SY-8217
B-H curve measurement over 10 Hz to 5 MHz. The standard core loss measurement tool replaces V-A-W meters. The SY-8217 measures the magnetization characteristics of soft magnetic materials with high accuracy and in a short time over a wide frequency range from 10 Hz to 5 MHz. It is a powerful tool for research, development, manufacturing and inspection of low-loss magnetic materials for use in compact motor yokes, flyback transformers for TV, highfrequency power transformer, etc. Incorporating an oscillator for measurements and digitizer in a compact, desktop type cabinet, the SY-8217 is the ideal repalcement for an existing V-A-W(well known as a U-function) meter as the standard tool for core loss measurement.
SY-8217 10 Hz to 5 MHz
Wide measurement frequency bandwidth
Color TFT display The bright color TFT display shows the characteristic values, B-H curves and excitation waveforms simultaneously.
The frequency range for measurement ranges from 10 Hz to 5 MHz.
Printer and 3.5-inch FDD provided as standard
High-precision measurement An FFT arithmetic circuit permits automatic compensation and calibration of frequency characteristics of the measurement system.
The measurement results can be printed out immediately. Convenient for data sorting. The results can also be output to a 3.5-inch floppy disk in the csv format, then loaded in a PC and sorted using tabulation software.
High-speed measurement High-speed measurement can be performed in approx. 3 seconds, considerably improving measurement efficiency.
GP-IB interface provided as standard An automated measurement system can be constructed with a PC.
Compliance with JIS C2514, JIS H7153 and EMAS 5003. Ideal for estimation of everything from yoke materials for compact motors to the materials for high-frequency transformers. JIS C2514 : E-shaped ferrite core. JIS H7153 : High-frequency core loss test method for amorphous metallic cores. EMAS 5003 : Test method for ferrite cores of power applications. IEC 62044-3 : Cross-power method.
Compact, desktop size (Rack mountable)
Easy operation Simple and easy to use design.
35
B-H Analyzer
SY-8217
Major Applications Development and inspection of high-frequency power cores, micromotor yokes, etc. Core loss measurement by replacing a U-function meter. Manufacturing and inspection of amorphous thin films and amorphous cores.
Example of Combination For R&D
+
Power amplifier IE-1119 Power amplifier : IE-1119
SY-8217+ IE-1125
For manufacturing & inspection
B-H analyzer SY-8217
For automated temperature measurement
B-H analyzer SY-8217
+
Power amplifier IE-1125
+
Power amplifier IE-1125
Scanner
+
(with oven)
IE-1123, IE-8I7
Specif i cations Oscillating section Oscillation frequency: 10 Hz to 5 MHz, sine wave Output voltage/resistance: 0 to ±2.5 V/50 Built-in excitation control in 3 steps of approximate value, asymptotic and demagnetization. Detection section (standard) Excitation current detection System: Voltage drop across two ends of non-inductive resistance. Detection sensitivity: ±5 mA to ±5 A (1-2-5 steps) Frequency bandwidth: DC to 50 MHz Magnetic flux detection System: Detection of voltage (induced voltage) across two ends of detection coil. Detection sensitivity: ±5 mV to ±200 V (1-2-5 steps) Frequency bandwidth: DC to 50 MHz Storage section Sampling rate: 512 times the set frequency (synchronized with the built-in oscillator). Resolution: 16 bits Display section System: Color 8.4-inch TFT Contents: Measuring condition, characteristic values, B-H curve, excitation current and induced voltage waveforms Printer (Built-in) Heat-sensitive dot line system. Paper width approx. 112 mm, printing time no more than approx. 5 sec. Floppy disk drive 3.5-inch FDD (DOS format) External interface GP-IB interface Power supply 100 to 240 V AC, approx. 400 VA (when the built-in printer is running) Dimensions & weight Main unit: Approx. 420W x 266H x 480L mm (excluding the pod), approx. 17 kg Environment Operating temperatures +5°C to +35°C (+10°C to +30°C when the printer is used) Performance guaranteed temperatures +18°C to +28°C (after power on 60 min.) Provided accessories Measurement pod (1 set), power cord (1), sample cores (1 set), adapters for calibration (1 set), thermal paper (2 rolls), fuses (2), instruction manual (1), dust cover (1)
Measurement method Cross power method Measurement modes Core loss measurement: Frequency (f) and maximum magnetic flux density (Bm) can be set. B-H curve measurement: Frequency (f) and maximum magnetic field (Hm) can be set. Measured items Core loss measurement: Max. magnetic flux density (Bm), residual magnetic flux density (Br), max. magnetic field (Hm), coercive force (Hc), core loss (Pc, Pcv, Pcm), phase ( ), amplitude permeability (µa). B-H curve measurement: Max. magnetic flux density (Bm), residual magnetic flux density (Br), max. magnetic field (Hm), coercive force (Hc), core loss (Pc, Pcv, Pcm), phase ( ), amplitude permeability (µa), differential permeability (µdiff), square ratio (Br/Bm), total magnetic flux change (2 m). Operation modes Single : One measurement step with one operation. Set frequency 10 Hz to 99 Hz, 2 digits effective 100 Hz to 5 MHz, 3 digits effective Accuracy: ±0.1% for the set value. Measurement accuracy (23°C ±5°C)* IF – 1))** H amplitude measurement (dH): = ± (1% + 0.2% x ( Ir VF – 1))** B amplitude measurement (dB): = ± (1% + 0.2% x ( Vr = ± (0.1° + 0.05° x ( IF + VF – 2) + 0.01°)** Phase accuracy (d ): Ir Vr Core loss measurement (dPc): = ± ( l dH l + l dB l + l
180°
x d x tan
l x100 )%
Measurement time Approx. 3 sec./step (when the measured frequency is 1 kHz or more)
* Defined as the basic accuracy. The measurement accuracy may differ depending on the D.U.T. and the measuring condition. For details, please contact to Iwatsu.
**