Transcript
Fisher Scientific FS2 Series Biological Safety Cabinets CLASS II Type A/B3 • NSF49 approved, Type A2 airflow design for excellent user, product and environmental protection • Suitable for use in Biosafety Levels 1, 2, and 3 • Large work volume with stainless steel interior liner • Digital touch tone control panel with on-board monitoring and diagnostics • 3 year Manufacturer’s warranty against defects in material or workmanship
One place One standard One call
Front Elevation
Side View
Class II, Type A2 (Recirculating) models come complete with fluorescent lights, UV light, one duplex electrical outlet, digital motor controller, static pressure monitor and one gas turret located on the left hand side wall. Console models include base support stand. Model
Cat. No.
Type
Hand Opening Dim. A
Work Volume WxDxH
Work Tray
4 ft Console
36-100-4616
A/B3
10” (254mm)
52.5” (1334mm)
46.5” x 27.5” x 28” (1181mm) x (699mm) x (710mm)
18” x 46.5” (457mm) x (1181mm)
Weights Actual Shipping 420 Lbs. 600 Lbs. (191Kg.) (272Kg.)
4 ft Bench Top 36-100-4618
A/B3
10” (254mm)
52.5” (1334mm)
46.5” x 27.5” x 28” (1181mm) x (699mm) x (710mm)
18” x 46.5” (457mm) x (1181mm)
420 Lbs. 600 Lbs. (191Kg.) (272Kg.)
6 ft Console
36-100-4617
A/B3
10” (254mm)
76.5” (1943mm)
70.5” x 27.5” x 28” (1791mm) x (699mm) x (710mm)
18” x 70.5” (457mm) x (1791mm)
540 Lbs. 725 Lbs. (245Kg.) (329Kg.)
6 ft Bench Top 36-100-4619
A/B3
10” (254mm)
76.5” (1943mm)
70.5” x 27.5” x 28” (1791mm) x (699mm) x (710mm)
18” x 70.5” (457mm) x (1791mm)
540 Lbs. 725 Lbs. (245Kg.) (329Kg.)
Class II, Type A2 (Exhausted) models are also available
Technical Specifications Air System Classification
CLASS II Type A/B3 as per NSF 49 Standard
Motor
1/2 hp direct drive capacitor start type
Design
Negative pressure plenum
Standard Filters
Exhaust and Supply HEPA filters are rated 99.99% efficient at 0.3 microns and larger
Electrical Line Feed
115 V, 12 A, 60 Hz, also available in 220V/50Hz models
Control Panel
Fully electronic type featuring soft touch keypad with on/off indicators on all switches, dual processor controls, LVS (Low Voltage Stabilization) circuitry and calibrated linear static pressure display (LED’s) complete with high/low audio and visual alarm feature and mute function. In addition, the B3 unit has an additional electronic system c/w audio/visual and mute functions that monitors the static pressure in the exhaust ductwork.
General Noise Level
Less than 62 dBA
Base Frame
The knock down base frame is 30” (standard) or 36” (optional)
Certification
NSF 49 certified, CSA/NRTL
Optional Equipment Cat. No.
Description
36-100-4729
Additional duplex outlet
36-100-4730
Additional gas turret
36-100-4731
Additional air turret*
36-100-4732
Additional vacuum turret**
36-100-4733
I.V. bar for 4’ models
36-100-4734
I.V. bar for 6’ models
36-100-4735
4’ knock down base frame, 30” high
36-100-4736
6’ knock down base frame, 30” high
36-100-4737
4’ knock down base frame, 36” high
36-100-4738
6’ knock down base frame, 36” high
36-100-4741
4’ hard connection transition canopy
36-100-4742
6’ hard connection transition canopy
36-100-4743
4’ thimble connection transition canopy
36-100-4744
6’ thimble connection transition canopy
* Unless otherwise stated additional air turrets will be located on the left hand side of hood ** Unless otherwise stated additional vacuum turrets will be located on the right hand side of hood.