Transcript
ANNEXURE TO TENDER NO : DPS/ MRPU/IGCAR/ MIA/7459 ANNEXURE for FUNCTION GENERATOR 1. Description of the item and quantity: Single channel, 10MHz Function /Arbitrary Waveform Generator capability and deep memory with LXI Class C compliant, Custom waveform generation, Digital readout and Over current protection with all accessories and software. 2. Detailed Technical Specification: S.NO Parameter Name 1 General Characteristics i Types of Output Waveform ii Built in Arbitrary waveform iii
User Defined Arbitrary waveform Number of Output Channels
2
3 i ii iii 4 i ii iii iv v 5 i ii iii 6 i 7 i ii iii iv v vi
Qty: 1 No.
Value Sine, Square, Ramp, Triangle, Pulse, Noise, DC Exponential rise, Exponential fall, Negative ramp, Sin(x)/x, Cardiac 14-bit, 50 MSa/s 8K point 1 (one) (Impedance- 50 Ω typical, Isolation - 42 Vpk maximum to earth, Protection - Short-circuit protected, overload automatically disables main output)
Sine waveform characteristics Frequency range 1 mHz to 10 MHz Amplitude Flatness (up to Max ±0.3 dB 10MHz) THD (DC to 20 kHz) <0.04% Square & Pulse characteristics Range 1 mHz to 10 MHz Rise/fall time 20 ns overshoot <2% Duty cycle Variable from 40% to 60% (till 10 MHz) 1 ns + 100 ppm of period (square) Jitter 300 ps + 0.1 ppm of period (pulse) Ramp & Triangle characteristics Frequency range 1 mHz to 100 kHz Linearity < 0.1% of peak output Variable symmetry 0.0% to 100.0% Noise characteristics Bandwidth 7 MHz typical Arbitrary Waveform Characteristics Frequency range 1 mHz to 3 MHz Waveform length 2 to 8 k points Sample Rate 50 MSa/s Voltage Resolution 14 bits (including sign) Min. rise/fall time 70 ns typical Settling Time < 500 ns to 0.5% of final value
vii Jitter (RMS) Common characteristics: 1
Frequency i Accuracy
ii Resolution 2 Amplitude i Range ii Units iii Resolution iv Accuracy (at 1 kHz) 3
6 ns + 30 ppm
±(20 ppm + 3 pHz) in 1 year 1 μHz (internal) 10 mVpp to 10 Vpp into 50 Ω 20 mVpp to 20 Vpp into open circuit Vpp, Vrms, dBm 3 digits ± 2% of setting ± 1 mVpp
DC offset Range i (peak AC + DC) ii Resolution
± 5 V into 50 Ω ± 10 V into open circuit 3 digits ± 2% of offset setting iii Accuracy ± 0.5% of amplitude ± 2 mV 4 Modulation Parameter Settings i AM, FM, PM modulations - Internal/External
5
ii AM modulation depth
0.0% to 120.0%
iii FM frequency deviation
DC to 5 MHz
iv PWM deviation Burst i Waveforms type ii Type iii Internal period
0% to 100% of pulse width
iv Start/Stop Phase
-360º to + 360º
v Gate Source
External trigger
vi Trigger source
Single, External or Internal
6 i ii iii iv
Trigger characteristics Input level Slope Pulse width Input impedance
v Jitter (rms) vi vii viii ix 7 i
Output Level Output impedance Maximum rate Pulse width Power Supply Voltage
Sine, Square, Ramp Counted (1 to 50,000 cycles), Infinite, Gated 1 μs to 500 s
TTL compatible Rising or Falling, Selectable > 100 ns > 10 kΩ, DC coupled 6 ns (3.5 ns for pulse) TTL compatible into ≥ 1 kΩ 50 Ω typical 1 MHz > 400 ns 100 V - 240 V 50/60 Hz -5%, +10%;
100 V - 120 V 400 Hz ±10% < 45 W, < 130 VA
ii Power consumption (typ) 8 Mechanical & Environmental i Size Approx 262mm W x 104 mm H x 304m D - Bench top ii Operating environment EN61010, pollution degree 2; indoor locations iii Operating temperature 0˚C to 55˚C iv Operating humidity 5% to 80% RH, non-condensing Shall Conform to UL-1244, CSA 1010 EN61010 v Safety & EMC Compliance MIL-461C, EN55011,EN50082-1 9 Interface LAN LXI-C Ethernet 10/100 USB 2.0, GPIB
NOTE:1. Quotation shall be offered with detailed product specification and catalog with compliance of all the technical specification. Any deviation in the technical specification may be clearly specified. The item shall be supplied with product manual, troubleshooting guide, service manual and quick reference guide. 2. The item must be supplied with standard calibration certificate & test report. Standard Accessories must be supplied with the item viz. Power cable, USB Cable , Signal cable for Input and Trigger/Sync, software in CD compatible to WinXP with all utilities and drivers for programming etc. 3. The item shall have standard ONSITE warranty of one year. The quotation which does not offer onsite warranty may not be considered. The firm must demonstrate detailed functionality of the item at site.
ANNEXURE for 6 ½ Digit Multimeter I. Description of the ITEM and Quantity: 6 ½ digit table top true RMS auto and manual ranging Digital Multimeter as per the following specification. Quantity: 1 No. II. Detailed SPECIFICATIONS & Quantity: : DCV, True RMS ACV, DC Current, AC Current, Resistance, Input Frequency, Continuity Test, Diode Test. I DCV: Range
: 100mV to 1000V (100mV, 1V, 10V, 100V, 1000V)
Accuracy
: less than 35ppm
Measurement method
: continuously integrating, multi-scope III ADC
Input resistance
: 0.1V, 1V, 10V with range – 10MΩ/10GΩ selectable For 100V, 1000V- 10MΩ
Input bias current
: <30pA at 25 °C
Input protection
: 1000V on all range.
II True RMS AC voltage: Range
: 100mV to 750V
Frequency range
: 3Hz to 300 KHz
Measurement method
: AC coupled true rms
Crest factor
: maximum 5:1 at full scale
Input impedance
: 1MΩ
Input protection
: 750V rms all range
Accuracy
: ±0.04%
III DC Current: Range
: <1mA to 10A
I/P protection
: with internal and external fuses
Accuracy
: ±0.01%
IV True RMS AC Current: Range
: <1mA to 10A
Frequency range
: 3Hz to 10 KHz
Measurement method
: AC coupled true rms measurement.
I/P protection
: with internal and external fuses
V Resistance Range
: 100 Ω to 1000M Ω
Measurement method
: 4 wire or 2 wire selectable
I/P protection
: 1000V on all ranges
Accuracy
: ±0.002%
VI Capacitance Range VII Continuity/Diode test Continuity Test
: 1 nF to 100 μF : 1000 Ω at 1mA
Diode Test
: 5 V (for 1mA)
Response time
: 300 samples/sec with audible tone
Accuracy
: ±0.002%
VIII Frequency Range
: 3Hz to 300 KHz, also square wave
AC voltage range
: 100mV to 750V(auto/manual range)
Measurement method
: reciprocal counting technique.
Accuracy
: 0.006%
Gate time
: 1ms, 10ms, 100ms or 1sec
IX Display No. of digits
: 6 ½ digit
Display type
: 7 segment vacuum fluorescent display
Display resolution
: 4 ½, 5 ½, 6 ½ selectable
X Math functions XI Desirable features
: Null, min/max/avg, dB, dBm, limit test : Automatic reading hold, 5k readings/second
XII General specifications Power supply
: 220 V AC ± 10 %, 50Hz
Power consumption
: 25VA max
Range selection
: auto/ manual
Manual range selection
: up/down key
I/P terminal
: Copper alloy
Safety
: EN61010 – 1:2010
RFI & ESD
: MIL-461C,FT21046,FCC
Operating environment
: upto 550C, 90% R.H
Storage environment
: up to 700C
Shock and vibration
: MIL-T-28800E type III class 5
Warranty
: 1 years
Triggering
: 1 to 1,00,000 samples per trigger
Remote interface
: RS-232C , USB, LAN/LXI core
Size
: 103mm H x 261mm W x 303mm D
Accessories
: test lead kit with probes, alligator and grabber, attachments, User guide, Service guide, Power cord and Software CDs.
NOTE: • • •
The calibration certificate and test data of the multimeter must be supplied along with the item. Necessary catalogues with technical specifications must be enclosed to the quotation. Without catalogue containing detailed technical specification, offer will not be considered. Software for multiple measurements simultaneously, for visualizing and performing data logging.