Transcript
MEA1060-Inv-(BC)-PGP Perfusion Ground Plate for MEA1060-Inv and MEA1060-Inv-BC Amplifier Adapter to use perforated MEAs (pMEA) together with MEA1060-Inv and MEA1060-Inv-BC amplifiers respectively.
O-Ring O-Ring-PGP
Perfusion In Perfusion Out
The external diameter of the cannula Perfusion In and Perfusion Out is 0.9 mm. It is recommended to connect flexible hoses with an internal diameter of 0.8 mm that enclose the cannula leakprooved. Please do not clean the plexiglass of the perfusion ground plate with alcohol! The plexiglass will get refractory. Please use distilled water only. Scope of delivery: 1 MEA1060-Inv-(BC)-PGP; 3 silicon sealing rings (O-Ring-PGP 40 x 1; available in sets of four rings). 1 Set of screws. Installation: Replace the standard ground plate of the MEA1060-Inv or MEA1060-Inv-BC amplifier by the MEA1060-Inv / MEA1060-Inv-BC perfusion ground plate. Please note that the MEA1060-Inv-PGP, respectively MEA1060-Inv-BC-PGP does NOT have a heating element.
Multi Channel Systems MCS GmbH Aspenhaustrasse 21 72770 Reutlingen Germany
Fon +49-7121-9 09 25- 0 Fax +49-7121-9 09 25-11
© 2012 Multi Channel Systems MCS GmbH
[email protected] www.multichannelsystems.com
Product information is subject to change without notice.