8'
CONTRACTOR SHOULD CALL HCMI BEFORE CONSTRUCTION OR ASAP! PHONE: 800-641-4107 FAX: 417-864-7394
10"
8'
12'
1" x 6" x 3ft.mounted to wall 92" AFF to bottom of board. For new construction,use 2 x 6 wall.
7' 8"
HF Power Cabinet circuit breaker per HCMI electrical requirements spec sheets. Interconnect between circuit breaker and power cabinet must be made by electrical contractor at time of equipment installation via flexible conduit. Circuit breaker should be mounted flush to the wall. Location of the circuit breaker must meet local, state and federal codes and regulations. Refer to electrical spec sheet for wire size and circuit breaker size. Refer to State approved radiation protection requirements for lead and drywall specification. Lighting on dimmer switch is recommended in room and control booth area. All construction (mechanical, electrical and plumbing) must meet local, state and federal codes and regualtions.
HF-300 ELECTRICAL SPECIFICATIONS WIRE SIZE LENGTH TO ROOM DISCONNECT (AWG)
HF 300 Power Cabinet
LINE VOLTAGE
110 volt outlet
240 VAC, 100 amp circuit breaker Dedicated line. See Barrier Wall Elevation drawing for more detail
50FT
100FT
150FT
200FT
208
1
00
000
220
1
0
00
0000
240
2
1
00
000
250
2
1
0
00
DISCONNECT TO GENERATOR 15FT MAXIMUM
1" x 6" mounted to wall 82" AFF to bottom of board. Board should run from Barrier wall to Buckystand wall. Board must be flat and level. Horizontal pull is 150 lbs.
BUCKYSTAND
250MCM 4AWG
CONTRACTOR SHOULD CALL HCMI BEFORE CONSTRUCTION OR ASAP! PHONE: 800-641-4107 FAX: 417-864-7394 1" x 6" x 3ft.mounted to wall 92" AFF to bottom of board. For new construction,use 2 x 6 wall.
1" x 6" mounted to wall 82" AFF to bottom of board. Board should run from Barrier wall to Buckystand wall. Board must be flat and level. Horizontal pull is 150 lbs. 10'
HF Power Cabinet circuit breaker per HCMI electrical requirements spec sheets. Interconnect between circuit breaker and power cabinet must be made by electrical contractor at time of equipment installation via flexible conduit. Circuit breaker should be mounted flush to the wall. Location of the circuit breaker must meet local, state and federal codes and regulations. Refer to electrical spec sheet for wire size and circuit breaker size. Refer to State approved radiation protection requirements for lead and drywall specification. Lighting on dimmer switch is recommended in room and control booth area. All construction (mechanical, electrical and plumbing) must meet local, state and federal codes and regualtions. ADDITIONAL NOTES: 110 VAC outlet needs to be located near digital detector for power supply. 110 VAC outlet and network jack should be located in control booth for coputer station. 110 VAC outlet and network jack should also be supplied in the location of additional workstations. Consult with doctor to determine these locations.
HF-300 ELECTRICAL SPECIFICATIONS
12" x 16" Lead Glass
LINE VOLTAGE
50FT
100FT
150FT
208
1
00
000
220
1
0
00
0000
240
2
1
00
000
250
2
1
0
00
HF 300 Power Cabinet
4'
240 VAC, 100 amp circuit breaker Dedicated line. See Barrier Wall Elevation drawing for more detail
110 V outlet Network Jack
WIRE SIZE LENGTH TO ROOM DISCONNECT (AWG)
DO NOT SCALE 8'
200FT
DISCONNECT TO GENERATOR 15FT MAXIMUM
14'
250MCM 4AWG