Preview only show first 10 pages with watermark. For full document please download

N5242 90012

   EMBED


Share

Transcript

Agilent Technologies N5241/42AS Option H30 and H31 User’s and Service Guide Agilent Technologies N5241/42AS Option H30 and H31 User’s and Service Guide Reference the PNA Series Network Analyzer On-line Help System, N5242A Service Guide: N5242-90001 and PNA Installation and Quick Start Guide: E8356-90001 Manufacturing Part Number: N5242-90012 Printed in USA: December 2010 Supersede October 2010 © Copyright 2010 Agilent Technologies, Inc. All rights reserved. Warranty Statement THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL. DFARS/Restricted Rights Notice If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data. ii Safety Notes The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before performing any of the procedures in this document. WARNING Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met. CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met. Definitions • Specifications describe the performance of parameters covered by the product warranty (temperature –0 to 55 °C, unless otherwise noted.) • Typical describes additional product performance information that is not covered by the product warranty. It is performance beyond specification that 80% of the units exhibit with a 95% confidence level over the temperature range 20 to 30 °C. Typical performance does not include measurement uncertainty. • Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty. • Characteristic Performance describes performance parameter that the product is expected to meet before it leaves the factory, but is not verified in the field and is not covered by the product warranty. A characteristic includes the same guard bands as a specification. iii iv Contents N5241/42AS Option H30/H31 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Rear Panel Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Operating the PNA-X as a Downconverter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Typeface Key Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Downconverting the Signal Using a Test Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 2-Port PNA-X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 4-Port PNA-X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Downconverting the Signal Using a External Receiver Input . . . . . . . . . . . . . . . . . . . . . . . . 7 Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 IF Output Verification Using N5241/42AS Internal Receivers . . . . . . . . . . . . . . . . . . . . . . . 7 2-Port Internal Receiver Measurement (B and D/R2 IF Outputs) . . . . . . . . . . . . . . . . . . . . . 9 4-Port Internal Receiver Measurement (B and D/R2 IF Outputs) . . . . . . . . . . . . . . . . . . . . 12 2-Port Internal Receiver Measurement (A and C/R1 IF Outputs) . . . . . . . . . . . . . . . . . . . . 15 4-Port Internal Receiver Measurement (A and C/R1 IF Outputs) . . . . . . . . . . . . . . . . . . . . 18 4-Port Internal Receiver Measurement (R IF Output, Opt H30 only) . . . . . . . . . . . . . . . . . 21 Measuring the IF Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Measuring Conversion Loss Using a Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 PNA-X Option 200 (2-Port) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 PNA-X Option 400 (4-Port) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Measuring Conversion Loss Using a Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . 25 PNA-X Option 200 (2-Port) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 PNA-X Option 400 (4-Port) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Measuring Bandwidth Using a Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 N5241/42AS Block Diagrams. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Safety and Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Before Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Connector Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Statement of Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 General Safety Considerations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Instrument Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Battery Collection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Compliance with German Noise Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 EMC Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Electrostatic Discharge Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Agilent Support, Services, and Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Service and Support Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Contacting Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Shipping Your Analyzer to Agilent for Service or Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Contents-1 Contents Contents-2 N5241/42AS Option H30/H31 User’s and Service Guide N5242-90012 1 N5241/42AS Option H30/H31 Introduction Introduction This document describes how to use and service the N5241AS and N5242AS Option H30 and H31 (IF Outputs) in the N5241/42AS PNA-X Microwave Network Analyzer. The Option H31 provides a wider IF bandwidth for the IF Outputs. The N5241/42AS is a N5242A PNA-X with custom options in addition to the standard options available. Refer to the standard N5242A User and Service Guide for further information. Description The N5241/42AS Option H30 and H31 provide IF Outputs on the rear panel and require Option 020 (External IF INPUTS) to verify the Option H30 and H31. This enables the PNA-X to be used as a downconverter. The Option H30 IF Outputs, A, B, C/R1, D/R2 and R are the result of the RF Input signal at the PNA-X test port, or RCVR IN mixing with the internal source. The R IF Output is only available on the 4-Port N5241/42AS Option H30. The Option H31 IF Outputs, A, B, C/R1 and D/R2 are direct from the mixer with an IF splitter/amplifier assembly installed to provide more gain and bandwidth over the RF frequency range of 3 to 300 MHz. R IF Output is not available with Option H31. The IF Outputs are terminated with a 50 Ohm load when not in use to reduce radiated emissions and protect the IF Outputs. 2 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Specifications Specifications Table 1 IF Output Characteristics Description 2-Port 4-Port Option H30 IF Output Characteristics Input Frequency Range 1 10 MHz to 26.5 GHz 10 MHz to 26.5 GHz Center IF Frequency Output2 PNA-X Freq − RF Input Freq + IF Offset = IF Freq Output PNA-X Freq − RF Input Freq + IF Offset = IF Freq Output IF Frequency Bandwidth 40 MHz max 40 MHz max IF Output (A, B, Conversion Loss) −13 dB (± 5 dB) −13 dB (± 5 dB) IF Output (C/R1, D/R2 Conversion loss) −25 dB (± 5 dB) −13 dB (± 5 dB) IF Output (REF) Conversion loss n/a −25 dB ((± 5 dB) Compression Input Level +10 dBm at 1 GHz RF Input +10 dBm at 1 GHz RF Input IF Output Impedance 50 Ohm 50 Ohm Option H31 IF Output Characteristics Input Frequency Range 10 MHz to 26.5 GHz 10 MHz to 26.5 GHz Center IF Frequency Output2 RF Input Freq + IF Offset − PNA-X Freq = IF Freq Output RF Input Freq + IF Offset − PNA-X Freq = IF Freq Output IF Frequency Bandwidth 3 dB, 200 MHz 6 dB, 300 MHz 3 dB, 200 MHz 6 dB, 300 MHz IF Output (A, B, Conversion Loss) −10 dB (± 5 dB) −10 dB (± 5 dB) IF Output (C/R1, D/R2 Conversion loss) −10 dB (± 5 dB) −10 dB (± 5 dB) Compression Input Level +10 dBm at 1 GHz RF Input +10 dBm at 1 GHz RF Input IF Output Impedance 50 Ohm 50 Ohm 1. The coupler in the test ports causes high conversion loss for frequency < 500 MHz. Use RCVR Input for best conversion loss. 2. Refer to Equation 1, “IF Offset Frequency” on Page 5. User’s and Service Guide N5242-90012 3 N5241/42AS Option H30/H31 Rear Panel Features Rear Panel Features Figure 1 Rear Panel (Multiport Test Set) Figure 2 Rear Panel Labeling External IF Outputs D/R2 C/R1 External IF Inputs 4 R D/R2 C/R1 R B B A A External IF Inputs User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Operating the PNA-X as a Downconverter Operating the PNA-X as a Downconverter The PNA-X test ports, or front panel receiver inputs (with jumper RCVR IN to CPLR ARM removed) can be used to downconvert an RF signal to a lower IF frequency. The downconverted IF signals are the difference between the RF signal frequency at the test port you are using, and the tuned frequency of the PNA-X. The amplitude of the IF Output signals will increase or decrease if the RF Input signal amplitude changes. The IF Output frequency will increase or decrease with the input frequency applied to the test port or RCVR Input as shown in Equation 1. To simplify the equation set the IF Offset to a usable integer, for example 1 MHz IF would replace the default IF Offset (7.60563380282 MHz). Reference the following commands: 1. sens:if:freq 1e6 (sets the desired IF frequency, in this example 1 MHz) 2. sens:if:freq:auto off (activates the selected IF frequency) Equation 1 IF Offset Frequency IF = ( RFSignalFreq + IFOffset ) – PNAXFreq CAUTION Do not apply DC or RF on the IF Outputs, the IF Outputs are DC coupled. Typeface Key Conventions The following key conventions are used throughout this document. • [HARDKEYS] are labeled front panel keys • SOFTKEYS are unlabeled key whose function is indicated on the instrument display. User’s and Service Guide N5242-90012 5 N5241/42AS Option H30/H31 Operating the PNA-X as a Downconverter Downconverting the Signal Using a Test Port Port 1 can be used to downconvert an RF Input signal using the PNA-X. Selecting S12 will enable the RF applied signal at Port 1 to be downconverted to the IF Offset ± the difference of the RF signal and the tuned frequency of the PNA-X. Refer to Equation 1, “IF Offset Frequency” on Page 5. CAUTION Do not exceed the maximum input power of the test ports or RCVR Inputs, or damage to the PNA-X will occur. The maximum input power on the PNA-X test port is 30 dBm (1 W) or 40 VDC. 1. Connect the RF signal to be downconverted to Port 1. 2. Select [Sweep] > Sweep Type > CW Time. 3. Input the frequency desired to downconvert your RF signal. Note: The IF signal will decrease approximately 3 dB when your RF signal is 40 MHz more than the PNA-X frequency. Example: Select [Freq] > [1 GHz]. 4. Set the PNA-X to measure S12. Select [Meas] > S12. This will set Port 1 to be the receiver port and Port 2 will be a source port. 2-Port PNA-X • Port 1 — Set the PNA-X to measure S12. A and D/R2 IF Outputs will have the downconverted IF signal. • Port 2 — Set the PNA-X to measure S21. B and C/R1 IF Outputs will have the downconverted IF signal. • There will be no signal on the IF Output R in a 2-Port PNA-X. 4-Port PNA-X • Port 1 — Set the PNA-X to measure S12. A and R IF (Option H30 only) Outputs will have the downconverted IF signal. • Port 2 — Set the PNA-X to measure S21. B and R IF (Option H30 only) Outputs will have the downconverted IF signal. • Port 3 — Set the PNA-X to measure S31. C/R1 and R IF (Option H30 only) Outputs will have the downconverted IF signal. • Port 4 — Set the PNA-X to measure S41. D/R2 and R IF (Option H30 only) Outputs will have the downconverted IF signal. 6 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Performance Verification Downconverting the Signal Using a External Receiver Input The external RCVR Input (A, B, C or D) bypasses the internal test port coupler, resulting in less loss of the RF signal to be downconverted. It is recommended that you use the external RCVR for input frequency < 500 MHz. The coupler in the test ports causes high conversion loss for frequencies < 500 MHz. To use the receiver input, remove the jumper (RCVR IN to CPLR ARM) and connect the RF signal to the RCVR IN female SMA on the front panel. Removing the jumper will disable the front panel test port. CAUTION Ensure the receiver input power level is < +15 dBm with 0 VDC if you are using RCVR A, B,C or D IN or damage may occur. Performance Verification The operational check allows the user to verify the IF Outputs are functioning. The N5241/42AS Option H30 and H31 can be verified in two different ways. Connect the IF Outputs to the IF Inputs and verify the trace on the PNA-X, or measure the IF Output using a separate instrument such as a power meter or spectrum analyzer. IF Output Verification Using N5241/42AS Internal Receivers This procedure verifies the Option H30 and H31 IF Outputs using Option 020 (External IF Inputs) by connecting the IF Outputs to the IF Inputs. The IF Outputs will be measured by the internal N5241/42AS receivers. • A and C/R1 external IF Inputs measure B and D/R2 IF Outputs. • B and D/R2 external IF Inputs measure A and C/R1 IF Outputs. 1. Connect the cables from the IF Outputs to the IF Inputs on the rear panel, using four 1 foot SMA cables as shown in Table 3 (Agilent RF cable part number 5062-6662 or equivalent). Table 3 Rear Panel RF Connections IF OUT IF IN A B C/R1 D/R2 B A D/R2 C/R1 User’s and Service Guide N5242-90012 7 N5241/42AS Option H30/H31 Performance Verification Figure 3 Cable Connections (1) C/R1 OUT D/R2 IN Figure 4 Cable Connections (2) C/R1 IN 8 A OUT B IN D/R2 OUT R OUT B OUT A IN User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Performance Verification 2-Port Internal Receiver Measurement (B and D/R2 IF Outputs) The procedure measures B and D/R2 IF Outputs using A and R1 IF Inputs. 1. Press [Preset]. 2. Select [Power] > −5 dBm. 3. Select [Sweep] > Sweep Type > CW Time. 4. Select [Freq] > [1 GHz]. 5. Select [Avg] > IF Bandwidth > [1 kHz]. 6. Select [Channel] > Hardware Setup > IF Config. 7. Select External: A and External: R1 > OK. Figure 5 IF Config - Channel 1 (2-Port) User’s and Service Guide N5242-90012 9 N5241/42AS Option H30/H31 Performance Verification 8. Select Trace/Chan > Trace > Delete Trace > Trace 1 > OK. 9. Select Trace/Chan > Trace > New Trace. Select the Receivers tab. 10.Select Activate for Numerator A and R1, Denominator 1.0 and Source Port, Port 2 > OK. Figure 6 10 Measure B and D/R2 IF Output with A and R1 Receivers (2-Port) User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Performance Verification 11.Select Tr 2 (A,2) > [Marker]. To measure B Output. 12.Select Tr 3 (R1,2) > [Marker]. To measure D/R2 Output. • Tr 2 (A,2) is the B IF Output level. Verify the power level is –18 dBm (± 5 dBm) for Option H30, or –13 dBm (± 5 dBm) for Option H31. • Tr 3 (R1,2) is the D/R2 IF Output. Verify the power level is –28 dBm (± 5 dBm) for Option H30, or –13 dBm (± 5 dBm) for Option H31. Figure 7 Option H30 B and D/R2 Output Measurement (2-Port) Figure 8 Option H31 B and D/R2 Output Measurement (2-Port) User’s and Service Guide N5242-90012 11 N5241/42AS Option H30/H31 Performance Verification 4-Port Internal Receiver Measurement (B and D/R2 IF Outputs) The procedure measures B and D/R2 IF Outputs using A and C IF Inputs. 1. Press [Preset]. 2. Select [Power] > −5 dBm. 3. Select [Sweep] > Sweep Type > CW Time. 4. Select [Freq] > [1 GHz]. 5. Select [Avg] > IF Bandwidth > [1 kHz]. 6. Select [Channel] > Hardware Setup > IF Config. 7. Select External: A and External: C > OK. Figure 9 12 IF Path Config - Channel 1 A and C (4-Port) User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Performance Verification 8. Select Trace/Chan > Trace > Delete Trace > Trace 1 > OK. 9. Select Trace/Chan > Trace > New Trace. Select the Receivers tab. 10.Select Activate for Numerator A, Denominator 1.0 and Source Port, Port 2. Select Activate for Numerator C, Denominator 1.0 and Source Port, Port 4 > OK. Figure 10 Measure B and D/R2 IF Outputs with A and C Receivers (4-Port) User’s and Service Guide N5242-90012 13 N5241/42AS Option H30/H31 Performance Verification 11.Select Tr 1 (A,2) > [Marker]. 12.Select Tr 2 (C,4) > [Marker]. • Tr 1 (A,2) is the B IF Output level.Verify the power level is –20 dBm (± 5 dBm) for Option H30, or –13 dBm (± 5 dBm) for Option H31. • Tr 2 (C,4) is the D/R2 IF Output. Verify the power level is –20 dBm (± 5 dBm) for Option H30, or –13 dBm (± 5 dBm) for Option H31. Figure 11 Option H30 B and D Output Measurement (4-Port) Figure 12 Option H31 B and D Output Measurement (4-Port) 14 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Performance Verification 2-Port Internal Receiver Measurement (A and C/R1 IF Outputs) The procedure measures A and R1 IF Outputs using B and R2 IF Inputs. 1. Press [Preset]. 2. Select [Power] > −5 dBm. 3. Select [Sweep] > Sweep Type > CW Time. 4. Select [Freq] > [1 GHz]. 5. Select [Avg] > IF Bandwidth > [1 kHz]. 6. Select [Channel] > Hardware Setup > IF Config. 7. Select External: B and External: R2 > OK. Figure 13 IF Path Config - B and R2 (2-Port) User’s and Service Guide N5242-90012 15 N5241/42AS Option H30/H31 Performance Verification 8. Select Trace/Chan > Trace > Delete Trace > Trace 1 and Trace 2 > OK. 9. Select Trace/Chan > Trace > New Trace. Select the Receivers tab. 10.Select Activate for Numerator B and R2, Denominator 1.0 and Source Port, Port 1 > OK. Figure 14 16 Measure A and C/R1 Outputs with B and R2 Receivers (2-Port) User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Performance Verification 11.Select Tr 2 (B,1) > [Marker]. To measure A IF Output. 12.Select Tr 4 (R2,1) > [Marker]. To measure C/R1 IF Output. • Tr 2 (B,1) is the A IF Output level. Verify the power level is –20 dBm (± 5 dBm) for Option H30, or –13 dBm (± 5 dBm) for Option H31. • Tr 4 (R2,1) is the C/R1 IF Output. Verify the power level is –30 dBm (± 5 dBm) for Option H30, or –13 dBm (± 5 dBm) for Option H31. Figure 15 Option H30 A and C/R1 Measurement (2 Port) Figure 16 Option H31 A and C/R1 Measurement (2 Port) User’s and Service Guide N5242-90012 17 N5241/42AS Option H30/H31 Performance Verification 4-Port Internal Receiver Measurement (A and C/R1 IF Outputs) The procedure measures the A and C/R1 IF Outputs using the B and D IF Inputs. 1. Press [Preset]. 2. Select [Power] > −5 dBm. 3. Select [Sweep] > Sweep Type > CW Time. 4. Select [Freq] > [1 GHz]. 5. Select [Avg] > IF Bandwidth > [1 kHz]. 6. Select [Channel] > Hardware Setup > IF Config. 7. Select External: B and External: D > OK. Figure 17 18 IF Config - B and D (4-Port) User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Performance Verification 8. Select Trace/Chan > Trace > Delete Trace > Trace 1 and Trace 2 > OK. 9. Select Trace/Chan > Trace > New Trace. Select the Receivers tab. 10.Select Activate for Numerator B, Denominator 1.0 and Source Port, Port 1. Select Activate for Numerator D, Denominator 1.0 and Source Port, Port 3 > OK. Figure 18 Measure A and C/R1 Using B and D Receivers (4-Port) User’s and Service Guide N5242-90012 19 N5241/42AS Option H30/H31 Performance Verification 11.Select Tr 1 (B,1) > [Marker]. 12.Select Tr 2 (D,3) > [Marker]. • Tr 1 (B,1) is the A IF Output level. Verify the power level is –20 dBm (± 5 dBm) for Option H30, or –13 dBm (± 5 dBm) for Option H31. • Tr 2 (D,3) is the C/R1 IF Output level. Verify the power level is –20 dBm (± 5 dBm) for Option H30, or –13 dBm (± 5 dBm) for Option H31. Figure 19 Option H30 A and C/R1 Measurement (4-Port) Figure 20 Option H31 A and C/R1 Measurement (4-Port) 20 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Performance Verification 4-Port Internal Receiver Measurement (R IF Output, Opt H30 only) 1. Remove the cable from B IF Output to A IF Input. 2. Select [Power] > −5 dBm. 3. Connect R IF Output to A IF Input using RF cable (5062-6662) or equivalent. 4. Press [Preset]. 5. Select [Sweep] > Sweep Type > CW Time. 6. Select [Freq] > [1 GHz]. 7. Select [Avg] > IF Bandwidth > [1 kHz]. 8. Select [Channel] > Hardware Setup > IF Config. 9. Select External: A and External: R1 > OK. Figure 21 IF Config - A and R1 (4-Port) User’s and Service Guide N5242-90012 21 N5241/42AS Option H30/H31 Performance Verification 10.Select Trace/Chan > Trace > Delete Trace > Trace 1 > OK. 11.Select Trace/Chan > Trace > New Trace. Select the Receivers tab. 12.Select Activate for Numerator A, Denominator 1.0 and Source Port, Port 2 > OK. Figure 22 Measure R Receiver Trace (4-Port) 13.Select Tr 1 (A,2) > [Marker]. • Tr 1 (A,2) is the R IF Output. Verify the power level is –30 dBm (± 5 dBm). Figure 23 22 R Measurement (4-Port) User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Measuring the IF Outputs Measuring the IF Outputs The IF Outputs can be verified using a power meter with a sensor that can measure a signal at the IF Offset frequency selected, such as the E4413A and sensor 8482A (or equivalent). A spectrum analyzer can also be used, such as the E4411B ESA-L or N9010A EXA. Measuring Conversion Loss Using a Power Meter Values are dependent on the accuracy of the power meter and sensor. It is recommended that the power meter be zeroed and the reference power level be verified before beginning. PNA-X Option 200 (2-Port) 1. Press [Preset]. 2. Select [Sweep] > Sweep Type > CW Time > [Freq] > CW > [1 GHz]. 3. Select [Power] > [0 dBm]. 4. Select [Meas] > S11. 5. Connect the power sensor to A IF Output. 6. Verify the output power. Option H30 = −15 dBm (± 5 dBm) Option H31 = −10 dBm (± 5 dBm) 7. Connect the power sensor to C/R1 IF Output. 8. Verify the output power. Option H30 = −25 dBm (± 5 dBm) Option H31 = −10 dBm (± 5 dBm) 9. Select [Meas] > S-Parameters > S22 > OK. 10.Connect the power sensor to B IF Output. 11.Verify the output power. Option H30 = −15 dBm (± 5 dBm) Option H31 = −10 dBm (± 5 dBm) 12.Connect the power sensor to D/R2 IF Output. 13.Verify the output power. Option H30 = −25 dBm (± 5 dBm) Option H31 = −10 dBm (± 5 dBm) User’s and Service Guide N5242-90012 23 N5241/42AS Option H30/H31 Measuring the IF Outputs PNA-X Option 400 (4-Port) 1. Press [Preset]. 2. Select [Sweep] > Sweep Type > CW Time > [Freq] > CW > [1 GHz]. 3. Select [Power] > [0 dBm]. 4. Select [Meas] > S11. 5. Connect the power sensor to A IF Output. 6. Verify the output power. Option H30 = −15 dBm (± 5 dBm) Option H31 = −10 dBm (± 5 dBm) 7. Select [Meas] > S22. 8. Connect the power sensor to B IF Output. 9. Verify the output power. Option H30 = −15 dBm (± 5 dBm) Option H31 = −10 dBm (± 5 dBm) 10.Select [Meas] > S-Parameters > S33 > OK. 11.Connect the power sensor to C/R1 IF Output. 12.Verify the output power. Option H30 = −15 dBm (± 5 dBm) Option H31 = −10 dBm (± 5 dBm) 13.Select [Meas] > S-Parameters > S44 > OK 14.Connect the power sensor to D/R2 IF Output. 15.Verify the output power. Option H30 = −15 dBm (± 5 dBm) Option H31 = −10 dBm (± 5 dBm) 16.Connect the power sensor to R IF Output (Option H30 only). 17.Verify the output power. Option H30 = −25 dBm (± 5 dBm) 24 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Measuring the IF Outputs Measuring Conversion Loss Using a Spectrum Analyzer Values are dependent on the accuracy of the spectrum analyzer and cable loss. PNA-X Option 200 (2-Port) 1. Press [Preset]. 2. Select [Sweep] > Sweep Type > CW Time > [Freq] > CW > [1 GHz]. 3. Select [Meas] > S11. 4. Select [Power] > [0 dBm]. Spectrum Analyzer 1. Press [Preset]. 2. Select [Amplitude] > Log Scale > [10 dB/div]. 3. Select [Frequency] > Center Frequency, and enter the frequency that you have selected as the IF Offset. Refer to Equation 1, “IF Offset Frequency” on Page 5. 4. Select Span > [10 MHz]. 5. Select [Reference Level] > [0 dBm]. 6. Connect the spectrum analyzer to the PNA-X A IF Output on the rear panel. 7. Select [Marker]. Place the marker on the IF Offset signal or use the search feature. 8. Verify the output power. Option H30 = −15 dBm (± 5 dBm) Option H31 = −10 dBm (± 5 dBm) 9. Connect the spectrum analyzer to the PNA-X C/R1 IF Output on the rear panel. 10.Select [Marker]. Place the marker on the IF Offset signal or use the search feature. 11.Verify the output power as in step 8. 12.Select [Meas] > S22 on the PNA-X. 13.Connect the spectrum analyzer to the PNA-X B IF Output on the rear panel. 14.Select [Marker]. Place the marker on the IF Offset signal or use the search feature. 15.Verify the output power as in step 8. 16.Connect the spectrum analyzer to the PNA-X D/R2 IF Output on the rear panel. 17.Select [Marker]. Place the marker on the IF Offset signal or use the search feature. 18.Verify the output power as in step 8. User’s and Service Guide N5242-90012 25 N5241/42AS Option H30/H31 Measuring the IF Outputs PNA-X Option 400 (4-Port) 1. Press [Preset]. 2. Select [Sweep] > Sweep Type > CW Time > [Freq] > CW > [1 GHz]. 3. Select [Meas] > S11. 4. Select [Power] > [0 dBm]. Spectrum Analyzer 1. Press [Preset]. 2. Select [Amplitude] > Log Scale > [10 dB/div]. 3. Select [Frequency] > Center Frequency, and enter the frequency that you have selected as the IF Offset. Refer to Equation 1, “IF Offset Frequency” on Page 5. 4. Select Span > [10 MHz]. 5. Select [Amplitude] > [Reference Level] > [0 dBm]. 6. Connect the spectrum analyzer to the PNA-X A IF Output on the rear panel. 7. Select [Marker]. Place the marker on the IF Offset signal or use the search feature. Adjust Video BW and/or Resolution BW as needed. 8. Verify the output power. Option H30 = −15 dBm (± 5 dBm) Option H31 = −10 dBm (± 5 dBm) 9. Select [Meas] > S22 on the PNA-X. 10.Connect the spectrum analyzer to the PNA-X B IF Output on the rear panel. 11.Select [Marker]. Place the marker on the IF Offset signal or use the search feature. 12.Verify the output power as in step 8. 13.Select [Meas] > S-Parameters > S33 > OK on the PNA-X. 14.Connect the spectrum analyzer to the PNA-X C/R1 IF Output on the rear panel. 15.Select [Marker]. Place the marker on the IF Offset signal or use the search feature. 16.Verify the output power as in step 8. 17.Select [Meas] > S-Parameters > S44 > OK on the PNA-X. 18.Connect the spectrum analyzer to the PNA-X D/R2 IF Output on the rear panel. 19.Select [Marker]. Place the marker on the IF Offset signal or use the search feature. 20.Verify the output power as in step 8. 21.Connect the spectrum analyzer to the PNA-X R IF Output on the rear panel (Option H30 only). 22.Select [Marker]. Place the marker on the IF Offset signal or use the search feature. 23.Verify the output power. Option H30 = −25 dBm (± 5 dBm) 26 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Measuring the IF Outputs Measuring Bandwidth Using a Spectrum Analyzer On the PNA-X. 1. Select [Measure] > S12. Port 1 is now the receiver. 2. Select [Sweep] > Sweep Type > CW Time > [Freq] > CW > [1 GHz]. On the Signal Source (ESG, MXG, or PSG). 3. Connect to Port 1 on the PNA-X. 4. Press [Preset]. 5. Select [Frequency] > [1.01 GHz] > [RF On]. 6. Select [Incr Set] > [5 MHz]. 7. Select [Amplitude] > [0 dBm]. On the Spectrum Analyzer (MXA, ESA, or PSA). 8. Connect the Input to A IF Output on the rear panel. 9. Press [Preset]. 10.Select [Frequency] > Start Freq > [0 MHz] and Stop Freq > [400 MHz]. 11.Select [Marker] > Marker 1. Place the marker on the 2.4 MHz signal or use the search feature. 12.Select [Amplitude] > Scale/Div > [1 dB]. 13.Select [Amplitude] > Ref Level > [−5 dB]. Set the signal to the top graticule on the display manually, or select [Marker] > Mkr ->Ref Lvl. 14.Select [View/Trace] > Max Hold. 15.Increase the frequency, in 5 MHz s, on the signal source to [1.410 GHz] tracing the bandwidth of the IF Output. 16.Select [Marker] > Marker 1 > Delta with the marker on the 2.4 MHz signal. 17.Marker 1 should read approximately 0 dBm. Using the RPG knob, move the Marker to read −3 dBm as shown in Figure 24 on page 28. 18.Using the RPG knob, move the Marker to read −6 dBm as shown in Figure 25 on page 28. 19.Connected the source to Port 2, and the spectrum analyzer connected to B IF Output. 20.Select [Measure] > S21. Port 2 is now the receiver. 21.Repeat step 2 through step 18. NOTE The step size of the source should not be greater than the resolution bandwidth of the spectrum analyzer your are using, or a measurement error may occur. User’s and Service Guide N5242-90012 27 N5241/42AS Option H30/H31 Measuring the IF Outputs Figure 24 3 dB Bandwidth Figure 25 6 dB Bandwidth 28 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Theory of Operation Theory of Operation The N5241/42AS Option H30 provides IF Output connectors on the rear panel that connect from the IF Mux assembly (N5240-60045) IF Outputs A (P2), B (P202), C (P602), D (P802) and R (R402). The test port or external receiver input is downconverted to an IF frequency (Input Freq + IF Offset - PNA set frequency). Refer to Figure 26, “N5241/42AS Option H30 RF Block Diagram.” The N5241/42AS Option H31 provides IF Output connectors on the rear panel. The IF splitter/amplifier assemblies (N5242-63160) connect the receiver outputs A, B, C and D from the mix brick assembly (5087-7727) to the IF Output connectors on the rear panel. R1, R2, R3 and R4 signals are also routed through the IF splitter/amplifier assemblies providing signal conditioning to match receivers A, B, C and D, required for IF gain calibration to operate correctly. The N5241/42AS Option H31 provides a wider IF Output bandwidth than Option H30, by bypassing the IF Mux assembly filtering. The IF splitter/amplifier assembly splits the power of each IF path. The first assembly splits the 2-Port, or 4-Port PNA-X outputs (A, R1, B and R2) from the mixerbrick. A second assembly splits C, R3, D and R4 for a 4-Port PNA-X only. In addition the splitter/amplifier assemblies amplify the IF signals to correct for the loss of the splitter, and provides isolation for each path. An additional 5 dB of gain is provided for the rear panel IF Outputs. The IF signals are routed from the IF splitter/amplifier assembly to the IF Mux assembly (N5240-63045) Inputs A (P1), B (P201), C (601) and D(801), and to the rear panel A, B, C and D IF Outputs. The REF IF signals are routed to the IF Mux assembly Inputs R1 (411), R2 (412), R3 (413) and R4 (R414). A power supply ribbon cable from J111 on the IF Mux assembly provides DC +15 V for the IF splitter/amplifier assemblies. User’s and Service Guide N5242-90012 29 N5241/42AS Option H30/H31 N5241/42AS Block Diagrams N5241/42AS Block Diagrams Figure 26 N5241/42AS Option H30 RF Block Diagram IF Mux Assembly Detail for Receiver A and R1 1 of 4 Mixer sets in the Mixerbrick (5087-7727) A Ext IF OUT Rear Panel A OUT 40 MHz LPF 50 Ohm A IN x2 Test Mixer 50 Ohm A IN Front Panel A Ext IF IN Mux ESD Diodes A, B, C, D Mixers 50 Ohm LO Aux R4 Input B IN Source Test Port B IN B Ext IF OUT B Ext IF IN PNA-X Test Port 1 B OUT B IN B-Channel IF Switch Ref Mixer Aux R3 C IN C IN C Ext IF IN C Ext IF OUT C OUT C IN R1, 2, 3 & 4 Mixers C-Channel IF Switch Aux R2 D IN LO D IN D Ext IF OUT D Ext IF IN R1 IN D OUT D IN IF Mux Channel R D-Channel IF Switch Aux R1 R IN R IN R2 IN R Ext IF IN R Ext IF OUT R OUT R IN R3 IN R-Channel IF Switch DC – 10 MHz R Out is only functional on the 4-Port PNA-X R4 IN 30 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 N5241/42AS Block Diagrams Figure 27 N5241/42AS Option H31 RF Block Diagram 1 of 4 Mixer sets in the Mixerbrick (5087-7727) A Ext IF OUT IF Mux Assembly Detail for Receiver A and R1 Rear Panel A OUT 40 MHz LPF 50 Ohm Test Mixer A IN Front Panel x2 50 Ohm A IN A Ext IF IN A, B, C, D Mixers 50 Ohm Test Port PNA-X Test Port 1 Mux ESD Diodes LO Source 1 of 2 Splitter Amp Bd. A, B, R1, R2 and C, D, R3, R4 Aux R4 Input B IN B Ext IF IN From H31 Splitter Amp Bd. From H31 Splitter Amp Bd. B OUT B IN B-Channel IF Switch Aux R3 Ref Mixer C IN C Ext IF IN From H31 Splitter Amp Bd. R1, 2, 3 & 4 Mixers From H31 Splitter Amp Bd. C-Channel IF Switch Aux R2 LO D IN From H31 Splitter Amp Bd. D Ext IF IN R1 IN IF Mux Channel R C OUT C IN From H31 Splitter Amp Bd. D OUT D IN D-Channel IF Switch Aux R1 From H31 Splitter Amp Bd. R2 IN From H31 Splitter Amp Bd. R IN R Ext IF IN R IN From H31 Splitter Amp Bd. R3 IN R-Channel IF Switch DC – 10 MHz From H31 Splitter Amp Bd. R4 IN User’s and Service Guide N5242-90012 No rear panel R OUT on Option H31 31 N5241/42AS Option H30/H31 Replaceable Parts Replaceable Parts Table 4 Qty N5241/42AS Option H30 Replaceable Parts Agilent Part Number Description1 5 RF Connector, SMA Male straight 50 ohm 1810-0118 5 Cable Assembly, A06/A30 170 mm 8121-0149 1 Rear Panel (IF Output) N5242-00035 1 N5242A H30 and H31 User’s Guide N5242-90012 1. Refer to the standard N5242A Service Guide for additional part information. Table 5 N5241/42AS Option H31 Replaceable Parts 2-Port Qty 4-Port Qty Agilent Part Number 4 4 50 Ohm Load, SMA 1810-0118 4 4 Cable Assembly, A06/A30 400 mm 8121-1271 2 4 Cable A06/A06 365 mm 8120-5022 1 2 PCA IF Splitter Amp Assembly N5242-60160 1 1 Rear Panel (IF Output) N5242-00035 1 1 N5242A H30 and H31 User’s Guide N5242-90012 1 1 Bracket N5242-00039 2 4 Cable A06/A06 510 mm 8120-6153 2 4 50 Ohm Load, SMB 1250-0676 Description1 1. Refer to the standard N5242A Service Guide for additional part information. Table 6 N5241/42AS Option H30 Internal Rear Panel Connections From: IF MUX (N5240-60045) 32 To: Rear Panel IF Out SMA Connector P2 A P202 B P402 R P602 C/R1 P802 D/R2 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Replaceable Parts Table 7 Splitter Amp Assembly Connections From: Splitter Amp Assembly (N5242-60160) To: IF MUX Board (N5240-60045) Board 1: 2-Port P11 P1, A IN P71 P201, B IN P31 P601, C IN P51 P801, D IN Board 1: 4-Port P11 P1, A IN P71 P201, B IN P31 P411, R1 IN P51 P412, R2 IN Board 2: 4-Port only P11 P601, C/R1 P71 P801, D/R2 P31 P413, R3 IN P51 P414, R4 IN From: Splitter AMP Assembly (N5242-60160) To: Mixerbrick Board (5087-7727) Board 1 P10 A P40 B P20 R1 P30 R2 Board 2: 4-Port only P10 C/R1 P40 D/R2 P20 R3 P30 R4 User’s and Service Guide N5242-90012 33 N5241/42AS Option H30/H31 Replaceable Parts Table 7 Splitter Amp Assembly Connections From: Splitter Amp Assembly (N5242-60160) To: Rear Panel IF Out SMA Connector Board 1: 2-Port P21 A, P81 B P41 C/R1 P61 D/R2 Board 1: 4-Port P21 A P81 B P41, 50 Ohm SMB n/a P61, 50 Ohm SMB n/a Board 2: 4-Port 34 P21 C/R1 P81 D/R2 P41, 50 Ohm SMB n/a P61, 50 Ohm SMB n/a User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Replaceable Parts Troubleshooting If the IF Output test fails, refer to troubleshooting the IF MUX Assembly (N5240-60045), PNA-X RF or IF paths, in the N5242A Service Guide (N5242-90001). See the “N5241/42AS Block Diagrams” on page 30, and Table 6 on page 32 and Table 7 on page 33. The IF MUX J111 supplies +15 VDC to connector J1, pins 28, 29, 62 and 63 on the Splitter Amp board. The splitter amplifier assembly provides 6 dB of gain for the following RF paths: • P10 to P11 • P20 to P31 • P30 to P51 • P40 to P71 The splitter amplifier assembly provides 11 dB of gain for the following RF paths: • P10 to P21 • P20 to P41 • P30 to P61 • P40 to P81 Figure 28 Splitter Amp Board User’s and Service Guide N5242-90012 35 N5241/42AS Option H30/H31 Replaceable Parts Figure 29 IF MUX Board P1 P201 P414 P413 P411 P412 P601 P801 J111 36 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Safety and Regulatory Information Safety and Regulatory Information Introduction Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition. Before Applying Power Verify that the premises electrical supply is within the range of the instrument. The instrument has an autoranging power supply. WARNING To prevent electrical shock, disconnect the Agilent Technologies N5241/42AS Option H30 and H31 from mains electrical supply before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally. Connector Care and Cleaning If alcohol is used to clean the connectors, the power cord to the instrument must be removed. All cleaning should take place in a well ventilated area. Allow adequate time for the fumes to disperse and moist alcohol to evaporate prior to energizing the instrument. WARNING Keep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly closed container. It is extremely flammable. In case of fire, use alcohol foam, dry chemical, or carbon dioxide; water may be ineffective. Declaration of Conformity A copy of the Declaration of Conformity is available upon request, or a copy is available on the Agilent Technologies web site at http://regulations.corporate.agilent.com/DoC/search.htm Statement of Compliance This instrument has been designed and tested in accordance with CAN/CSA 22.2 No. 61010-1-04, UL Std No. 61010-1 (Second Edition), and IEC 61010-1 (Second Edition). User’s and Service Guide N5242-90012 37 N5241/42AS Option H30/H31 Safety and Regulatory Information General Safety Considerations Cautions Cautions applicable to this instrument. CAUTION The Mains wiring and connectors shall be compatible with the connector used in the premise electrical system. Failure, to ensure adequate earth grounding by not using the correct components may cause product damage, and serious injury. CAUTION Always use the three prong AC power cord supplied with this product. Failure to ensure adequate earth grounding by not using this cord may cause product damage and the risk of electrical shock. CAUTION This product is designed for use in Installation Category II and Pollution Degree 2. CAUTION Verify that the premise electrical voltage supply is within the range specified on the instrument. CAUTION Ventilation Requirements: When installing the instrument in a cabinet, the convection into and out of the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the instrument by 4 °C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, forced convection must be used. 38 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Safety and Regulatory Information Servicing Warnings applicable to this instrument. WARNING Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended. Discard used batteries according to manufacturer’s instructions. WARNING This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall be inserted only into a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the product is likely to make the product dangerous. Intentional interruption is prohibited. WARNING For continued protection against fire hazard replace line fuse only with same type and rating. Fuse 3.15A/250V (2110-0655). The use of other fuses or material is prohibited. WARNING These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so. WARNING The opening of covers or removal of parts is likely to expose the user to dangerous voltages. Disconnect the instrument from all voltage sources before being opened. WARNING This product is designed for use in Installation Category II and Pollution Degree 2. WARNING No operator serviceable parts inside. Refer servicing to qualified personnel. WARNING If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only. User’s and Service Guide N5242-90012 39 N5241/42AS Option H30/H31 Safety and Regulatory Information Regulatory Information This section contains information that is required by various government regulatory agencies. Instrument Markings The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation. This symbol indicates that the instrument requires alternating current (ac) input. This symbol indicates separate collection for electrical and electronic equipment, mandated under EU law as of August 13, 2005. All electric and electronic equipment are required to be separated from normal waste for disposal (Reference WEEE Directive, 2002/96/EC). This symbol indicates that the power line switch is ON. This symbol indicates that the power line switch is in the STANDBY position. This symbol indicates that the power line switch is in the OFF position. This symbol is used to identify a terminal which is internally connected to the product frame or chassis. The CE mark is a registered trademark of the European Community. (If accompanied by a year, it is when the design was proven.) The CSA mark is a registered trademark of the CSA International. This instrument complies with Canada: CSA 22.2 No. 61010-1-04. This is a symbol of an Industrial Scientific and Medical Group 1 Class A product. ICES/NMB-001 This is a marking to indicate product compliance with the Canadian Interference-Causing Equipment Standard (ICES-001). Direct Current. IP 2 0 The instrument has been designed to meet the requirements of IP 2 0 for egress and operational environment. This is a required mark signifying compliance with an EMC requirement. The C-Tick mark is a registered trademark of the Australian Spectrum Management Agency. China RoHS regulations include requirements related to packaging, and require compliance to China standard GB18455-2001. This symbol indicates compliance with the China RoHS regulations for paper/fiberboard packaging. 40 User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Safety and Regulatory Information Battery Collection Do not throw batteries away but collect as small chemical waste, or in accordance with your country’s requirements. You may return the battery to Agilent Technologies for disposal. Refer to “Contacting Agilent” on page 43 for assistance. Compliance with German Noise Requirements This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordnung-3. GSGV Deutschland). Acoustic Noise Emission/Geraeuschemission LpA<70 dB Lpa<70 dB Operator Position am Arbeitsplatz Normal Operation normaler Betrieb per ISO 7779 nach DIN 45635 t. 19 EMC Information Complies with European EMC Directive 2004/108/EC • IEC/EN 61326-1 • CISPR Pub 11 Group 1, class A • AS/NZS CISPR 11 • This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB du Canada. User’s and Service Guide N5242-90012 41 N5241/42AS Option H30/H31 Electrostatic Discharge Protection Electrostatic Discharge Protection Protection against electrostatic discharge (ESD) is essential while removing assemblies from or connecting cables to the instrument. Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. To prevent damage to the instrument: • always have a grounded, conductive table mat (9300-0797) in front of your test equipment. • always wear a grounded wrist strap (9300-1367) with grounding cord (9300-0980), connected to a grounded conductive table mat, having a 1 MΩ resistor in series with it, when handling components and assemblies or when making connections. • always wear a heel strap (9300-1126) when working in an area with a conductive floor. If you are uncertain about the conductivity of your floor, wear a heel strap. • always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly. • always ground the center conductor of a test cable before making a connection to the analyzer test port or other static-sensitive device. This can be done as follows: 1. Connect a short (from your calibration kit) to one end of the cable to short the center conductor to the outer conductor. 2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector. 3. Connect the other end of the cable to the test port and remove the short from the cable. Figure 8 42 ESD Protection Setup User’s and Service Guide N5242-90012 N5241/42AS Option H30/H31 Agilent Support, Services, and Assistance Agilent Support, Services, and Assistance Service and Support Options The standard warranty is a one-year return to Agilent Technologies service warranty. NOTE There are many other repair and calibration options available from the Agilent Technologies support organization. These options cover a range of service agreements with varying response times. Contact Agilent for additional information on available service agreements for this product. Contacting Agilent Assistance with test and measurements needs and information or finding a local Agilent office are available on the Web at: http://www.agilent.com/find/assist You can also purchase accessories or documentation items on the Internet at: http://www.agilent.com/find If you do not have access to the Internet, contact your field engineer. NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number. With this information, the Agilent representative can determine the warranty status of your unit. Shipping Your Analyzer to Agilent for Service or Repair IMPORTANT Agilent Technologies reserves the right to reformat or replace the internal hard disk drive in your analyzer as part of its repair. This will erase all user information stored on the hard disk. It is imperative, therefore, that you make a backup copy of your critical test data located on the analyzer’s hard disk before shipping it to Agilent for repair. If you wish to send your instrument to Agilent Technologies for service or repair: • Include a complete description of the service requested or of the failure and a description of any failed test and any error message. • Ship the analyzer using the original or comparable antistatic packaging materials. • Contact Agilent for instructions on where to ship your analyzer. User’s and Service Guide N5242-90012 43 N5241/42AS Option H30/H31 Agilent Support, Services, and Assistance 44 User’s and Service Guide N5242-90012