Transcript
24-Well Plate 24W700/100F 24-Well Plate with Epoxy Base for Use with the Multiwell-MEA-System
Technical Specifications Advantages
Base material
Epoxy
•
Isolator
Epoxy
Track material
Au (Gold)
Contact pads
Au (Gold)
Electrode material
Au (Gold)
Diameter of the electrodes
100 μm
Interelectrode distance (center to center)
700 μm
Electrodes per well
12 + 1 reference
•
• • •
Ideal for medium throughput recordings from cardiac and neuronal cell cultures. Comes with lid which can be kept in place during recording to enable repeated recordings under sterile conditions. 1 circular ground electrode per well. Each electrode selectable for stimulation. ANSI/SLAS compliant well plate, compatible with traditional plate readers and liquid handling devices.
Applications The 24-well design includes an optimum of 12 electrodes and one circular ground electrode per well. The 24-well plate with epoxy base is perfect for primary, ES or iPS derived cells, for neurons and cardiomyocytes to work on toxicology and - with cardiomyocytes on toxicology and safety pharmacology research. The gold electrodes with 100 μm diameter are ideally suited for recordings and stimulation. Every single electrode can be selected for stimulation.
Warning: Do not autoclave multiwell plates. Do not expose to temperatures over 60° C.
Multi Channel Systems MCS GmbH Aspenhaustrasse 21 72770 Reutlingen Germany
Fon +49-7121-9 09 25- 0 Fax +49-7121-9 09 25-11
© 2016 Multi Channel Systems MCS GmbH a division of Harvard Bioscience, Inc.
[email protected] www.multichannelsystems.com
Product information is subject to change without notice.
24-Well Plate 24W700/100F Pin Layout of the 24-Well Plate with Epoxy Base for Use with the Multiwell-MEA-System Dimensions of one Well Height: 17.3 mm Diameter base: 6.4 mm Diameter opening: 16 mm Total volume: 2.43 ml
J2
J1
J3
J4 Channel 144
Channel 1
85.48 mm
A1
A2
A3
A4
A5
A6
B1
B2
B3
B4
B5
B6
C1
C2
C3
C4
C5
C6
D1
D2
D3
D4
D5
D6 Channel 145
Channel 288
127.76 mm Height of the plate: 19 mm
J7
J8
J6
J5
Electrode ID per Well 21 12
22
13
23
32
42
33
43
700 μm
24
34
100 μm
Multi Channel Systems MCS GmbH Aspenhaustrasse 21 72770 Reutlingen Germany
Inside each well the numbering of MEA electrodes in the 4 x 4 grid follows the standard numbering scheme for square grids: The first digit is the column number and the second digit is the row number. These electrode IDs are displayed in the channel map of the Multiwell-Screen software.
Cleaning Please rinse the well plate thoroughly with distilled water. Fill the wells with Terg-A-Zym for min. 12 hours at room temperature. Wash the wells twice for min. 30 minutes with distilled water again. Optional: Treat Epoxy well plates in a plasma cleaner before use. Fon +49-7121-9 09 25- 0 Fax +49-7121-9 09 25-11
© 2016 Multi Channel Systems MCS GmbH a division of Harvard Bioscience, Inc.
[email protected] www.multichannelsystems.com
Product information is subject to change without notice.