holes in it.) 6. Slide the original Pump Body over the Riser and thread onto the Bottom Platen. 7. Take the original Top Platen and while lining up the top of the Riser into the centre of the Top Platen, thread the Pump Body to the Top Platen.
V-014 103492
Pump Body
103522
Bottom Retainer
(#109205)
O-ring
106130
3/8" Check Balls
V-124 101133
3/8" Check Ball
101399
Bottom Platen 106140
O-ring
V-116 103493
Model 407 1.66" dia. x 2 ft Bladder Pump Assembly
DVP Valve Body
103529
V-012 100291
O-ring
101399
O-ring
V-012 100291
Pump Body 103522
DVP Bottom Platen
50 Mesh Filter Screen
103525
103603
Filter Retainer 103581
O-ring
V-124 101133
O-ring
V-116 103493
Model 408 1.66" dia. x 2 ft Double Valve Pump Assembly
Stainless Steel Bladder Pump Conversion to a Double Valve Pump Note: See separate instructions for Double Valve Pump operation.
®Solinst is a registered trademark of Solinst Canada Ltd.
Printed in Canada July 6, 2011
V-124 101133
Centre Rod
2. Once loosened, unscrew the Top Platen from the Pump Body. Be careful not to lose the 3/8" OD Check Ball. Unscrew the Pump Body from the Bottom Platen and unscrew the Filter Assembly from the Bottom Platen.
Assembling the Double Valve Pump
O-ring
DVP Riser
V-015 100457
Disassembling the Bladder Pump
5. Check all O-rings for wear or damage. Use the O-rings from the Conversion Kit, to replace as necessary.
3/8" Drive Fitting
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
DVP to Bladder Pump Conversion Instructions Model 408 to 407 1.66" x 2 ft SS
Tools and Materials Needed
3/8" Drive Fitting
1. Model 408 1.66" x 2ft. SS Double Valve Pump (#106082)
1/4"Sample Fitting
Top Platen O-ring
O-ring
5. Unscrew the Filter Retainer with Filter Screen from the Bottom Platen and remove. 6. Check all O-rings for wear or damage. Use the O-rings from the Conversion Kit, to replace as necessary.
Assembling the Bladder Pump Note: If the Bladder Cartridge (Bladder, Bladder Sleeve, and Top and Bottom Retainers) comes already assembled, skip to Step 6.
1. Manually slide the new Bladder into the Bladder Sleeve. It should extend about 1/4" (6 mm) beyond the Bladder Sleeve at both ends. Using your finger, slightly flare both ends of the Bladder. 2. Hold both ends of the Bladder to minimize slip within the Sleeve. Liberally lubricate a Retainer with water, and insert it first at a 45° angle and rotate gently back and forth until the Bladder has been pushed about 1/4" (6 mm) past the second O-ring on the Retainer. 3. Insert the Centre Rod into the Sleeve until it rests in the Retainer. ®Solinst is a registered trademark of Solinst Canada Ltd.
Printed in Canada December 12, 2008 (#106535)
Bladder Sleeve 106364
V-015 100457
Riser
103522
Note: Be careful not to lose either of the Check Balls, as they will be required for assembling the Bladder Pump.
106364
V-012 100291
O-rings
Disassembling the Double Valve Pump
4. Remove the Valve Body and then the 3/8" OD Check Ball from the Bottom Platen.
Top Retainer 106131
103528
3. Remove the Riser and then the 3/8" Check Ball from the Valve Body.
Teflon Bladder
101399
V-012 100291
Note: These instructions assume that the Model 408 Double Valve Pump has been completely decontaminated and is fully assembled before beginning the conversion.
2. Once loosened, unscrew the Top Platen from the Pump Body. Unscrew and remove the Pump Body from the Bottom Platen.
3/8" Check Ball
V-124 101133
• 407 1.66" Replacements Part Kit (#106898) • 407 2 ft. Replacement Bladder (#106364) • 407 2 ft. Bladder Sleeve (#106185) • 407 2 ft. Riser (#106376) • 4 x #V115 O-rings (#100457) • 407 1.66" Top Retainer (#106131) • 407 1.66" Bottom Retainer (#106130) • 407 1.66" Bottom Platen (#106140)
1/4" Drive Fitting
O-ring
103526
2. 1.66" x 2 ft. 408 to 407 Conversion Kit (#106534) Includes:
1. With each hand, hold the Top and Bottom Platens and rotate each counter clockwise, opposite to each other.
3/8"Sample Fitting
Centre Rod 106376
Pump Body
DVP Valve Body
103529
O-ring
O-rings Bottom Retainer 106130
O-ring
V-012 100291
O-ring
3/8" Check Balls
V-124 101133
101399
V-012 100291
Bottom Platen
DVP Bottom Platen O-ring
V-124 101133
O-ring
V-116 103493
V-015 101133
106140
103525
50 Mesh Filter Screen 10360
O-ring
V-124 101133
3/8" Check Ball
101399
Pump Body
103522
O-ring
V-116 103493
Filter Retainer 103581
Model 408 1.66" dia. x 2 ft Double Valve Pump Assembly
Model 407 1.66" dia. x 2 ft Bladder Pump Assembly
Stainless Steel Double Valve Pump Conversion to a Bladder Pump 4. Lubricate the second Retainer with water, align it onto the Centre Rod and working gently to avoid twisting or crumpling of the Bladder, push the Retainer into the Bladder until it is about 1/4" (6 mm) past the second O-ring. 5. Trim off any excess bladder material if necessary. 6. Drop the 3/8" OD Check Ball into the newly provided Bottom Platen. Fit the Bottom Retainer of the assembly firmly into the Bottom Platen. 7. Slip the Pump Body over the assembly and screw it onto the Bottom Platen. 8. Drop a 3/8" OD Check Ball into the Top Retainer and screw the Top Platen into the Pump Body. 9. To ensure the Check Balls are positioned correctly, shake the Pump to ensure the Check Balls rattle. If not, repeat steps 6 to 7. 10. Thread the Filter Retainer with the Filter Screen into the Bottom Platen.
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
407 Adaptor Assembly Instructions 407 1.66" x 2 ft. SS to 800 3.7" Single Point
Tools and Materials Needed 1. 407 Adaptor Assembly - 1.66" Pump to 3.7" Single Point Packer (#110097), includes: • Pump to Packer Adaptor (#110096) • Packer to Filter Retainer Adaptor (#110095)
407 Bladder Pump (1.66" x 2 ft SS) Pump to Packer Adaptor (#110096)
2. Model 407 1.66" x 2 ft. SS Bladder Pump (#106138) 3. Model 800 3.7" x 3 ft. Single Point Packer (#107969)
1" NPTM Threading
Note: New Packers come with Centralizers and Bushings - these are not to be used with this assembly.
Instructions 1. Unscrew the Filter Retainer from the Bottom Platen of the Bladder Pump. 2. Unscrew the Bottom Platen from the Bladder Pump. Be careful not to lose the Check Ball. 3. Screw the Filter Retainer into Adaptor #110095 until finger-tight. 4. Ensure there is white Teflon tape around the 1" NPTM threading on the bottom end of the Packer, and thread Adaptor #110095 with Filter Retainer onto the Packer. 5. Screw the Bottom Platen of the Bladder Pump onto the stem of Adaptor #110096 until finger-tight. 6. Ensure there is white Teflon tape around the 1" NPTM threading on the top end of the Packer, and thread Adaptor #110096 with Bottom Platen onto the Packer.
800 Packer (3.7" Single Point)
Note: It is important to screw the Bottom Platen and Adaptor to the Packer separate from the rest of the Bladder Pump to avoid cross-threading of the 1" NPTM and Adaptor threading. 1" NPTM Threading
7. Drop the check ball into the Bottom Platen, making sure it seats properly in the Bottom Platen (the check ball should not rattle when the assembled Bladder Pump is shaken). 8. Screw the remaining Bladder Pump assembly back onto the Bottom Platen. Ensure all components are threaded tightly.
Check Ball
Bottom Platen
Note: See separate operating instructions for the Model 800 Packers, and Model 407 Bladder Pumps.
Printed in Canada February 3, 2011 (#110098)
Packer to Filter Retainer Adaptor (#110095)
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
Filter Retainer
12 Volt Compressor Instructions For Use with Model 407 Bladder Pumps & 408 Double Valve Pumps Fuse
Technical Specifications Frontal Width:
15" (38 cm)
Depth:
9" (23 cm)
Stand Height:
16" (41 cm)
Weight:
21 lbs (9.5 kg)
Tank size:
2 US gallons (7.6 litres)
Pump time to fill tank from 0 - 125 psi:
4 minutes
Pump Rates:
1.12 cfm @ zero head to 0.4 cfm @ 150 psi
Power Draw:
18 - 20 amps (25 amp fuse)
Max. Operating Pressure
125 psi
Auto on/off Feature Setting:
90 psi - on 125 psi - off
Cord Length:
7 ft (2 m)
ON/OFF Switch
Ball Valve & Handle
7 ft (2 m) Cable
Operating Instructions Notes: • The air tank is rated to 150 psi.
•
If a car battery is used, be sure the vehicle is running as the compressor will drain the battery quickly. If a stand alone battery is used, the battery will drain in approximately 20 minutes.
• It is recommended not to alter the power cord for use with a cigarette lighter adapter.
• The compressor is thermally protected, therefore it will automatically shut off if it overheats. Once the compressor has cooled sufficiently, it will power up once again.
1. Connect the pressure line from the Solinst pump control unit to the pressure outlet on the 12 Volt Compressor. 2. Ensure that the ON/OFF switch is in the OFF position before connecting to the power source. The ball valve handle on the side of the compressor should also be in the OFF position before connecting. 3. Connect the compressor to a 12 volt DC power source (such as a car battery) using the alligator clips. The red clip should connect to the positive terminal on the power source, and the black should connect to a proper ground or to the negative terminal on the power source. 4. Once the compressor has been connected to the power source and to all pressure lines, switch the compressor ON. This will begin to charge the tank to 125 psi, which takes approximately 4 minutes. 5. Once the tank has been fully charged to 125 psi, switch the ball valve to the ON position. 6. Once pumping begins, the air pressure in the tank will drop to 90 psi. The compressor will then recharge to 125 psi or to the selected pumping rate. Printed in Canada May 9, 2014 (#109486)
Pressure Outlet
Draining Valve
Maintenance After every 20 hours of operation, the compressor tank should be vented first, then drained. If the compressor is being operated in more humid conditions, it may need to be drained more often. Note: If not draining the tank, it is not necessary to vent the tank, before disconnecting the supply and drive lines. All of these air fittings are designed to safely connect and disconnect under pressure. If the tank is empty, or at a very low pressure, it will be easier to connect/disconnect the lines.
To vent/release the air pressure from the compressor, turn the compressor OFF and disconnect the clips from the power source. Ensure the ball valve handle is in the ON position. With the drive and sample lines still connected to the pump control unit, select “Drive” on the pump control unit until the compressor’s pressure gauge shows 0 psi. Note: Do not open the draining valve until all air in the compressor tank has been released (gauge shows 0 psi).
To drain the compressor tank, turn the draining valve on the underside of the tank counterclockwise to open. This allows condensation/water to exit the tank.
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
12 Volt Compressor Assembly Replacement 3. Pull the red wire on top of the pressure switch to the old Compressor Assembly. Leave the adjacent red wire on the pressure switch. 4. Cut the black wire of the old Compressor Assembly, at a point about 3" (7.6 cm) away from the old Assembly. Strip the wire about 1/2" (1.25 cm). 5. Remove the old Compressor Assembly from the Compressor Unit Frame. 6. Remove the air filter from the old Compressor Assembly. It will be used with the new Compressor Assembly. 7. Use the 9/16" wrench to remove the check valve from the output of the old Compressor Assembly. It will be used with the new Compressor Assembly.
Tools and Materials Needed 1. 12 Volt Compressor Assembly (#109483) 2. 9/16" Wrench 3. 3/8" Wrench 4. 3/16" Allen Key 5. Wire Cutters and Strippers 6. Pliers or Crimping Tool 7. Heat Gun 8. Teflon® Tape
Instructions CAUTION: Ensure the power supply is disconnected from the 12V Compressor and the pressure has been fully vented from the tank.
Black Wire from Compressor Assembly
Connects to top-front of Pressure Switch Butt Splice Crimp Connection Heat Shrink
Butt Splice Crimp Terminal
Check Valve 9/16" Brass Hex Air Line
Red Wire from Compressor Assembly Pressure Switch Air Filter
12 Volt Compressor Assembly Connections to the Compressor Unit Remove the Old Compressor Assembly 1. Use the 9/16" wrench to undo the 9/16" brass hex and remove the air line from the silver coloured check valve. 2. Use the 3/8" wrench and 3/16" allen key to undo and remove the three bolts holding the Compressor Assembly to the frame of the Compressor Unit.
Replacement 12 Volt Compressor Assembly Connect the New Compressor Assembly 8. By hand, screw the air filter into the new Compressor Assembly. Do not over-tighten. Note: Do not over-tighten the air filter when connecting to the new Compressor Assembly.
9. Remove any old Teflon tape and re-tape the threads of both ends of the check valve. Screw the check valve into the output of the new Compressor Assembly and tighten. 10.Ensure the piece of heat shrink is over the black wire of the new Compressor Assembly. Insert the stripped black wire into the butt splice terminal on the black wire of the new Compressor Assembly. 11. Use crimps or pliers to crimp the butt splice terminal to secure the wire in the butt splice connection. 12. Center the heat shrink over the butt splice connection and apply heat with a heat gun. 13. Connect the red wire from the new Compressor Assembly to the terminal on the pressure switch. 14. Use the three bolts to attach the new Compressor Assembly to the Compressor Unit Frame. 15.Use the 9/16" wrench to connect the air line to the check valve of the new Compressor Assembly. ® Teflon is a registered trademark of DuPont Corp.
Printed in Canada November 24, 2009 (#109484)
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
Dedicated Wellhead Instructions Model 407 and 408 Sample Line 3/8" (9.5 mm) Push Fit
1/4" (6 mm) Push Fit Adaptor
Drive-Line Quick Connect 1/4" (6 mm) Push Fit 7/8" (22 mm) Access Hole
Suspension Hook 1/4" (6 mm) Push Fit Adaptor 407/408 Dedicated 2" Well Cap Assembly (#110227)
Instructions The Dedicated 2" (50 mm) Well Cap Assembly come with two 1/4" (6 mm) Push Fit Adaptors, so that either 1/4" (6 mm) or 3/8" (9.5 mm) tubing can be used for the sample return line. If 1/4" (6 mm) tubing is desired for the sample return line, simply push the Push Fit Adaptors into the existing 3/8" (9.5 mm) Push Fit Connectors until they seat (one Push Fit Adaptor for the top Connector and one for the underside). To attach tubing just push into Push Fit Connector. To release tubing push down on both sides of the Top Ring and pull tubing out.
Note: See separate operating instructions for the Model 407 Bladder Pumps and Model 408 Double Valve Pumps.
Printed in Canada July 29, 2011 (#110229)
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023
35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
407 and 408 Pump Curves
Ground Surface
0
0
20
1.66" x SS Bladder Pump (1/4" D x 3/8" S)
100 Standard Tubing (OD inches) D = Drive, S = Sample 1" x 2 ft SS Bladder Pump (1/4" D x 1/4" S)
200
60
5.8" x 1 ft SS Double Valve Pump (1/4" D x 1/4" S)
300
0
0.5
1 Flow Rates (L / min)
80
1.66" x 2 ft SS Double Valve Pump (3/8" D x 1/4" S)
1.5
Pumping Comparison of Solinst Model 407 vs 408 Integra Bladder Pump vs Double Valve Pump Flow Rate vs. Depth to Pump (Depth to Water at 50 ft (15m))
February 11, 2009
40
For further information contact: Solinst Canada Ltd. Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
2
Depth to Pump (Meters Below Ground Surface)
Depth to Pump (Feet Below Ground Surface)
Models 407 and 408
Standard Pump Configuration Model 407 and 408 408 Double Valve Pump
407 Bladder Pump
Model Number Material Type Maximum Depth Nominal Size (inches)
PVC
316 Stainless Steel
PVC
316 Stainless Steel
100 ft / 30 m Below Grade
500 ft / 150 m Below Grade
100 ft / 30 m Lift
500 ft / 150 m Lift
1.66" x 2 ft
1" x 2 ft
1.66" x 2 ft
1.66" x 2 ft
5/8" x 1 ft
1.66" x 2 ft
103868
108207
106138
105398
103104
106082
1/4" D & 3/8" S
1/4" D & S
1/4" D & 3/8" S
3/8" D & 1/4" S
1/4" D & S
3/8" D & 1/4" S
200
110
200
450
30
450
Viton® O-Rings (Sold in sets of 10)
(2) V012 (1) V116 (2) V124
(1) V010 (4) V111 (1) V012 (2) V116
(1) V012 (1) V014 (4) V115 (1) V116 (2) V124
(3) V012 (1) V116 (2) V124
(2) V010 (2) V013
(3) V012 (1) V116 (2) V124
Teflon® Check Balls (Set of 2) (Part No.)
3/8” dia (106393)
5/16” dia (106394)
3/8” dia (106393)
3/8” dia (106393)
5/16” dia (106394)
3/8” dia (106393)
Part No. Pump Intake
Pump Connection Standard Tubing (OD inches) (D = Drive, S = Sample) Pump Volume (mL) (See Note 1.)
SS Stem for 1/4" OD tubing
X
X
1/4" Barb Polyethylene
X
1/4" Comp Polyethylene 3/8" Comp Nylon
2 (1 as spare)
X
X
1/4" Comp 316 SS w/tubing inserts
X
3/8" Comp 316 SS w/tubing inserts
X
X
1/4" Comp x 3/8" SS Tube Adaptor
X
X
1/4" Eye Bolt w/0.57" ID hole
X
Safety Stem w/ 0.18" ID hole Vyon Filter 60 micron (Part No.)
X
X
X 103515
103515
SS 50 Mesh Filter (Part No.)
103115
103603
Teflon Bladder (Part No.)
104891
106364
Teflon Bladder Assembly (Part No.)
X
103989
103603
108713 108521 (Each) 108525 (Set of 10)
LDPE Bladder (Part No.)
103115
108523 (Each) 108526 (Set of 10)
Notes: 1. Please click here to view our groundwater samplers: http://www.solinst.com/Prod/Lines/Samp/Samp.html
2. Standard bladder is Teflon. Replacement bladders available in LDPE or Teflon
3. Model 103 Replacement Cable is suitable as an optional safety support cable. Includes one aluminum crimp for top connection.
®Solinst is a registered trademark of Solinst Canada Ltd. Printed in Canada July 29, 2011
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
Sampling Groundwater using Model 407/408 Pumps
Introduction
Drive Gas Volume
Retrieving a groundwater sample using a Bladder or Double Valve Pump, requires answers to two questions how much pressure and how much volume of drive gas do I need? The drive gas is usually air or nitrogen delivered through portable gas cylinders or an air compressor. Portability to the site and the pressure requirement for the application, are the main deciding factors. The selection of either compressed nitrogen or air, to use as the drive gas, is made based on the availability, as well as the potential for affecting groundwater sample quality.
Gas volume is usually expressed as CFM (Cubic Feet per Minute). As a 'rule of thumb', it takes about 1 CFM of gas (air or nitrogen) to deliver 1 USGPM (US Gallon per Minute) of water, at 0 psi. If you opt to sample with an air compressor, you should also consider using a 50% duty cycle or, 'factor of safety', for the compressor. This 50% duty cycle ensures that the air compressor won't be running all the time.
Pumping Pressure Determining the amount of applied pressure (in psi) to retrieve a sample, is simple. 1 psi of pressure can raise a 2.3 ft column of water, which is about half of the column height of water in feet, expressed as psi. e.g. If the Bladder Pump's intake is at 100 ft below ground surface, you will require approximately 50 psi of pressure to bring a sample to ground surface. If you are sampling with a Double Valve Pump, then this calculation is made from ground level to static water level (as a minimum, but can be made from total pump depth to maximize purging rates). Therefore, if a Double Valve Pump is 100 ft below ground surface, and static water level is at 50 ft, you can select a pressure of between 25 and 50 psi. With both Bladder Pumps and Double Valve Pumps, add an extra 10 psi to allow for “line loss”.
Tip: What do you do when your pnuematic pump is properly connected, yet there is no sample discharge? A simple 'trick', is to submerge the sample discharge line into a clear container of water. During the drive cycle, you should see bubbles. An aggressive blast of air bubbles can mean that there is no water available, while a steady mild bubbling indicates that the pump is operating and the sample water is on the way up!
Printed in Canada May 5, 2008
e.g. To collect a sample from a depth to water of 100 ft (30 m), using a Double Valve Pump at a rate of 0.05 USGPM (200 mL/min), the calculation is: 0.05 USGPM (200 mL/min) x 1.5 (includes a 50% duty cycle) = a drive gas requirement of 0.075 CFM, or about 0.1 CFM. Now the user needs to select an air compressor that will supply a minimum of 0.1 CFM at 50 psi pressure. This is done by checking the manufacturer's data sheet to see that the CFM needed will meet the output delivery for the required pressure. Considering a 50% duty cycle, the Solinst 12 Volt Air Compressor (pt.#106009) will deliver 0.1 CFM at 50 psi, which is equivalent to a 100 ft (30 m) pumping depth. A larger compressor should be used for deeper applications.
Pump Controller When selecting a pump controller, look for easy to follow preset pumping options. This helps take the guesswork out of determining suitable drive and vent times. For example, if you anticipate that the monitoring well has good recharge and 'makes water', then select a 'fast' cycle rate (~6 sec/cycle of drive & vent times); in poorly producing, or 'slow to recharge' wells, a longer cycle rate of ~115 sec may be more suitable. To help protect the pump controller from damage due to moisture, always position the pump controller physically higher than the sample discharge and wellhead. This helps prevents a syphoning effect, where the gravity back-flow of sample water can enter into the pump controller. Note that there will be a time delay from when the pump controller is turned 'ON', to when an actual water sample is discharged at ground surface. This time delay could be as much as several minutes.
For further information contact: Solinst Canada Ltd. Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario, Canada L7G 4R8 Web Site: www.solinst.com E-mail:
[email protected]
SOLINST TECHNICAL BULLETIN Best Practice Low Flow Methods for Highest Quality Samples Low Flow Sampling Since 1996, low flow sampling has become an increasingly approved method for obtaining high quality groundwater samples. Through the work of Puls and Barcelona, the US EPA released standard operating procedures for low flow sampling (EPA/540/S-95/504). Following such guidelines ensures the collection of samples that are representative of actual in-situ conditions. Low flow purging and sampling involves extracting groundwater at rates comparable to ambient groundwater flow (typically less than 500 ml/min), so that the drawdown of the water level is minimized, and the mixing of stagnant water with water from the screened intake area in a well is reduced. Stabilization of parameters (pH, D.O., conductivity, temperature, etc.) and turbidity of the purged water are monitored before a sample is taken, thus low flow methods facilitate equilibrium with the surrounding formation and produce samples that are truly representative of the formation water.
Automating the pumping rate or the drawdown is often not a good idea for wells with slow recharge. When the rate of recharge is slower than the pumping rate, an unwanted instantaneous purge can occur rather than a preferred slow and steady pumping rate that approximates the well's recovery rate. Rapid removal of water in low hydraulic conductivity formations increases the velocity of water flowing back into a well, creating turbulence and turbidity as the well recharges. Some automated pumping control devices stop a drive cycle as soon as the water level goes below the sensor; the drive cycle is restarted after the sensor detects water again (the well has recovered). This may lead to incomplete drive cycles and inconsistent flow rates rather than the preferred slow gentle pumping rate. A water level drawdown indicator used to monitor drawdown beyond a set level, may not operate as intended, and may not be ideal for low flow sampling. A suitable device should allow the water to fall slowly to a maximum drawdown of less than 0.1 meter (or 1% of the standing column of water) while monitoring the entire pumping process. In fast recovery/recharge wells, setting an automatic pump controller to maintain a minimal drawdown (less than 0.1 meter) is quite feasible. Proper drive and vent times are easy to determine and set. There is no need to use extra equipment such as a water level drawdown sensor to regulate the pump controller.
Low flow methods allow the collection of high quality, representative groundwater samples in 40 ml glass vials for VOC analysis.
Automation May Not Be Best Pneumatic gas drive pumps operated by automated pump controllers are often chosen as the desired equipment for low flow purging and sampling; however, automating these samplers may not be the best approach.
In the end - as with any field equipment - the best practice is to have the field sampler or technician adjust the equipment to suit the pumping characteristics of each well, rather than rely on automated devices.
References Puls, R.W., and M.J. Barcelona. 1996. Lowflow (minimal drawdown) ground-water sampling procedures. U.S. EPA Ground Water Issue: EPA/540/S-95/504. Sevee, J.E., C.A. White, and D.J. Maher. 2000. An analysis of low-flow ground water methodology. Groundwater Monitoring and Remediation Spring 2000: 87-93.
® Solinst and Integra are registered trademarks of Solinst Canada Ltd.
Solinst Integra Bladder Pumps are ideal for low flow sampling. Pumps are either Stainless Steel or PVC, and are 1.66" (42 mm) or 1" (25 mm) in diameter. Bladders are available in Teflon® or LDPE.
Solinst Low Flow Instruments Solinst offers a variety of equipment for use in low flow applications. Double Valve Pumps and Integra® Bladder Pumps are able to provide flow rates as low as 100 ml/min, using the Solinst Electronic Pump Control Unit. The Integra Bladder Pump allows very slow, steady compression of the bladder during a drive cycle, unlike some other samplers, as it can be set to provide a consistent rate comparable to ambient groundwater flow. This results in representative, high quality undisturbed VOC samples. Using low flow techniques, turbulence is reduced and off-gassing minimized, providing more accurate and reliable VOC sample collection. A submerged Levelogger Gold water level data logger, used with a Leveloader Gold or laptop, can provide viewing of real-time water level readings in the field and allows actual drawdown to be monitored during purging and sampling. Pumps can be operated manually based on the Levelogger readings.
® Teflon is a registered trademark of DuPont Corp.
High Quality Groundwater and Surface Water Monitoring Instrumentation Solinst Canada Ltd., 35 Todd Road, Georgetown, ON L7G 4R8 Fax: +1 (905) 873-1992; (800) 516-9081 Web: www.solinst.com E-mail:
[email protected] Tel: +1 (905) 873-2255, (800) 661-2023
SOLINST TECHNICAL BULLETIN High Quality Groundwater Sampling
Representative Samples Sample Water
A project manager’s goal for groundwater sampling is to collect samples that are representative of actual in-situ conditions, and to maintain sample integrity by minimizing the changes in sample quality and chemistry during collection. Sampling has evolved from earlier techniques of bailing and high rate pumping, to techniques of minimal drawdown and low flow pumping. Low flow pumping has been proven to provide consistent, highly representative groundwater samples. By maintaining flow rates of 500 ml/min or less, sample extraction is kept at a rate comparable to ambient groundwater flow and drawdown is minimized, thus reducing turbidity and disturbance in the well. Reduced agitation is particularly important for accurate measurement of dissolved volatile organic compound (VOC) concentrations in groundwater; too much agitation can cause degassing of VOCs to the atmosphere.
High Quality Pumps Solinst Bladder Pumps meet US EPA guidelines for low flow and VOC groundwater sampling. Chemical concentrations of VOCs can be easily lost through vaporization if the sample comes into contact with air. Bladder Pumps provide assurance that there is no contact between the sample water and the drive gas. When in a well, hydrostatic pressure fills the pump’s bladder. Drive gas is applied, which fills the area around the bladder and closes the bottom check valve. Compression of the bladder forces water up into the sample tubing. During a vent cycle, the top check valve closes, keeping water in the sample tubing. The bottom check valve opens and water again enters the bladder. The drive gas never comes in contact with the sample.
12 Volt Compressor
1" SS Bladder Pump
Model 464
Portable Sampling Setup
Application Solinst Bladder pumps are suitable for use in various environments and applications, including municipal, industrial, and general environmental site assessments. As an example, a Brownfield site with contaminated groundwater is set for redevelopment. It is imperative to acquire accurate samples of VOC concentrations from an on-site groundwater monitoring well in order to properly address exposure and remediation concerns. The depth to water in the well is 95 ft. (29 m) with the screened interval at 100 to 105 ft. (30 m to 32 m) below ground surface. A stainless steel Bladder Pump (1.66" (42 mm) ø or 1" (25 mm) ø) with a Teflon® bladder is ideal for this application. The Solinst Model 464 Pump Control Unit is used to control the supply of compressed gas to the pump. The Control Unit, along with a Solinst 12 Volt Air Compressor,
® Solinst is a registered trademark of Solinst Canada Ltd.
allows very slow compression of the bladder, which creates ideal sampling conditions for representative samples. The Model 464 provides 125 psi output and can be set to sample at 100 ml/min or less. To obtain a sample from the middle of the screened interval, or slightly above (about 102 ft. (31 m)), the Control Unit should be set to output at least 50 psi, as it takes about 1 psi to raise a column of water 2 ft (0.6 m). To achieve a low flow rate, careful attention is required to choose the correct drive/vent cycle. The Model 464 has preset sampling modes to choose from. The low flow preset has a 50 second drive time that results in slow compression of the bladder, and a 25 second vent period to allow steady refill of the bladder. Alternatively, to adjust the flow rate to better suit the site, a unique drive/vent cycle can be created and saved in the Control Unit memory. The saved flow rate setting can be used each time that well is sampled. Another well on site is known to contain corrosive contaminants of concern. To get accurate samples here, it is best to use a PVC pump (1.66" (42 mm) ø) and a disposable LDPE bladder. The bladder is lower in cost, so decontamination and reuse is not an issue. In both cases, dedicating a pump to each well avoids cross contamination, and also reduces column disturbance that can be caused by repeated lowering of the pumps.
Conclusion Overall, high quality groundwater sampling requires preparation, the proper technique and the right equipment. To meet the needs of project managers, Solinst offers convenient sampling solutions for various applications, especially where VOC analysis is the focus. ®Teflon is a registered trademark of DuPont Corp.
High Quality Groundwater and Surface Water Monitoring Instrumentation Solinst Canada Ltd., 35 Todd Road, Georgetown, ON L7G 4R8
Tel: +1 (905) 873-2255; (800) 661-2023 Fax: +1 (905) 873-1992; (800) 516-9081
Visit our website: www.solinst.com
E-mail:
[email protected]