Transcript
Function table: Model Function
KJ610x Series Safety rules and precautions · Please read the manual carefully before using the instrument. · Please check the instrument hasn’t any damage with good meter pen insulating layer before using the instrument. · Every measuring operation is under direction of drawing, on which each number indicate one operation, it may standard your measuring action and impel you to form a good operation habit. · Before opening the battery cover, please pull off the meter pen first; after closing the battery cover, you should screw on the bolts firmly and then start to measure; if not, electric shock may occur. · The input voltage and current cannot exceed the range stipulation. · It is regarded to be dangerous voltage if DC exceeds 60V or AC exceeds 42V, please operate the device carefully and cautiously to avoid electric shock. · The common mode voltage between COM (public end of meter measurement) and the voltage to earth cannot exceed 1000V. · The fuse inside the instrument should be replaced only with these fuses having same specification. · If the LCD displays the battery icon, it indicates the battery voltage is insufficient and it should be replaced with new one to ensure the normal working of the instrument. · If the instrument shall be not used for a long period, please take out the battery to avoid the battery leakage to damage the device. · Please keep the instrument in a dry place and away from damp. · Do not dismantle, alter or rebuild the instrument by yourself. Overview Features of KJ61 series digital multimeter: ★ The casing and protective sheath are made by antiflaming engineering plastic and equipped with magnetic hook. ★Sound alarm will be automatically activated if the meter pen is inserted wrongly; the corresponding relation is the same to mechanical baffle. ★NCV (non-contact voltage test) is attached for free.
KJ6101
KJ6105
KJ6108
KJ6103
2V 20V 200V 1000V Range(6103) 200mV
Max. Value Auto Power Off HOLD DC voltage AC voltage DC Current AC Current Resistance Capacitance Frequency Temperature (℃ ) Diode test hFE Circuit Break-make test Battery test (Analog overload) Backlight
1999 ▲ ▲ ▲ ▲ ▲ ▲
▲ ▲
1999 ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲
1999 ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲
▲ ▲ ▲
19999 ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲
▲ ▲
Description of symbols Safety prompt
Fuse Compliance with EU standard
High voltage danger Doubleinsulating
Earth
AC
Buzzer
DC
Battery
AC/DC
Diode
Technical Indexes Degree of accuracy: ± (a%×reading + word count) Temperature (accuracy can be guaranteed at this temp): (23±5℃); Relative humidity <70%. DC Voltage Range
Accuracy
Resolution
200mV
±(0.5%+2)
100μV
±(1%+2) Accuracy
1mV 10mV 100mV 1V Resolution
±(0.05%+3)
10μV
±(0.5%+2)
2V 100μV 20V ±(0.05%+5) 1mV 200V 10mV 1000V ±(0.5%+2) 100mV Input impedance: All ranges 10MΩ Over-load protection: the ranges of 200mV are DC 300V, other ranges are DC 1000V;
AC Voltage Range Accuracy Resolution 200mV 100μV 2V 1mV ±(1%+2) 20V 10mV 200V 100mV 750V ±(1.5%+5) 1V Range(6103) Accuracy Resolution 2V 100μV 20V ±(1%+2) 1mV 200V 10mV 750V ±(1.5%+2) 100mV Input impedance: All ranges 10MΩ(KJ6103 2V ranges 1 MΩ) Over-load protection: the ranges of 200mV are AC 220V, other ranges are AC 750V; Frequency range: 40Hz—400Hz(the ranges of 750V are:40Hz—100Hz) Display: average value response (effective value calibration for sine wave)
DC Current Range 200μA 2mA 20mA 200mA 10A Range(6103) 2mA 200mA 10A
Accuracy ±(1%+2) ±(1.5%+2) ±(2%+2) Accuracy ±(1.5%+2) ±(2%+2)
Resolution 0.1μA 1μA 10μA 0.1mA 10mA Resolution 0.1μA 10μA 1mA
AC Current Range Accuracy Resolution 200μA 0.1μA 2mA ±(1.5%+2) 1μA 20mA 10μA 200mA ±(2%+2) 0.1mA 10A ±(2%+5) 10mA Range(6103) Accuracy Resolution 2mA 0.1μA ±(2%+2) 200mA 10μA 10A ±(2%+5) 1mA Protection: 250mA fuse for other current range Frequency range: 40Hz—400Hz Display: average value response (effective value calibration for sine wave) Resistance Range Accuracy Resolution 200Ω ±(1%+4) 0.1Ω 2KΩ 1Ω 20K 10Ω ±(1%+2) 200K 100Ω 2MΩ 1KΩ 20MΩ ±(1%+8) 10KΩ 200MΩ ±(2%+10) 100KΩ Range(6103) Accuracy Resolution 200Ω ±(1%+4) 0.01Ω 2KΩ 0.1Ω 20K 1Ω ±(1%+2) 200K 10Ω 2MΩ 100Ω 20MΩ ±(1%+8) 1KΩ 200MΩ ±(2%+10) 10KΩ Limit protection capability when voltage is input wrongly: AC 220V; Capacitance Range Accuracy Resolution 20nF 10pF 200nF 0.1nF ±(3%+2) 2μF 1nF 20μF 10nF 200μF ±(5%+5) 0.1μF Limit protection capability when voltage is input wrongly: AC 220V; Frequency Range Accuracy Resolution
20kHz ±(0.5%+2) Input signal range: 700mVpp20Vp-p Limit input: AC220V Diode Measurement Reverse Display voltage Approx 3V Forward voltage drop of diode
10Hz
Forward current Approx 1mA
Limit protection capability when voltage is input wrongly: AC 220V;
Break-make Measurement of Speaker Open circuit Threshold value Output current voltage of alarm sound Approx 3V Approx 50Ω Approx 1.0 mA Limit protection capability when voltage is input wrongly: AC 220V; Battery Voltage Measurement Range(6101) Load Input end current protection 1.5V battery Approx mA fuse 150 mA 9V battery Approx 8 PTC thermistor mA Description: 1. The difference of battery voltage measurement and general voltage measurement is that, the general voltage measurement has high input impedance; and it’s only used to measure the open circuit voltage of battery and check the battery and electricity conditions, it must be measured under load state. 2. mA jack input is employed for 1.5V battery has low voltage but big analog load current. mA fuse shall be blown if the voltage is input wrongly. 3. 9V battery employs voltage jack input, limit protection capability: AC220V. Temperature Range(6108) Accuracy Resolution -40℃~1000℃ ±(3%+5) 1℃ Temperature transducer: K type thermocouple (nickel-chromium—nickel-aluminum) Open circuit characteristics: display ambient temperature Limit protection capability when voltage is input wrongly: AC 220V;
Temperature conversion: Conversion relation of Celsius degree (℃) and Fahrenheit degree (℉): ℃=5(℉-32)/9 Comparison table of Celsius degree and Fahrenheit degree: Celsius -40 -20 0 20 25 30 degree Fahrenheit -40 -4 32 68 77 86 degree Celsius 35 40 100 300 500 1000 degree Fahrenheit 95 104 212 572 932 1832 degree