Ultra-Low Temperature V.I.P.™ Freezer
MDF-U33V
Temperature (Ambient temp. 30°C)
-86˚C Effective capacity
333L/11.8cu.ft. ™
Patented V.I.P. V V.I.P .I.P.
(Vacuum Insulation Panel)
Ultra slim, space saving UltraLow Freezer with Improved energy consumption • V.I.P.™ technology maximises storage capacity • Safe operation with Status Alert continuous condition monitoring • The inner doors are removable for cleaning and defrosting • Enhanced security • Improved accessibility • Quiet operation • Improved energy consumption • Can hold up to 216 pcs of 2in boxes or 144 pcs of 3in boxes
MDF-U33V
MDF-U33V Features EVOLUTIONARY DESIGN • SANYO V.I.P.™ Evolution Series -86˚C ultra-low temperature freezers use patented revolutionary vacuum insulation cabinet construction that reduces the wall thickness from 14cm to 7cm (5.5” to 2.7”) and achieves up to 30% more storage capacity than a conventionally insulated freezer without increasing the footprint. • The newly developed refrigeration system and the freezer structure offer a quiet experimental environment. • Achieves 12% higher energy conservation than previous model. (115V, 60Hz) • The newly developed back cover combined with new aerodynamically designed and placed components in the refrigeration compartment provide superior air flow, making it possible to drastically reduce the stress on the freezer and contributing to excellent durability.
Model Temperature range Maximum cooling performance Exterior dimensions (W x D x H) Interior dimensions (W x D x H) Net weight Effective capacity Shelves Access port
MDF-U33V –50˚C to –86˚C (1˚C increments) –86˚C (Ambient temp. 30˚C) 670 x 867 x 1860mm (26.4” x 34.1” x 73.2”) 490 x 600 x 1140mm (19.3” x 23.6” x 44.9”) Approx. 255kg (562 lbs.) 333 L (11.8 cu.ft.) Stainless steel, 3 shelves 17mm diameter, 3 locations (back, bottom left/right corner) Compressor Hermetic type, Output: 450 W (high stage side), 750 W (low stage side) Refrigerant HFC Alarm High/low temperature, Power failure, Filter check, Self diagnostics, Door check, Remote alarm contact, Part replacement notification Remote alarm contact Allowable contact capacity: DC 30V, 2A Accessories 1 set of keys, 1 scraper
• Alarm lamp and buzzer offer secure warning of power failure or abnormal temperature increase. • High and low temperature warning provides an audible and visual alarm when the temperature deviates more than ±5˚C to ±20˚C (adjustable) from the set point. • Alarm ring-back function ensures that buzzer will resume operation, should alarm conditions continue after it is silenced. • Microprocessor-controlled filter-clog check function protects the refrigeration circuit. • The rugged, one-handed outer door latch has a hole to allow a padlock to securely protect valuable samples. • Control panel with digital display for easy operation.
*Cooling performance is indicated by the temperature reached at the center of the freezer (at ambient temperature of 30°C with no load). In order to use the freezer at a stable temperature for a long time, it is recommended that the temperature be set to at least 5°C higher than the indicated lowest temperature. In addition, depending on the usage conditions, it may not be possible to reach the indicated lowest temperature. Caution: SANYO guarantees the product under certain warranty conditions. SANYO in no way shall be responsible for any loss of content or damage to content.
Compressors Back cover
Air
0 1 2 3 4 5 6 7 8 9 10 11 12
40 20 0 –20 –40 –60 –80 –100 –120 –140 –160
20
10
5
Time (hours)
Dimensions 105 (4.1)
Ambient temp. and maintenance of –70˚C by injection of LCO2 (30kg)
15
5
0 1 2 3 4 5 6 7 8 9 10 11 12
Time (hours)
• A beak style inner door latch tightly closes the inner door against the freezer frame. It also helps make opening and closing the door smoother.
[Unit : mm (inch)]
10
15
45 (1.8)
795 (31.3) 600 (23.6)
30 (1.2)
670 (26.4)
25
30
35
Options Rack Capacity
733 (28.9)
867 (34.1)
20
Ambient temperature (˚C)
897 (35.3)
For 2-inch box 6 IR-216U 6 IR-220U For 3-inch box 12 IR-312UN
18 (0.7)
490 (19.3)
450 (17.7)
1639 (64.5)
• The condenser filter is situated at the bottom right side of the front panel to make filter removal and cleaning easier.
Central air temp. in cabinet Ambient temp. 30˚C, No-load
1410 (55.5)
• Status Alert (Condition monitor) Monitors ambient and system conditions continuously and notifies of any abnormalities before a problem happens.
Pull-up characteristics during power failure CO2 backup system operation characteristics
Inventory rack
Inventory rack
IR-216U
IR-220U Slip chart recorder
1860 (73.2)
40 20 0 –20 –40 –60 –80 –100 –120 –140 –160
Time (hours)
Pull-down characteristics
RELIABILITY
ACCESSIBILITY
•Appearance and specifications are subject to change without notice.
Performance Data Temperature (˚C)
• Two independent and insulated inner doors with gaskets prevent cold air leakage.
1140 (44.9)
Rear
Temperature (˚C)
Front
Specifications SAFETY
MTR-85H Recorder mounting bracket Circular recorder
MDF-S3085
MTR-G85
(for MTR-85H)
SANYO DAQ (Data Acquisition) system
LCO2 backup system
MTR-480/MTR-2000
LN2 backup system
CVK-UB2 CVK-UBN2
SANYO Electric Co., Ltd., Biomedical Division, Gunma is certified for: Quality management system: ISO9001 / Medical devices quality management system: ISO13485 / Environmental management system: ISO14001 RoHS (European Restriction of Hazardous Substances directives) compliant Distributed by:
SANYO Electric Co., Ltd. Biomedical Division http://www.sanyo.co.jp/cmg/biomedical ©2008 SANYO Printed in Japan 2008.9 MA
SHR128
Ultra-Low Temperature V.I.P.™ Freezer
MDF-U53V
Temperature (Ambient temp. 30°C)
-86˚C Effective capacity
519Liter/18.3
cu.ft.
™
Patented V.I.P. V V.I.P .I.P.
(Vacuum Insulation Panel)
Large capacity, space saving Ultra-Low Freezer with Improved energy consumption • V.I.P.™ technology maximises storage capacity • Safe operation with Status Alert continuous condition monitoring • The inner doors are removable for cleaning and defrosting • Enhanced security • Improved accessibility • Quiet operation • Improved energy consumption • Can hold up to 352 pcs of 2-inch boxes or 224 pcs of 3-inch boxes
MDF-U53V
MDF-U53V Features EVOLUTIONARY DESIGN • SANYO V.I.P.™ Evolution Series -86˚C ultra-low temperature freezers use patented revolutionary vacuum insulation cabinet construction that reduces the wall thickness from 14cm to 7cm (5.5” to 2.7”) and achieves up to 30% more storage capacity than a conventionally insulated freezer without increasing the footprint. • The newly developed refrigeration system and the freezer structure offer a quiet experimental environment. • Achieves 10% higher energy conservation than previous model. (For model rated 220V, 50Hz) • The newly developed back cover combined with new aerodynamically designed and placed components in the refrigeration compartment provide superior air flow, making it possible to drastically reduce the stress on the freezer and contributing to excellent durability.
SAFETY
Model Temperature range Maximum cooling performance Exterior dimensions (W x D x H) Interior dimensions (W x D x H) Net weight Effective capacity Shelves Access port
MDF-U53V –50˚C to –86˚C (1˚C increments) –86˚C (Ambient temp. 30˚C) 770 x 875 x 1990mm (30.3” x 34.4” x 78.3”) 630 x 600 x 1380mm (24.8” x 23.6” x 54.3”) Approx. 299kg (659 lbs.) 519 L (18.3 cu.ft.) Stainless steel, 3 shelves 17mm diameter, 3 locations (back, bottom left/right corner) Compressor Hermetic type, Output: 1100 W (high stage side), 1100 W (low stage side) Refrigerant HFC Alarm High/low temperature, Power failure, Filter check, Self diagnostics, Door check, Part replacement nortification Remote alarm contact Allowable contact capacity: DC 30V, 2A Accessories 1 set of keys, 1 scraper
• Alarm lamp and buzzer offer secure warning of power failure or abnormal temperature increase. • High and low temperature warning provides an audible and visual alarm when the temperature deviates more than ±5˚C to ±20˚C (adjustable) from the set point. • Alarm ring-back function ensures that buzzer will resume operation, should alarm conditions continue after it is silenced. • Microprocessor-controlled filter-clog check function protects the refrigeration circuit. • The rugged, one-handed outer door latch has a hole to allow a padlock to securely protect valuable samples. • Control panel with digital display for easy operation.
*Cooling performance is indicated by the temperature reached at the center of the freezer (at ambient temperature of 30°C with no load). In order to use the freezer at a stable temperature for a long time, it is recommended that the temperature be set to at least 5°C higher than the indicated lowest temperature. In addition, depending on the usage conditions, it may not be possible to reach the indicated lowest temperature.
Compressors Back cover
• Two independent and insulated inner doors with gaskets prevent cold air leakage.
Performance Data Pull-down characteristics
Pull-up characteristics during power failure CO2 backup system operation characteristics 20
• Status Alert (Condition monitor) Monitors ambient and system conditions continuously and notifies of any abnormalities before a problem happens.
Central air temp. in cabinet Ambient temp. 30˚C, No-load
20 0
Temperature (˚C)
RELIABILITY
Temperature (˚C)
40
–20 –40 –60 –80
–100
0
5
10
15
12
0
Time (hour)
Rear
Caution: SANYO guarantees the product under certain warranty conditions. SANYO in no way shall be responsible for any loss of content or damage to content. •Appearance and specifications are subject to change without notice.
Air
–20 –40 –60
–100
0 1 2 3 4 5 6 7 8 9 10 11 12
Time (hour)
21 (0.8)
795 (31.3)
5 (0.2)
600 (23.6) 4 5 (0.16) (0.2)
59 (2.3)
450 (17.7)
329 (13.0)
655 (25.8)
1380 (54.3)
630 (24.8) 5 (0.2)
875 (34.4)
10
0
5
4327651 81 453276 81 2453 678 23451 678 3241 4532671 81 465327 81 2453 786 4327651 81 432
15
20
25
30
35
Options
Inventory rack
IR-220U IR-224U
11 (0.4)
28 (1.1)
10
Ambient temperature (˚C)
105 (4.1)
1545 (60.8)
770 (30.3)
660 (26.0)
• The condenser filter is situated at the bottom right side of the front panel to make filter removal and cleaning easier.
45 (1.8)
[Unit : mm (inch)] 33 (1.3)
836 (32.9)
15
Time (hour)
Dimensions
1540 (60.6)
• A beak style inner door latch tightly closes the inner door against the freezer frame. It also helps make opening and closing the door smoother.
1990 (78.3)
ACCESSIBILITY
Ambient temp. and maintenance of –70˚C by injection of LCO2 (30kg)
–80
1770 (69.7)
Front
Specifications
3 drawers (for lower chamber)
MDF-50R Strip chart recorder
MTR-85H Recorder mounting bracket
MDF-S3085 (for MTR-85H) Circular recorder SANYO DAQ MTR-G85 (Data Acquisition) system
LCO2 backup system
CVK-UB2 LN2 backup system
CVK-UBN2
MTR-480/MTR-2000
SANYO Electric Co., Ltd., Biomedical Division, Gunma is certified for: Quality management system: ISO9001 / Medical devices quality management system: ISO13485 / Environmental management system: ISO14001 RoHS (European Restriction of Hazardous Substances directives) compliant Distributed by:
SANYO Electric Co., Ltd. Biomedical Division http://www.sanyo.co.jp/cmg/biomedical ©2008 SANYO Printed in Japan 2008.12 MA
SHR139