Transcript
CMB551 1A: Microscopy and Image Analysis in Cell Biology Sam Johnson http://microscopy.duke.edu/learn/CMB551.html
[email protected]
August 2015
Fluorescence
High-contrast, multi-channel imaging.
Relies on intrinsic contrast, little specificity
Allows many modes of imaging – confocal, TIRF, 2-photon, singlemolecule, fluorescent proteins= live cell imaging, FRAP, photoconversion, FRET, FCS, super resolution. . .
Fluorescence can be . . . 488
2˚ 1˚
Basis of all fluorescence scopes
Photo-bleaching Chemical reaction causing irreversible loss of fluorescence Reactive Reactive and long-lived
Photo-toxicity Same principle as photo-bleaching but this time it’s bad for the cell not just the fluorophore
Oxygen
Radicals
Generally UV light is more toxic than far-red wavelength, with or without any exogenous fluorophore around
Modalities
|
Photon budget
The insides of a microscope: Fluorescence
Adapted from http://www.microscopyu.com/articles/formulas/formulasconjugate.html
Microscope configurations
The objectives point in different directions but the optical principles are the same
The objective Collects light from the sample and forms an image up the microscope near the eyepieces
Intermediate image
SAMPLE
Objectives are the most important parts
Three important concepts in microscopy
• Magnification • Resolution • Contrast
Magnification “How many times bigger the image is than the object”
Magnification is not a very useful concept on its own
Resolution “The smallest distance between two objects that can be observed as two objects”
It doesn’t mean anything else (e.g. how nice the image looks)
Does the resolution also limit the smallest object we can see?
Imaging a tiny fluorescent object
Imaging a tiny fluorescent object
In 3D this is a point-spread function XZ XY
Resolution in fluorescence terms Object
Airy disk
http://www.olympusmicro.com/primer/java/imageformation/rayleighdisks/index.html
Resolution limit of Light Microscopy About 200 nm • Naked eye = 100 mm • Electron microscopy = <1 nm
http://learn.genetics.utah.edu/content/begin/cells/scale/
Reference sizes • • • • • • •
Diameter of an adherent fibroblast 100 mm Mammalian cell nucleus 10 mm Red blood cell 7 mm Bacteria 1 mm Virus 50 nm Ribosome 20 nm Globular protein 2 nm
Resolution in fluorescence terms Imaging green fluorescent beads of different sizes
1000 nm
Objects
Images
750 nm
500 nm
250 nm
125 nm
Resolution in transmitted light terms With brightfield TL you don’t see things below the resolution
Ok
Small objects diffract light at a greater angle, not captured by the lens
Brightfield illumination
Contrast Contrast is needed in the image to be able to resolve anything
Contrast =
DI I
Signal to noise is a related concept
Numerical Aperture
NA = n sin (m)
Numerical Aperture determines. . .
Brightness
= NA4 / Mag2 (epi) = NA2 / Mag2 (trans)
Resolution
= 0.61 l / NA
Example set objectives for fluorescence Brightness
Magnification
NA
= NA4 / Mag2 = NA2 / Mag2 Immersion
(epi) (trans)
Resolution (nm)
Relative brightness
5x
0.15
DRY
2033
1.0
10x
0.30
DRY
1017
4.0
20x
0.50
DRY
610
7.7
40x
0.75
DRY
407
10
63x
1.40
OIL
218
48
100x
1.40
OIL
218
19
This formula ignores transmission efficiency
Numerical Aperture and working distance
Low NA
High NA
Working distance
NA and depth of . . . • Depth of field – Z- range of image that is sharp • (Depth of focus – Range focus can be moved and image still sharp, more a photography term but sometimes used interchangeably)
Optical cost comparison
$3.29
$11,822
Aberrations Simple lenses
Problems . . . chip
Field curvature
curved image plane
Objectives correct for some aberrations
Good optics are relatively free of aberrations. Microscopes are expensive because high magnification requires the best optics.
Many types of objective Achromats – Fluorite Apochromats – Plan Apochromats –
Limited color correction Good all round objectives, good transmission, NA up to 1.3 Highly color corrected Additional correction for field curvature
$
$$$
http://www.olympusmicro.com/primer/anatomy/objectives.html
Many types of objective immersion
Dry
Dipping
Oil
Silicon n=1.4
Water
Glycerol
Multi
Why use oil or water objectives? Higher NA = more resolution and brightness
NA = n sin (m) (sin (90) = 1) n = refractive index Effective limit of NA Air = 1.0003 ~0.95 Water = 1.33 ~1.2 Glass = 1.515 ~1.4
Why use oil or water objectives? Higher NA = more resolution and brightness
NA = n sin (m) (sin (90) = 1) n = refractive index Effective limit of NA Air = 1.0003 ~0.95 Water = 1.33 ~1.2 Glass = 1.515 ~1.4
Why use oil or water objectives?
Virtual fish Fish
Matching the refractive index of the sample and the objective immersion helps keep aberrations and degradation to a minimum
Quiz 5 – (sorry about the numbering)
Coverslips
Number
Ideal thickness
Range
#0
100 μm
80-130 μm
#1
150 μm
130-170 μm
#1.5
170 μm
160-190 μm
#2.0
220 μm
190-250 μm http://microscopy.duke.edu/coverslip.html
Coverslip thickness is very important
Worst for high NA dry objectives
Adjustable correction collar to minimize spherical aberration and light loss
http://www.microscopyu.com/articles/formulas/formulascoverslipcorrection.html
Infinity optics A missing bit from a few slides ago
Infinity advantages -
Easier to add extra components (eg filters) Can focus by moving the objective
http://www.olympusmicro.com/primer/anatomy/infinityintro.html
/The objective Collects light from the sample and forms an image up the microscope near the eyepieces
Intermediate image
SAMPLE
The eyepiece (s)
Intermediate image
Magnifies the intermediate image and allows us to see the image
SAMPLE
Total magnification = Mobjective x Meyepeice
Attaching a camera to a scope
1.4 Mpixels
12 Mpixels
~$12 000 from Photometrics
~$80 from BestBuy
The details required for microscopy cameras • Quantum Efficiency – Chance of a photon being recorded
• Dynamic range – Full well capacity (how many electrons each pixel holds)/noise • Cooling – Low light levels, noise is significant and cooling helps. • Noise – Cooling reduces dark noise a lot. Electronics optimized for low read noise.
• Frame rate – 100 fps can be useful sometimes. • Grade – No (or few dead pixels)
CCDs, charge-coupled devices
Photons Charge Pixels
http://www.microscopyu.com/articles/digitalimaging/ccdintro.html
CCD sensors Photons Charge Pixels
http://www.microscopyu.com/articles/digitalimaging/ccdintro.html
Binning Combining pixels on the CCD chip to make superpixels
Bin 8:
Bin 4:
Bin 2: 672x512
Bin 1: 1344x1024
Binning
More photons collected in each output pixel
Electron Multiplying-CCD cameras
Also tend to be built for highest QE – peak ~95% (cf ~70% normal)
Good for very fast, dim or photosensitive samples http://learn.hamamatsu.com/articles/emccds.html
CMOS CCD
Each column in a CMOS chip has own readout structure Possible to read the image faster (and use a larger chip = larger fov or more resolution) 2048 x 2048 px resolution at 100 fps
~50 GB/min!
Fluorescence in more detail
http://www.olympusmicro.com/primer/java/jablonski/jabintro/
Excitation and emission
Stokes shift
Reducing energy
Notice the “tails” http://www.invitrogen.com/site/us/en/home/support/Research-Tools/Fluorescence-SpectraViewer.html
Emission is excitation-independent
Always back to here before fluorescing
Illumination sources Arc lamp
Metal Halide http://zeiss-campus.magnet.fsu.edu/articles/lightsources/mercuryarc.html
Regulation of exposure to excitation Arc lamps tend to bright, unregulatable and slow to turn on/off (30 min minimum to extend life)
•Fast electromechanical shutter •About 10 mSec cycle
• • •
Neutral Density Usually expressed as a transmission ND20 = 20% of light goes through
Illumination sources: Arc lamps and similar Mercury Arc lamp
Spectra uneven and much in UV
Xenon
Spectra more even, weak in UV
Arc lamp
Visible
Illumination sources: LEDs and similar Last ~forever (50 000 hr)
Fast switching and power regulation Taking off for the future - cheaper, brighter, more l . . .
Filters and dichroics
T
Filter terminology Bandpass (BP) – 480/40 CWL/bandwidth Full-Width at Half Maximal (FWHM) Central wavelength Longpass – LP600 (Defined around center of cut-on l)
(guess what, say, Short Pass 670 means) http://micro.magnet.fsu.edu/primer/techniques/fluorescence/filters.html
Filters and dichroics Even better filter set
T
Filter choice and efficiency Choose filter to match your fluorophore (tools to help)
More photons = better Does this help?
Filter choice and efficiency A fluor’ with a fairly small stokes shift
Best excitation
Better emission
Multi-channel imaging More than one color; more than one fluorescent protein, probe, antibody . . .
AF488
AF568
Multi-channel imaging Filter cubes are an easy way of switching excitation filter, dichroic and emission filter
Spectraviewer quiz http://www.invitrogen.com/site/us/en/home/support/Research-Tools/FluorescenceSpectraViewer.html
Use a spectra viewer to choose which of these filter sets you would prefer to use for a sample labelled with AF488 and AF568 Set A: Cube1 - x480/25 dichroic 495 m520/40 Cube2 - x570/20 dichroic 585 mLP590 Set B: Cube1 - x500/25 dichroic 515 m540/30 Cube2 - x595/10 dichroic 610 mLP615
Bleedthrough
DAPI
AF488
• Down the spectrum (1st law of thermodynamics) • Worse when intensities are unbalanced
Other multi-channel methods DAPI
Can be faster Better registration Some disadvanatges
GFP
AF568
Exciters Dichroic Emitters
The insides of a microscope: transmitted light
Adapted from http://www.microscopyu.com/articles/formulas/formulasconjugate.html
Light source: Halogen bulb
The condenser The condenser collects the light and concentrates it onto to the specimen
May also have some special parts in for phase contrast or DIC etc
Kohler illumination
Bright and even illumination with good contrast and resolution Optimal alignment of the condenser How do we do this alignment in practice? http://www.microscopyu.com/tutorials/java/kohler/
Kohler illumination in practice Open the apertures and focus on your specimen as best you can in brightfield. Don’t change the focus through out. You can adjust the lamp brightness at any stage
• Close down the field diaphragm/aperture so it becomes visible - you should see an octagon shaped aperture appear but it may be very blurry
• Focus the condenser with the knobs that raise/lower entire condenser - the octagon shaped field aperture should be made as sharp as possible
• Then center the condenser using the two centering pins • Open the field diaphragm until it is just out of view – now the whole area is evenly illuminated • Adjust the condenser aperture so the contrast of the image is good – you can do this empirically or remove the eyepiece and adjust so 2/3 to 5/6 of the pupil is filled
Components on an inverted scope
Components on an upright scope
Back focal plane of the objective Image
Diffraction
Object
Conjugate planes Image, object or field planes
Illumination, aperture, or diffraction planes
D = Retina of the eye
Eye
C = Intermediate image
4 = Pupil of the eye
Eyepiece
3 = Objective pupil B = Specimen plane
Objective
Condenser
2 = Condenser aperture
A = Field diaphragm
Lamp
1 = Lamp filament
If that sounds useful to you please find a simple scope and try the alignment process (or use this virtual scope) http://www.microscopyu.com/tutorials/java/kohler/
Contrast in transmitted light images Cells don’t absorb much light, there isn’t much contrast based only on that
Staining isn’t ideal
Many methods and variations to increase the contrast over brightfield. . .
Phase contrast Brightfield
Phase
Simple to setup, good depth of field, copes with plastic (ideal for TC cells)
. . . but not much contrast is produced
Surround wave
Diffracted waveWhy is it smaller amplitude?
Combined wave
Much more contrast in this condition . . .
Surround wave
Diffracted waveNote the similar amplitude?
Combined wave
Phase contrast Interference at BFP of objective
Objective
Scattered light has its phase shifted by 1/4l
Condenser
Phase contrast alignment Kohler alignment with the condenser aperture fully open
DIC (Normarski) Brightfield
DIC
More complex and $$
Pseudo-3D
Two beams of different polarization pass through the sample and are recombined where they interfere and produce an image with high contrast
Differential Interference Contrast microscopy
0˚
90˚
http://en.wikipedia.org/wiki/Differential_interference_contrast_microscopy
Differential Interference Contrast microscopy
http://en.wikipedia.org/wiki/Differential_interference_contrast_microscopy
The problem with widefield microscopes and biology
Sectioning Vibratome
Microtome
Cryostat
Non-physical sectioning – optical sectioning by . . . Confocal | Multiphoton | Spinning disk | TIRF | SPIM
The confocal principle
How a laser scanning confocal microscope works
http://www.olympusfluoview.com/theory/index.html
The confocal advantage
Optical sectioning of thick samples
3D reconstruction
What a confocal looks like Scanhead
A microscope (inverted or upright)
Lasers and electronics
Probably on an air table
Computer
LASERs are used for excitation
Ideal for point scanning: •Narrow collimated beam, low divergence •Powerful
http://www.olympusfluoview.com/theory/laserintro.html
Many lasers available . . . Gas lasers
Diode lasers
(Kr/Ar
488 568 647)
405
Ar/Ar
458 488
488
HeNe
543 594
633
514
561 635
(Color coding refers to the color of the fluorophore for which the laser line is most commonly used)
Adjusting the laser power Relatively fixed output Rapid control 0-100% T
Laser
AOTF
Laser
AOTF
l selection and intensity control
Fluorophore saturation Bleaching is proportionately worse here
Emission intensity
Image will suffer loss of contrast and quantification issues
Excitation intensity
Widefield is normally in the linear range, the concentrated laser spot in confocal may not be. Start low and increase to the minimum necessary
Scanning mirrors
The relative position of the two mirrors can point the spot anywhere in the field http://www.olympusconfocal.com/theory/confocalscanningsystems.html
Scan speed
Fast scans
• Fast processes • Useful for focusing and adjustments – eg 1 fps
Slower scans
• More light gathered • Better images • More damage
Averaging Scanning each pixel multiple times and averaging improves the noisy signal . . .
Averaging: how much do you need? 1
2
8
4
16
Decreasing rate of improvement, empirically determine a good balance between final SNR and time/damage in acquisition
Pinhole adjustment
Axial (z) or lateral (xy) resolution
Is the pinhole . . .
A. A big waste of signal? B. Only good as we didn’t want that light? C. A bit of both
Signal : Noise
Signal : background
Photomultiplier tubes (PMTs) are used as detectors
• • • • •
Fast (good for scanning) Large collection area Good SNR Very large dynamic range (with gain&offset adjustment) Adequate dynamic range at a single gain&offset position
• QE <30% (not as good as CCD)
Gain and offset adjustments 100 msec
200 msec
400 msec
600 msec
800 msec
1000 msec
2000 msec
Camera:
Confocal:
Increasing gain (voltage on PMT)
Offset to set the background to black
Optimal gain and offset Confocals have special display modes to highlight saturated and 0 intensity pixels
Should you have no, a few or many saturated and zero intensity pixels?
Multi-channel confocals Most confocals have several laser lines and PMTs
2˚ dichroic
1˚ dichroic
Why don’t we in general do this with widefield systems?
Simultaneous or sequential acquisition Blue Green Red Faster
Blue then Green then Red Less bleedthrough
Line vs frame switching
Blue line1 then Green line1 then Red line1 Blue line2 then Green line2 then Red line2
The AOTFs switch the laser light on and off very rapidly Good to see all the channels appearing together
Blue image then Green image then Red image
Good when something physical happens between colours (eg change dichroic) Use this when you have more channels than PMTs
Spectral imaging: serial Leica spectral detector Select l Prism
PMT
Lamda-scan over time
Spectral imaging: parallel
Zeiss META/QUASAR detector
Array of 32 PMTs
Unmixing overlapping signals Reference spectra (or ACE)
Overlapping signals
Separate signals
Ideally, avoid having to do this. But maybe . . . • • •
You’ve only GFP-YFP mice You have a strong autofluorescence (eg chloroplasts) You need many colours
Scan area: zoom The area swept by the galvo mirrors can be adjusted . . .
Is this meaningful zoom or just digital zoom? . . .
Scan area and number of pixels Any particular frame can have different numbers of pixels. . .
Zoom and number of pixels are obviously related
How many pixels do I need in my image?
Same area: 225 mm across
225 mm/512 = 440 nm
225 mm/1024 = 220 nm
The most possible, right? The Leica SP5 goes up to 8K by 8K, shall we have 64 Mpx for every scan?
How many pixels do I need in my image for the best resolution? Nyquist sampling theorem: Sample at twice the resolution
Resolution
=
0.61 l / NA
Increasing number of pixels per area
Signal under-sampled
Not capturing all the resolution of the system
Signal Well sampled
Just right, The Nyquist rate
Signal Over-sampled
Not gaining any more resolution, more bleaching waste of time and disk space
Sampling rate
Aliasing
Over-sampled
How do frequencies relate to resolution? But I don’t image wavy green lines or brick walls
Our object has a spatial frequency in the distribution of the contrast we are imaging
Nyquist calculator for microscopy http://support.svi.nl/wiki/NyquistCalculator
Q
Consider this principle and a few more subtle factors
A
Do I really need to listen to Nyquist? • You might not always be seeking the best resolution • You might need to under-sample for speed, phototoxicity . . . • Over-sampling is effectively averaging
But optimal sampling is important and beneficial in many cases
3D acquisition Confocals are good for acquiring stacks of images because of the optical sectioning ability
Z-stack acquisition control
The point-spread function Calibration of the microscope, the 3D image of a point source XY
XZ
Point spread function 3D
3D
2D
Widefield
Confocal
FWHM for 63x/1.4 NA is about 200 * 500 nm
Blur degradation in 3D wf images Too many PSFs to draw
Image: Sharp image
+dimmer blur
Overwhelmed by blur
Lateral and axial resolution XY
rxy ~ 0.61 l / NA
YZ
rz l / NA2
Resolution is always worse in Z than XY
Optical section thickness 10x/0.3
20x/0.45
Optical section (micron)
Smaller with higher NA
63x/1.2
40x/1.3 63x/1.4 100x/1.4
Larger with more open PH
Pinhole size (AU)
Sampling in the z-axis The same principle as in XY
Some regions not imaged
Covered
Covered and well sampled
Transmitted image For transmitted light
For fluorescent light
It is NOT confocal (why?), beware of the overlay
Setting up a transmitted image 1. Setup Kohler TL
2. Confocal forms a TL image
You can use whichever lasers you are using for fluorescence, but it tends to be best with far-red
(Gain and offset controls are physical dials on a Leica)
http://www.olympusfluoview.com/java/confocalsimulator/index.html
Other ways of doing something like this
Without these problems . . . Point based consequences: • Speed (eg 260,000 fold for 512 image) • SNR (if quicker than 7 hr per image) • Photodamage (bright spot to compensate) Low detector QE, few photons
Without these problems . . . Point based consequences: • Speed (eg 260,000 fold for 512 image) • SNR (if quicker than 7 hr per image) • Photodamage (bright spot to compensate) Low detector QE, few photons Fixed
Live
What can we do about this? Improve confocals GaAsP detectors improve QE
Drive confocals wisely
• • • •
Keep laser power to a minimum Open the pinhole a bit Under sample Get ok-ish images
• Use the best reagents • Keep live samples otherwise as happy as possible
Resonant scanners Smaller max amplitude so zoom is >1
8000 Hz resonant scanner vs say 400 Hz standard scanner Leica SP5 and SP8 have this feature
Everything else is the same except the scanner
Emission intensity
Consequences of scanning faster: reduced photobleaching
Excitation intensity
Faster scanning = lower fluorophore saturation Phototoxicity and photobleaching are reduced
Consequences of scanning faster: reduced photobleaching t ~3 ns
t ~30 ms
Example pixel dwell times: ~2 mSec standard, ~100 nSec resonant The longer you illuminate, the greater the % of GFP accumulates in T Triplet state is effectively permanent within the scale of pixel dwell time
But the time between scans of the same spot is >tT http://www.leica-microsystems.com/science-lab/brighter-fluorescence-by-resonant-scanning/
The spinning disk principle
1. Pinholes are in an image plane so project many spots onto the sample
Defocused laser
4. Image captured on a CCD
3. Disk descans the light – motion is trivial relative to speed of light
2. Spots scan the sample as the disk spins
The spinning disk principle
1. Pinholes are in an image plane so project many spots onto the sample
Defocused laser
4. Image captured on a CCD
3. Disk decans the light – motion is trivial relative to speed of light
2. Spots scan the sample as the disk spins
Detector comparison CCD
PMT
EMCCD
Photodetectors
hv
QE
Something Results dark noise
PMT
20%
CCD
70%
EMCCD 95%
An electron
Amplification (perhaps) multiplicative noise
Something bigger
A2D Read noise
Gain regulated cascade
More electrons
Charge, e in well
No
Same
Charge, e in well
em gain register
wf 1,000-20,000 PSC 20-100 Disk 10-3,000
Pixels
More electrons
S:B over SNR
Shot Noise Statistical noise in photon arrival, not from the detector How many photons emitted in 1 second?
kf
kf
100,000 photons/sec
~100,000 (normal distribution)
2 photons/sec
Probably not 2, Poisson distribution
SNR =
N N
=
N
NSR: 5% error 400 with 1% error with 10,000 photons
Photobleaching advantages t ~3 ns Emission intensity
t ~30 ms
Excitation intensity
Each spot is less intense than in a point scanner
The stroboscopic avoidance of triplet
How does a spinning disk work?
99% disk
~40% T
No zoom
Synch required above ~20 fps
Sectioning in spinning disks Ideal pinhole diameter = 0.5 l M/NA • 100x/1.4 = 20 mm • 20x/0.5 = 11 mm
Trade-off for excitation intensity and emission confocality: 50-70 microns and fixed The pinhole is > AU1 (often much more at low power lenses) so the sectioning and z-axis resolution will be less good than a point scanner
Sectioning in spinning disks
Pinhole cross talk
(Both these factors would be much worse without the microlens disk= increased transmission without having bigger pinholes)
What works well on a spinning disk Living things that need sectioning . . . • Things which match the high magnification, high NA optimizations (eg subcellular imaging) • Photosensitive samples • Fast imaging
Not the dimmest samples Not thick homogenous samples
Things quite like a Yokogawa spinning disk . . .
Slits not spots Changeable Swept field Visitech Infinity
Prairie/Nikon
Similar to the microlens spinning disk in terms of strengths and weaknesses
Sectioning by excitation Excite a defined region of the z-axis and image pretty much all the fluorescence available
• TIRF
• Multi-photon excitation • SPIM
Total internal reflection Some reflected, some refracted
Evanescent wave
Evanescent wave
Exponential decay of intensity Iz = I0e-bz
Evanescent wave
Evanescent wave
Plane of excitation ~100 nm thick
This is much thinner than a confocal slice
Widefield vs TIRF
Myosin Actin
Two ways of generating and imaging TIRF Prism-based
A bit more to align Better SNR, lower background Slight constraint on imaging objective Sample access is difficult in some setups Have to build your own
Objective-based
More convenient Needs >1.45 NA objective SNR still very good
Components of a TIRF system Inverted microscope with a special TIRF objective
Opaque incubator
TIRF angle autoalignment optics round the back
Fiber-coupled
lasers EMCCD
What is TIRF good for? Anything at the edge of the cell/tissue • Exocytosis/endocytosis • Vesicle dynamics • Cytoskeletal activity at the membrane – Focal adhesions
• Signalling in the membrane – translocation
Relatively distinct subset of samples gain from TIRF imaging
Single molecule imaging Single molecule sensitivity allows single molecule biochemistry
TIRF imaging is the highest SNR fluorescence imaging modality so good for single molecule studies, for which you need . . . • A sensitive way of imaging • A way of only having a few molecules in your imaging volume • (also the basis of some super-resolution techniques . . .)
Two photon excitation Single photon
Two-photon
Two photon excitation Single photon
Two-photon
Excitation is limited to a small focal volume where photons are most concentrated
Differences to a single photon confocal
Can use all the light, no pinhole needed
Non-descanned detector Since we don’t need to go through a pinhole, we don’t even need to descan and the PMT can be close to the objective and efficient
Is it a confocal?
(this arrangement makes it very sensitive to room light)
2-photon advantages Main advantage: Imaging thicker specimens The longer wavelength excitation penetrates further into the sample. The scattered excitation light doesn’t cause background fluorescence. The excitation is also not attenuated by fluorophore absorption above the plane of focus
Tissue absorption of light of l . . . Tissue
H20
700-1200 nm is a good window between absorption/scatter by tissue and absorption by water
Emission advantage Because we are able to image all the light (no pinhole) this is less affected by scatter With pinhole
No pinhole
The NDD is closer and more efficient for a scattered beam path (which is hard to move efficiently through several lenses)
1P
2P
Pretty pictures from the Olympus FV1000 MPE brochure
A Pulsed laser is required for MPE For efficient MPE we need photon concentration in space and time . . .
http://www.olympusmicro.com/primer/techniques/fluorescence/multiphoton/multiphotonintro.html
A pulsed laser Chameleon Ultra II Femtosecond pulsed laser
3-4 W of power at peak
Tunable 680-1080 nm @40 nm/sec
This gives flexibility, since these laser cost about $200,000 each its not normally practical to have several per machine like with 1 photon
Excitation spectra Most good 1P fluor’s are ok for 2P (weak ones are normally even worse in 2P) Not exactly double 1photon Tends to be broader and blue shifted This makes multichannel imaging easier in some ways and more difficult in others
Photobleaching and toxicity Good luck!
Extremely toxic l Won’t go through glass
Much less phototoxicity associated with these l
Multi-photon excitation provides a means of exciting UV and blue fluorophores with less phototoxity
Photobleaching and toxicity Photodamage in 2P is confined to the thin layer being imaged Power (photons/mm2
Wide-field
X
1 Photon
105X
2 Photon (average)
106X
2 Photon (peak)
1011X
But the damage in that area can be worse
In general, for thick samples 2P has an advantage over 1P For thin samples, 2P is often worse than 1P There may be strange forms of damage due to very high field strength – ROS, DNA breaks, tweezing
2-photon vs. 1-photon • Improved SNR with thick samples • The IR 2PE is less phototoxic in many cases, especially for UV dyes.
• Photobleaching/damage restricted to plane being imaged. Photobleaching or uncaging is possible with fine z-axis resolution
• Resolution is slightly less good
• Multi-channel acquisition is harder (excitation cross-sections are normally much broader) and limited in excitation l • No control of optical section thickness • Lasers are expensive and require exact alignment and can produce heating and other damage
Don’t use a 2-photon system unless you need the advantages
Final sectioning by excitation technique for today . . . A brief mention
Selective Plane Illumination Microscopy (SPIM) Light sheet fluorescence microscopy Widefield CCD detection
Sheet illumination from the side
Components of an example SPIM system Often a low-ish mag low NA dipping objective
For TL
Laser(s) for excitation shaped into a sheet
Plane illumination photodamage advantage
Reduced region of illumination and photodamage
n times better for a z-stack of n slices The excitation intensities are overall much lower than confocal Selective plane illumination microscopy techniques in developmental biology Jan Huisken and Didier Y. R. Stainier Development 2009
Samples good for SPIM etc The sectioning, speed, efficiency and low photodamage make it ideal for study of living samples up to a few mm across
Thin-sheet laser imaging microscopy for optical sectioning of thick tissues, Santi et al 2009 Reconstruction of zebrafish early embryonic development by Scanned Light Sheet Microscopy Keller et al 2009
SPIM systems are good for a range of samples imaged poorly by other techniques
Optical clearing: SCALE reagent
• • • •
4 M urea 10% (wt/vol) Glycerol 0.1% (wt/vol) Triton X-100 pH of 7.7
Refractive index of 1.387 and 486 nm
Soak your sample in it for a couple of weeks . . . http://www.nature.com/neuro/journal/v14/n11/full/nn.2928.html
It clears fixed samples by removing refractive index changes
Doesn’t destroy fluorescence
Preserves tissue structure
And allows . . .
SCALE reagent
24 tiled stacks =9000 images
Movies of data
Tools available for you Light Microscopy Core Facility DUKE UNIVERSITY AND DUKE UNIVERSITY MEDICAL CENTER
Choose wisely
Modality comparison Modality
Sensitivity
Speed
Photodamage
Sectioning
Samples
WF Fluorescence
Good
Camera fast
Not bad
No
Cells, thin sections
Not bad
“100 nm”
Things close to the membrane, single molecules
A concern
Adjustable >500 nm
Nearly anything, only helps with thick samples/sectioning
Better than normal confocal
Adjustable >500 nm
Living samples that need high-speed/ low photodamage
Good
Fixed by disk, depends on objective
Live samples, cells to embryos
It depends
By excitation
Things too thick for 1P
TIRF
Good
Camera fast
Confocal point scanning
Poor
Slow as point based
Resonant PS confocal
Poor
Point-based still
Spinning Disk
Good (but goes through a disk)
2 Photon
Poor
Camera fast
Slow as point based
But testing a few is often a good idea
Commercial choices: the big four
The best microscopes are definitely from (Redacted by the LMCF legal department)
Super resolution How to improve the resolution of fluorescence microscopy Resolution
= 0.61 l / NA
It should be said that it’s really quite good already ~200 nm
1. STED: STimulated Emission Depletion More strange things about the quantum mechanics of fluorophores
A long wavelength photon can deplete the fluorescence
How STED can get us better resolution Normal 1P excitation spot
Doughnut of depletion
Bigger doughnut, smaller spot
STED system
Strengths and weaknesses of STED Point scanning confocal with improved XY resolution ~3X Fast – pretty much as a standard confocal
• Fluorophore limitations, multiple fluors difficult • Requires precise alignment • The power required for depletion is not ideal for living cells • Corruption with depth • Z-resolution not improved
2. Structured illumination extends the passband
We can image the moire fringes and with knowledge of the illumination structure we can capture the object in finer detail than ordinarily possible
3D Structured illumination
Subdiffraction Multicolor Imaging of the Nuclear Periphery with 3D Structured Illumination Microscopy
SIM in practice • • • •
100 nm XY by 200-300 nm Z resolution Essentially a widefield technique Normal dyes Two fold resolution improvement still pretty useful
3. Core principle of the next approach We can’t separate two objects beyond our diffraction limited resolution but . . .
. . . If you only have one object, you can position the centroid with very high accuracy, say 1 to a few nm
Dark sample with PAfluorophore
Weak photoactivation to produce a few isolated fluorphores
Image and bleach individual fluorophores and map their position
Many iterations to produce super-resolution map/image
TIRF image
PALM Super-resolution
Resolution can be improved more than 10 fold Imaging Intracellular Fluorescent Proteins at Nanometer Resolution
Processing the images
• 10,000 images, ~10 min, few GB • Fiducials help correct drift • Individual molecules must be separated by >resolution • But sequential so >105 molecules per mm2 possible • Accuracy of centroid fitting is l/2NA √#photons about 50 photons per point (STORM more than FPs)
Astigmatism: 3D STORM
iPALM
Which super resolution approach is best? Well they do pretty different things . . . STED– Confocal like Limited fluorophores, damage, no z-axis improvement Structured illumination Smallest improvement in resolution, but 3D improvement Most versatile and like “normal” microscopy Probably live cell compatible
PALM/STORM family Best resolution Temporal constraint How easy are the 3D versions? Unlike “normal” imaging http://www.nature.com/nmeth/collections/superresmicroscopy/index.html http://www.annualreviews.org/doi/full/10.1146/annurev-cellbio-100109-104048
What should I understand? • How fluorescence works • How all those spectra, filters, lamps, objectives add up to a photon efficient imaging combination. How to choose a filter for a particular fluorophore. • Resolution (in fluorescence terms) – what it means, what it doesn’t • What factors into a wise choice of an objective • NA and its consequences • How CCD cameras work and are used in microscopy. When to use an EMCCD. • What all the components in the TL path do and HOW TO KOHLER A SCOPE (maybe the conjugate planes explanation) • Contrast – principles of brightfield, Phase contrast, DIC • The confocal principle, advantages and disadvantages • How a confocal works, what the components do and how to adjust them • Resolution and sampling in 2D and 3D • Advantages and disadvantages of Spinning disk • TIRF and the type of samples it works for • Advantages and disadvantages of multiphoton
More information This is a fairly comprehensive collection of review articles and interactive tutorials about the optics involved in microscopes http://www.olympusmicro.com/primer/anatomy/anatomy.html This book has a very good for transmitted light and optical basics: Fundamentals of Light Microscopy and Electronic Imaging – Douglas B Murphy (Duke has an eBook) Optical Microscopy by Davidson and Abramowitz is a 40 page review article you can download here http://www.olympusmicro.com/primer/opticalmicroscopy.html Review articles about fluorescence microscopy http://www.olympusmicro.com/primer/techniques/fluorescence/fluorhome.html
Spinning disks TIRF
http://zeiss-campus.magnet.fsu.edu/articles/spinningdisk/introduction.html
http://www.microscopyu.com/articles/fluorescence/tirf/tirfintro.html
SPIM http://dev.biologists.org/content/136/12/1963 Multiphoton articles/tutorials http://micro.magnet.fsu.edu/primer/techniques/fluorescence/multiphoton/multiphotonhome.html Reviews on multiphoton imaging http://zeiss-campus.magnet.fsu.edu/referencelibrary/multiphoton.html