INFRARED RADIATION THERMOMETER KT15 II
Operating Instructions
95582944 36/10/16e
(valid from Serial Number 10301)
All rights reserved! The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Heitronics GmbH reserves the right to make modifications and improvements in their design without prior notice. (c) HEITRONICS Infrarot Messtechnik GmbH
HEITRONICS Infrarot Messtechnik GmbH Kreuzberger Ring 40 D-65205 Wiesbaden Tel.: +49 (0)611 97393-0 Fax: +49 (0)611 97393-26 E-Mail:
[email protected] Internet: www.HEITRONICS.com
SAFETY INSTRUCTIONS Please read carefully the instructions given in TECHNICAL SPECIFICATIONS, especially the connection and operating instructions as well as the connection instructions and configurations dealt with in SYSTEM START-UP.
ATTENTION In case of wrong connection, the device might be destroyed.
The Infrared Radiation Thermometer is an optical measuring instrument. Dirty lenses will lead to measuring errors. Please pay attention to the instruction given in MAINTENANCE AND CALIBRATION. HEITRONICS production meets the latest technical development thereby applying sophisticated components. Nevertheless, functional errors may occur in exceptional cases. The failure of an instrument may cause a measuring value being issued which seems to be correct, but is wrong. Please pay attention to the instructions given in MAINTENANCE AND CALIBRATION. Only when using a female connector with cable or a dummy plug attached to the plug of the device - in addition to the equipment sealing - the specified system of protection can be achieved. Operation with Laser:
Devices completely mounted and equipped with integrated laser meet the safety requirements of Laser class 2.
Before handling the lens in any way, the voltage supply is to be switched off or the connecting plug is to be pulled out to ensure that the laser will not switch on automatically.
ATTENTION The Laser must not be switched on, if the lens of the Infrared Radiation Thermometer is removed.
The integrated laser can only operate up to the upper limit of the ambient temperature of the Infrared Radiation Thermometer (60 °C).
ATTENTION The Laser will be switched off at an ambient temperature ≥ 60 °C to defend it from destruction.
95582944 03/03/14e
SAFETY INSTRUCTIONS
KT 15 II HEITRONICS Infrarot Messtechnik GmbH
Erklärung über die Konformität DECLARATION OF CONFORMITY
Diese Erklärung gilt für folgende Erzeugnisse: This declaration is valid for the following products:
Geräteart:
Infrarot Strahlungspyrometer
Type of instrument:
Infrared Radiation Pyrometer
Typenbezeichnung:
KT15 II Serie KT15 II Series
Designation of model:
Diese Erklärung wird abgegeben durch This declaration is issued by
HEITRONICS Infrarot Messtechnik GmbH Kreuzberger Ring 40 65205 Wiesbaden, Germany
Hiermit wird bestätigt, dass die Produkte gemäß den Richtlinien des Rates zur Angleichung der Rechtsvorschriften der Mitgliedsstaaten über die elektromagnetische Verträglichkeit (89/336/EWG) mit den unten genannten Normen übereinstimmen: In accordance with the EU-Directive of Electro-magnetic-compatibility (89/336/EWG) the manufacturer declare, that the device described above is conform to the essential requirements of the EU-Directives:
EN 55011 Class B EN 61326 Wiesbaden, 15. Juli 2003 Wiesbaden, July 15, 2003
PREFACE The remarkable features of the HEITRONICS devices are a design suitable for specific application as well as an uncomplicated operation. Nevertheless, it is recommended to read the entire instruction manual before putting the instruments into operation. Please observe the safety instructions as well as the particularly emphasized annotations when reading the pertinent instruction to the operation. The operating instructions are primarily directed to the user. They contain the information and instructions necessary to successfully use the instruments including all options. If you still have any questions after having read this instruction manual, please contact us. Our staff will be at your disposal.
RoHS-CONFORMITY This product is in conformity with Directive 2011/65/EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment.
95582944 02/01/15e
PREFACE
KT15 II HEITRONICS Infrarot Messtechnik GmbH
CONTENTS
SAFETY INSTRUCTIONS DECLARATION OF CONFORMITY PREFACE RoHS-CONFORMITY
CONTENTS ...................................................................................................... 0-1 DATA SHEET ................................................................................................... 1-1 GENERAL ........................................................................................................ 2-1 Measuring temperatures using Infrared Radiation Thermometers .............. 2-1 Infrared Radiation Thermometers for GENERAL APPLICATION ....................... 2-2 Infrared Radiation Thermometers for measuring METALS ............................ 2-3 Infrared Radiation Thermometers for measuring PLASTICS.......................... 2-3 Infrared Radiation Thermometers for measuring GLASS.............................. 2-4 Infrared Radiation Thermometers for measuring GASES ............................. 2-4 TECHNICAL SPECIFICATIONS ...................................................................... 3-1 Basics ......................................................................................................... 3-1 Technical Information 'Protective Cooling Jacket' ....................................... 3-3 Assignment for 12-pin connector for RS232-Interface ................................ 3-4 Assignment for 12-pin connector for RS485-Interface ................................ 3-6 Assignment for 7-pin connector .................................................................. 3-7 Legend table "Temperature resolution" ....................................................... 3-8 Temperature resolution KT15.01 II / KT15.02 II .............................................................................. 3-8 KT15.21 II ................................................................................................... 3-9 KT15.23 II / KT15.24 II ............................................................................ 3-10 KT15.25 II ................................................................................................. 3-11 KT15.41 II / KT15.42 II ............................................................................ 3-12 KT15.43 II ................................................................................................. 3-13 KT15.45 II ................................................................................................. 3-14 KT15.62 II ................................................................................................. 3-15 KT15.63 II ................................................................................................. 3-16 KT15.69 II ................................................................................................. 3-17 KT15.81 II / KT15.82 II ............................................................................ 3-18 KT15.83 II / KT15.85 II ............................................................................ 3-19 START-UP ........................................................................................................ 4-1 Mounting ..................................................................................................... 4-1 Fig. Mounting KT15 II Protective Cooling Jacket ........................................ 4-1 Electrical Connection .................................................................................. 4-2 Operation without serial interface................................................................ 4-2 Operation with serial interface..................................................................... 4-2 Optical Adjustment ...................................................................................... 4-3 Aiming via focus laser ................................................................................. 4-3 Aiming via pilot laser ................................................................................... 4-3 Aiming via Laserpointer LP15 ..................................................................... 4-3 Installation Protective Cooling Jacket WK15 ............................................... 4-4
0-1 95582944 17/01/15e
CONTENTS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
OPERATION AND APPLICATION ................................................................... 5-1 Application ................................................................................................ 5-1 Check routines ..................................................................................... 5-1 Continuous measurement of ambient temperature .............................. 5-2 Use of digital input ............................................................................... 5-3 Use of digital outputs ........................................................................... 5-4 Monitoring the device function during operation ................................... 5-4 Use of maximum and minimum value memories ................................. 5-5 Operation via keyboard ............................................................................ 5-8 Operation of the Menu ......................................................................... 5-8 Main Menu ........................................................................................... 5-9 Submenu: Temp / Info / Emi ............................................................ 5-10 Submenu: Tamb ............................................................................... 5-11 Submenu: Resp / Memo ................................................................... 5-12 Submenu: HiAI / LoAI ....................................................................... 5-13 Submenu: Laser / Unit ...................................................................... 5-14 Submenu: Aout ................................................................................. 5-15 Submenu: Uart.................................................................................. 5-16 Submenu: Cal ................................................................................... 5-17 Submenu: Conf ................................................................................. 5-18 Submenu: Test ................................................................................. 5-19 Communication via interface ................................................................. 5-20 Preparing the device for operation with interface RS232C ................ 5-20 Using the interface ........................................................................... 5-20 Control of communication ................................................................. 5-20 Data transfer from KT15 II for evaluation purposes........................... 5-21 Commands ....................................................................................... 5-21 Delimiter............................................................................................ 5-21 Input buffer........................................................................................ 5-21 Description of commands ......................................... separate instructions MAINTENANCE AND CALIBRATION ............................................................. 6-1 General information .................................................................................... 6-1 Cleaning the lens ........................................................................................ 6-1 Checking the display accuracy.................................................................... 6-1 FIGURES AND TABLES .................................................................................. 8-1 Spectral emissivity of various materials ...................................................... 8-2 Total emissivity of some materials at 20 °C ................................................ 8-2 Emissivity of foils ........................................................................................ 8-3 Spectral emissivity of Infrared Radiation Thermometers and transmission curves of various synthetic materials ..................................... 8-4 Spectral emissivity ε, transmissivity τ and reflectivity ρ of glass ................................................................................... 8-5 Housing dimensions KT15II ........................................................................ 8-6 Field of view ................................................................................................ 8-8 Transmission curves (illustrated in a diagram) .......................................... 8-9ff WARRANTY CONDITIONS SERVICE ADDRESSES
0-2 95582944 17/01/15e
CONTENTS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
1
DATA SHEET
Type .......................................................................................................... Serial No.
Calibration factor ........................................................
....................................................... Lens ................................................ Spectral response Spacer ............................................ ....................................................... Temperature range
Detector ..........................................
...............................................
Digital Interface ...............................
Options .................................................................................................................................... .................................................................................................................................... Accessories .................................................................................................................................... ....................................................................................................................................
Code ...........................................................................................................................
Further data ................................................................................................................. ....................................................................................................................................
Hereby it is confirmed that the Infrared Radiation Thermometer specified above conforms to the data indicated in the specifications.
Tester:
Wiesbaden,
1-1 95582944 03/01/15e
DATA SHEET
KT15 II HEITRONICS Infrarot Messtechnik GmbH
2 2.1
GENERAL Measuring temperatures using Infrared Radiation Thermometers
Above a temperature of absolute zero, which is approx. - 273 °C or 0 K, all bodies emit electromagnetic radiation, the wavelength and density of which depend on the temperature. Up to a temperature of approx. 600 °C, the wavelength of the radiation can be found exclusively within the infrared range (thermal radiation). It is only at higher temperatures where part of this radiation is emitted in the form of visible light. The radiation emitted by the body (radiation density) also depends on the surface of the individual body. At a fixed temperature, the maximum radiation density is emitted by a "blackbody" source of radiation. All real bodies only emit a fraction of this radiation density at the same temperature. The ratio of this fractional part to the maximum radiation density is called the emissivity ε. Naturally, the emissivity is always < 1. It depends on the nature of the surface of the material, on the material itself and on the wavelength. If the emissivity is known, the temperature of the object can be determined by measuring the infrared radiation emitted by the very object. Devices used to measure this kind of radiation are called Infrared Radiation Thermometers. As the measurement is effected without the Infrared Radiation Thermometer being in contact with the object, no distortion of the temperature field due to heat dissipation need to be feared, as is the case, e.g., with probe-type thermometers. The Infrared Radiation Thermometer is a measuring transducer, which receives the infrared radiation emitted by the measuring object itself and transforms it into a standardized output signal. All optic and electronic components are located in a small housing of a solid construction, which is made from die-cast metal so that the installation of a Infrared Radiation Thermometer is possible even in narrow spaces. By selecting a number of various lenses and detectors, it is possible to generously modify the measuring field, if the measuring distance is given. For the application of the Infrared Radiation Thermometer under high ambient temperature, water-cooling, air purge fittings and vacuum tight lenses can be delivered as accessories.
2-1 95582944 02/03/14e
GENERAL
KT15 II HEITRONICS Infrarot Messtechnik GmbH
2.2 Infrared Radiation Thermometers for GENERAL APPLICATION The Infrared Radiation Thermometers series KT15.81 II / KT15.82 II /KT15.83 II are used to measure surface temperatures within a temperature range of - 50 °C to 1000 °C. The spectral sensitivity can be found within the range of the atmospheric window, which is between 8 and 14 µm. As the atmospheric transmissivity is very high in this spectral range, no weakening of the infrared radiation due to CO2 or to water vapor contained in the air is to be expected. To meet especially high requirements as to the transmission of the atmosphere as it is a case e.g. when carrying out meteorological measurements Infrared Radiation Thermometer KT15.85 II is suitable. The KT15.81 II / KT15.82 II / KT15.83 II can be employed in the measuring of temperatures during the production and processing of plastics, rubber, paper, textiles, paint, ceramics etc. The emissivity of various metals is shown in figure 10 (chapter 8: Illustrations). The devices can also be used to measure the temperature of objects exposed to infrared radiation, since the radiation maximum of infrared radiators is within the range of shorter wavelengths, which do not cause the Infrared Radiation Thermometers KT15.81 II / KT15.82 II / KT15.83 II to react. The Infrared Radiation Thermometer KT15.81 II is used in the spectral range of 8 to 10 μm. Besides being used for large-surface measurement, this Infrared Radiation Thermometer type is also suitable for measuring thick foils. The Infrared Radiation Thermometer KT15.82 II is applied for the standard measuring range of 8 to 14 µm thereby guaranteeing a good signal resolution. The Infrared Radiation Thermometer KT15.83 II provides a particularly high signal resolution owing to its large spectral range. Therefore, it is very well suited for measuring low temperatures. For high requirements on the transmission of the atmosphere, such as found in meteorological measurements, the KT15.85 II type is available. This device type operates within the spectral ranges of 9.6 to 11.5 µm, where the transmission of the atmosphere is extremely high.
2-2 95582944 02/03/14e
GENERAL
KT15 II HEITRONICS Infrarot Messtechnik GmbH
2.3.
Infrared Radiation Thermometers for measuring METALS
The spectral emissivity of the Infrared Radiation Thermometers KT15.01 II / KT15.02 II is 2 to 2.7 µm and 2 to 4.5 µm. Within this spectral range, metals and metal-oxides show a relatively high spectral emissivity. This is why the device is particularly well suited to measure the temperatures of these materials. Since the atmospheric transmissivity is high within the spectral range of 2 to 2.7 µm, distortions of the measured value owing to absorption of water vapor and CO2 in the air can be ignored (Fig. 11, ILLUSTRATIONS). The Infrared Radiation Thermometer type KT15.01 II is used for measuring temperatures from 300 °C onwards; the Infrared Radiation Thermometer type KT15.02 II is suitable for measuring temperatures from 200 °C onwards. Temperatures from 500 °C respectively 700 °C onwards are preferably measured by means of the spectral pyrometers of the series KT18S or the quotient pyrometers of the series KT18R. 2.4
Infrared Radiation Thermometers for measuring PLASTICS
These types of Infrared Radiation Thermometers are employed for the measuring of surface temperatures of plastics within a temperature range of 0 to 600 °C. These Infrared Radiation Thermometers are equipped with narrow-band infrared filters. KT15.21 II measures radiation at 3.43 µm, KT15.23 II measures radiation at 6.8 µm, KT15.24 II measures radiation at 7.93 µm, KT15.25 II measures radiation at 8.05 µm. Within these spectral ranges, virtually all synthetic materials have strong absorption bands so that even thin foils show high emissivity. The Infrared Radiation Thermometer KT15.21 II is suitable for most foils from 100 °C onwards. The Infrared Radiation Thermometer KT15.23 II measures thin foils made from polyethylene, polypropylene, polyisobutane, polynitrile, polystyrene and similar materials from 0 °C onwards. The Infrared Radiation Thermometer KT15.24 II and KT 19.25 II measure thin foils polyester as well as foils made from - acrylic, - cellulose, - fluorine compounds, - polycarbonate, - polyamide, - polyurethane, - polyvinylchloride and similar materials after analysis from 0 °C onwards.
2-3 95582944 02/03/14e
GENERAL
KT15 II HEITRONICS Infrarot Messtechnik GmbH
2.5 Infrared Radiation Thermometers for measuring GLASS KT15.42 II is suitable for measuring glass and quartz. The spectral sensitivity is 4.9 to 5.5 µm. The emissivity of glass is very similar to the emissivity of a blackbody radiator within this spectral range. Furthermore, this spectral range eliminates any disturbing influences resulting from the strong absorption bands of water vapor at 6.2 µm. The Infrared Radiation Thermometers KT15.01 II and KT15.41 II are not measuring the surface but the average value of temperatures up to a certain penetration depth (glass volume). The lower limits are 300 °C for KT15.01 II and 400 °C for KT15.41 II. The types KT15.42 II and KT15.43 II measure the surface temperature from 100 °C respectively from 0 °C onwards.
2.6 Infrared Radiation Thermometers for measuring GASES By means of narrow-band filters, the selective measurement of temperature of certain gases is possible. A rather long response time is to be chosen since the radiation energy is very low. KT15.21 II measures hydrocarbon from 100 °C onwards. KT15.62 II measures materials within the absorption range of carbon dioxide and carbon monoxide, KT15.63 II measures carbon monoxide selectively, KT15.64 II is provided for nitrogen oxide and KT15.69 II has been especially developed for determination of combustion temperatures. Using KT15.41 II, objects can be measured through gases and flames from 400 °C onwards.
2-4 95582944 02/03/14e
GENERAL
KT15 II HEITRONICS Infrarot Messtechnik GmbH
3
TECHNICAL SPECIFICATIONS
Spectral sensitivity:
DATA SHEET
Temperature measuring range:
DATA SHEET
Temperature resolution:
Tables (Page 3-7ff)
Accuracy (provided that the emissivity has been correctly set, after a warm-up period of 15 min.): ± 0.5 °C plus 0.7 % of the temperature difference between the housing containing the measuring instruments and the object to be measured or: value of temperature resolution. The higher value shall prevail. Long-time stability:
better than 0.1 ‰ of the absolute measuring temperature in Kelvin/month
Lens used:
DATA SHEET
Possible lenses:
OPTIONS AND ACCESSORIES
Target diameter (95 %):
The target diameter depends on the lens and the detector used in each individual case. When short-focus lenses are used, the distance within which the minimum measuring field can be found varies by ± 4%.
Field of view:
Marking the field of view is carried out by means of different accessories. OPTICAL ADJUSTMENT (chapter 4.5)
Radiation detector:
HEITRONICS pyroelectric detector
Permissible ambient temperature:
- 20 ... + 60 °C For higher ambient temperatures, cooling and protecting jackets are available. OPTIONS AND ACCESSORIES and TECHNICAL INFORMATION (page 3-3)
Storage temperature:
- 20 … + 70 °C
Weight:
approx. 0.55 kg
Dimensions:
FIGURES: Fig. 15
Operating voltages:
AC: 12 (-10 %) ... 24 V (+ 10 %) 48 ... 400 Hz DC: 10.5 ... 30 V power consumption: ≤ 3.5 W
Type of protection:
IP65
3-1 95582944 21/03/12e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Analog output:
possible signal outputs (variable by programming) 0 ... 20 mA, 4 ... 20 mA 0 ... 1 V, 0 ... 10 V
Lower temperature value (T.low): Upper temperature value (T.end):
minimum temp. value for analog signal maximum temp. value for analog output
Minimum temperature difference:
depends on final temperature (T.end) The following values will result: Minimum temperature difference 50 °C 100 °C 200 °C 400 °C
Final temperature ≤ 150 °C ≤ 200 °C ≤ 1000 °C > 1000 °C
Load of analog output:
current output: voltage output:
Resolution of analog output:
16 bit
load ≤ 550 Ohm load ≥ 10 kOhm
Digital interface V24 (RS232C):
9.6 ... 115.2 kBaud
Response time (90 %):
variable by programming KT15 IIP: 0,005; 0,01; 0,03; 0,1; 0,3; 1; 3; 10; 30; 60; 120; 240; 360, 480 und 600 s KT15 IID: 0,03; 0,1; 0,3; 1; 3; 10 s
Protection against oscillation:
DIN 40046 Bl. 8; test: Fc vibration strength: A B1 E frequency range: 10 to 55 Hz amplitude: +/- 0.2 mm duration of test/position: 30 min
Connector type:
miniature circular connector, Franz Binder GmbH, type 99-5630-15-12, series 423, 12 pin and type 99-5626-15-07, series 423, 7 pin
Options: Thermal switch:
switching temperature: > 70 °C load: voltage: ≤ 48 V, current: ≤ 0.5 A
alternative: Analog input:
0 … 10 V
alternative: Digital input:
floating contact or
Laser:
SAFETY INSTRUCTIONS
0 … 30 V (low 1 V, high 4 V)
Laser type
Protection class
Output
Wavelength
Form
Diameter
Max. ambient temperature
Pilot laser
2
< 1 mW
650 nm
round beam
Ø 3 mm @ 1 m distance
60.0 °C
Focus laser
2
< 1 mW
650 nm
cross lines with centering circle
adapted to field of view in the focus distance
60.0 °C
3-2 95582944 21/03/12e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Technical Information Cooling Jacket WK15
The HEITRONICS Infrared Radiation Thermometers can be operated at an environmental temperature of up to 60°C without any additional coolants. If the environmental temperatures are higher than that mentioned above, cooling fittings can be used. Basically, the cooling jacket WK15 can be cooled with air or water. The possible maximum environmental temperatures are higher for water cooling than for air cooling ( individual Data Sheet). If the cooling process is too strong, i.e. if the cooling air / cooling water is too cold, this may result in condensate as soon as cooling medium falls below the dew point. To prevent this from occurring, the cooling air / cooling water must have a minimum temperature as a function of the relative humidity. The following table specifies the minimum temperature of the cooling air / cooling water as a function of the environmental temperature and the relative humidity of the environmental air.
Minimum temperature of the cooling air / cooling water Relative air humidity (of the environmental air) Environmental air temperature /°C
2%
4%
10%
20%
30%
50%
70%
30,0
5,0
5,0
5,0
6,0
11,0
19,0
25,0
°C
40,0
5,0
5,0
5,0
13,0
20,0
28,0
34,0
°C
50,0
5,0
5,0
10,0
21,0
28,0
38,0
45,0
°C
60,0
5,0
5,0
18,0
28,0
38,0
47,0
54,0
°C
70,0
5,0
9,0
24,0
38,0
45,0
57,0
nm
°C
80,0
5,0
15,0
32,0
45,0
55,0
nm
nm
°C
90,0
10,0
21,0
38,0
52,0
nm
nm
nm
°C
>100,0
15,0
27,0
45,0
60,0
nm
nm
nm
°C
Legend: nm * Operation not possible since the minimum temperature is above 60°C
Table: Minimum temperature of the coolant
3-3 95582944 25/01/15e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Assignment for 12-pin connector - for RS232-Interface
ATTENTION In case of wrong connection, the device might be destroyed.
Cores color
Code DIN IEC 757
plug-in contacts
red
RD
A
CTS
white
WH
B
− supply voltage (AC or DC)
green/white - alternative: grey/pink or colorless
grey
+ thermal switch analog input digital input
GNWH GYPK colorless
GY
Function
C
RS232 9-pin
RS232 25-pin
7
4
6
6
(note 1)
isolated digital input (DI) 24V-logic / 5V-logic (note 2) open-collector output 2 (DO2)
(note 3)
− thermal switch analog input digital input
(note 1)
DI/DO1-zero (Gnd)
(note 2)
D DI/DO1/DO2-zero (Gnd) (note 3)
yellow brown/white - alternative: red/blue or orange
YE
E
BNWH RDBU OR
+ analog output DTR
F
open-collector output (DO1)
(note 2)
pink
PK
G
TXD
2
3
violet
VT
H
RTS
8
5
blue
BU
J
RXD
3
2
black
BK
K
− Datacom (data ground)
5
7
brown
BN
L
+ supply voltage (AC or DC)
green
GN
M
− analog output
Notes: 1: Hardware-Programming see Data Sheet page 1-1 2: Valid for Interface map with open-collector output. 3: Valid for Interface map with 2 open-collector output. Further information next page
12-pin plug connector
12-pin socket-outlet
3-4 95582944 25/01/15e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Assignment valid for RS232-Interface Only 1 variant possible (Version Data Sheet page 1-1)
3-5 95582944 25/01/15e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Assignment for 12-pin connector - for RS485-Interface
ATTENTION In case of wrong connection, the device might be destroyed.
Cores color red white
green/white - alternative: grey/pink or colorless
grey
yellow brown/white - alternative: red/blue or orange pink
violet
blue
Code DIN IEC 757
plug-in contacts
RD
A
WH
B
GNWH GYPK colorless
GY
YE
C
Function BUS + input (RXD+)
(note 1, 4)
not connected
(note 1, 5)
− supply voltage (AC or DC) + thermal switch analog input digital input
(note 1)
isolated digital input (DI) 24V-logic / 5V-logic
(note 1, 2)
open-collector output 2 (DO 2)
(note 1, 3)
− thermal switch analog input digital input
(note 1)
DI/DO1-zero (Gnd)
(note 1, 2)
DI/DO1/DO2-zero (Gnd)
(note 1, 3)
D
E
+ analog output
F
open-collector output (DO1)
(note 1, 2)
BUS - output (TXD-)
(note 1, 4)
BUS - input / BUS - output (DATA-)
(note 1, 5)
BUS + output (TXD+)
(note 1, 4)
BUS + input / BUS + output (DATA+)
(note 1, 5)
BUS - input (RXD-)
(note 1, 4)
not connected
(note 1, 5)
BNWH RDBU OR PK
VT
BU
G
H
J
black
BK
K
BUS-GND
brown
BN
L
+ supply voltage (AC or DC)
green
GN
M
− analog output
Notes: 1: Hardware-Programming see Data Sheet 1-1 2: Valid for Interface map with isolation and open-collector output. 3: Valid for Interface map with isolation and 2 open-collector output. 4: Valid for RS485, Full-duplex 5: Valid for RS485, Half-duplex
12-pin plug connector
12-pin socket-outlet
3-6 95582944 25/01/15e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Assignment for 7-pin connector
ATTENTION When connecting the Infrared Radiation Thermometer to the Power Supply Unit T24 please note the connection conditions in the manual.
Cores color
Code DIN IEC 757
plug-in contacts
brown
BN
4
+
white
WH
2
−
yellow
YE
5
+
supply voltage (AC or DC)
analog output
green
GN
1
−
blue
BU
3
housing
pink
PK
6
+ thermal switch analog input digital input (note 1)
grey
GY
7
− thermal switch analog input digital input (note 1)
Note 1: Hardware-Programming see Data Sheet page 1-1
7-pin plug connector
7-pin socket-outlet
The load resistors for the analog outputs are: ► for voltage output 0 to 1 V ≥ 10 kOhm, ► for voltage output 0 to 10 V ≥ 10 kOhm, ► for current output 0 to 20 mA ≤ 550 Ohm ► for current output 4 to 20 mA ≤ 550 Ohm For the signal evaluation, differential inputs are recommended. 3-7 95582944 25/01/15e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.01 II / KT15.02 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses.
Strahlertemperatur Radiation temperature
200 °C
300 °C
500 °C
700 °C
1000 °C
KT15.01 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
KT15.02 II
Detektortyp
Einstellzeit
Detector type
A
B
C
G
** ** ** 10.00 6.70 2.80 1.65 1.10
** ** ** ** ** 10.30 6.20 4.10
** ** ** ** ** ** 15.00 10.00
** ** ** ** ** ** ** **
13.30 10.80 3.95 1.60 1.10 0.45 0.25 0.20
** ** 14.65 6.00 4.00 1.65 1.00 0.65
** ** ** 14.55 9.70 4.05 2.45 1.60
** ** ** ** ** 13.50 8.10 5.40
1.45 1.15 0.45 0.20 0.10 0.05 0.05 0.05
5.25 4.25 1.55 0.65 0.45 0.20 0.10 0.05
12.80 10.40 3.80 1.55 1.05 0.45 0.25 0.15
** ** 12.70 5.20 3.45 1.45 0.85 0.60
0.65 0.40 0.20 0.05 0.05 0.05 0.05 0.05
1.60 1.30 0.50 0.20 0.15 0.05 0.05 0.05
3.80 3.05 1.15 0.45 0.30 0.15 0.10 0.05
12.55 10.15 3.75 1.50 1.00 0.40 0.25 0.15
0.80 0.45 0.30 0.10 0.05 0.05 0.05 0.05
0.95 0.60 0.35 0.10 0.05 0.05 0.05 0.05
1.55 1.15 0.50 0.20 0.10 0.05 0.05 0.05
4.45 3.55 1.35 0.55 0.35 0.15 0.10 0.05
Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
4.50 3.65 1.35 0.55 0.35 0.15 0.10 0.05
16.70 13.55 4.95 2.05 1.35 0.55 0.35 0.25
** ** 12.10 4.95 3.30 1.35 0.80 0.55
** ** ** 16.45 11.00 4.60 2.75 1.85
1.35 1.05 0.40 0.15 0.10 0.05 0.05 0.05
4.85 3.95 1.45 0.60 0.40 0.15 0.10 0.05
11.80 9.60 3.50 1.45 0.95 0.40 0.25 0.15
** ** 11.70 4.80 3.20 1.35 0.80 0.55
0.50 0.30 0.20 0.05 0.05 0.05 0.05 0.05
1.15 0.90 0.35 0.15 0.10 0.05 0.05 0.05
2.60 2.10 0.80 0.30 0.20 0.10 0.05 0.05
8.60 7.00 2.55 1.05 0.70 0.30 0.15 0.10
0.60 0.35 0.25 0.05 0.05 0.05 0.05 0.05
0.75 0.45 0.25 0.10 0.05 0.05 0.05 0.05
1.20 0.90 0.40 0.15 0.10 0.05 0.05 0.05
3.50 2.80 1.05 0.40 0.30 0.10 0.05 0.05
0.95 0.50 0.35 0.10 0.05 0.05 0.05 0.05
1.00 0.50 0.35 0.10 0.05 0.05 0.05 0.05
1.05 0.60 0.40 0.10 0.10 0.05 0.05 0.05
1.80 1.35 0.60 0.20 0.15 0.05 0.05 0.05
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.01 II / KT15.02 II 3-8 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.21 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature
20 °C
100 °C
200 °C
300 °C
400 °C
KT15.21 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
** ** ** ** ** ** ** **
** ** ** ** ** ** ** **
** ** ** ** ** ** ** **
** ** ** ** ** ** ** **
** ** ** ** ** 9.10 5.45 3.65
** ** ** ** ** ** ** 13.50
** ** ** ** ** ** ** **
** ** ** ** ** ** ** **
** ** 12.55 5.15 3.40 1.45 0.85 0.55
** ** ** 19.00 12.65 5.25 3.15 2.10
** ** ** ** ** 12.85 7.70 5.15
** ** ** ** ** ** ** 17.10
13.10 10.60 3.90 1.60 1.05 0.45 0.25 0.20
** ** 14.40 5.90 3.90 1.65 1.00 0.65
** ** ** 14.30 9.55 4.00 2.40 1.60
** ** ** ** ** 13.25 7.95 5.30
6.05 4.90 1.80 0.75 0.50 0.20 0.10 0.10
** 18.15 6.65 2.75 1.80 0.75 0.45 0.30
** ** 16.20 6.65 4.40 1.85 1.10 0.75
** ** ** ** 14.75 6.15 3.70 2.45
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.21 II 3-9 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.23 II / KT15.24 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses.
Strahlertemperatur Radiation temperature
20 °C
100 °C
200 °C
300 °C
400 °C
KT15.23 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
KT15.24 II
Detektortyp
Einstellzeit
Detector type
A
B
C
G
** ** ** 10.20 6.80 2.85 1.70 1.15
** ** ** ** ** 10.50 6.30 4.20
** ** ** ** ** ** 15.30 10.20
** ** ** ** ** ** ** **
** ** 8.65 3.55 2.35 1.00 0.60 0.40
** ** ** 13.05 8.70 3.65 2.20 1.45
** ** ** ** ** 8.85 5.30 3.55
** ** ** ** ** ** 17.65 11.75
** ** 4.15 1.70 1.15 0.45 0.30 0.20
** ** 15.30 6.25 4.15 1.75 1.05 0.70
** ** ** 15.25 10.15 4.25 2.55 1.70
** ** ** ** ** 14.10 8.45 5.65
** ** 2.70 1.10 0.75 0.30 0.20 0.10
** ** 10.00 4.10 2.75 1.15 0.70 0.45
** ** ** 9.95 6.65 2.75 1.65 1.10
** ** ** ** ** 9.25 5.55 3.70
** ** 2.05 0.85 0.55 0.25 0.15 0.10
** ** 7.55 3.10 2.05 0.85 0.50 0.35
** ** 18.40 7.50 5.00 2.10 1.25 0.85
** ** ** ** 16.70 6.95 4.20 2.80
Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
** ** 13.70 5.60 3.75 1.55 0.95 0.60
** ** ** ** 13.85 5.75 3.45 2.30
** ** ** ** ** 14.00 8.40 5.60
** ** ** ** ** ** ** 18.70
** ** 6.00 2.45 1.65 0.70 0.40 0.25
** ** ** 9.05 6.05 2.50 1.50 1.00
** ** ** ** 14.70 6.10 3.65 2.45
** ** ** ** ** ** 12.25 8.15
** ** 3.30 1.35 0.90 0.40 0.25 0.15
** ** 12.25 5.00 3.35 1.40 0.85 0.55
** ** ** 12.20 8.10 3.40 2.05 1.35
** ** ** ** ** 11.25 6.75 4.50
** ** 2.40 1.00 0.65 0.25 0.15 0.10
** ** 8.90 3.65 2.45 1.00 0.60 0.40
** ** ** 8.85 5.90 2.45 1.50 1.00
** ** ** ** 19.70 8.20 4.90 3.30
** ** 2.00 0.80 0.55 0.20 0.15 0.10
** ** 7.30 3.00 2.00 0.85 0.50 0.35
** ** 17.70 7.25 4.85 2.00 1.20 0.80
** ** ** ** 16.10 6.70 4.00 2.70
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.23 II / KT15.24 II 3-10 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.25 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses.
Strahlertemperatur Radiation temperature
20 °C
100 °C
200 °C
300 °C
400 °C
KT15.25 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
** ** 13.25 5.40 3.60 1.50 0.90 0.60
** ** ** ** 13.35 5.55 3.35 2.25
** ** ** ** ** 13.55 8.15 5.40
** ** ** ** ** ** ** 18.05
** ** 5.80 2.40 1.60 0.65 0.40 0.25
** ** ** 8.80 5.85 2.45 1.45 1.00
** ** ** ** 14.25 5.95 3.55 2.40
** ** ** ** ** 19.80 11.90 7.90
** ** 3.30 1.35 0.90 0.35 0.20 0.15
** ** 12.20 5.00 3.30 1.40 0.85 0.55
** ** ** 12.10 8.10 3.35 2.00 1.35
** ** ** ** ** 11.20 6.75 4.50
** ** 2.35 0.95 0.65 0.25 0.15 0.10
** ** 8.75 3.60 2.40 1.00 0.60 0.40
** ** ** 8.70 5.80 2.40 1.45 0.95
** ** ** ** 19.35 8.05 4.85 3.20
** ** 1.95 0.80 0.55 0.20 0.15 0.10
** ** 7.20 2.95 1.95 0.80 0.50 0.35
** ** 17.45 7.15 4.75 2.00 1.20 0.80
** ** ** ** 15.85 6.60 3.95 2.65
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.25 II 3-11 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.41 II / KT15.42 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature
20 °C
100 °C
300 °C
700 °C
1000 °C
KT15.41 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
KT15.42 II
Detektortyp
Einstellzeit
Detector type
A
B
C
G
** ** ** ** ** ** ** 13.90
** ** ** ** ** ** ** **
** ** ** ** ** ** ** **
** ** ** ** ** ** ** **
** ** ** 13.10 8.75 3.65 2.20 1.45
** ** ** ** ** 13.45 8.10 5.40
** ** ** ** ** ** 19.65 13.10
** ** ** ** ** ** ** **
7.90 6.40 2.35 0.95 0.65 0.25 0.15 0.10
** ** 8.65 3.55 2.35 1.00 0.60 0.40
** ** ** 8.60 5.75 2.40 1.45 0.95
** ** ** ** 19.15 8.00 4.80 3.20
1.65 1.25 0.55 0.20 0.15 0.05 0.05 0.05
5.50 4.45 1.65 0.65 0.45 0.20 0.10 0.05
13.25 10.75 3.95 1.60 1.05 0.45 0.25 0.20
** ** 13.10 5.35 3.55 1.50 0.90 0.60
1.60 1.00 0.55 0.20 0.10 0.05 0.05 0.05
3.75 2.90 1.15 0.45 0.30 0.10 0.05 0.05
8.60 6.95 2.60 1.05 0.70 0.30 0.15 0.10
** ** 8.45 3.45 2.30 0.95 0.55 0.40
Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
** ** ** 12.60 8.40 3.50 2.10 1.40
** ** ** ** ** 12.95 7.75 5.20
** ** ** ** ** ** 18.90 12.60
** ** ** ** ** ** ** **
** 17.40 6.40 2.60 1.75 0.70 0.45 0.30
** ** ** 9.65 6.45 2.70 1.60 1.05
** ** ** ** 15.65 6.50 3.90 2.60
** ** ** ** ** ** 13.05 8.70
3.80 3.05 1.15 0.45 0.30 0.15 0.10 0.05
14.00 11.35 4.15 1.70 1.15 0.45 0.30 0.20
** ** 10.15 4.15 2.75 1.15 0.70 0.45
** ** ** 13.80 9.20 3.85 2.30 1.55
1.70 1.20 0.55 0.20 0.15 0.05 0.05 0.05
5.10 4.05 1.55 0.60 0.40 0.15 0.10 0.05
12.15 9.80 3.60 1.45 1.00 0.40 0.25 0.15
** ** 12.00 4.90 3.25 1.35 0.80 0.55
1.95 1.20 0.70 0.20 0.15 0.05 0.05 0.05
4.20 3.25 1.30 0.50 0.35 0.15 0.10 0.05
9.55 7.65 2.85 1.15 0.75 0.30 0.20 0.15
** ** 9.35 3.80 2.55 1.05 0.65 0.40
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.41 II / KT15.42 II
3-12 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.43 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses.
Strahlertemperatur Radiation temperature
20 °C
100 °C
300 °C
700 °C
1000 °C
KT15.43 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
15.60 12.65 4.65 1.90 1.25 0.55 0.30 0.20
** ** 17.15 7.00 4.65 1.95 1.15 0.80
** ** ** 17.05 11.35 4.75 2.85 1.90
** ** ** ** ** 15.80 9.45 6.30
6.30 5.10 1.90 0.75 0.50 0.20 0.15 0.10
** 18.95 6.95 2.85 1.90 0.80 0.45 0.30
** ** 16.90 6.90 4.60 1.90 1.15 0.75
** ** ** ** 15.35 6.40 3.85 2.55
2.55 2.05 0.75 0.30 0.20 0.10 0.05 0.05
9.30 7.50 2.75 1.15 0.75 0.30 0.20 0.15
** 18.25 6.70 2.75 1.85 0.75 0.45 0.30
** ** ** 9.15 6.10 2.55 1.50 1.00
2.00 1.40 0.70 0.25 0.15 0.05 0.05 0.05
5.65 4.50 1.70 0.70 0.45 0.20 0.10 0.10
13.45 10.90 4.00 1.65 1.10 0.45 0.25 0.20
** ** 13.30 5.45 3.60 1.50 0.90 0.60
2.45 1.50 0.90 0.25 0.20 0.05 0.05 0.05
5.30 4.10 1.65 0.65 0.40 0.15 0.10 0.05
12.05 9.65 3.60 1.45 0.95 0.40 0.25 0.15
** ** 11.75 4.80 3.20 1.35 0.80 0.55
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.43 II 3-13 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.45 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature
20 °C
100 °C
300 °C
700 °C
1000 °C
KT15.45 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
** ** 12,55 5,15 3,4 1,45 0,85 0,55
** ** ** 19 12,65 5,25 3,15 2,1
** ** ** ** ** 12,85 7,7 5,15
** ** ** ** ** ** ** 17,1
17 13,75 5,05 2,05 1,4 0,55 0,35 0,25
** ** 18,65 7,65 5,1 2,1 1,25 0,85
** ** ** 18,6 12,4 5,15 3,1 2,05
** ** ** ** ** 17,2 10,3 6,9
6,4 5,15 1,9 0,8 0,5 0,2 0,15 0,1
** 19,1 7 2,85 1,9 0,8 0,5 0,3
** ** 17,05 6,95 4,65 1,95 1,15 0,75
** ** ** ** 15,5 6,45 3,85 2,6
4 3,15 1,25 0,5 0,3 0,15 0,1 0,05
14,05 11,4 4,2 1,7 1,15 0,45 0,3 0,2
** ** 10,15 4,15 2,75 1,15 0,7 0,45
** ** ** 13,8 9,2 3,85 2,3 1,55
3,95 2,9 1,25 0,45 0,3 0,15 0,1 0,05
12,5 10,05 3,75 1,5 1 0,4 0,25 0,15
** ** 8,95 3,65 2,45 1 0,6 0,4
** ** ** 12,15 8,1 3,4 2,05 1,35
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.45 II 3-14 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.62 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses. Strahlertemperatur Radiation temperature
400 °C
500 °C
700 °C
1000 °C
1500 °C
KT15.62 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
** ** 1.25 0.50 0.35 0.15 0.10 0.05
** ** 4.65 1.90 1.25 0.55 0.30 0.20
** ** 11.30 4.60 3.10 1.30 0.75 0.50
** ** ** 15.40 10.25 4.25 2.55 1.70
** ** 0.90 0.35 0.25 0.10 0.05 0.05
** ** 3.20 1.30 0.90 0.35 0.20 0.15
** ** 7.80 3.20 2.15 0.90 0.55 0.35
** ** ** 10.65 7.10 2.95 1.75 1.20
** ** 0.65 0.25 0.15 0.05 0.05 0.05
** ** 2.10 0.85 0.55 0.25 0.15 0.10
** ** 5.05 2.05 1.40 0.55 0.35 0.25
** ** 16.85 6.90 4.60 1.90 1.15 0.75
** ** 0.70 0.20 0.15 0.05 0.05 0.05
** ** 1.60 0.60 0.40 0.15 0.10 0.05
** ** 3.65 1.45 1.00 0.40 0.25 0.15
** ** 11.95 4.90 3.25 1.35 0.80 0.55
** ** 1.05 0.30 0.20 0.10 0.05 0.05
** ** 1.55 0.55 0.35 0.15 0.10 0.05
** ** 3.00 1.20 0.80 0.35 0.20 0.15
** ** 9.55 3.90 2.60 1.10 0.65 0.45
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.62 II
3-15 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.63 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses.
Strahlertemperatur Radiation temperature
400 °C
500 °C
700 °C
1000 °C
1500 °C
KT15.63 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
** ** 1.20 0.50 0.35 0.15 0.10 0.05
** ** 4.45 1.80 1.20 0.50 0.30 0.20
** ** 10.80 4.40 2.95 1.25 0.75 0.50
** ** ** 14.70 9.80 4.10 2.45 1.65
** ** 0.90 0.35 0.25 0.10 0.05 0.05
** ** 3.20 1.30 0.85 0.35 0.20 0.15
** ** 7.75 3.15 2.10 0.90 0.55 0.35
** ** ** 10.55 7.05 2.95 1.75 1.15
** ** 0.65 0.25 0.15 0.05 0.05 0.05
** ** 2.10 0.85 0.55 0.25 0.15 0.10
** ** 5.10 2.10 1.40 0.60 0.35 0.25
** ** 16.95 6.95 4.65 1.95 1.15 0.75
** ** 0.70 0.25 0.15 0.05 0.05 0.05
** ** 1.65 0.65 0.45 0.20 0.10 0.05
** ** 3.75 1.55 1.00 0.45 0.25 0.15
** ** 12.45 5.10 3.40 1.40 0.85 0.55
** ** 1.05 0.30 0.20 0.10 0.05 0.05
** ** 1.60 0.55 0.40 0.15 0.10 0.05
** ** 3.15 1.25 0.85 0.35 0.20 0.15
** ** 10.00 4.05 2.70 1.15 0.70 0.45
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.63 II 3-16 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.69 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses.
Strahlertemperatur Radiation temperature
400 °C
500 °C
700 °C
1000 °C
1500 °C
KT15.69 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
** ** 1.20 0.50 0.35 0.15 0.10 0.05
** ** 4.45 1.80 1.20 0.50 0.30 0.20
** ** 10.80 4.40 2.95 1.25 0.75 0.50
** ** ** 14.70 9.80 4.10 2.45 1.65
** ** 0.90 0.35 0.25 0.10 0.05 0.05
** ** 3.20 1.30 0.85 0.35 0.20 0.15
** ** 7.75 3.15 2.10 0.90 0.55 0.35
** ** ** 10.55 7.05 2.95 1.75 1.15
** ** 0.65 0.25 0.15 0.05 0.05 0.05
** ** 2.10 0.85 0.55 0.25 0.15 0.10
** ** 5.10 2.10 1.40 0.60 0.35 0.25
** ** 16.95 6.95 4.65 1.95 1.15 0.75
** ** 0.70 0.25 0.15 0.05 0.05 0.05
** ** 1.65 0.65 0.45 0.20 0.10 0.05
** ** 3.75 1.55 1.00 0.45 0.25 0.15
** ** 12.45 5.10 3.40 1.40 0.85 0.55
** ** 1.05 0.30 0.20 0.10 0.05 0.05
** ** 1.60 0.55 0.40 0.15 0.10 0.05
** ** 3.15 1.25 0.85 0.35 0.20 0.15
** ** 10.00 4.05 2.70 1.15 0.70 0.45
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.69 II 3-17 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.81 II / KT15.82 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses.
Strahlertemperatur Radiation temperature
─ 25 °C
20 °C
100 °C
300 °C
700 °C
KT15.81 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
KT15.82 II
Detektortyp
Einstellzeit
Detector type
Response time
A
B
C
G
7.70 6.25 2.30 0.95 0.65 0.25 0.15 0.10
** ** 8.50 3.45 2.30 0.95 0.60 0.40
** ** ** 8.45 5.65 2.35 1.40 0.95
** ** ** ** 18.75 7.80 4.70 3.15
4.35 3.50 1.30 0.55 0.35 0.15 0.10 0.05
16.05 13.00 4.75 1.95 1.30 0.55 0.35 0.20
** ** 11.60 4.75 3.15 1.30 0.80 0.55
** ** ** 15.80 10.55 4.40 2.65 1.75
2.20 1.80 0.65 0.25 0.20 0.05 0.05 0.05
8.10 6.55 2.40 1.00 0.65 0.25 0.15 0.10
19.70 16.00 5.85 2.40 1.60 0.65 0.40 0.25
** ** 19.55 8.00 5.35 2.20 1.35 0.90
1.20 0.90 0.40 0.15 0.10 0.05 0.05 0.05
4.05 3.25 1.20 0.50 0.35 0.15 0.10 0.05
9.75 7.90 2.90 1.20 0.80 0.35 0.20 0.15
** ** 9.65 3.95 2.65 1.10 0.65 0.45
1.65 0.95 0.60 0.15 0.10 0.05 0.05 0.05
3.05 2.30 0.95 0.35 0.25 0.10 0.05 0.05
6.65 5.35 2.00 0.80 0.55 0.20 0.15 0.10
** 17.60 6.45 2.65 1.75 0.75 0.45 0.30
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
3.70 3.00 1.10 0.45 0.30 0.15 0.10 0.05
13.70 11.10 4.05 1.65 1.10 0.45 0.30 0.20
** ** 9.90 4.05 2.70 1.10 0.65 0.45
** ** ** 13.50 9.00 3.75 2.25 1.50
2.15 1.75 0.65 0.25 0.15 0.05 0.05 0.05
7.90 6.40 2.35 0.95 0.65 0.25 0.15 0.10
19.25 15.60 5.70 2.35 1.55 0.65 0.40 0.25
** ** 19.10 7.80 5.20 2.15 1.30 0.85
1.20 0.95 0.35 0.15 0.10 0.05 0.05 0.05
4.35 3.50 1.30 0.55 0.35 0.15 0.10 0.05
10.55 8.55 3.15 1.30 0.85 0.35 0.20 0.15
** ** 10.45 4.30 2.85 1.20 0.70 0.50
0.85 0.60 0.30 0.10 0.05 0.05 0.05 0.05
2.35 1.90 0.70 0.30 0.20 0.10 0.05 0.05
5.60 4.55 1.65 0.70 0.45 0.20 0.10 0.10
18.60 15.05 5.55 2.25 1.50 0.65 0.40 0.25
1.60 0.85 0.60 0.15 0.10 0.05 0.05 0.05
2.20 1.50 0.75 0.25 0.15 0.05 0.05 0.05
4.15 3.25 1.30 0.50 0.35 0.15 0.10 0.05
13.05 10.55 3.90 1.60 1.05 0.45 0.25 0.20
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.81 II / KT15.82 II 3-18 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Temperaturauflösung/Temperature resolution KT15.83 II / KT15.85 II Temperaturauflösung (NET) in ± K (bei Emissionsgrad = 1; σ = 2) für Standardobjektive. Temperature resolution (NET) in ± K (emissivity-setting = 1; σ = 2) for standard lenses.
Strahlertemperatur Radiation temperature
─ 25 °C
20 °C
100 °C
300 °C
700 °C
KT15.83 II Einstellzeit Response time
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
KT15.85 II
Detektortyp
Einstellzeit
Detector type
Response time
A
B
C
G
3.00 2.45 0.90 0.35 0.25 0.10 0.05 0.05
11.10 9.00 3.30 1.35 0.90 0.40 0.25 0.15
** ** 8.05 3.30 2.20 0.90 0.55 0.35
** ** ** 10.95 7.30 3.05 1.85 1.20
1.80 1.45 0.55 0.20 0.15 0.05 0.05 0.05
6.70 5.40 2.00 0.80 0.55 0.25 0.15 0.10
16.25 13.15 4.85 2.00 1.30 0.55 0.35 0.20
** ** 16.10 6.60 4.40 1.85 1.10 0.75
1.00 0.80 0.30 0.10 0.10 0.05 0.05 0.05
3.70 3.00 1.10 0.45 0.30 0.10 0.10 0.05
9.00 7.30 2.65 1.10 0.75 0.30 0.20 0.10
** ** 8.90 3.65 2.45 1.00 0.60 0.40
0.80 0.55 0.30 0.10 0.05 0.05 0.05 0.05
2.05 1.60 0.65 0.25 0.15 0.05 0.05 0.05
4.80 3.90 1.45 0.60 0.40 0.15 0.10 0.05
15.95 12.95 4.75 1.95 1.30 0.55 0.30 0.20
1.55 0.85 0.60 0.15 0.10 0.05 0.05 0.05
2.05 1.35 0.70 0.25 0.15 0.05 0.05 0.05
3.70 2.85 1.15 0.45 0.30 0.10 0.05 0.05
11.30 9.15 3.40 1.35 0.90 0.40 0.25 0.15
5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s 5 ms 10 ms 30 ms 100 ms 300 ms 1s 3s 10 s
Detektortyp Detector type
A
B
C
G
19.55 15.85 5.80 2.40 1.60 0.65 0.40 0.25
** ** ** 8.80 5.85 2.45 1.45 1.00
** ** ** ** 14.25 5.95 3.55 2.40
** ** ** ** ** 19.80 11.90 7.90
11.75 9.50 3.50 1.45 0.95 0.40 0.25 0.15
** ** 12.90 5.30 3.50 1.45 0.90 0.60
** ** ** 12.85 8.55 3.55 2.15 1.45
** ** ** ** ** 11.90 7.15 4.75
6.70 5.45 2.00 0.80 0.55 0.25 0.15 0.10
** ** 7.40 3.00 2.00 0.85 0.50 0.35
** ** 18.00 7.35 4.90 2.05 1.25 0.80
** ** ** ** 16.35 6.80 4.10 2.70
3.85 3.10 1.15 0.45 0.30 0.15 0.10 0.05
14.05 11.40 4.20 1.70 1.15 0.50 0.30 0.20
** ** 10.15 4.15 2.75 1.15 0.70 0.45
** ** ** 13.85 9.25 3.85 2.30 1.55
3.25 2.45 1.05 0.40 0.25 0.10 0.05 0.05
10.60 8.50 3.15 1.30 0.85 0.35 0.20 0.15
** ** 7.60 3.10 2.05 0.85 0.50 0.35
** ** ** 10.30 6.90 2.85 1.70 1.15
** Value is > 20 °C. This setting is not recommended. Table: Temperature resolution Infrared Radiation Thermometer KT15.83 II / KT15.85 II 3-19 95582944 26/01/16e
TECHNICAL SPECIFICATIONS
KT15 II HEITRONICS Infrarot Messtechnik GmbH
4
START-UP
4.1
Mounting
There are four ways of mechanically mounting the Infrared Radiation Thermometer: ►
direct mounting via two tapped holes located on the bottom side of the device ( figure below)
►
direct mounting via the tripod thread BSW 1/4" located on the bottom side of the device ( figure below)
►
mounting by ball joint via thread BSW 1/4" or spring flange. The device permits an easy adjustment of the Infrared Radiation Thermometer to the measured object. ( OPTIONS AND ACCESSORIES)
►
mounting via adapter ( OPTIONS AND ACCESSORIES)
4-1 95582944 05/03/14e
START-UP
KT15 II HEITRONICS Infrarot Messtechnik GmbH
4.2
Electrical Connection
The Infrared Radiation Thermometer is delivered with PVC- or PTFE-cable with free ends. The Infrared Radiation Thermometer is delivered with a connector which is watertight when closed (IP65). The 12-pole connector is intended for the connection of the supply voltage of the analog output signal, the thermal switch, as well as in addition the connection of a serial interface (RS232C) as well as two alarm contacts.
ATTENTION In case of wrong connection, the device might be destroyed.
Connection of supply voltage T24 MS30 MS35
Connecting cable with free ends Connecting cable with free ends Connecting cable with plug connector
Assignment of the individual cores, respectively plug-in contacts ► tables "Assignment" Option: 7-pole connector for the connection of the supply voltage of the analog output signal and the thermal switch
4.3 Operation without serial interface When operating without a serial interface the emissivity will be set as described in chapter 5.1.
4.4 Operation with serial interface Chapter 5.2.3 : Description of the commands ATTENTION: 12-pole connection required
4-2 95582944 05/03/14e
START-UP
KT15 II HEITRONICS Infrarot Messtechnik GmbH
4.5 Optical Adjustment By means of Infrared Radiation Thermometer KT15 II, the temperature of target can be measured at any distance on the condition that the desired target area is larger than the measured target diameter of the Infrared Radiation Thermometer ( ILLUSTRATIONS, Field of view diagram). For applications desiring a minimum target area to be measured, the measuring field diameter will define at what distance the smallest measured target diameter will be. For optical adjustment of the measuring object further accessories are available ( OPTIONS AND ACCESSORIES).
4.5.1 Aiming via focus laser The illumination of the measured target diameter via the focus laser is switched on/off the "keyboard". At works the focus laser is set for the mounted lens. This means that there will be a reticule with circle at the minimum target diameter for the mounted objective indicating the target size and position. In case other objectives are attached to the Infrared Radiation Thermometer, the focus laser is no longer adjusted, i.e. neither the position nor the size of the target diameter is indicated correctly.
4.5.2 Aiming via pilot laser
ATTENTION Please observe the SAFETY INSTRUCTIONS.
A pilot laser can be integrated into the Infrared Radiation Thermometer KT15 II in case a lens is used which is transparent to visible light. It marks the centre of the target diameter for the objective which has been mounted at works. It is not possible to show the minimum target diameter with this option.
4.5.3 Aiming via Laserpointer LP15 For lenses of type Kx, a laserpointer is required ( OPTIONS AND ACCESSORIES), which is to be attached to lens Kx. The laser marks the centre of the measuring field within a distance between 0.5 m and 5 m.
4-3 95582944 05/03/14e
START-UP
KT15 II HEITRONICS Infrarot Messtechnik GmbH
4.6
Installation of Protective Cooling Jacket WK15
Installation of protective cooling jacket WK15 with integrated air purge for different applications in connection with adapter B4-B2 and B4-B7-B2
Fitting for air purge *
Fitting for cooling jacket *
Please, use only the marked connection for the respective combination! 4-4 95582944 04/07/13e
START-UP
KT15 II HEITRONICS Infrarot Messtechnik GmbH
5
OPERATION AND APPLICATION Configuration
The Infrared Radiation Thermometers series KT15 II can be adapted to the process by means of parameterization. The adaptation is carried out via the keyboard or via the digital interface. The operation via keyboard is described in the following. The communication via interface is carried out analog to it. 5.0
Application
5.0.1
Check routines
Check routines detect errors and indicate such errors. ►
Interface error Errors arising with the communication via the serial interface, are automatically output via the latter as text.
►
Error of measuring range Errors arising due to the overflow or underflow of the range limits are also given via the analog output. Overflow and underflow is indicated at the display. If KT15 II interface is programmed for temperature output, an error message is output instead of a temperature value. The configured signal of the analog output is switched to the maximum value in case of overflow (OF) (approx. 5 % above range limit value), and to the minimum value in case of underflow (UF) (approx. 5 % below range limit value at voltage output, approx. 0.396 % at current output). The following cases are monitored: Overflow (OF)
1.
5.
The measuring temperature is above the instrument measuring range. The internal temperature is above the permissible sensor temperature. The external ambient temperature is above the instrument measuring range. The voltage at the external temperature input is above the operating range (0 to 10 V). Data memory is partly destroyed.
6.
Initialization error of the instrument.
2. 3.
4.
95582944 07/02/15e
Underflow (UF) 1. The measuring temperature is below the instrument measuring range. 2. The internal temperature is below the permissible sensor temperature. 3. The external ambient temperature is below the instrument measuring range. 4. The voltage at the external temperature input is below the operating range (0 to 10 V). 5. Rotational speed of the modulator is outside the limit. 6. Internal miscalculation. 5-1
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
5.0.2
Continuous measurement of ambient temperature
For continuous measurement of the ambient temperature, a 0 to 10 V signal which is proportional to the ambient temperature must be fed in at terminal C and D of the 12-pole connector. The configuration of the input must be programmed with EasyConfig: Minimum temperature value corresponding to 0 V, Maximum temperature value corresponding to 10 V. For the transmission of the measured temperature, a module which is recommended for the sensor is needed. OPTIONS AND ACCESSORIES. The figure shows the connection of a module with PT100 in three-wire technique to an Infrared Radiation Thermometer KT15 II.
ATTENTION The device must be provided for 'Analog input'.
5-2 95582944 06/03/14e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
5.0.3
Use of the digital input
If the device is equipped with the option 'digital input', this option can be used for remote control of one of the programmable functions as described below, or a connecting cable with integrated switch can be used. The option 'digital input' must be programmed at the factory. ( Data Sheet Page 1-1 )
5.0.3.1 Remote control functions via the digital input Remote control is done by means of a floating contact 'open collector' or a voltage. Standard is status control. The input is active (ON) at 0 V to 1 V and inactive (OFF) from 4 V to 30 V respectively in open condition.
5.0.3.2 Switch integrated in connecting cable Alternatively for the remote control a switch can be integrated in the connecting cable. Then remote control via the contact is no longer possible.
5.0.3.3 Programmable functions Turning on the laser If the laser is installed and enabled by the interface command 'LASER...' (ENABLE), the laser can be switched on via digital input. Example: 0 V or short circuit switches the laser on. Reset of the peak values The device continuously updates a maximum and a minimum value memory. With the digital input the values can be reset to the actual value. The function is helpful in case the analog output is programmed by the command 'OUT…' for reproduction of a peak value. Example: 0 V resets the maximum value memory to the actual value. Reset of the digital outputs If it is configured, the digital outputs can be reset via the digital input to the inactive state. 5.0.4 Use of the digital outputs DESCRIPTION OF COMMANDS: Command 2.14: Configuration of the digital outputs Example: 0 V will reset the digital output.
5-3 95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
5.0.4
Use of the digital outputs
The digital outputs display the alarm status of the Infrared Radiation Thermometer via the active level. The digital outputs can be configured that either 'Low' alarm or 'High' alarm or both alarms are transmitted. The digital outputs are also configurable as a synchronization output for the function 'Self-reset' of the maximum and minimum value memories. 'Inverse Logic' or 'Not Inverse Logic' can also be selected. For the reset of the digital output the following configurations can be selected: • Reset by command • Reset by digital input • Reset by lower deviation of alarm level • Reset by expiration of a programmable time • Reset after lower deviation or exceed of threshold temperature 5.0.3 Use of the digital input DESCRIPTION OF COMMANDS: Command 2.12: Configuration of digital input Command 2.14: Configuration of digital outputs
5.0.5
Monitoring the device function during operation
Monitoring the operating temperature The series KT15 II Infrared Radiation Thermometers have a built-in temperature switch which opens at ≥ 70 °C. This switch is connected to the flange-plug ( chapter 3). Monitoring the analog output signal The type of the output signal can be selected for the series KT15 II Infrared Radiation Thermometers. For monitoring the output signal the '4 - 20 mA' type is recommended. At this setting a measuring temperature within the set temperature range will supply a signal between 4 and 20 mA. If the value falls below the calibrated temperature range, refer to the type sheet, ≤ 3.95 mA is supplied, and ≥ 21 mA is supplied if it is exceeded. Monitoring the power consumption The series KT15 II Infrared Radiation Thermometers draw a current of 90 mA to 150 mA at 24 V during operation. If this value is out of range, this indicates misbehaviour. Monitoring during operation with a serial interface For monitoring the Infrared Radiation Thermometer we recommend that the serial interface should be switched to the 'send the measured temperature value repeatedly'. As long as the Infrared Radiation Thermometer functions, the temperature value will be sent at the defined time interval. If it is out of range, the Infrared Radiation Thermometer will provide an error message.
5-4 95582944 06/03/14e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
5.0.6
Use of the maximum and minimum value memories
The configuration of these parameters can be done by the software EasyConfig or EasyMeas. The maximum and minimum value memories are continually updated by the Infrared Radiation Thermometer. The stored data can be read by command, and also switched to the analog output. → Commands for 'Memory functions' → Commands for 'Configuration of analog output' If cyclically the temperature of moving material has to be determined by using the maximum or minimum value memory, the relevant memory must be reduced by a decay rate to the current measured value, or reset. 5.0.6.1
Decay rate With the command 'OUT..' the decay rate can be programmed.
5.0.6.2
Reset of memory (manually controlled)
5.0.6.2.1
Reset with the command 'MEM RESET'
5.0.6.2.2
Reset via digital input For this the digital input must be programmed at the factory. Furthermore, the function memory reset must be programmed with the command 'CONFig DIGin..'.
5.0.6.3
Self-reset If an external reset signal is not available, the Infrared Radiation Thermometer generates this automatically. It offers alternatively the self-reset after a reset time, as well as after falling below a temperature threshold.
5.0.6.3.1
Reset after lower deviation of threshold temperature If there is a space (gap) between the targets, and the background temperature deviates to a sufficient extent of the temperature of the target, the Infrared Radiation Thermometer can generate the reset signal automatically. Therefore in the case of the maximum value memory the background temperature must be lower than the temperature of the target, in the case of the minimum value memory the background temperature must be higher than the temperature of the target. For synchronization of the evaluation, the generated reset signal can be switched to the digital output. The edge change occurs by exceeding (reset point maximum value memory) and falling below Tlim; the minimum value memory accordingly reversed (see figures 1 + 2). By selecting 'MANUAL' or 'AUTO' mode, the self-reset of the memories is activated. The device detects a 'temperature gap' between two targets and resets maximum or minimum value memory at the end of the gap. 5-5
95582944 08/07/16e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
● Mode 'MANUAL' uses a programmable temperature threshold (Tlim ) for detecting the gap. ● Mode 'AUTO' determines Tlim automatically with the average of the maximum and minimum values. In this mode a reset time tres can be configured additionally. If during that time no target has been detected, the maximum resp. minimum value memory will be reset. If the time-controlled reset is not needed, the maximum reset time of 600 s should be adjusted. If the maximum value memory is enabled, the reset is done when Tlim is exceeded again. If the minimum value memory is enabled, the reset is done when Tlim has fallen below (see figures 1 + 2). →
Commands 'Configuration of digital output' in Universal Protocol.
Please note: The time constant of the device may be set to auto-reset function only as large as the measurement is not distorted by integration! The extended memory configuration can be provided both via the digital interface and the analog output. Via the extended configuration of the analog output, the maximum or minimum value memory can be switched to the analog output. →
Commands 'Configuration of analog output' in Universal Protocol.
5-6 95582944 08/07/16e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Example: Self-reset with maximum value memory, MANUAL mode
Background
Thick product
Thin product
Figure 1: Measurement
Figure 2: Signal curve
5.0.6.3.2
Reset after reset time
The maximum or minimum value memories can be reset alternatively after the expiration of a programmable time. The reset time is retriggered automatically, when the value of the activated maximum value memory rises, resp. minimum value memory falls. This feature can be selected for only one of the memory types.
5-7 95582944 08/07/16e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
5.1
Operation via keyboard
With the configuration menu, the most important parameters are set or interrogated. For full access to all parameters the software EasyConfig or EasyMeas is needed. LED on: up-/down-button is active
page up menu increase the value
menu- and value-display
divide into submenu confirm the change
back to the previous menu back without change
Enter
Temp ESC
Change
RS
232 RS 232 serial interface
change someything
page down menu reduce the value
Operation of the Menu To activate the display press any button. The main menu starts (Temp). The LED between the up-/down-buttons shines. That means: these buttons are activated. With these buttons you can choose other menus or change values. Press Enter to go to a submenu in order to indicate settings and values. Press ESC to go back to the main menu without any change. Press Change to change settings or values. The submenu is flashing and the LED shines. Use the up-/down-buttons for selecting and changing. Confirm the selection with Enter. Feed the Code if required. Confirm the correctness of the Code with ok. Press ESC several times to go back to main menu or to switch off the menu. The menu automatically switches off after two minutes. The Code is valid until you leave the menu (display off). 5-8 95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Main Menu 1.1
Temp ESC
Enter
1.9
Change
Lase ESC
Change
RS 232
1.2
Info ESC
Enter
RS 232
2.0
Change
Unit ESC
Emi ESC
Enter
RS 232
2.1
Change
AOut ESC
TAmb ESC
Enter
RS 232
2.2
Change
Uart ESC
Resp ESC
Enter
RS 232
2.3
Change
Cal ESC
Memo ESC
Enter
RS 232
2.4
Change
Conf ESC
HiAl ESC
Enter
Change
1.8
LoAl ESC
RS 232
2.5
Code ESC
RS 232
Enter
Change
RS 232
1.7
Enter
Change
RS 232
1.6
Enter
Change
RS 232
1.5
Enter
Change
RS 232
1.4
Enter
Change
RS 232
1.3
Enter
Enter
Change
RS 232
Enter
Change
RS 232
5-9 95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Submenu Temp
Temp ESC
Enter
Enter
350
ESC
C
Change
RS 232 Temp Temperature indicator
Submenu Info
Info ESC
Enter
Enter
Change
KT15
SeNr
-50
1000
.82A
1111
C
C
.953
1.00
ESC
RS 232 Info device information KT15 .82A -50 1000
pyrometer series spectral range, detector type temperature range beginning temperature range end
Submenu Emi
Emi ESC
Enter
Enter
Emi
Change
.953
Change
Enter
Emi
ESC
1.00 ESC
RS
232
Emi emissivity
5-10 95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
5-11
95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
internal temperature
manual input
external sensor
Int
Man
Ext
RS 232
ambient temperature
Change
Enter
Tamb
ESC
TAmb
Enter 35
Int
Man
ESC
Enter
ESC
ESC
Valu
C
150
Ext
Enter
ESC
Enter
Only with option ext. temp. input Ext
Int
C
ESC
Change
C
35
Int
Submenu Tamb
ESC
Enter
ESC
ESC
C
25
Man
or
ESC
Change
Ext
no
ESC
flash
75
Enter
C
75
Man
Configuration of the external input with HEITRONICS Software EASY CONFIG via RS232 serial interface.
ESC
5-12
95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Change
ESC
Change
Max
Hold
Hold momentary value Max maximal value Min minimum value Diff difference value Aktu actual value (In this menu the button ENTER is also button for HOLD and RESET)
Actu
Diff
C
100
Hold
. . 600s
.
1s
.
.
.
.
5ms
Memo memory
RS 232
Enter
ESC
Change
Min
Change
Enter
5ms
Hardware-reset only with option Digital Input
ESC
Memo
Enter
Submenu Memo
RS 232
Enter
Resp response time
ESC
Resp
Submenu Resp
Enter
ESC
Enter
C
100
Max
1s
ESC
ESC
5-13
95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Change
Change
RS 232
Enter
On switched on Off switched off Valu temperature value
LoAl low alarm
ESC
LoAl
Enter
on
Enter
on
Submenu LoAl
RS 232
Enter
On switched on Off switched off Valu temperature value
HiAl high alarm
ESC
HiAl
Submenu HiAl
ESC
Change
ESC
Change
Valu
off
on
Valu
off
on
Enter
Enter
Enter
Enter
C
20
off
C
20
off
ESC
Change
ESC
ESC
Change
ESC
20
20
45
55
ESC
Enter
ESC
Enter
C
45
C
55
ESC
ESC
5-14
95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Change
switched on switched off
Change
C K F
° Celsius ° Kelvin ° Fahrenheit
Unit Units
ESC
Unit
on off
Lase Laser
ESC
Lase
RS 232
Enter
Enter
on
Enter
C
Submenu Unit
RS 232
Enter
Submenu Laser
ESC
Change
ESC
Change
F
flasch
K
C
off
on
ESC
Enter
ESC
Enter
K
off
ESC
ESC
95582944 05/03/12e
Change
RS 232
Enter
OPERATION AND APPLICATION
actual value decay rate high temperature range Kelvin/second low temperature range current-/voltage output maximum value minimum value output radiation temperature range decay rate value temperature radiant temperature radiant
Enter
DeRa
Out
mA/V
Rang
TeRa
Enter
ESC
Enter
ESC
Enter
ESC
Enter
ESC
Enter
K/s
0
Rate
Actu
mA
0_20
C
500
0
Temp
ESC
Change
ESC
Change
ESC
Change
ESC
Change
ESC
Change
0
Min
Max
Actu
0_10
0_1
4_20
0_20
high
low
Rad
Temp
2
Enter
Enter
Change
Change
ESC
Enter
Enter
Min
V
0_1
500
0
Rad
K/s
2
Rate
ESC
ESC
ESC
ESC
400
20
Enter
Enter
C
400
0
C
500
20
ESC
ESC
Final temperature ≤ 150 °C ≤ 200 °C ≤ 1000 °C > 1000 °C
Actu DeRa high K/s low mA/V Max Min Out Rad Rang Rate Temp TeRa
Aout analogue output
ESC
AOut
Submenu Aout
Minimum temperature difference: depends on final temperature The following values will result: Minimum temperature difference 50 °C 100 °C 200 °C 400 °C
5-15
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Submenu Uart
Uart ESC
Enter
Enter
BauR
Change
9,6
ESC
19
9,6
Change
Enter
19,2
ESC
8
ESC
1
ESC
even
ESC
RS 232 ESC 38
57
115
Data
Change
8
Enter
ESC
7
ESC
Change
1
Enter
ESC
2
ESC
Change
even
Enter
ESC
odd
ESC
8
Stop 1
Pari Uart
digital output
BauR Data even Hand odd Pari Stop Hand
baudrate data bits even handshake odd parity stopbits with hardware handshake
even
Hand
ESC
flasch
yes
5-16 95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
5-17
95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Change
RS 232
Enter
Enter
show Temp indication of temperature value
show CalF indication of cal-factor
Cal calibration
ESC
Cal
Submenu Cal
Temp
show
CalF
show
Enter
Enter
C
348o
CalT
1254
CalF
ESC
Change
ESC
Change
348o
1254
350o
1250
ESC
Enter
ESC
Enter
350o
1250
ESC
ESC
5-18
95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Change
Mode Stnd SC12 Tamb Pmtr Dflt
RS 232
Enter
mode standard mode scanner mode ambient temperature parameter delivery data
Conf Configuration
ESC
Conf
Enter
Submenu Conf
Pmtr
Mode
Enter
Enter
ESC
????
Dflt
ESC
Stnd
Mode
Enter
Change
SC12
Stnd
ESC
Enter
SC12
Mode
ESC
5-19
95582944 05/03/12e
OPERATION AND APPLICATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Change
RS 232
Enter
Enter
0
1
Enter
10
12
120
back to menu
Enter
123
and confirm with ENTER. Then the next numeral and so on.
Such as example: Code 1234
Press ENTER when Code is questioned. At first feed the left numeral (f.e.: 1) of the code with the buttons To go back use button ESC.
Code Code input
ESC
Code
Code test Enter
1230
1234
Enter
OK
5.2
Communication via interface
5.2.1
Preparing the device for operation with interface RS232C
The interface signals are available at the 12-pin plug of the Infrared Radiation Thermometer KT15II. The signals are connected to the computer interface as follows: -----------KT15II------------Name plug-in contact RXD J TXD G RTS H CTS A DTR F GND K
-------computer-----------9 pole 25 pole name 3 2 TXD 2 3 RXD 8 5 CTS 7(4) 4(20) RTS(DTR) 6 6 DSR 5 7 GND
The transfer is controlled by means of hardware handshake (RTS/CTS). Some RS232 drivers use DTR instead of RTS for the handshake. In this case, CTS of Infrared Radiation Thermometer KT15II must be connected to DTR of the computer. The transfer parameters baud rate, number of data bits, parity, stop bits and delimiter are set by means of the operating keys at the Infrared Radiation Thermometer KT15II. These parameters are to be set to equal values as for the Infrared Radiation Thermometer KT15II and the computer. Selectable baud rates: 9.6 kBaud 19.2 kBaud 38.4 kBaud 57.6 kBaud 115.2 kBaud Possible data formats: Number of data bits: 7 or 8 Number of stop bits: 1 or 2 Parity: even, uneven or none The Infrared Radiation Thermometer KT15II is able to receive data with one stop bit independent of the setting. For transmission, the selected number of stop bits is inserted.
5.2.2
Using the interface
5.2.2.1
Control of communication
Data transfer from evaluation to Infrared Radiation Thermometer KT15II The ready-to-receive state of the Infrared Radiation Thermometer KT15II is signalized via line RTS. The coding shown in the following table shall apply.
5-20 95582944 15/03/14e
OPERATION AND APPLICATION
KT15II HEITRONICS Infrarot Messtechnik GmbH
Function
RTS-state
ready-to-receive
on (high)
evaluation of a command
off (low)
receive buffer overflow
off (low)
Data which are sent during the RTS-state is OFF are lost. Data transfer from Infrared Radiation Thermometer KT15II for evaluation purposes Before the Infrared Radiation Thermometer KT15II sends a message, the CTS input is checked. If the CTS-state is on (high), the whole message including the selected delimiter is sent. ● Commands The individual commands are transmitted as a text in ASCII characters. As for instruction characters, only the first three characters will be checked, numerical values are accepted with or without fractional parts, blanks must not be filled in between operational sign and numerical value. An answer is only given in case of interrogation (?). If a command cannot be read, an error message will be output. If parameters are not found to be within the permissible range of values (e.g. left-hand end of extent < right-hand end of extent), the command will not be executed and an error message will be output. ● Delimiter ↵ Infrared Radiation Thermometer KT15II identifies the ASCII control characters carriage return (0D) and line feed (0A) as end of instruction. As soon as the end of instruction has been recognized, KT15II will reset the ready-toreceive state (RTS) until the command is evaluated. The answer of the Infrared Radiation Thermometer KT15II is (0D). ● Input buffer Infrared Radiation Thermometer KT15II has an input buffer of 40 characters. When a longer command has been received, KT15II issues an error message.
5.2.3
Description of the commands
The commands for programming the parameters are described in separate instructions: Universal Protocol Communication via Serial Interface for all digital HEITRONICS Infrared Radiation Thermometers
5-21 95582944 15/03/14e
OPERATION AND APPLICATION
KT15II HEITRONICS Infrarot Messtechnik GmbH
6
MAINTENANCE AND CALIBRATION
6.1
General information
The design of HEITRONICS Infrared Radiation Thermometers provides for a long period of service without the necessity of special maintenance. It is advisable to check the display accuracy approximately every three months. Measuring errors may arise, for instance, owing to pollution of the lens. For this reason, this component should first of all be cleaned. 6.2
Cleaning the lens
For this purpose a cleaning- and servicing kit is available (→ OPTIONS AND ACCESSORIES). Fine dust can be brushed off by means of a lens brush or blown off the lens carefully. Do not use compressed air or canned compressed air which could cause residues on the objective. If the lens is heavily soiled and has greasy deposits, a microfiber cloth, cotton tabs and the special lens cleaner will help cleaning it, after having used the lens brush. 6.3
Testing the display accuracy
The display accuracy of the Infrared Radiation Thermometer can be checked by means of a blackbody calibration source. This test is carried out within the range indicated on the type plate. It is advisable to carry out the test under high temperature. For this purpose, the Infrared Radiation Thermometer is placed in front of a black body calibration source so that the integrated optical equipment is focused on the radiator. The temperature of the black body calibration source must be determined. If a black body calibration Source of type HEITRONICS SW15 (→ OPTIONS AND ACCESSORIES) is used, it has to be mounted to the lens. The Infrared Radiation Thermometer has to be connected to an appropriate power supply. Approximately 15 minutes after appliance of the supply voltage the Infrared Radiation Thermometer is ready for calibration. For checking we recommend to use the following minimum temperatures: Model KT15.01 II KT15.02 II KT15.21 II KT15.23 II KT15.24 II KT15.25 II KT15.41 II KT15.42 II KT15.43 II KT15.62 II KT15.63 II KT15.69 II KT15.81 II KT15.82 II KT15.83 II KT15.85 II
Recommended minimum temperature °C 580 to 600 580 to 600 280 to 300 130 to 150 130 to 150 130 to 150 580 to 600 330 to 350 130 to 150 580 to 600 580 to 600 580 to 600 80 to 100 80 to 100 80 to 100 80 to 100
6-1 95582944 03/07/16e
MAINTENANCE AND CALIBRATION
KT15 II HEITRONICS Infrarot Messtechnik GmbH
8
FIGURES
Nr.
Page
10a
Spectral emissivity of various materials............................................................. 8-2
10b
Total emissivity of some materials at a temperature of 20 °C (spectral range 8 ... 14 µm) ............................................................................... 8-2
12
Emissivity of various foils .................................................................................. 8-3
13
Spectral emissivity of Infrared Radiation Thermometers and transmission curves of various synthetic materials..................................... 8-4
14
Spectral emissivity ε, transmissivity ε and reflection ρ of glass ......................... 8-5
15
Dimensions KT15 II .......................................................................................... 8-6
31
Target Diameter ................................................................................................ 8-8 Transmission curves (illustrated in a diagram): Spectral Response
2.0 … 2.7 µm ................................................................. 8-9
Spectral Response
2.0 … 4.5 µm ................................................................. 8-9
Spectral Response
3.43 µm ....................................................................... 8-10
Spectral Response
6.8 µm ......................................................................... 8-10
Spectral Response
7.93 µm ....................................................................... 8-11
Spectral Response
8.05 µm ....................................................................... 8-11
Spectral Response
3.87 µm ....................................................................... 8-12
Spectral Response
4.9 … 5.5 µm ............................................................... 8-12
Spectral Response
7.5 … 8.2 µm ............................................................... 8-13
Spectral Response
4.24 µm ....................................................................... 8-13
Spectral Response
4.48 µm ....................................................................... 8-14
Spectral Response
4.66 µm ....................................................................... 8-14
Spectral Response
8 … 10 µm ................................................................... 8-15
Spectral Response
8 … 14 µm ................................................................... 8-15
Spectral Response
8 … 20 µm ................................................................... 8-16
Spectral Response
9.6 … 11.5 µm ............................................................. 8-16
8-1 95582944 04/03/14e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Fig. 10a
Spectral emissivity of various materials
Fig. 10b
Total emissivity of some materials at a temperature of 20 °C (spectral range 8 ... 14 µm)
8-2 95582944 02/06/08e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Fig. 12
Emissivity of various foils
8-3 95582944 02/06/08e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
PE: PTFE:
Polyäthylen Polytetrafluoräthylen
PET: PVC:
Polyester Polyvinylchlorid
Fig. 13
Spectral emissivity of Infrared Radiation Thermometers and transmission curves of various synthetic materials 8-4
95582944 04/03/14e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Fig. 14
Spectral emissivity ε, transmissivity ε and reflection ρ of glass
8-5 95582944 02/06/08e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Fig. 15
Housing dimensions Infrared Radiation Thermometer KT15 II
8-6 95582944 04/03/14e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Fig. 30
no longer required
8-7 95582944 03/07/13e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Example: Detector Type A 8 ... 14 µm 8 ... 20 µm 7 ... 16 µm 1 2 3 4 Fig. 31
lens K6 lens L6 lens M6 lens N6 Target Diameter KT15.8x II – Detector Type A
8-8 95582944 02/06/08e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Spektrale Empfindlichkeit SP01 (2.0…2.7 µm) Spectral Response SP01 (2.0…2.7 µm)
Relative Empfindlichkeit [%] Relative Sensitivity [%]
30
25
20
15
10
5
0 0
0,5
1
1,5
2
2,5
3
3,5
Wellenlänge [µm] Wavelength [µm]
Spektrale Empfindlichkeit SP02 (2.0…4.5 µm) Spectral Response SP02 (2.0…4.5 µm) 40
Relative Empfindlichkeit [%] Relative Sensitivity [%]
35 30 25 20 15 10 5 0 0
1
2
3
4
5
6
Wellenlänge [µm] Wavelength [µm]
8-9 95582944 02/06/08e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Spektrale Empfindlichkeit SP21 (3.43 µm) Spectral Response SP21 (3.43 µm) 35
Relative Empfindlichkeit [%] Relative Sensitivity [%]
30 25 20 15 10 5 0 2,5
2,7
2,9
3,1
3,3
3,5
3,7
3,9
4,1
4,3
4,5
7,6
7,8
8
Wellenlänge [µm] Wavelength [µm]
Spektrale Empfindlichkeit SP23 (6.8 µm) Spectral Response SP23 (6.8 µm)
Relative Empfindlichkeit [%] Relative Sensitivity [%]
30
25
20
15
10
5
0 6
6,2
6,4
6,6
6,8
7
7,2
7,4
Wellenlänge [µm] Wavelength [µm]
8-10 95582944 02/06/08e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Spektrale Empfindlichkeit SP24 (7.93 µm) Spectral Response SP24 (7.93 µm) 70
Relative Empfindlichkeit [%] Relative Sensitivity [%]
60 50 40 30 20 10 0 7
7,2
7,4
7,6
7,8
8
8,2
8,4
8,6
8,8
9
8,6
8,8
9
Wellenlänge [µm] Wavelength [µm]
Spektrale Empfindlichkeit SP25 (8.05 µm) Spectral Response SP25 (8.05 µm) 40
Relative Empfindlichkeit [%] Relative Sensitivity [%]
35 30 25 20 15 10 5 0 7
7,2
7,4
7,6
7,8
8
8,2
8,4
Wellenlänge [µm] Wavelength [µm]
8-11 95582944 02/06/08e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Spektrale Empfindlichkeit SP41 (3.87 µm) Spectral Response SP41 (3.87 µm) 100
Relative Empfindlichkeit [%] Relative Sensitivity [%]
90 80 70 60 50 40 30 20 10 0 3
3,2
3,4
3,6
3,8
4
4,2
4,4
4,6
4,8
5
5,6
5,8
6
Wellenlänge [µm] Wavelength [µm]
Spektrale Empfindlichkeit SP42 (4.9…5.5 µm) Spectral Response SP42 (4.9…5.5 µm) 35
Relative Empfindlichkeit [%] Relative Sensitivity [%]
30 25 20 15 10 5 0 4
4,2
4,4
4,6
4,8
5
5,2
5,4
Wellenlänge [µm] Wavelength [µm]
8-12 95582944 02/06/08e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Spektrale Empfindlichkeit SP43 (7.5…8.2 µm) Spectral Response SP43 (7.5…8.2 µm) 50
Relative Empfindlichkeit [%] Relative Sensitivity [%]
45 40 35 30 25 20 15 10 5 0 7
7,2
7,4
7,6
7,8
8
8,2
8,4
8,6
8,8
9
4,6
4,8
5
Wellenlänge [µm] Wavelength [µm]
Spektrale Empfindlichkeit SP61 (4.24 µm) Spectral Response SP61 (4.24 µm) 40
Relative Empfindlichkeit [%] Relative Sensitivity [%]
35 30 25 20 15 10 5 0 3
3,2
3,4
3,6
3,8
4
4,2
4,4
Wellenlänge [µm] Wavelength [µm]
8-13 95582944 02/06/08e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Spektrale Empfindlichkeit SP62 (4.48 µm) Spectral Response SP62 (4.48 µm) 40
Relative Empfindlichkeit [%] Relative Sensitivity [%]
35 30 25 20 15 10 5 0 3,5
3,7
3,9
4,1
4,3
4,5
4,7
4,9
5,1
5,3
5,5
5,1
5,3
5,5
Wellenlänge [µm] Wavelength [µm]
Spektrale Empfindlichkeit SP63 (4.66 µm) Spectral Response SP63 (4.66 µm) 90
Relative Empfindlichkeit [%] Relative Sensitivity [%]
80 70 60 50 40 30 20 10 0 3,5
3,7
3,9
4,1
4,3
4,5
4,7
4,9
Wellenlänge [µm] Wavelength [µm]
8-14 95582944 02/06/08e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Spektrale Empfindlichkeit SP81 (8…10 µm) Spectral Response SP81 (8..10 µm)
Relative Empfindlichkeit [%] Relative Sensitivity [%]
60
50
40
30
20
10
0 6
7
8
9
10
11
12
13
14
15
16
Wellenlänge [µm] Wavelength [µm]
Spektrale Empfindlichkeit SP82 (8…14 µm) Spectral Response SP820 (8…14 µm) 80
Relative Empfindlichkeit [%] Relative Sensitivity [%]
70 60 50 40 30 20 10 0 6
8
10
12
14
16
18
20
Wellenlänge [µm] Wavelength [µm]
8-15 95582944 02/06/08e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
Spektrale Empfindlichkeit SP83 (8…20 µm) Spectral Response SP83 (8…20 µm) 90
Relative Empfindlichkeit [%] Relative Sensitivity [%]
80 70 60 50 40 30 20 10 0 6
8
10
12
14
16
18
20
Wellenlänge [µm] Wavelength [µm]
8-16 95582944 05/10/16e
FIGURES
KT15 II HEITRONICS Infrarot Messtechnik GmbH
GARANTIEBEDINGUNGEN Die HEITRONICS Infrarot Messtechnik GmbH (nachfolgend HEITRONICS genannt) haftet unter Ausschluss weitergehender Ansprüche für Mängel an den von ihr gelieferten Infrarot-Strahlungsthermometern und deren Zubehör, und zwar für die Dauer von 24 Monaten nach Maßgabe folgender Bedingungen: 1. Die Mängelhaftung erstreckt sich ausschließlich auf kostenlosen Ersatz fehlerhafter Teile im Hause HEITRONICS. Die Mängelhaftung bezieht sich insbesondere nicht auf natürliche Abnutzung und nicht auf Schäden, die auf unsachgemäßer Bedienung oder Beanspruchung oder sonstigen von HEITRONICS nicht verschuldeten Umständen beruhen. Die Mängelhaftung gilt nicht für Batterien. Das Gerät ist in der Originalverpackung frachtfrei an HEITRONICS zu senden. Kosten für Steuern, Gebühren und Zölle trägt der Versender. Transportschäden gehen zu Lasten des Versenders. 2. Die Frist für die Mängelhaftung beginnt mit dem Tage des Geräteversandes aus dem Hause HEITRONICS. 3. Etwa auftretende Mängel sind HEITRONICS unverzüglich zu melden, um weitergehende Auswirkungen möglichst zu vermeiden. 4. Ersetzte Teile gehen in das Eigentum von HEITRONICS über. Für Ersatzteile leistet HEITRONICS bis zum Ablauf der für den ursprünglichen Liefergegenstand geltenden Frist in der vorgenannten Weise Gewähr. 5. Alleiniger Gerichtsstand für alle sich aus der Mängelhaftung ergebenden Streitigkeiten ist Wiesbaden, Deutschland.
WARRANTY CONDITIONS Temperature measuring equipment delivered by HEITRONICS Infrarot Messtechnik GmbH (hereinafter referred to as HEITRONICS) is warranted against defects, excluding consequential liability, notably for a period of 24 months subject to the following conditions: 1. Warranty is limited to the free replacement of defective parts at its works. In particular, warranty does not cover normal wear and tear or damage due to improper use or overloading or other circumstances for which HEITRONICS is not responsible. Warranty does not include batteries. The device must be returned to HEITRONICS in original packing, carriage paid. Costs for taxes, fees and customs duties are to be paid by sender. Shipping damage is borne by the sender. 2. The warranty period starts from the date of delivery from its works. 3. Information concerning eventually encountered defects has to be forwarded to HEITRONICS immediately to preclude possible consequential damage. 4. Replaced parts or components are returned to the property of HEITRONICS. Replacements are warranted on the conditions mentioned above until the expiration of the warranty period for the originally delivered equipment. 5. Jurisdiction for any legal dispute arising from this warranty shall be limited to the Court District of Wiesbaden, Germany.
CONDITIONS DE GARANTIE La garantie de HEITRONICS Infrarot Messtechnik GmbH (ci-après mentionnée HEITRONICS) couvre les défauts des radiomètres et accessoires livrés par elle, à l'exclusion de toute autre réclamation, pour une durée de 24 mois dans les conditions suivantes: 1. La responsabilité de HEITRONICS est limitée au remplacement gratuit des pièces défectueuses dans les usines de HEITRONICS. La garantie ne couvre pas le cas d'usure normale, non plus les dommages provoqués par fausse manoeuvre, par des conditions de travail trop dures ou des circonstances dont HEITRONICS n'est pas responsable. Les batteries ne sont pas sous garantie. L'appareil doit être retourné à HEITRONICS dans son emballage d'origine, port payé. L’expéditeur paye pour les frais des impôts, des taxes et des droits de douane. Dommages de transport sont à la charge de l'expéditeur. 2. La période de garantie commence le jour d'expédition des appareils par les usines HEITRONICS. 3. Tout défaut doit être signalé à HEITRONICS de toute urgence pour éviter des conséquences plus graves. 4. Les pièces échangées deviennent propriété de HEITRONICS. Les pièces de rechange bénéficient de la garantie dans les conditions mentionnées ci avant, jusqu'à l'expiration de la période prévue pour la livraison d'origine. 5. Pour tous litiges qui pourraient naître de l'application de la garantie, la seule juridiction compétente sera celle de Wiesbaden, R.F.A.
Garantiebedingungen 04/08/14 def
HEITRONICS Infrarot Messtechnik GmbH
HEITRONICS INFRAROT MESSTECHNIK GMBH IRM SERVICE Lieferanschrift / Delivery address / Adresse de livraison / Dirección de entrega: HEITRONICS Infrarot Messtechnik GmbH Kreuzberger Ring 40 65205 WIESBADEN GERMANY Tel.: +49 611 97393-0 Fax: +49 611 97393-26 E-Mail:
[email protected] Internet: www.heitronics.com ________________________________________________________________________ Vertriebsorganisation Angaben zu unseren regionalen Vertriebspartnern finden Sie im Internet. Sales Network For details about our regional representatives, please, refer to the internet. Réseau des ventes Vous pouvez trouver les coordonnées de nos représentants régionaux sur Internet. Organizacion de la venta Informaciónes referente a nuestros regionales colaboradores de venta encuentran en el internet.