Transcript
Operating Manual
I3031-5.0 en
d
3
Content
1
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6
2
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12
2.1 2.2 2.3 2.4
About the QuantumX documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The QuantumX family . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Channel overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sychronization of QuantumX modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12 13 15 16
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20
3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8
QuantumX Assistant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . catmanAP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Lab VIEW library . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Programming interface (API) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Firmware update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TEDS Editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DIAdem driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DASYLab driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20 21 22 22 22 25 25 25
Housings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
26
4.1 4.2 4.3 4.4 4.5
Mounting housing clips on modules with degree of protection IP20 . . . . . . . . . . . . . . . . . . . . . . . Mounting housing clips on modules with degree of protection IP65 . . . . . . . . . . . . . . . . . . . . . . . Connecting housings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CASE-FIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . BPX001 backplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5.1 Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5.2 Wall mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5.3 Mounting the modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
28 31 33 35 36 37 38 39
Connecting individual QuantumX modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
42
5.1 5.2
42 43 43 44 45 46 54 55 56 58 60 62 63
3
4
5
QuantumX
Connecting the supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Connection to host PC or notebook . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.1 Single Ethernet connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.2 Multiple Ethernet connection without synchronization . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.3 Multiple Ethernet connection with synchronization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.4 Ethernet setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.5 Connection via FireWire (IEEE 1394b) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.6 Multiple Ethernet connection with synchronization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.7 FireWire setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.8 Firmware update via Ethernet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.9 Connecting more than 12 modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.10 Bridging greater distances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.11 FireWire with opto-hub and glass fiber cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HBM
4
6
Transducer connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
64
6.1 6.2 6.3
64 66 67 69 70 74 78 79 82 85 89 90 90 91 92 93 94 95 96 97 97 98 99 100 101 102 103 104 105 106 109 110 111 112 113 114 116 117 118 119 120
Shielding design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Adjustable sensor supply for active transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TEDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3.1 Retrofitting the transducer plug with TEDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4 MX840 universal amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.5 MX840A universal amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.6 MX440A universal amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.7 MX410 highly dynamic universal amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.8 MX460 frequency measuring amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.9 MX1609 / 1609Pmodule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.9.2 MX1609/MX1609-P status display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.10 MX471 CAN module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.10.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.10.2 MX471 pin assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.10.3 Receiving CAN messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.10.4 LED status display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.11 MX1601 universal amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.11.1 MX1601pin assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.11.2 MX1601status display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12 Transducder technologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.1 Strain gages full bridges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.2 Strain gage full bridges, inductive full bridges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.3 Piezoresistive full bridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.4 Strain gage half bridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.5 Inductive half bridges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.6 Quarter bridge, strain gage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.7 Connection of transducers with double shield technique . . . . . . . . . . . . . . . . . . . . . . . . 6.12.8 Potentiometric transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.9 LVDT transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.10 Piezoelectric sensors IEPE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.11 DC voltage sources 100 mV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.12 DC voltage sources 10 V and 60 V . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.13 DC power sources 20 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.14 Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.15 Resistance thermometers Pt100, Pt1000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.16 Thermocouples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12.17 Frequency measurement without directional signal (RS 485: Differential signals) . . . 6.12.18 Frequency measurement with directional signal (RS 485: Differential signals) . . . . . . 6.12.19 Frequency measurement without directional signal (single−pole mode) . . . . . . . . . . . 6.12.20 Frequency measurement with directional signal (single−pole mode) . . . . . . . . . . . . . . 6.12.21 Pulse counting, (RS 485: Differential signals) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HBM
QuantumX
5
6.12.22 6.12.23 6.12.24 6.12.25 6.12.26
7
Pulse counting (single−pole mode) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SSI protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Passive inductive rotary encoder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PMW − Pulse width, Pulse duration, Period duration . . . . . . . . . . . . . . . . . . . . . . . . . . . CANbus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
121 122 123 124 125
Functions and outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
126
7.1 7.2 7.3 7.4
8
MX410 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MX460 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MX878 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MX471 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
127 130 131 135
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
136
8.1
System accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1.1 BPX001 backplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1.2 Housing connection elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Voltage supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.2.1 Power pack NTX001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.2.2 Supply cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . FireWire . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.3.1 FireWire cable (module to module; IP20) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.3.2 FireWire cable (module to module; IP65) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.3.3 Connection cable (PC to module; IP20) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4.1 Plug kit with TEDS chip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4.2 Port saver SubHD 15pol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4.3 Adapter D−Sub−HD 15-pin to D−Sub 15-pin. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MX840 accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.5.1 Cold junction for thermocouples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.5.2 Adapter cable (CAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Accessories MX410 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MX1609 and 1609P accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.7.1 Thermo-connector with integrated RFID chip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.7.2 RFID chip for sticking onto thermo-connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
139 139 139 140 140 141 141 141 142 142 143 143 143 144 145 145 146 146 147 147 147
Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
147
8.2
8.3
8.4
8.5
8.6 8.7
9
QuantumX
HBM
6
1
Safety instructions
Safety instructions Appropriate use The module and the connected transducers may be used for measurement and directly related control tasks only. Any other use is not appropriate. To ensure safe operation, the module may only be used as specified in the operating manual. It is also essential to follow the respective legal and safety regulations for the application concerned during use. The same applies to the use of accessories. Each time, before starting up the modules, you must first run a project planning and risk analysis that takes into account all the safety aspects of automation technology. This particularly concerns personal and machine protection. Additional safety precautions must be taken in plants where malfunctions could cause major damage, loss of data or even personal injury. In the event of a fault, these precautions establish safe operating conditions. This can be done, for example, by mechanical interlocking, error signaling, limit value switches, etc. Safety rules A module must not be connected directly to a power supply network. The maximum permissible supply voltage is 10...30 V (DC). The supply connection, as well as the signal and sense leads, must be installed in such a way that electromagnetic interference does not adversely affect device functionality (HBM recommendation: ”Greenline shielding design”, downloadable from the Internet at http://www.hbm.com/Greenline). Automation equipment and devices must be covered over in such a way that adequate protection or locking against unintentional actuation is provided (such as access checks, password protection, etc.). When devices are working in a network, these networks must be designed in such a way that malfunctions in individual nodes can be detected and shut down. Safety precautions must be taken both in terms of hardware and software, so that a line break or other interruptions to signal transmission, such as via the bus interfaces, do not cause undefined states or loss of data in the automation device.
HBM
QuantumX
Safety instructions
7
Conditions on site For modules in IP20 protected housing: − Protect the modules from humidity or effects of the weather such as rain, snow, etc. − The permissible relative humidity at 31°C is 80% (non condensing); linear reduction down to 50% at 40°C. − Ensure that the ventilation openings on the sides are not covered. For all modules: − Do not expose the instrument to direct sunlight. − Observe the maximum permissible ambient temperatures given in the specifications. Maintenance and cleaning The modules are maintenance-free. Please note the following when cleaning the housing: − Before cleaning, disconnect the equipment completely. − Clean the housing with a soft, slightly damp (not wet!) cloth. Never use solvents, since these could damage the labelling on the front panel and the display. − When cleaning, ensure that no liquid gets into the module or connections.
General dangers of failing to follow the safety instructions The module is a state of the art device and as such is failsafe. The module may give rise to further dangers if it is inappropriately installed and operated by untrained personnel. Any person instructed to carry out installation, commissioning, maintenance or repair of the module must have read and understood the Operating Manuals and in particular the technical safety instructions.
Remaining dangers The scope of supply and performance of the module covers only a small area of measurement technology. In addition, equipment planners, installers and operators should plan, implement and respond to the safety engineering considerations of measurement technology in such a way as to minimize remaining dangers. Prevailing regulations must be complied with at all times. There must be reference to the remaining dangers connected with measurement technology. After making settings and carrying out activities that are password−protected, you must make sure that any controls that may be connected remain in safe condition until the switching performance of the module has been tested.
Working safely Error messages should only be acknowledged once the cause of the error is removed and no further danger exists.
QuantumX
HBM
8
Safety instructions
Conversions and modifications The module must not be modified from the design or safety engineering point of view except with our express agreement. Any modification shall exclude all liability on our part for any resultant damage. In particular, any repair or soldering work on motherboards (replacement of components is prohibited. When exchanging complete modules, use only original parts from HBM. The module is delivered from the factory with a fixed hardware and software configuration. Changes can only be made within the possibilities documented in the manuals. Outputs It is essential to take safety precautions when using a module’s digital, analog or CANbus outputs. Make sure that status or control signals do not perform any actions that result in damage to human health or the environment. Qualified personnel Qualified persons means persons entrusted with the installation, fitting, commissioning and operation of the product who possess the appropriate qualifications for their function. This module is only to be installed and used by qualified personnel, strictly in accordance with the specifications and the safety rules and regulations. This includes people who meet at least one of the three following requirements: • Knowledge of the safety concepts of automation technology is a requirement and as project personnel, you must be familiar with these concepts. • As automation plant operating personnel, you have been instructed how to handle the machinery and are familiar with the operation of the modules and technologies described in this documentation. • As commissioning engineers or service engineers, you have successfully completed the training to qualify you to repair the automation systems. You are also authorized to activate, to ground and label circuits and equipment in accordance with safety engineering standards. It is also essential to comply with the legal and safety requirements for the application concerned during use. The same applies to the use of accessories.
NOTE The safety instructions in this document also apply to the NTX001 power supply and the BPX001 active back plane.
HBM
QuantumX
Safety instructions
9
In this manual, the following symbols are used to point out residual dangers:
Symbol:
DANGER
Meaning: Maximum danger level Warns of an imminently dangerous situation in which failure to comply with safety requirements will result in death or serious bodily injury.
Symbol:
WARNING
Meaning: Dangerous situation Warns of a potentially dangerous situation in which failure to comply with safety requirements can result in death or serious bodily injury.
Symbol:
CAUTION
Meaning: Potentially dangerous situation Warns of a potentially dangerous situation in which failure to comply with safety requirements could result in damage to property or some form of bodily injury.
Symbol:
Meaning: Electrostatic Sensitive Devices Devices marked with this symbol can be destroyed by electrostatic discharge. Please observe the precautions for handling electrostatic sensitive devices.
Symbol: Meaning:
QuantumX
On the equipment Observe information provided in the operating manual.
HBM
10
Safety instructions
Symbols pointing out notes on use and waste disposal as well as useful information:
Symbol:
NOTE
Points out that important information about the product or its handling is being given.
Symbol: Meaning: CE mark The CE mark enables the manufacturer to guarantee that the product complies with the requirements of the relevant EC directives (the declaration of conformity is available at http://www.hbm.com/hbmdoc).
Symbol: Meaning: Statutory marking requirements for waste disposal National and local regulations regarding the protection of the environment and recycling of raw materials require old equipment to be separated from regular domestic waste for disposal. For more detailed information on disposal, please contact the local authorities or the dealer from whom you purchased the product.
HBM
QuantumX
Safety instructions
11
Conditions on site For modules in housings with IP20 degree of protection: − Protect the modules from moisture and dampness or weather such as rain, snow, etc. − The permissible relative humidity at 31oC is 80% (non condensing); linear reduction to 50% at 40 oC. − Make sure that you do not cover the ventilation openings at the side. For all modules: − Do not expose the modules to direct sunlight − Comply with the maximum permissible ambient temperatures as stated in the specifications. − Ensure sufficient ventilation during installation into the BPX001 backplane. Maintenance and cleaning The modules are maintenance−free. Please note the following points when cleaning the housing: − Before cleaning, disconnect the device from all connections. − Clean the housing with a soft, slightly damp (not wet!) cloth. Never use solvent as this could damage the labeling or the housing. − When cleaning, ensure that no liquid gets into the module or connections.
QuantumX
HBM
Introduction
12
2
Introduction
2.1
About the QuantumX documentation The QuantumX family documentation consists of
•
A printed quick start guide for initial start up
•
The datasheets in PDF format
•
This operating manual in PDF format
•
The EtherCAT / CX27 Ethernet gateway operating manual in PDF format
•
A comprehensive online help with index and easy search options which is available after the installation of a software packet (e.g. QuantumX Assistant, catmanEASY)
These documents are available
HBM
•
on the QuantumX system CD supplied with the instrument
S
after installation of QuantumX−Assistant on your PC’s hard disk
S
always up to date on the Internet at http://www.hbm.com/hbmdoc
QuantumX
Introduction
2.2
13
The QuantumX family The QuantumX family is a modular and universally applicable measurement system. The modules can be individually combined and intelligently connected according to the measurement task. The distributed operation makes it possible to position individual modules close to the measuring points, resulting in short sensor lines. The QuantumX family currently consists of the following modules:
S
MX840 universal amplifier The module has 8 universal inputs and supports more than 10 transducer technologies
S
MX840A universal amplifier As MX840, however it is additionally possible to connect half bridges and ohmic resistors.
S
MX440A universal amplifier As MX840A, however with 4 inputs (without CAN).
S
MX410 highly−dynamic universal amplifier The module has 4 universal inputs and supports commonly used transducer technologies (with a sampling rate of up to 96000 measuring values per second per channel).
S
MX460 rotation specialist frequency amplifier The module has 4 individually configurable inputs for connecting HBM torque transducers (T10, T40), incremental encoders, frequency signal sources and digital signals.
S
MX1609 thermocouple amplifier The module has 16 inputs for type K thermocouples
S
MX1609−P thermocouple amplifier As MX1609, however with IP65 degree of protection.
S
MX878 analog output module The module has 8 analog outputs to which a system or source signal can be assigned. In addition, signals can be computed in real time.
S
MX471 CAN module The module has 4 CANbus nodes that can be configured for receiving or transmitting messages.
S
EtherCAT/Ethernet gateways CX27 The module is used for connecting QuantumX modules to the Ethernet fieldbus.
S
MX1601 universal amplifier The module has 16 inputs that can be individually configured for voltage or current measurement or for connecting current-fed, piezoelectric transducers.
S
Data recorder CX22 / CX22−W (WLAN) The module is used for local recording of measured data.
The modules have the following in common:
S S QuantumX
Low voltage connection Ethernet interface for data communication with an operating PC
HBM
Introduction
14
S
2 FireWire interfaces
− For optional voltage supply (see data sheet) − For optional data communication via PC − For medule synchronization − For internal transmission of measurement data
S
Connector (VG strip) for installation on an active module carrier (not applicable for IP65 variants)
S S
Status LEDs for displaying general system and channel states
S
AutoBoot (module configuration is retained)
A working standard calibration certificate has been stored on every amplifier. It can be read out via QuantumX Assistant.
With amplifiers, the following applies for each measurement channel:
S S
Electrical isolation
S S S S
Support for TEDS*) technology (read, write)
Configurable supply voltage (signal inputs/outputs, voltage supply, communication) for active sensors Configurable sampling rate Configurable active digital filter (Bessel, Butterworth) Scaling a. Zero span and gain linearization (constant and slope) b. Multipoint−/table-based linearization (data points) c. Polynomial linearization
Sensors assigned via the sensor database can be measured (calibrated) via the channel and rewritten to the sensor database. Variants a. and b. too can be saved to the sensor data sheet (TEDS). When the transducer is connected, its characteristic curve is automatically downloaded into the amplifier and the signal is linearized accordingly. The device configuration is permanently retained, even if the device is switched off. *)
HBM
TEDS = Transducer Electronic Data Sheet
QuantumX
Introduction
15
2.3
Channel overview
QuantumX Modules MX840A Number of channels (total)
MX440A
MX410
MX460
MX1609 MX1609-P
MX1601
MX878
MX471
CX27
CX22W
8
4
4
4
16
16
-
-
-
-
Data rate (Samples/s)
19200
19200
96000
96000
300
19200
-
-
-
-
Bandwidth (Hz)
3200
3200
38000
38000
14
3000
-
-
-
-
Full-bridge strain gages
•
•
•
Half-bridge strain gages
•
•
•
Inductive full-bridge
•
•
•
Inductive half-bridge
•
•
•
LVDT
•
•
Voltage
•
•
•
•
Current (± 20 mA)
•
•
•
•
•1)
•1)
•1)
•
Piezoresistive transducer
•
•
•
Resistance
•
•
Potentiometers
•
•
PT100 and PT1000 resistance thermometers
•
•
Thermocouples
•
•
Digital IN (static)
•
•
Digital OUT (static)
•
•
Current fed piezoelectric transducer (IEPE)
•
(Type K)
•
Indctive rotary encoder Incremental encoder
•
•
SSI
•
•
Frequency measurement, pulse counting
•
•
• • •
PWM Torque / rotary speed CANbus Analog outputs
• •
•
•
•
(Input/Output)
(Input/Output)
•
•
• Mathematics
•
•
Local recording of measured data
•
• •
¹) A Smart module (1-EICP-B-2) is required for connecting current-fed piezoelectric transducers.
See data sheets for precise technical specifications. The pin assignments can be found in the following chapters. QuantumX
HBM
Introduction
16
2.4
Sychronization of QuantumX modules To enable measurement signals to be synchronized for processing and analysis, they need to be synchronously measured. All QuantumX modules can be synchronized among themselves. This ensures simultaneous measurement on all channels. All the analog-digital converter rates, measuring rates and bridge excitation voltages are also synchronized. Synchronization methods: Synchronization via FireWire All the modules are synchronized automatically when they are connected via the FireWire cable. This is the recommended method. No CX27 module present in the system: The module with the highest serial number takes over the Master function. CX27 module present in the system: If a CX27 module is connected, this automatically becomes the synchronization Master. When starting the system, the system time is set once to the actual time. If only QuantumX modules are being used, internal synchronization is sufficient. However, if synchronous measurements are to be performed by different measurement systems, an external Master must be used for synchronization. This is also a requirement if the QuantumX modules are a long distance away from one another, and a FireWire connection would be too costly and complex.
Synchronization via EtherCAT The CX27 gateway supports the EtherCAT ”Distributed Clocks” enhancement. In an EtherCAT group, the time is distributed to all the EtherCAT nodes. The CX27 module can be synchronized to the EtherCat time. This will mean that all the QuantumX module clocks are synchronized to the specified time.
Synchronization via an NTP server Each QuantumX module can synchronize its internal clock with an NTP server. The NTP time is distributed to all other modules via FireWire. It is possible to achieve accuracies in the 100 µs range, although this depends on the relevant Ethernet utilization. Modules that are in proximity to each other should be synchronized by FireWire. If the synchronization source for a module is changed to NTP, the system must be restarted once. catmanEASY software from HBM includes an NTP software package. Parameters: IP address of the NTP server Threshold in µs, from which the time deviation to NTP time is tolerated HBM
QuantumX
Introduction
17
Synchronization using IRIG−B IRIG−B is a standardized time code. The digitally or analogously modulated time signal is externally fed to an arbitrary analog voltage input of type MX840A or MX440A amplifiers to synchronize the QuantumX system (see assignment, chapter 6.5.1). The amplifiers can record type B000 through B007 and B120 through B127 IRIG−B signals. All modules connected via FireWire are automatically synchronized as well. The code includes the time, the year and, optionally, the seconds of the current day.
QuantumX
Characteristic Synchronization with other device types
FireWire QuantumX only
Ethernet (NTP) QuantumX, MGCplus other interrogators
EtherCat all EtherCAT nodes
IRIG−B all IRIG−Bnodes
Max. distance between QuantumX modules
5 m (40 m with FireWire extender, 500 m via fiber-optics
100 m (Ethernet)
100 m
−
Number of modules to be synchronized
24
unlimited
CX27 required, unlimited
unlimited MX440A, MX840A required
Synchronization accuracy
< 1 µs
100 µs to 10 ms
< 1 µs
< 1 µs
Synchronization setting time
immediately
approx. 2 h on first start, approx. 10 min on restart
immediately
immediately
Synchronization Master
Auto 1 QuantumX module
external SyncMaster , e.g. PC
external SyncMaster
external IRIG−B master
Voltage supply
< 1.5 A, looped through
−
−
−
HBM
Introduction
18
Synchronizing via FireWire Auto
Auto
Auto
Auto
Time base : automatic (factory setting)
Synchronizing via CX27 (EtherCAT) Auto
Auto
Auto
Time base : EtherCat Master
CX27
EtherCAT
Synchronizing via NTP (with FireWire) Auto
Auto
Auto
Time base : NTP server
NTP *)
Ethernet
NTP
Synchronizing via NTP (without FireWire) NTP
NTP
NTP
NTP
Time base : NTP server Ethernet switch
Fig. 2.1: *) HBM
Different methods of time synchronization
CX27 or the module with the highest serial number QuantumX
Introduction
19
Successful synchronization: For accurate synchronization, the corresponding channels should be parameterized using identical filter settings. The delay is not automatically corrected. Filter delays are given in the datasheet. After booting and successful synchronization, the system LED shows green. If synchronization is disturbed, or not yet established, the system LED shows orange. Example: MX840
System LED
Time format used: Basis:
1.1.2000
Time stamp: 64 bit 32 bit seconds 32 bit fractions of a second, resolution (1/232) These time stamps are appended to the measured values. There are several synchronization methods to choose from (also see Fig. 2.1 page 18): S synchronizing via FireWire S synchronizing via EtherCAT (CX27) S synchronizing via NTP (Network Time Protocol) with FireWire S synchronizing via NTP without FireWire
QuantumX
HBM
20
3
Software
Software The supplied QuantumX System CD contains a powerful software package consisting of the QuantumX Assistant, Lab View library, programming library for .NET/COM, TEDS-Editor, FireWire driver as well as a program for module firmware updating. catmanEASY and DIAdem driver are software products that can be obtained as separate product packages.
3.1
QuantumX Assistant The HBM software ”QuantumX Assistant” offers the following functions: System..
•
Create survey (modules, host PC)
Modules:
• • • • •
Search and configuration (e.g. TCP/IP communication), naming Reset to factory settings Read factory calibration certificate Analysis (information, status, log file) Save configuration to operating PC
Channels/sensors:
• • • •
Configuration (name, connection type, TEDS, semi-automatic assignment) Measurement (alphanumeric display) Open the TEDS editor and read/write to TEDS Activate/deactivate isochronous operation via FireWire
Individual signals:
•
Set sampling rate and filter (type, cut-off frequency)
Measured values (Scope):
• • •
Start/Stop continuous graphic measurements (time frames, trigger, zoom) Basic signal analysis (X-Y cursor) Record measurements
Functions and outputs
•
Generate new signals using mathematics functions (peak value, rms value, adding and multiplying, rotation)
Sensor database
• • HBM
Output signals (scaled, filteresd) Modify and expand existing sensor databases (e.g. own sensors, dbc data files) QuantumX
Software
3.2
21
catmanAP HBM’s catmanAP software is optimally suited for the following tasks:
•
Configuring communication and measurement channels (integrated TEDS Editor and expandable sensor database)
•
Configuring the test or measurement task (channels, sampling rates, triggers, comments, interaction)
•
Setting up virtual, online−computed channels (algebra, FFT, logics, strain−gage rosette analysis, differential, integral, etc.)
•
Setting up limit value or event monitoring (activation of digital output, acoustic alarm, log entry)
•
Individual graphical visualization options (line recorder, analog meter, digital or bar graph indicator, table, status LED, etc.)
• •
Various storage options (all data, cyclic, ring buffer, long duration measurement, etc.)
• • •
Graphical analysis of recorded data
Exporting measurement data into commonly used data formats (catmanBIN, Excel, ASCII, DIADEM, MDF) Automating measurement sequences (AutoSequence and EasyScript) Reporting (using graphical displays, analyses, comments)
The software package consists of various modules:
QuantumX
S
EasyMath for mathematical evaluation
S
AutoSequence automates repeating measurement or analysis steps
S
EasyLog records the measurement data onto a storage medium
S
EasyPlan enables preparatory parametrization and configuration without amplifier connection
S
EasyScript is based on current VBA standard (Visual Basic for Applications) and enables writing of own scripts for individual measurement tasks
HBM
22
3.3
Software
Lab VIEW library Lab VIEW is a graphic programming system from the company ”National Instruments”. The acronym stands for ”Laboratory Virtual Instrumentation Engineering Workbench”. The main application areas for Lab VIEW are in measurement, control and automation technology. A Lab VIEW device driver is a VI (Virtual Instrument) or sub-program that is used in Lab VIEW programs for easy operation of devices. Device drivers are used to initialize, open and close interfaces when required, to initialize and configure devices, to make settings, trigger and query measurements, etc. A setup program is available on the QuantumX System CD with which you can install the Lab VIEW driver on your computer.
3.4
Programming interface (API) The abbreviation API stands for ”Application Programming Interface” and designates so-called programming interfaces. Programmers can directly access functions of other programs via APIs and use these in their own programs. With the API, you have full access to all QuantumX functions through an individually programmed application, e.g. your own operator interface. The API can be used in the form of programming libraries in .NET or COM technologies. The libraries enable the creation of own applications in programming languages such as e.g. Visual Basic, C++, C# or Delphi. Functions such as communication connection, configuration of measurement channels, implementation of measurements and troubleshooting are components of the library. The API can be easily installed from the QuantumX System CD. Application-based examples and practice-oriented documentation enable rapid a quick start.
3.5
Firmware update You can easily check the firmware status of the modules and update them with the ”QuantumX firmware update” software. Before updating your firmware, please check whether your PC software needs to be updated. We recommend that you check and, if necessary, update the firmware • if you want to use a new PC software package • if you expand your system with new modules QuantumX Assistant also enables you to check the firmware status of your modules: • Right−click on a module −> Details −> System parameters
HBM
QuantumX
Software
23
• Compare your version with the actual firmware version on the Internet under: www.hbm.com\quantumX
QuantumX
HBM
24
Software
If the firmware version number of your module is lower than the actual number on the Internet, an update is necessary: • Download the actual firmware from the HBM website and save it in the download directory of the firmware updater (in most cases: C:\Programs\HBM\QuantumX Firmware Update\Download) • Download the actual software package from the HBM website • Close all open HBM software, install the new software and start the QuantumX Firmware Update • Click on the ”Find modules” symbol or press the function key F4.
• Select the module • Select the required version in the drop down menu ”New firmware” • Activate the modules whose firmware you want to update by checking the boxes in the column ”Update” and then click on the ”OK” button.
• Press the ”Start” button and wait until the update is complete (do not interrupt the process/do not switch off the modules/do not interrupt the connection) HBM
QuantumX
Software
3.6
25
TEDS Editor HBM’s TEDS Editor enables TEDS data to be read, edited and written directly via a measurement channel or the HBM TEDS dongle. When retrofitting TEDS, the editor provides corresponding templates for the different transducer types. Some templates can be saved and loaded. Chapter 6.3 describes TEDS in general.
3.7
DIAdem driver DIAdem is a graphic programming system from the company ”National Instruments”. DIAdem provides corresponding library modules for the complete data flow, ranging from measurement acquisition through analysis to reporting. HBM’s DIAdem driver enables measurement data acquisition using the following QuantumX amplifiers: MX840, MX840A, MX440A, MX410, MX460 and MX1609. Latest information can be found in the corresponding release notes at www.hbm.com.
3.8
DASYLab driver DASYLab is a graphic programming system from the company ”National Instruments”. The acronym stands for ”Data Acquisition System Laboratory”. The main application areas for DASYLab are in measurement, control and automation technology. The graphic library modules enable interfaces to be controlled, for example, inputs and outputs, signal conditioning and analysis, control, visualization and data storage. The partner company IMP sells a DASYLab driver for measurement data acquisition using the follwing QuantumX modules: MX840, MX840A, MX440A and MX1609. Latest information can be found at www.impkoeln.de.
QuantumX
HBM
26
4
Housings
Housings The degree of protection given in the technical data indicates the suitability of the housings for various ambient conditions and also the protection of persons against potential risks when used. The letters IP (International Protection), which are always present in the designation, are followed by two digits. These indicate which degree of protection a housing offers against contact or foreign bodies (first digit) and moisture (second digit). The QuantumX modules are available in housings with degree of protection IP20 and, in some cases, with IP65 (as per DIN EN 60529).
IP IP Code index
2 6 Degree of protection against
0 5 Code index
Degree of protection against water
contact and foreign bodies 2
Protection against contact with fingers, protection against foreign bodies with ∅ > 12 mm
0
No water protection
6
Complete protection against contact, protection against penetration of dust
5
Protection against water jet(nozzle) from any angle
Both housing types can be connected together with the aid of two lateral housing clips (1-CASECLIP, not included in scope of delivery). To do this, the existing lateral covers must be removed and the housing clips screwed on.
Housing clip
Fig. 4.1:
HBM
Two connected IP20 housings
QuantumX
Housings
27
Fig. 4.2:
QuantumX
Amplifier MX1609-P in IP65 housing
HBM
28
4.1
Housings
Mounting housing clips on modules with degree of protection IP20 The module electronics are integrated in a metal housing that is surrounded by a case protection (CASEPROT). This also serves for centering when several devices are stacked on top of each other and offers a certain degree of protection against mechanical damage. Case protection MX840 housing
Cover
Fig. 4.3:
Universal amplifier MX840 with case protection
The mounting of the housing clips shown in the following pictures must be implemented on both sides of the housing.
a.f. 2.5
Fig. 4.4:
HBM
Removing the CASEPROT
QuantumX
Housings
29
Cover
Fig. 4.5:
Removing the cover
Housing clip
a.f. 2.5
Fig. 4.6:
QuantumX
Mounting the housing clip
HBM
30
Housings
a.f. 2.5
Fig. 4.7:
HBM
Mounting the CASEPROT
QuantumX
Housings
4.2
31
Mounting housing clips on modules with degree of protection IP65 The mounting shown in the following pictures must be implemented on both sides of the housing.
Cover
Fig. 4.8:
Removing the cover
Housing clip
a.f. 2.5
Fig. 4.9:
QuantumX
Mounting the housing clip
HBM
32
Housings
Fig. 4.10:
HBM
Attaching the covers
QuantumX
Housings
4.3
33
Connecting housings The following pictures show the connection of two IP20 housings. The procedure is identical for IP65 housings.
Press
Fig. 4.11:
Unclip the housing clip
Lever
Catch
Fig. 4.12:
QuantumX
Unclip the lever and catch
HBM
34
HBM
Housings
Fig. 4.13:
Close the lever
Fig. 4.14:
Connected housings
QuantumX
Housings
4.4
35
CASE-FIT A fitting panel (CASE−FIT) allows flexible mounting of IP20 protected modules of the QuantumX series. A lashing strap or CASE−CLIP can be used for fastening the modules.
Fig. 4.15:
QuantumX
22
11,4
Laschen für die zusätzliche fixierung mit Spanngurten
Ø 5,6
approx. 30
169,5
132
Mounting of the IP20 version
HBM
36
4.5
BPX001 Backplane
BPX001 backplane
You can connect up to 9 modules with each other using the BPX001 backplane without the need for complicated cabling and you can link these via the two FireWire interfaces of the backplane with other modules or module groups. You can also set up a direct connection to a PC via one of the FireWire interfaces of the backplane. The FireWire interfaces of the individual modules are actively connected to each other. The supply voltage (18 V ... 30 V DC) for the modules must be supplied externally. The circuits of the FireWire connections and the modules are protected with a total of four fuses with control displays (arrangement see Tab. 4.1). The modules can be positioned anywhere in the backplane. The backplane is designed for wall or control cabinet installation and has boreholes for attachment.
Slot 9
Slot 1
Fig. 4.16:
HBM
Example for setup with 6 modules
QuantumX
BPX001 Backplane
4.5.1
37
Connection
VG strip module connection FireWire
X1 / X2
Fuses with control LEDs 4 x 4 A/T
Supply voltage 18 V ... 30 V DC 5 A max.
+ −
Ground
Fig. 4.17: No.
QuantumX
BPX001 connections Protection
1
FireWire X1 connection
2
FireWire X2 connection
3
Slots 1 to 4
4
Slots 5 to 9
Tab. 4.1:
Fuse assignments
HBM
38
4.5.2
BPX001 Backplane
Wall mounting A total of 10 boreholes are provided in the backplane for wall mounting (∅ 6.5 mm). We recommend using the outer 4 boreholes for wall mounting.
NOTE Only use countersunk head bolts for fastening, otherwise the modules cannot be mounted correctly.
Fig. 4.18:
BPX001 drilling pattern and dimensions
Note the following information when installing one or more backplanes in a control cabinet: • When installing in a control cabinet, the temperature limits given in the technical data of the backplanes must be complied with • Depending on the installation situation, sufficient ventilation (vertical air flow) or cooling must be provided (the maximum total output on a backplane is approx. 150 Watt) • The ventilation slits of the modules must not be covered (e.g. by cable ducts)
HBM
QuantumX
BPX001 Backplane
4.5.3
39
Mounting the modules
TOOLS We recommend a T-handle socket spanner 4x150 (across flats 4 mm, length 150 mm).
NOTE The modules can only be fastened in backplanes in housings with IP20 protection without case protection, case clips or lateral covers If these are present, remove as shown in Chapter 4.
Mounting sequence: 1. Remove the cover of the connecting plug (rear of module).
Cover
Fig. 4.19:
Remove cover
2. Unscrew the upper and lower screwed clamping glands of the backplane up to the stop (the screws are secured against falling out!).
QuantumX
HBM
40
BPX001 Backplane
3. Position the module vertically on the backplane and push it in carefully on the lower guide rail back up to the stop.
Upper screwed clamping gland a.f. 4.0
Guide rails Lower screwed clamping gland
HBM
Fig. 4.20:
Mounting the module
Fig. 4.21:
Centering above the connection plug QuantumX
BPX001 Backplane
41
4. Tighten the lower then the upper screwed clamping gland.
2.
1.
Fig. 4.22:
QuantumX
Tightening the screwed clamping glands, sequence
HBM
42
Connection
5
Connecting individual QuantumX modules
5.1
Connecting the supply voltage Connect the modules to a DC voltage of 10 V ... 30 V (recommended 24 V), see the table for the power consumption per device.
CAUTION
A rule of thumb for voltage distribution via FireWire says: ”An external supply voltage of equal potential is required at every third module”. Defects in the device cannot be excluded if a supply voltage > 30 V is used. If the supply voltage drops below 10 V, the modules switch off. In the event of battery operation in a vehicle, we recommend incorporating an uninterrupted voltage supply between battery and module to compensate for voltage drops during starting processes. Module
Typical power consumption, including transducer excitation (Watt)
MX840
13
MX840A
12
MX440A
10
MX1601
13
MX410
15
MX460
9
MX1609
6
MX1609-P
6
CX22
12
CX27
7
MX878
7
MX471
6
If several modules are connected to each other via FireWire for time−synchronous data acquisition (see Fig.5.4), the supply voltage can be looped through. The power pack used must be able to provide the appropriate output. The maximum permissible current on the FireWire connection cable is 1.5 A.If the chain is longer repeating the supply connection is mandatory. If several amplifiers are operated non−synchronously (see Fig.5.3), they must be supplied separately. NTX001 Oder
X104
1−Kab271−3 1−Kab269 FireWire
Fig.5.1: HBM
X101/X102
Connecting socket for supply voltage QuantumX
Connection
43
5.2
Connection to host PC or notebook
5.2.1
Single Ethernet connection
10 V ... 30 V DC 1-NTX001 or 1-KAB271-3
X104
TCP/IP, 100 Mbps X100 Cross Over
Fig.5.2:
Single Ethernet connection
NOTE You must use an Ethernet crossover cable with older computers. Newer PCs/laptops have Ethernet interfaces with autocrossing function. You can also use Ethernet patch cables here.
QuantumX
HBM
44
5.2.2
Connection
Multiple Ethernet connection without synchronization
10 V ... 30 V DC
10 V ... 30 V DC
10 V ... 30 V DC
Patch cable
Patch cable
TCP/IP, 100 Mbps
Fig.5.3:
Multiple connection via Ethernet
Modules can be connected to the PC with standard Ethernet switches. We recommend patch cables. With the star structure displayed here, measurement data from other modules is not lost in case of a cable break in the ethernet cable.
HBM
QuantumX
Connection
5.2.3
45
Multiple Ethernet connection with synchronization
10 V ... 30 V DC
Patch cable
TCP/IP, 100 Mbps Patch cable
FireWire connection 1-Kab269-x: Connection cable with various lengths (x m)
Fig.5.4:
Example for multiple connection via Ethernet with synchronization
The supply voltage for the modules is looped through FireWire in the configuration shown above (max. 1.1 A through FireWire; power consumption for 1 module, see specifications in data sheet). Advantage of this connection structure: The other modules remain active in case of a cable break in the ethernet cable.
QuantumX
HBM
46
5.2.4
Connection
Ethernet setup
Settings for direct connection with a PC (peer-to-peer)
NOTE Ensure that your PC has a valid IP address. Install and start up the latest QuantumX Assistant version on your PC. (All screenshots shown in this quick start guide display the menus in the operating system WindowsXP). • Click on the icon
HBM
(Find modules) or press the function key F4.
QuantumX
Connection
47
The next dialog window offers you the several options for network searching. For the first startup we recommend: • Search the whole network and select from the result (recommended setting)
Recommended
• If your modules are not yet displayed, click on the button.
NOTE The network connection can be influenced by: w An activated WiFi connection on your PC: Switch this connection off and start the network search again w The use of a standard patch cable in a direct connection (peer-to-peer) If the module you have selected appears in black in the list, immediate operation is possible.
QuantumX
HBM
48
Connection
If the module appears gray in the list, mark it and click on the ”Edit network settings” button.
Check your settings and adjust if necessary as follows:
Adapt the IP address of the module: • Activate DHCP/APIPA if you want automatic configuration. Please set any PC directly connected to QuantumX to DHCP as well. • Manual configuration: Deactivate DHCP and enter the same subnet mask address as used with your PC. Change the IP address of your module so that it matches communication (see example below) Example: Manual IP setting − module side Settings
IP address
Subnet mask
Module before
169.1.1.22
255.255.255.0
PC / Notebook
172.21.108.51
255.255.248.0
Module after
172.21.108.1
255.255.248.0
The first three digit groups of the PC and module IP addresses shall be the same. The subnet mask address digit groups must be identical in the module and PC!
HBM
QuantumX
Connection
49
Automatic Configuration
Module settings
PC settings
Manual Configuration
Module settings
172.21.108.1 255.255.248.0
PC settings
Fig.5.5: QuantumX
Example settings for direct connection of one module HBM
50
Connection
• Click on ”OK” • Confirm the settings with the ”Yes” button, the module is then restarted with the current settings.
After approx. 45 seconds click on the
button.
The system LED of the module should now light up green, if not please check your network settings again! If the network settings are ok, the module names are shown in black. • Mark the relevant modules by checking the relevant checkbox • Confirm with ”OK” and you are now ready for your first measurement job Connector pin-outs can be found in the operating manual. Channel configuration can be found in the online help of the software.
Ethernet settings: adapt your PC’s IP address If you use your computer in various networks (IP address changes), but your modules have a fixed IP address, you should use the ”Alternative Configuration” in the TCP/IP properties (fixed IP address and subnet mask, user-defined)! Edit the PC settings as follows: • Open the network connections (Start/Settings/Network connections). • Mark your LAN connection with a right−click and select ”Properties” in the context menu. • Select the tab ”General” and mark the Internet (TCP/IP) under ”This connection uses the following items”. Click on the ”Properties” button.
HBM
QuantumX
Connection
51
• In the tab ”Alternate Configuration”, select the option ”User-defined” and enter your data in the ”IP address line” and ”subnet mask”.
Example: Manual IP setting − PC side
QuantumX
Settings
IP address
Subnet mask
Module before
169.1.1.22
255.255.255.0
PC / Notebook
172.21.108.51
255.255.248.0
PC / Notebook after
169.1.1.1
255.255.255.0
HBM
52
Connection
• Confirm twice with ”OK”. In future, your computer will use the “Alternate Configuration” for the direct connection.
Integrating modules in an Ethernet Network • Activate the checkbox DHCP and click on ”OK”, the following confirmation window then appears:
• Confirm the settings with the ”Yes” button, the module is then restarted with the current settings. • After approx. 45 seconds click on the HBM
button. QuantumX
Connection
53
The system LED of the module should now light up green, if not please check your network settings again! If the network settings are ok, the module names are shown in black. • Mark the relevant modules by checking the relevant checkbox • Confirm with ”OK” Channel configuration can be found in the online help of the software.
QuantumX
HBM
54
5.2.5
Connection
Connection via FireWire (IEEE 1394b) General information
S S
Baud rate of 400 MBaud (approx. 50 MByte/s)
S S
Data synchronization
Asynchronous (all nodes) or isochronous (in real time to a specific node) data transmission Supply voltage via FireWire connection cable (max. 1.5 A)
10 V ... 30 V DC 1-NTX001 or 1-KAB271−3
X104
X101 1-KAB270-3
Adapter PC: PCI or PCIe card Notebook: PC card (previously PCMCIA)
Fig.5.6:
Single FireWire connection
NOTE Please check in advance whether a firmware or software update is required. Software/firmware downloads can be found on the HBM website: www.hbm.com\downloads
HBM
QuantumX
Connection
5.2.6
55
Multiple Ethernet connection with synchronization
10 V ... 30 V DC (e.g. NTX001)
10 V ... 30 V DC (e.g. NTX001)
X102 X101
FireWire connection 1-Kab269-2 2 m connection cable
Fig.5.7:
1−Kab269-0.2 0.2 m connection cable
1-Kab270-3 3 m connection cable
Example for multiple connection via FireWire with synchronization
Data are transferred, modules are synchronized in timing and voltage is supplied via the FireWire connections. You can connect a maximum of 12 modules in series with each other.
NOTE Different voltage sources need to supply the same voltage level.
QuantumX
HBM
56
5.2.7
Connection
FireWire setup • Integrate the FireWire PC adapter into your computer. • Install the ”t1394bus_installwizard.exe” driver Wizard provided by HBM from the QuantumX system CD or from the catmanAP CD (target directory, for example, c:\Programme\HBM\FireWire). Double−click to start the program.
NOTE For trouble shooting purposes you can switch to the original FireWire driver with the tool “t1394bus_installwizard.exe”. You can find the tool after having installed the driver on your hard disc. • After installation and configuration, connect the FireWire cable first to the PC adapter and then to the first module. Activation is confirmed acoustically by Windows. Whenever you connect a new module via FireWire to your computer, the operating system will prompt you to register the module. Always refer here to the driver “hbm1394.sys”. • Install and then start the latest QuantumX Assistant on your PC. • Click on the icon
(Find modules) or press the function key F4.
The field ”Modules found” displays all modules connected via FireWire. • If your modules are not yet displayed, mark ”Search the complete network” and click again on the
button.
• Mark the relevant modules by checking the relevant checkbox • Confirm with ”OK” This completes the settings for your connection. Connect your transducer now, connection information can be found in the ”QuantumX operating instructions”. Information about further channel configuration can be found in the online help of the HBM software you are using.
HBM
QuantumX
Connection
57
NOTE If no modules are found via FireWire this may be caused by the following: − The modules have not been properly registered. Click on the FireWire driver in Systray, check the driver behind the modules and reinstall it, if necessary (hbm1394.sys). − Check all connections between the modules.
QuantumX
HBM
58
5.2.8
Connection
Firmware update via Ethernet
We recommend updating the module firmware • When you want to use new HBM software • When you want to use a new module with a different firmware version Your PC software must also be updated • When you want to update the firmware of your module to improve its performance Use QuantumX Assistant as follows to determine which firmware version your module is working with: • Right-click on Module −> Details −> System parameters
• Compare your version with the latest firmware version on the Internet under: www.hbm.com\quantumX HBM
QuantumX
Connection
59
If the firmware version number of your module is lower than the current number on the Internet, an update is necessary: • Download the latest software package from the HBM website (QuantumX firmware downloader, QuantumX Assistant, etc.) • Close all running HBM software, install the new software and start QuantumX Firmware Updater • Download the latest firmware from the HBM website and save it in the download directory of the firmware updater (in most cases: C:\Programs\HBM\QuantumX Firmware Update\Download) • Click on the ”Find modules” symbol or press function key F4 (see chapter 2.1).
• Select the module • Select the required version in the dropdown menu ”New firmware” • Activate the modules whose firmware you want to update by checking the boxes in the column ”Update” and then click on the ”OK” button.
• Click on the ”Start” button and wait until the update is complete (do not interrupt the process/do not switch off the modules/do not interrupt the connection)
NOTE You can update the firmware of the modules via FireWire, directly via Ethernet or via the Ethernet connection of the Gateway CX27. QuantumX
HBM
60
5.2.9
Connection
Connecting more than 12 modules
1
2
3 Hub P3
P4 6
7
8
9
10
11
5 12
1
2
3
4
Displayed: Total number of modules: Total number of Hops:
15 12 Data sink
QuantumX module Longest chain to data sink (max. 12) P3 = Port 3
Fig.5.8:
Example of a star topology with two chains and one hub
The number of modules connected in series (daisy chain) is limited to 12. If you want to connect more modules (maximum 24), you must use hubs. Hubs are devices that connect network chains together in star configurations. This connection mode is again limited to 14 hops. A hop is the transition from one module to another (this means n−1 hops for n QuantumX modules in a chain). Depending on the connection situation, 1 to 2 hops are counted in one hub (see Fig.5.9). To count the total number of hops, the longest chain to the data sink must be counted (worst case). In FireWire systems that consist of several hubs with several module series, only the hops of the longest series are counted, the others are not taken into account.
HBM
QuantumX
Connection
61
Port 1 Port 2 Port 3
Port 1
Port 2 Phy 1
Port 3 Phy 2 Port 4
Port 4
Connection to Port 3 − Port 4: 1 Hop Connection to Port 1 and/or 2 − Port 4: 2 Hops
Fig.5.9:
Connection situation at AVT 1394b hub
Note Always connect the chain with the most modules to Port 3 or Port 4.
QuantumX
HBM
62
5.2.10
Connection
Bridging greater distances Greater distances (>5m) in FireWire networks can be bridged with opto-hubs that enable distances of up to 500 m with the use of a glass fiber cable.
1
1 Hop
2
OptoHub P3
1
2
3
4
Max. 500 m P4
P4
OptoHub
P3 7
8
6
5 1 Displayed: Total number of modules: Total number of hops:
11 8
QuantumX module Longest chain to data sink
Fig.5.10:
HBM
Example for the use of opto-hubs
QuantumX
Connection
5.2.11
63
FireWire with opto-hub and glass fiber cable
10 V ... 30 V DC
Glass fiber cable Max. length 500 m
Opto-Hub Opto-Hub
Fig.5.11:
QuantumX
FireWire and glass fiber cable
HBM
64
Connection
6
Transducer connection
6.1
Shielding design Sources of interference may cause electromagnetic fields which couple interference voltages into the respective measuring circuits inductively or capacitively through connecting cables and equipment housings and thus affect the way the equipment works. It is essential to ensure that the equipment used in the system itself does not emit electromagnetic disturbances. Electromagnetic compatibility (EMC) which combines the concepts of required electromagnetic immunity or susceptibility (EMS) and permitted electromagnetic interference (EMI) has become increasingly more important over the past years. The HBM Greenline shielding design The complete measuring chain is fully enclosed in a Faraday cage through the unique design of the cable shield. The cable shield is flattened, connected to the transducer housing, and routed to the amplifier housing through conductive connectors. These measures substantially reduce the effects of electromagnetic disturbances.
Conductive housing ensuring connection to plug and equipment housing Signal-carrying contacts
Fig.6.1:
HBM
Cable shield connected to conductive housing through strain relief
Cable shield connected to plug
QuantumX
Connection
65
Note All components of the measuring chain (including all cable joints such as connectors and couplings) have to be surrounded by a closed, EMC-tested shield. Shield transitions must be realized in the form of flattened, closed and low-impedance joints. This is ensured when original HBM connectors are used. Grounding and earthing EMC−compliant wiring requires that signal ground and shield are separate, therefore the shielding can be connected to earth at several points, for example, through the transducers (metal housing) and the amplifier (housing connected to protective earth conductor). In the case of potential differences in the measuring system, a potential equalization conductor has to be laid (recommended: very flexible stranded wire, 10 mm2 line diameter). Ensure that signal and data lines are run separately from current−carrying power lines. Ideally use metal cable ducts with internal separating plate. Signal ground, earth and shield should be separate from one another. To minimize the effect of electromagnetic disturbances and potential differences, the signal ground and earth (or shield) are separated in HBM equipment. The mains protective earth conductor or a separate earth potential line should be used as connection to earth. This is common practice, for instance, for potential equalization in buildings. Avoid connecting the earth lead to a radiator, water pipe or similar.
QuantumX
HBM
66
6.2
Connection
Adjustable sensor supply for active transducers Some modules can supply active transducers with an excitation voltage of 5…24 V. When using the adjustable transducer excitation, electrical isolation from the supply voltage of the amplifier is not required. The maximum permissible power consumption is 700 mW per channel, in total not more than 2 W. If the power consumption is more than 700 mW on one channel, the transducer excitation of this channel will switch off. If the power consumption exceeds a total of 2 W, the device may switch off.
MX840
1 ... 8 Connection as per measurement principle
Sensor
4 9 Supply voltage 5V ... 24V1) Power supply 0V Cable shield
1)
Adjustable via software
12 3 kΩ
11 Hsg.
Hsg. = housing
CAUTION The output voltage during idling is up to 20% higher than set when the voltage <8 V. If a current of 2 mA is drawn, the voltage falls back to the set value. The condition is met in sensors with a power consumption of min. 20 mW. This can also be met with an additional resistance of 3 kΩ if the power consumption is lower or if the sensors are sensitive. Check the correct voltage setting when connecting a sensor. Too high a voltage can destroy the sensor. The voltage value is a part of the MX840 parameterization and can only be changed with a new parameterization. The sensor supply is switched off in the delivery condition.
HBM
QuantumX
Connection
6.3
67
TEDS The acronym TEDS stands for ”Transducer Electronic Data Sheet” and denotes a transducer or sensor’s electronic data sheet which is saved to a small electronic chip or a corresponding module and is intrinsically linked to the transducer. In addition, valuable meta data is supplied, for example, calibration data providing essential information for the traceability of measurements or tests. The electronic data sheet can be located in the transducer housing, in the inseparable cable or connector plug. The function and principle of operation are defined in Standard IEEE1451.4.
Transducer with TEDS module e.g. U93
Data mode Measurement mode
QuantumX MX840A
Transducer information stored in TEDS includes:
S
The physical unit of the measured quantity (N for force, for example) and its measuring range
S S
The unit of the electrical output signal (mV/V for bridge transducers, for example)
S
If applicable, the transducer’s required excitation and electrical power supply
The linear characteristics as the relation between the measured quantity and the electrical signal
Additional information that can be read out using corresponding software, for example:
S S
Transducer manufacturer, type, serial number etc. Calibration date, recalibration interval, calibrator’s initials, etc.
Amplifiers of the QuantumX series enable transducer information stored in the electronic data sheet to be read out and automatically implemented in the correct amplifier settings; this allows fast and reliable measurement. The electronic data sheet is automatically loaded, when the transducer is connected to the device. Transducers are identified via the bridge between two pins in the connector. After the digital identification mode, the amplifier automatically switches to the configured measurement mode. TEDS data can also be loaded via a software command, e.g. using catmanAP. TEDS Editor enables all TEDS data to be read out and edited; see chapter 3.6.
QuantumX
HBM
68
Connection
QuantumX supports multiple options for reading or writing TEDS data:
S
One TEDS module can be accessed via two separate cable wires (”one-wire circuit”) or the transducer plug can be retrofitted with TEDS.
S
Amplifiers enabling IEPE transducers to be directly connected support TEDS version 1.0.
S
Some HBM transducers feature a special integrated TEDS module which enables TEDS data to be transmitted via the transducer’s feedback line (patented ”zero-wire circuit”). After digital communication (data mode), the amplifier switches to the measurement mode. The U93 force transducer is an example for such a transducer.
S
Thermocoulpe amplifiers with an RFID chip integrated in the transducer plug use TEDS technology, for example, to electronically connect the measuring point with the transducer.
The data sheet to the respective amplifier provides more specifications relating to TEDS, for example the maximum permitted cable length to the transducer. If no TEDS is used, the permitted cable length can be substantially longer.
Note Further information about TEDS topics can be found at http://www.hbm.com/teds
HBM
QuantumX
Connection
6.3.1
69
Retrofitting the transducer plug with TEDS The IEEE standard 1451.4 defines a generally accepted process which enables sensors to be identified. The sensor is identified via the respective data sheet which is stored in electronic format in the sensor, cable or plug on a 1−wire EEPROM (TEDS − Transducer Electronic Data Sheet). The amplifier uses a serial 1−wire interface for communication with this EEPROM, reads out the data sheet and configures the amplifier accordingly. Retrofilling TEDS in the plug is shown in the figure below. 6 1
11
5
15 10
4 Bridge 9
6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (Maxim DS2433+)
QuantumX
View from below
HBM
70
6.4
Connection
MX840 universal amplifier You can connect up to eight transducers to the universal amplifier MX840. The transducers are connected via 15-pin D-SUB-15HD device connectors. All measuring channels are electrically isolated von one another and from the mains. When using the adjustable transducer excitation, electrical isolation from the supply voltage of the amplifier is not required. MX840 connectable transducers
HBM
Transducer type
Connector sockets
see page
Full bridge strain gages
1 ... 8
97
Inductive transducers
1 ... 8
98
Inductive transducers
1 ... 8
82
LVDT
1 ... 8
85
Voltage
1 ... 8
87, 88
Current
1 ... 8
89
Current fed piezoelectric transducer (IEPE, ICP)
1 ... 8
106
Piezoresistive transducer
1 ... 8
80
PT100, PT1000 resistance thermometers
1 ... 8
91
Potentiometers
1 ... 8
84
Thermocouples
1 ... 8
92
Incremental encoders
5 ... 8
from 94
SSI protocol
5 ... 8
100
QuantumX
Connection
71
MX840 connectable transducers (continued)
QuantumX
Torque/rotary speed
5 ... 8
117, 118
Frequency measurement, pulse counting
5 ... 8
from 94
CANbus
1
103
HBM
72
6.4.1
Connection
MX840 pin assignment So that insertion or removal of a transducer connection can be unmistakably identified, Pin 4 and Pin 9 in the connector plug must be bridged! If this bridge is missing, no measurement values will be recorded at the connection!
Bridge
6 1
11
9 4 5
Fig.6.2:
Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
15
Pin arrangement of connection plug, view from the solder side
Connection TEDS (+) Bridge Excitation voltage (−), 0° reference pulse (zeroing pulse) (−) Bridge Excitation voltage (+), 0° reference pulse (zeroing pulse) (+) Always connect with Pin 9! (Plug identification) Measurement signal (+), potentiometer measurement signal(+), voltage input 100 mV (+), f1(−) signal differential, SSI data (−) TEDS (−), ground frequency measurement Sense lead (−), f2(−) signal differential, CAN-High, SSI clock (−) Sense lead (+), f2(+) signal differential, CAN-Low, SSI clock (+) Signal ground Measurement signal (−), f1(+) signal differential, SSI data (+) Active sensor supply 5 ... 24 V (0 V) Active sensor supply 5 ... 24 V (+) Current input "30 mA (+) Voltage input 10 V (+), 60 V (+) Vacant
NOTE Many HBM transducers are fitted with 15-pin D-SUB connectors (2 rows). The adapter cable 1-KAB416 can be used for connection to the 3 row D-SUB-15HD device connectors of the MX840. Pins 4 and 9 are already bridged in this adapter cable (see chapter 8.4.3).
HBM
QuantumX
Connection
6.4.2
73
MX840 status display The front panel of the universal amplifier has a system LED and 8 connection LEDs. The system LED indicates the status of the device, the connection LEDS the states of the individual connections. Connection LED
System LED
Fig.6.3:
MX840 front view
System LED Green
Error-free operation
Orange
System is not ready, boot procedure running
Flashing orange
Download active, system is not ready
Red
Error
Connection LEDs All LEDs are orange
Boot procedure running (system is not ready)
All LEDs are flashing orange
Firmware download active (system is not ready)
Orange
Connection newly assigned, transducer identification running (calibration)
Green
Error-free operation (”Ignore TEDS” or ”if available” set, channel is manually configured)
Flashing green (5s), then green
Reading TEDS data
Flashing orange (5s), then green
Manual configuration running (ignore TEDS)
Red
No sensor connected Channel error (incorrectly parameterized, connection error, invalid TEDS data)
CAN LEDs Green
CAN bus activated, CAN data can be received
Orange
CAN bus in ”WARNING” state, CAN data received but bus is occasionally disturbed; buffer overflow, individual data lost
Red
CAN bus in ”ERROR” or ”BUS-OFF” state, CAN data cannot be received or processed
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
QuantumX
HBM
74
6.5
Connection
MX840A universal amplifier The MX840A universal amplifier corresponds to the MX840 amplifier, with the additional option to connect strain gage half bridges and variable ohmic resistors. MX840A connectable transducers
HBM
Transducer type
Connector sockets
see page
Full-bridge strain gages
1 ... 8
97
Inductive transducers
1 ... 8
98
Half-bridge strain gages
1 ... 8
100
Inductive transducers
1 ... 8
82
LVDT
1 ... 8
85
Voltage
1 ... 8
87, 88
Current
1 ... 8
89
Current fed piezoelectric transducer (IEPE, ICP)
1 ... 8
106
Piezoresistive transducer
1 ... 8
80
Resistance
1 ... 8
90
Potentiometers
1 ... 8
84
PT100, PT1000 resistance thermometers
1 ... 8
91
Thermocouples
1 ... 8
92
Incremental encoders
5 ... 8
from 94
SSI protocol
5 ... 8
100
QuantumX
Connection
75
MX840A connectable transducers (continued)
QuantumX
Frequency measurement, pulse counting
5 ... 8
from 94
Pulse width modulation (PWM)
5 ... 8
124
Torque/Rotational speed
5 ... 8
117, 118
CANbus
1
103
HBM
76
6.5.1
Connection
MX840A pin assignment So that insertion or removal of a transducer connection can be unmistakably identified, Pin 4 and Pin 9 in the connector plug must be bridged! If this bridge is missing, no measurement values will be recorded at the connection!
Bridge
6 1
11
9 4 5
Fig.6.4:
Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
15
Pin arrangement of connection plug, view from the solder side
Connection TEDS (+) Bridge Excitation voltage (−), 0° reference pulse (zeroing pulse) (−) Bridge Excitation voltage (+), 0° reference pulse (zeroing pulse) (+) Always connect with Pin 9! (Plug identification) Measurement signal (+), potentiometer measurement signal(+), voltage input 100 mV (+), f1(−) signal differential, SSI data (−) TEDS (−), ground frequency measurement Sense lead (−), f2(−) signal differential, CAN-High, SSI clock (−) Sense lead (+), f2(+) signal differential, CAN-Low, SSI clock (+) Signal ground Measurement signal (−), f1(+) signal differential, SSI data (+) Active sensor supply 5 ... 24 V (0 V) Active sensor supply 5 ... 24 V (+) Current input "30 mA (+) Voltage input 10 V (+), 60 V (+) Calibration signal T10F(S) and T40, 5 V/max. 10 mA
NOTE Many HBM transducers are fitted with 15-pin D-SUB connectors (2 rows). The adapter cable 1-KAB416 can be used for connection to the 3 row D-SUB-15HD device connectors of the MX840. Pins 4 and 9 are already bridged in this adapter cable (see chapter 8.4.3).
HBM
QuantumX
Connection
6.5.2
77
MX840A status display The front panel of the universal amplifier has a system LED and 8 connection LEDs. The system LED indicates the status of the device, the connection LEDS the states of the individual connections. Connection LED
System LED
Fig.6.5:
MX840A front view
System LED Green
Error-free operation
Orange
System is not ready, boot procedure running
Flashing orange
Download active, system is not ready
Red
Error
Connection LEDs All LEDs are orange
Boot procedure running (system is not ready)
All LEDs are flashing orange
Firmware download active (system is not ready)
Orange
Connection newly assigned, transducer identification running (calibration)
Green
Error-free operation (”Ignore TEDS” or ”if available” set, channel is manually configured)
Flashing green (5s), then green
Reading TEDS data
Flashing orange (5s), then green
Manual configuration running (ignore TEDS)
Red
No sensor connected Channel error (incorrectly parameterized, connection error, invalid TEDS data)
CAN LEDs Green
CAN bus activated, CAN data can be received
Orange
CAN bus in ”WARNING” state, CAN data received but bus is occasionally disturbed; buffer overflow, individual data lost
Red
CAN bus in ”ERROR” or ”BUS-OFF” state, CAN data cannot be received or processed
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
QuantumX
HBM
78
6.6
Connection
MX440A universal amplifier You can connect up to eight transducers to the universal amplifier MX440A. The transducers are connected via 15-pin D-SUB-15HD device connectors. All measuring channels are electrically isolated von one another and from the mains. The transducer types that can be connected and the status display are identical to those of the MX840A universal amplifier (see page 74).
System LED
Connection LED Fig.6.6:
HBM
MX440A front view
QuantumX
Connection
6.7
79
MX410 highly dynamic universal amplifier You can connect up to four transducers to the highly dynamic universal amplifier MX410. The transducers are connected via 15-pin D-SUB-15HD device connectors. You will need BNC adapters (accessories 1-IEPE-MX410) for the connection of IEPE transducers. All measuring channels are electrically isolated von one another and from the mains. When using the adjustable transducer excitation, electrical isolation from the supply voltage of the amplifier is not required. MX410 Connectable transducers
QuantumX
Transducer type
Connector sockets
see page
Full-bridge strain gages
1 ... 4
97
Half-bridge strain gages
1 ... 4
100
Current fied resistiv transducer
1 ... 4
80
Inductive transducers
1 ... 4
98
Inductive transducers
1 ... 4
82
Voltage
1 ... 4
87, 88
Current
1 ... 4
89
Current fed piezoelectric transducer (IEPE, ICP)
1 ... 8
106
HBM
80
6.7.1
Connection
MX410 pin assignment So that insertion or removal of a transducer connection can be unmistakably identified, Pin 4 and Pin 9 in the connector plug must be bridged! If this bridge is missing, no measurement values will be recorded at the connection!
Bridge
6 1
11
9 4 5
Fig.6.7:
15
Pin arrangement of connection plug, view from the solder side
Pin 1 2 3 4 5 6 7 8 9 10 11 12 13
Connection TEDS (+) Bridge Excitation voltagen (−) Bridge Excitation voltage (+) Always connect with Pin 9! (Plug identification) Measurement signal (+) TEDS (−) Sense lead (−) Sense lead (+) Signal ground Measurement signal (−) Active sensor supply (−) Active sensor supply (+)
14 15
Voltage input 10 V, IEPE (+) Reset external charge amplifier
Current input "30 mA (+)
The analog output can be tapped via BNC. Please see chapter 7 ”Functions and outputs” for information about the configuration.
NOTE Many HBM transducers are fitted with 15-pin D-SUB connectors (2 rows). The adapter cable 1-KAB416 can be used for connection to the 3 row D-SUB-15HD device connectors of the MX840. Pins 4 and 9 are already bridged in this adapter cable.
HBM
QuantumX
Connection
6.7.2
81
MX410 status display
Connection LED
System LED
Analog output LED
Fig.6.8:
MX410 front view
System LED Green
Error-free operation
Orange
System is not ready, boot procedure running
Flashing orange
Download active, system is not ready
Red
Error
Connection LEDs All LEDs are orange
Boot procedure running (system is not ready)
All LEDs are flashing orange
Firmware download active (system is not ready)
Orange
Connection newly assigned, transducer identification running (calibration)
Green
Error-free operation (”Ignore TEDS” or ”if available” set, channel is manually configured)
Flashing green (5s), then green
Reading TEDS data
Flashing orange (5s), then green
Manual configuration running (ignore TEDS)
Red
No sensor connected Channel error (incorrectly parameterized, connection error, invalid TEDS data)
Red
Sensor supply overload
Analog output LEDs Green
Error-free operation (”Ignore TEDS” or ”if available” set, channel is manually configured)
Orange
System is not ready, boot procedure running
Rot
Over current at analog output Orange
Input signal overload
Rot
Overload resulting from invalid scaling of analog output
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
QuantumX
HBM
82
6.8
Connection
MX460 frequency measuring amplifier You can connect up to four transducers to the frequency measuring amplifier MX460. The transducers are connected via 15-pin D-SUB-15HD device connectors. All measuring channels are electrically isolated von one another and from the mains. When using the adjustable transducer excitation, electrical isolation from the supply voltage of the amplifier is not required. MX460 connectable transducers
HBM
Transducer type
Connector sockets
see page
Torque/rotary speed
1 ... 4
100
Frequecy mesurement, pulse counting
1 ... 4
from 94
Puls width, Pulse duration, Period duration
1 ... 4
102
Passive indctive rotary encoder
1 ... 4
101
Incremental encoders
5 ... 8
from 94
QuantumX
Connection
6.8.1
83
MX460 pin assignment So that insertion or removal of a transducer connection can be unmistakably identified, Pin 4 and Pin 9 in the connector plug must be bridged! If this bridge is missing, no measurement values will be recorded at the connection!
Bridge
6 1
11
9 4 5
Fig.6.9:
Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
QuantumX
15
Pin arrangement of connection plug, view from the solder side
Connection TEDS (+) Reference pulse 0° (zeroing pulse) (−) Reference pulse 0° (zeroing pulse) (+) Always connect with Pin 9! (Plug identification) Frequency input f1 (−) TEDS (−), signal ground Frequency input f2 (−) Frequency input f2 (+) Reference voltage Vref (2.5 V) Frequency input f1 (+) Active sensor supply 5 ... 24 V (−) Active sensor supply 5 ... 24 V (+) No function f1 AC+ (for passive inductive transducers) Calibration signal T10F(S) and T40, 5 V/max. 10 mA
HBM
84
6.8.2
Connection
MX460 status display
Connection LED
System LED
Fig.6.10:
MX460 front view
System LED Green
Error-free operation
Orange
System is not ready, boot procedure running
Flashing orange
Download active, system is not ready
Red
Error
Connection LEDs All LEDs are orange
Boot procedure running (system is not ready)
All LEDs are flashing orange
Firmware download active (system is not ready)
Orange
Connection newly assigned, transducer identification running (calibration)
Green
Error-free operation (”Ignore TEDS” or ”if available” set, channel is manually configured)
Flashing green (5s), then green
Reading TEDS data
Flashing orange (5s), then green
Manual configuration running (ignore TEDS)
Red
No sensor connected Channel error (incorrectly parameterized, connection error, invalid TEDS data)
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
HBM
QuantumX
Connection
6.9
85
MX1609 / 1609Pmodule Up to 16 type K thermocouples (Ni-CrNi) can be connected to the module MX1609 for measuring temperatures. MX1609/MX1609P connectable transducers Transducer type
Connector sockets
see page
Thermocouple, type K
1 ... 16
114
MX1609 − +
Thermocouple
white
green
−
−
+
+
Type
Thermocouple material 1 (+)
Thermocouple material 2 (−)
K
Nickel-chrome (color code green)
Nickel-aluminum (color code white)
Post−scaling The MX1609 has a post−scaling function. A table that converts degree oC to degree oC enables errors resulting from thermocouples and mounting conditions to be minimized. The MX1609 processes a maximum of 64 pairs of values. The ”Calibration Table” TEDS template enables 14 pairs of values to be saved, if no additional optional templates are being used. This function provides the best results when the ambient temperature of the MX1609 and thus the temperature of the cold junction is kept constant.
QuantumX
HBM
86
Connection
Connection of the thermocouple plug in the miniature design.
HBM
QuantumX
Connection
6.9.1
87
Measuring point identification with RFID RFID1) for measuring point identification An RFID chip in or on the thermocouple plug ensures wireless transducer identification by the amplifier. RFID technology enables contactless reading and writing of data, for example of a specific measuring point or the desired physical unit (5C or 5K). Data is input using TEDS Editor provided by HBM. Then data is written onto the RFID chip using an appropriate RFID transponder in the amplifier. The chip is reusable and works without batteries.
Conditions for using RFID chips for measuring point identification: − All channels can read/write via RFID − The neighboring channel must not be occupied in the MX1609 during writing − Maximum distance chip to housing: 1 mm − For self-assembly: Check position of chip on plug
Thermocouple plug with integrated RFID chip from HBM
1-THERMO-RFID (integrated by HBM)
The chip for measuring point identification is already integrated in the HBM THERMO-MINI.
1)
QuantumX
RFID = Radio Frequency Identification: Method for communication between transponder and read/write device with magnetic fields or electromagnetic waves.
HBM
88
Connection
RFID for self−assembly on thermocouple plugs
Installation position of RFID Wide side
Glue 1-THERMO-RFID on here
+
Top view +
Glue the RFID chip as shown above with an adhesive (recommended: Two-component adhesive).
HBM
QuantumX
Connection
6.9.2
89
MX1609/MX1609-P status display
Connection LED
Fig.6.11:
System LED
MX1609 front view
System LED Green
Error-free operation
Orange
System is not ready, boot procedure running
Flashing orange
Download active, system is not ready
Red
Error
Connection LEDs All LEDs are orange
Boot procedure running (system is not ready)
All LEDs are flashing orange
Firmware download active (system is not ready)
Orange
Connection newly assigned, transducer identification running (calibration)
Green
Error-free operation (”Ignore TEDS” or ”if available” set, channel is manually configured)
Flashing green (5s), then green
Reading TEDS data
Flashing orange (5s), then green
Manual configuration running (ignore TEDS)
Red
No sensor connected Channel error (incorrectly parameterized, connection error, invalid TEDS data)
Red
No sensor connected Channel error (incorrectly parameterized, connection error, invalid TEDS data)
Red
Sensor supply overload
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
QuantumX
HBM
90
Connection
6.10
MX471 CAN module
6.10.1
General information The MX471 module provides four independent CANbus nodes that are all electrically isolated from each other and the mains. MX471 busses that can be connected Typ
Connector sockets / nodes
see page
CANbus (High-Speed CAN)
1 ... 4
125
Connected devices are not directly addressed during data transmission on a CANbus. A unique identifier denotes the contents of a message (e.g. rotational speed or engine temperature). The identifier also signals the priority of the message. Message = identifier + signal + additional information Devices connected to the bus = nodes Every node on the MX471 can be parameterized either as receiver or as sender (gateway). Parameterization as receiver is described in chapter 6.10.3. Parameterization as sender is described in chapter 7. The online help coming with the respective software package provides detailed information about parameterization.
Note To ensure normal operation the CANbus needs to be terminated at both ends (and only there) using appropriate termination resistors. A 120−ohm termination resistor can be individually connected in the module by software. Termination is also required when short cables with low bit rates are used. Please refer to the data sheet for the relation between bit rate and maximum bus line length. The configuration of a node is retained after switching the modules off and on. For decoding signals at a rate greater than 2000/s, please set up signal inputs 1 to 8 on the MX471. The signal buffers of these signal inputs have been expanded accordingly.
HBM
QuantumX
Connection
6.10.2
91
MX471 pin assignment
1
5
Fig.6.12:
Pin 1 2 3 4 5 6 7 8 9
QuantumX
6
9
Pin arrangement of connection plug, view from the solder side
Connection no function CAN−Low GND no function CAN Shield GND CAN−High no function no function
HBM
92
6.10.3
Connection
Receiving CAN messages To enable CAN messages to be received, it is essential that the relevant messages can be identified by the node. This may be done directly at the node or, in a reproducible way, by previously generated messages in the sensor database. Individual messages can be linked to the node by drag & drop from the sensor database. Type *.dbc CAN databases, too, can be loaded into the sensor database. Should a CAN database not be available, it can be generated by the user. Editors for this purpose are provided by different companies. Received CAN messages are instantly ”time−stamped” in the measurement mode. This enables directly acquired measured quantities and CAN messages to be acquired and analyzed in parallel and synchronously in the entire system.
Note MX471 is not a classical data logger that logs the complete CAN data stream on the bit level. The parameterized node ”listens” on the CANbus and extracts the signals from the relevant CAN messages to transmit them as measured values.
HBM
QuantumX
Connection
6.10.4
93
LED status display
CAN LEDs “BUS”
System LED
CAN LEDs “Channel”
System LED: Green
Error−free operation
Yellow
System is not ready, boot procedure running
Flashing yellow
Download active, system is not ready
Red
Error, faulty synchronization
CAN LEDs (BUS): Green flickering
No bus error and CAN activity
Constant green
No bus error and no CAN activity
Yellow flickering
Intermittent bus errors (Warning) and CAN activity
Constant yellow
Intermittent bus errors (Warning) and no CAN activity
Red on
Bus error, CAN interface in ”Bus−OFF” status
CAN LEDs (Channel): Green flickering
No bus error and CAN activity
Constant green
No bus error and no CAN activity
Yellow flickering
Intermittent bus errors (Warning) and CAN activity
Constant yellow
Intermittent bus errors (Warning) and no CAN activity
Red on
Bus error, CAN interface in ”Bus−OFF” status
Ethernet LED:
QuantumX
Green on
Ethernet link status is ok
Flashing yellow
Ethernet data transmission ongoing
HBM
94
6.11
Connection
MX1601 universal amplifier Up to 16 freely configurable inputs for voltage (10 V, 100 mV) or current (20 mA) or current−fed piezoelectric sensors (IEPE) can be connected to the MX1601 universal amplifier. The transducers are connected via 8−pin plug terminal connectors (Phoenix Contact FMC 1,5/8−ST−3,5−RF (order no. 1952089)). All measuring channels are electrically isolated von one another and from the mains. When using the adjustable transducer excitation, electrical isolation from the supply voltage of the amplifier is not required. MX1601 connectable transducers
HBM
Transducer type
Connector sockets
see page
Voltage
1 ... 16
109, 110
Current
1 ... 16
111
Current fed piezoelectric transducer (IEPE, ICP)
1 ... 16
106
QuantumX
Connection
6.11.1
95
MX1601pin assignment So that insertion or removal of a transducer connection can be unmistakably identified, Pin 2 and Pin 5 in the connector plug must be bridged! If this bridge is missing, no measurement values will be recorded at the connection!
1 2 Bridge
3 4 5
1 2 Data 3 no function
1
2 3
1-Wire-EEPROM (Maxim DS2433+) Bottom view
6 7 8
Fig.6.13:
Pin 1 2 3 4 5 6 7 8
Pin assignment of the connector plug, view from the connection side
Connection Voltage input 10 V (+), 100 mV (+), IEPE (+) Signal ground, TEDS (−) Current input 20 mA (+) TEDS (+) Always connect with Pin 2! (Plug identification) Active sensor supply (+) Active sensor supply (−) Housing (shield connection)
NOTE The transducer excitation voltage can be set in the range of 5 ... 24 V and is only available at channels 1 ... 8. At channels 9 ... 16, the supply voltage (10 V ... 30 V) is output less approx. 1 V. A current drain of max. 30 mA is possible; the current limiter cuts the transducer excitation voltage at a higher current drain. QuantumX
HBM
96
6.11.2
Connection
MX1601status display The front panel of the universal amplifier has a system LED and 16 connection LEDs. The system LED indicates the status of the device, the connection LED the status of the individual connections. Connection LED
System LED
Connection LED Fig.6.14:
MX1601 front view
System LED Green
Error-free operation
Orange
System is not ready, boot procedure running
Flashing orange
Download active, system is not ready
Red
Error
Connection LEDs All LEDs are orange
Boot procedure running (system is not ready)
All LEDs are flashing orange
Firmware download active (system is not ready)
Orange
Connection newly assigned, transducer identification running (calibration)
Green
Error-free operation (”Ignore TEDS” or ”if available” set, channel is manually configured)
Flashing green (5s), then green
Reading TEDS data
Flashing orange (5s), then green
Manual configuration running (ignore TEDS)
Red
Amplifier overload, No sensor connected Channel error (incorrectly parameterized, connection error, invalid TEDS data)
Red
Sensor supply overload
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
HBM
QuantumX
Connection
97
6.12
Transducder technologies
6.12.1
Strain gages full bridges Supported by modules: MX840, MX840A, MX440A, MX410
6 1
11
5
15 10
wh
2 1
4 3
bk
Measurement signal (+) Excitation (−)
5 2
bu
Excitation (+)
3
rd
Measurement signal (−)
10
ye
Cable shield
gn
Sense lead (+)
8
gy
Sense lead (−)
7
Hsg.
4 9 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (view from below)
Hsg. = housing
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
QuantumX
HBM
98
6.12.2
Connection
Strain gage full bridges, inductive full bridges Supported by modules: MX840, MX840A, MX440A, MX410
6 1
11
5
15 10
Measurement signal (+) Excitation (−)
5 2
bu
Excitation (+)
3
rd
Measurement signal (−) Cable shield
10
wh bk
2 1
4 3
ye
Hsg.
gn
Sense lead (+)
8
gy
Sense lead (−)
7 4 9 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (view from below)
Hsg. = housing
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
HBM
QuantumX
Connection
6.12.3
99
Piezoresistive full bridge Supported by modules: MX840, MX840A, MX440A
1 ... 8 6 1
11
5
15 10
Measurement signal (+) Excitation (−) Excitation (+) Measurement signal (−) Cable shield
2 1
4 3
5 2 3 10 Hsg.
Sense lead (+)
8
Sense lead (−)
7 4 9 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (view from below)
QuantumX
Hsg. = housing
HBM
100
6.12.4
Connection
Strain gage half bridge Supported by modules: MX840A, MX440A, MX410
6 1
11
5
15 10
wh bk bu
ye gn gy
Measurement signal (+) Excitation (−)
5
Excitation (+)
3
Cable shield
2
Hsg.
Sense lead (+)
8
Sense lead (−)
7 4 9 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (view from below) Hsg. = housing
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
HBM
QuantumX
Connection
6.12.5
101
Inductive half bridges Supported by modules: MX840, MX840A, MX440A, MX410
6 1
11
5
15 10
wh bk bu
ye
Measurement signal (+) Excitation (−)
5
Excitation (+)
3
Cable shield
gn gy
2
Hsg.
Sense lead (+)
8
Sense lead (−)
7 4 9 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (optional) view from below
Hsg. = housing
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
QuantumX
HBM
102
6.12.6
Connection
Quarter bridge, strain gage For connecting single quarter−bridge strain gages in three−wire configuration, an adapter can be plugged onto the following modules: MX840, MX840A, MX440A, MX410 Available adapters: Strain gage with 120 ohm: order number: SCM−SG120 Strain gage with 350 ohm: order number: SCM−SG350
2
1
1’
Strain gage 120 / 350
See leaflet: QuantumX/SCM−SG120/350 for technical details.
HBM
QuantumX
Connection
6.12.7
103
Connection of transducers with double shield technique
6 1
11
5
15 10
Measurement signal (−) Measurement signal (+) Excitation (−) 2 1
Excitation (+) 4
3
Cable shield
10 5 2 3 Hsg.
Sense lead (+)
8
Sense lead (−)
7 4
RB / 2
9
(on transducer) Hsg. = housing
We recommend this connection technique for very small measurement ranges, in particularly disturbed environments and when using long cables. This applies for all bridge connections. With cable lengths >50 m, a resistor with half the value of the bridge resistance (RB/2) must be connected in each sense lead of the transducer.
QuantumX
HBM
104
6.12.8
Connection
Potentiometric transducers Supported by modules: MX840, MX840A, MX440A
6 1
11
5
15 10
2
1 3
Measurement signal (+)
5
Bridge excitation voltage (−)
2
Bridge excitation voltage (+)
3
Cable shield 1)
Case
Sense lead (+)
8
Sense lead (−)
7 4 9 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (optional) (view from below)
HBM
QuantumX
Connection
6.12.9
105
LVDT transducers Supported by modules: MX840, MX840A, MX440A
6 1
11
5
15 10
LVDT transducer
Measurement signal (+) Excitation (−) Excitation (+) Measurement signal (−) Cable shield
5 2 3 10 Hsg.
Sense lead (+)
8
Sense lead (−)
7 4 9 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (view from below) Hsg. = housing
QuantumX
HBM
106
6.12.10
Connection
Piezoelectric sensors IEPE Current fed piezoelectric transducers are supplied with a constant current of e.g. 5.5 mA and feed a voltage signal to the amplifier. This transducer type is also called IEPE or ICP transducer. IEPE means ”Integrated Electronics Piezo−Electric”. ICP is a registered trademark of PCB Piezotronics Is supported by the following modules: MX410 (direct to SubHD or via BNC adapter), MX1601 (directly to the connector) MX840, MX840A, MX440A with 10 V analog input and 24 V supply via a Smart module
6 1
11
5
15 10
Option BNC adapter (1-IEPE-MX410)
(−)
4 9
(+)
14
IEPE
Cable shield
Hsg.
Hsg. = housing
HINWEIS IEPE transducers with TEDS version 1.0 are supported.
HBM
QuantumX
Connection
107
Connection diagram for MX840, MX840A, MX440A with external Smart module:
6 1
11
5
15 10
Maximum input voltage to housing and signal ground : "60 V
PIN: 2 (+) 6 BNC
white
green
3
Cable shield
5
24 V
4
4
Supply voltage zero (−)
IEPE
14
9 Hsg.
red
12
black
0V
Smart module (1−EICP−B−2)
11 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (optional) (view from below)
Hsg. = housing
Accessories for connecting the Smart module: Product
QuantumX
Description
Order No.:
Smart module
External 24 V signal conditioning module, supplying IEPE with constant current (BNC socket) and feeding standardized " 10 V voltage signal.
1−EICP−B−2
Connection cable
Cable between Smart module and SubHD plug
1−SAC−EXT−MF−x−2 (x = length in meter)
Male device connector
QuantumX connector
1−SubHD15−MALE
HBM
108
Connection
Retrofilling the TEDS chip in the transducer plug when using the Smart module: The Smart module cannot read TEDS data directly saved to the IEPE transducer. The QuantumX plug can be retrofitted with TEDS to enable the Smart module to be read and to automate channel configuration according to the IEPE transducer. TEDS Editor enables settings to be made.
S
TEDS−specific settings: f ”High Level Voltage Output Sensor” f Physical Measurand: Acceleration (m2/s or g) f Electrical range: standard +/− 10 V f Power requirements: required
S S
Complete the data sheet according to the transducer. Excitation level: 24 V nominal
CAUTION Replacing the IEPE transducer on the Smart module may result in incorrect device settings.
HBM
QuantumX
Connection
6.12.11
109
DC voltage sources 100 mV Supported by modules: MX840, MX840A, MX440A, MX1601 Pin assignment for the MX1601 module see chapter 6.11.1
6 1
11
5
Maximum input voltage to housing and signal ground : "60 V
U
15 10
(+)
5
(−)
9 10 4
Cable shield
Hsg.
6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (optional) (view from below)
Adjustable sensor supply:
QuantumX
Pin 12: Pin 11:
Hsg.=Housing
5 V ... 24 V 0V
HBM
110
6.12.12
Connection
DC voltage sources 10 V and 60 V Supported by modules: MX840, MX840A, MX410, MX440A Voltage ranges supported by the amplifiers: 10 V and 60 V: MX840, MX840A, MX440A 10 V: MX410, MX1601 Pin assignment for the MX1601 module see chapter 6.11.1
6 1
11
5
15 10
Maximum input voltage to housing and signal ground: "60 V
U
14
(+)
4
Supply voltage zero (−)
9
Cable shield
Hsg.
6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (optional) (view from below)
Adjustable sensor supply:
Pin 12: Pin 11:
Hsg.=Housing
5 V ... 24 V 0V
Two measuring ranges (10 V or 60 V) can be selected, depending on the parameterization. Wrong parameterization does not result in destruction of the amplifier.
HBM
QuantumX
Connection
6.12.13
111
DC power sources 20 mA Supported by modules: MX840, MX840A, MX440A, MX410, MX1601 Pin assignment for the MX1601 module see chapter 6.11.1
6 1
11
5
Maximum input voltage to housing and signal ground: "60 V
(−)
15 10
9 4
I (+)
13 Hsg.
Cable shield
6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (view from below)
Adjustable sensor supply:
Hsg. = housing
Pin12: Pin 11:
5 V ... 24 V 0V
Maximum current "30 mA with an internal burden of 10 Ω.
QuantumX
HBM
112
6.12.14
Connection
Resistance Supported by modules: MX840A, MX440A
6 1
11
5
15 10
Four-wire circuit
ϑ
1)
Excitation (−)
2
Measurement signal (−) 1)
10
Cable shield Measurement signal (+) Excitation (+)
Hsg.
When connecting a two−wire sensor, wire bridges must be soldered in the connector (between measurement line and supply)
5 3 4 9 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (optional) view from below
HBM
Hsg. = housing
QuantumX
Connection
6.12.15
113
Resistance thermometers Pt100, Pt1000 Supported by modules: MX840, MX840A, MX440A
6 1
11
5
15 10
Four-wire circuit
ϑ
Excitation (−)
2
Measurement signal (−)
10
Cable shield Measurement signal (+) Excitation (+)
Hsg.
5 3 4 9 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (optional) view from below
QuantumX
Hsg. = housing
HBM
114
6.12.16
Connection
Thermocouples Supported by modules: MX840, MX840A, MX440A, MX1609, MX1608−P MX1609/MX1609P modules support only type K thermocouples with mini thermocouple plugs (see page 85). Cold−spot compensation is integrated here. 6
Connection with MX840, MX840A, MX440A 1
11
5
15 10
Maximum input voltage to housing and signal ground : "60 V Thermocouple −
+
1−THERMO−MX BOARD (soldered into connector plug and contains a cold spot compensation element and TEDS)
Type
Thermocouple material 1 (+)
Thermocouple material 2 (−)
J K T S
Iron Nickel-chrome (color−code green) Copper Rhodium-platinum (10%)
Copper-Nickel Nickel-aluminum (color code white) Copper-nickel Platinum
E
Nickel-chrome
Copper-nickel
B
Rhodium-platinum (30%)
Rhodium-platinum (6%)
N
Nickel-chrome-silicone1)
Nickel-silicone
R
Rhodium-platinum (13%)
Platinum
When recording temperatures with thermocouples using MX840, MX840A, MX440A, you must use a connector with integrated cold spot compensation element (1−THERMO−MX BOARD).
1)
HBM
Nicrosil
QuantumX
Connection
115
When recording temperatures with thermocouples, you must solder the ”1-THERMO-MX BOARD” board into the connector plug.
Position of board in plug 6
10 1
PT1000
5
PINs to be soldered
•
Insert the 1-THERMO-MX BOARD in the correct position between the plug pins
Note Check the position with the plug shape (see picture above). In this position, the PT1000 of the cold spot compensation element is on top.
PT1000
•
Solder the connector pins to the connections on the board
•
PIN 1
TEDS
PIN 6
TEDS
PIN 5
Thermocouple (+)
PIN 10 Thermocouple (−)
QuantumX
PIN 9
Signal ground
PIN 7
PT1000 cold junction
PIN 8
PT1000 cold junction
PIN 2
Supply (−)
PIN 3
Supply (+) HBM
116
6.12.17
Connection
Frequency measurement without directional signal (RS 485: Differential signals) Supported by modules: MX840, MX840A, MX440A, MX460 Schematic diagram 5V f1 (+)
6
200 mV
f1 (−) 0V
1
11
5
15 10
HBM torque transducer Connector 1 Md 1 4
wh rd
5
f1 (−)
10
f1 (+) Cable shield
5 gy
Hsg.
Ground
6 9 4 6 1
1 2 Data 3 No function
1
2 3 Hsg. = housing
1-wire EEPROM (view from below) Adjustable sensor supply:
Pin12: Pin 11:
5 V ... 24 V 0V
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
HBM
QuantumX
Connection
6.12.18
117
Frequency measurement with directional signal (RS 485: Differential signals) Supported by modules: MX840, MX840A, MX440A, MX460 Schematic diagram 5V f1 (+)
6
200 mV
f1 (−)
1
11
5
15
0V 10
HBM torque transducer Plug 2 n
bk and bu
8
6
1 7
6
Ground f1 (−)
wh rd
5 10
f1 (+)
gn
7
f2 (−) Cable shield
3 gy
Hsg.
f2 (+)
f1 (−) = Measurement signal speed, f1 (+) = Measurement signal speed, f2 (−) = Measurement signal speed, f2 (+) = Measurement signal speed,
8 9
0° (−) 0° (+) 90° (−) 90° (+)
4 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (view from below) Hsg. = housing
Adjustable sensor supply:
Pin12: Pin 11:
5 V ... 24 V 0V
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
QuantumX
HBM
118
6.12.19
Connection
Frequency measurement without directional signal (single−pole mode) Supported by modules: MX840, MX840A, MX440A, MX460 Schematic diagram; Single−pole mode f1
5V 3.5 V
Thresholds
6
Thresholds
1.5 V 0V
1
11
5
15 10
HBM torque transducer Connector 1 Md 1 4
wh rd
5
10
f1 (+) Cable shield
gy
Hsg.
6
Ground
9 4 5 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (optional) view from below Adjustable sensor supply:
Hsg. = housing
Pin12: Pin 11:
5 V ... 24 V 0V
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
HBM
QuantumX
Connection
6.12.20
119
Frequency measurement with directional signal (single−pole mode) Supported by modules: MX840, MX840A, MX440A, MX460 Schematic diagram; single−pole mode 5V 3.5 V
f1
6
Thresholds
1
Thresholds
1.5 V
11
5
0V
15 10
f1 (+)
10
f2 (+)
Industrial pulse generator
8
Cable shield
Hsg.
6 9 4 5 7 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (optional) view from below
Adjustable sensor supply:
QuantumX
Hsg. = housing
Pin12: Pin 11:
5 V ... 24 V 0V
HBM
120
6.12.21
Connection
Pulse counting, (RS 485: Differential signals) Supported by modules: MX840, MX840A, MX440A, MX460 Schematic diagram 6
5V f1 (+) f1 (−)
1
11
200 mV 5
0V
15 10
Zeroing pulse +
3
Zeroing pulse −
2
f1 (+) Industrial pulse generator
10 5
f1 (−) Cable shield f2 (−)
Hsg.
7 8
f2 (+)
6 9 4 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (optional) view from below Adjustable sensor supply:
HBM
Hsg. = housing
Pin12: Pin 11:
5 V ... 24 V 0V
QuantumX
Connection
6.12.22
121
Pulse counting (single−pole mode) Supported by modules: MX840, MX840A, MX440A, MX460 Schematic diagram, Single−pole mode f1
5V 3.5 V
6
Thresholds Thresholds
1.5 V 0V
1
11
5
15 10
6 8
f2 (+)
Industrial pulse generator
f1 (+) Cable shield Zeroing pulse
10
Hsg.
3 9 4 7 5 2 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (optional) view from below
Adjustable sensor supply:
QuantumX
Hsg. = housing
Pin12: Pin 11:
5 V ... 24 V 0V
HBM
122
6.12.23
Connection
SSI protocol 1) Supported by modules: MX840, MX840A, MX440A
1)
Schematic diagram
6 1
11
5
15 10
6
Ground f1 (−) SSI
5
f1 (+)
10
f2 (−)
7
Cable shield
Hsg.
f2 (+)
8 9
f1 (−) = Data (−) f1 (+) = Data (+) f2 (−) = Shift clock (−) f2 (+) = Shift clock (+)
4
Hsg. = housing
Adjustable sensor supply:
HBM
Pin12: Pin 11:
5 V ... 24 V 0V
QuantumX
Connection
6.12.24
123
Passive inductive rotary encoder1) Supported by module: MX460 1)
Schematic diagram
U 6 1
11
5
15 10
Maximum input voltage to housing and signal ground : "60 V
14
f1 AC+ Cable shield
Hsg.
6
Ground
9 4 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (view from below) Hsg. = housing
Adjustable sensor supply:
QuantumX
Pin12: Pin 11:
5 V ... 24 V 0V
HBM
124
6.12.25
Connection
PMW − Pulse width, Pulse duration, Period duration Supported by module: MX460
6 1
11
5
15 10
6 Transducer f1 (+) Cable shield
10
Hsg.
9 4 7 5 2 6 1
1 2 Data 3 No function
1
2 3
1-wire EEPROM (view from below) Adjustable sensor supply:
HBM
Pin12: Pin 11:
Hsg. = Housing
5 V ... 24 V 0V
QuantumX
Connection
6.12.26
125
CANbus The following modules enable CAN messages to be acquired: Channel 1 of MX840 or MX840A. The following modules enable CAN messages to be acquired and sent: MX471, MX840A (only module internal measurands). QuantumX Assistant enables a dbc data file to be created. MX471
MX840 / MX840A
Chanel 1 SubHD-15pol.
SubD-9pol.
6 1 1
11
5
15
6 9
5
10
Adjustable sensor supply: (does not apply for MX471)
:
Pin 12: Pin 11:
CAN-High
7
7
CAN-Low
8
2
CAN-GND
6
6
5 V ... 24 V 0V
Note Ensure correct termination with termination resistors is made, as shown in Fig.6.1. The MX840 does not have any termination, the MX471 features internal termination that can be activated by software.
Termination resistance 120Ω
Termination resistance 120Ω
Node 1
Fig.6.15:
...
...
Node n
Bus termination resistors
The adapter cable 1-KAB418 is used to connect the D-SUB-15HD device connectors of the MX840 to standard CAN plugs (D-SUB-9). QuantumX
HBM
126
7
Functions and outputs
Functions and outputs The MX878, MX410 and MX460 modules enable mathematics functions to be performed in real time and their results to be provided as standard system signals. These system signals can be used like actual measurement signals for subsequent tasks (analog output, EtherCAT signal, source signal for mathematics function, data visualization and storage). The MX878 and MX410 modules feature analog outputs that can be connected to a system or source signal, for example an actual measurement signal (additionally scaled, filtered) or the result of a mathematics function. Measurement channels that are used in mathematics functions or directly for analog outputs need to be activated for ”isochronous data transfer” (e.g. in the QuantumX−Assistant software, ”Signals” tab).
Note The module configuration is immediately active after system reboot (Auto Startup). No operating PC is required for configured signal outputs to run stand−alone. Mathematics functions at a glance: min
S
S
max
+ x
Peak values (Peak)
− =
Add & Multiply
S
Root mean square (RMS)
S
Vibration analysis (Rotational Analysis) and angle difference (Angle diff)
Mathematics functions provided by the modules MX878 + x
min
MX460 min
− =
max
HBM
MX410
max
min max
QuantumX
Functions and outputs
7.1
127
MX410 The MX410 module is a module with eight analog outputs that can be accessed on the front panel via BNC sockets. The outputs are directly assigned to the above located inputs.
Fig. 7.1:
MX410 front view
Note After configuring an analog output its function (configuration, scaling) continues to be available, even if the computer is disconnected. No connection to a PC is required. The MX410 supports 8 peak detection channels and 4 RMS channels. These functions can be used to generate so−called virtual signals that can also be output at the analog output and made available to the QuantumX system. This also makes the signals visible to the software. The device parameterization is implemented by the software (e.g. QuantumX Assistant or catmanAP). Keep in mind the following notes when working with peak monitor channels: S The maximum output (sample rate) is restricted to 4800 Hz S Resetting of PEAK values only via PC software (system input signals will follow) S The output rate of the peak monitor channel must not be higher than the sample rate of the input channel S Filters set for the MX410 do not apply to peak monitor channels S These channels are always unfiltered. However, the input signal is filtered. S The peak units do not accept other peak units or RMS as input − only the 4 analog inputs are allowed
QuantumX
HBM
128
Functions and outputs
min max
Peak Detection function Each peak detection unit can monitor either the Min or the Max of one of the module’s 4 analog input channels. A peak unit may operate in different modes: S RUN: peak value will be continuously updated S HOLD: last peak value will be ”frozen” S PEAK: peak detection enabled S FOLLOW: peak detection disabled, i.e. the channel yields the original signal of the input channel The following combinations are possible: Graph for MAX−PEAK−RUN Graph for MAX−PEAK−HOLD Graph for MAX−Follow−HOLD This also applies to the minimum values. Graph for Peak Detection
Vi, Vo
Funktion Betriebsart
HBM
Output
Run Peak
Measuring signal
t Hold Run Follow
QuantumX
Functions and outputs
129
RMS (root mean square) function RMS is computed from one of the module’s 4 analog input channels according to the formulaL
Where f(x) denotes the input channel signal and T the time window (in ms). Keep in mind the following notes when working with RMS channels: S The maximum sample rate is restricted to 4800 (2400) Hz S The output (sample) rate of the RMS channel must not be higher than the sample rate of the input channel S Filters set for the MX410 do not apply to RMS channels. These channels are always unfiltered. However, the input signal is filtered
QuantumX
HBM
130
7.2
Functions and outputs
MX460 The MX460 supports 4 special math channels for analysis of rotational parts: Rotational vibration and angle difference.
Fig. 7.2:
MX460 front view
MX460 Math channels Keep in mind the following notes when working with these channels: S The maximum sample rate is restricted to 4800 (2400) Hz S The sample rate of the channel must not be higher than the sample rate of the input channel S Filters set for the MX460 do not apply to math channels. These channels are always unfiltered. However, the input signal is filtered.
HBM
QuantumX
Functions and outputs
7.3
131
MX878 The MX878 module is a module with eight analog outputs that can be accessed on the front panel via BNC sockets.
Fig. 7.3:
MX878 front view
Two analog outputs each (1 and 2, 3 and 4, etc.) have the same ground potentials, for the others electrical isolation applies between them and the voltage supply ground. The MX878 can receive all signals that are isochronously available on the Firewire. The settings for this are implemented with the QuantumX Assistant. Before output at the analog output, the signal passes through an output characteristic curve parameterized by the user (2 point scaling) and a filter also parameterized by the user. In addition, the rate for DAC is reduced to 96 kS/s through interpolation. MX878 Math channels The MX878 is a dedicated module designed for analog output and math channels. The MX878 supports 4 math channels and 4 peak detection channels. In contrast to other modules, the MX878 has no analog sensor inputs − instead it receives data from other modules via the so called ”isochronous FireWire transfer” from any source within the system configured for this data transfer mode. The module routes this data to an analog output or performs a math computation on this data (which may also be output on one of the analog outputs). You need to connect all modules via FireWire (or use a backplane) for the MX878 to become operational! You can configure several channels to ”isochronous FireWire transfer” using QuantumX Assistant software or catmanAP 3.1 or higher. Please note: Providing the data via isochronous transfer may use up significant computing power on the module (in particular on the MX410 and MX460 high−speed modules). Do not activate the isochronous transfer needlessly!
QuantumX
HBM
132
Functions and outputs
+
−
x
=
Add & Multiply function The MX878 currently provides this type of computation: Result = a0 + a1* Input1 + a2 * Input2 + a3 * Input1 * Input2 Where Input1 and Input2 denote the two input channels used for this computation. These channels will reside on other modules and must have their isochronous FireWire transfer enabled! Keep in mind the following notes when working with math channels: S The maximum sample rate is restricted to 2400 Hz S The sample rate of the channel must not be higher than the sample rate of the input channels S Filters do not apply to math channels. These channels are always unfiltered.
..... ..... .....
Matrix computation function The MX878 enables 4 parallel matrix computations each with a maximum of 6 input and output quantities and 36 constants to be performed. Generic formula: Fx = a1*Ufx + a2*Ufy + a3*Ufz + a4*Umx + a5*Umy + a6*Umz etc. for Fy, Fz, Mx, My, Mz The ”Matrix computation” function allows mathematical compensation of interdependencies (crosstalk) of multi−component transducers for force and torque measurement. The maximum data rate of the input and output quantities is 1200 Hz (< 1 ms computation time). The computed output signals can be scaled and output as filtered analog voltages by the same module. The computed signals can also be distributed (isochronously) to the FireWire bus in real time and output via CAN bus or EtherCAT (MX471: CAN bus. MX878: EtherCAT bus). Scaling of the input and output quantities must always be ensured. An EXCEL compensation matrix can be copied directly to the matrix parameterization (Ctrl + C, Ctrl + V).
HBM
QuantumX
Functions and outputs
min max
133
Peak Detection function Keep in mind the following notes when working with peak monitor channels: S The maximum sample rate is restricted to 4800 Hz S The sample rate of the peak monitor channel must not be higher than the sample rate of the input signal S The peak units do not accept other peak units or RMS as input Each peak detection unit can monitor either the Min or the Max of one of 4 signals ”visible as isochronous” within the system. A peak unit may operate in different modes: S RUN: peak value will be continuously updated S HOLD: last peak value will be ”frozen” S PEAK: peak detection enabled S FOLLOW: peak detection disabled, i.e. the channel yields the original signal of the input channel The following combinations are possible: Graph for MAX−PEAK−RUN Graph for MAX−PEAK−HOLD Graph for MAX−Follow−HOLD This also applies to the minimum values.
Graph for Peak Detection
Vi, Vo
Funktion Betriebsart
Output
Run Peak
Measuring signal
t Hold Run Follow
MX878 analog outputs The MX878 is a dedicated module designed for analog output and math channels. In contrast to other modules, the MX878 has no analog sensor inputs − instead it receives data from other modules via the so called ”isochronous FireWire transfer”. The module then routes this data to an analog output. QuantumX
HBM
134
Functions and outputs
You need to connect all modules via FireWire (or use a backplane) for the MX878 to become operational! Refer to the ”ISO” column in the catman channel configuration window. This column indicates if a channel provides its data via the isochronous link (indicated by a symbol). Click the column or use the column’s context menu to enable or disable isochronous transfer for a channel. Providing the data via isochronous transfer may use up significant computing power on the module (in particular on the MX410 and MX460 high−speed modules). Do not activate the isochronous transfer needlessly! MX878 signal generator The MX878 has eight signal generators. The signals (e.g. set profiles for controlling uni− or multiaxial actuators) can be individually generated and assigned to analog outputs. The following signal forms are available (to be defined in an ASCII file): Constant, sine, rectangle, triangle Depending on their type, the signal forms are described by the following parameters: Level, frequency, duty ratio The signals are buffered and described as follows: Repetition cycle (continuous, triggered) Point in time A buffer that has been filled before can be output with a defined number of repetition cycles, continuously and triggered, starting at a specific point in time. In addition, a second buffer is available. While one buffer is output the second buffer can be filled. Output of the second buffer can be activated immediately or when the first buffer has been output. At the end of the sequence, the last output value is held.
HBM
QuantumX
Functions and outputs
7.4
135
MX471 The modular QuantumX data acquisition system can be individually assembled. FireWire enables quantities measured using amplifiers or computed values to be transmitted to specific modules in real time (isochronous signals). For example: To send a measured temperature as a so−called ”Process Data Object” (PDO) over the CANbus via a specific node of the MX471, the temperature signal simply has to be dragged onto the node using drag & drop. Parameterization is then performed automatically. The CAN message needs to be completed in the corresponding dialog. A parameterized MX471 node enables any signals or measured quantities to be cyclically sent as CAN messages. The CAN messages are sent to the CANbus at the data rate set for the source signal. With a data rate > 4.800 Hz, the transmission rate is limited. A configurable divider enables the transmission rate on the CANbus to be generally reduced. It is essential to specify via software the baud rate and, if necessary, activation of a termination resistor (bus termination) to ensure correct bus parameterization. Parameterizing the CAN message requires: S Definition of an identifier (decimal CAN−ID) and the corresponding format (11 or 29 bit) S If necessary, divider to reduce the transmission rate. Signals are sent using the 4−Byte−Float format. When parameterization of the node has been completed, a type *.dbc CAN database can be generated. This database makes configuring a receiver node easier.
Note In general, a CAN network enables both message types (standard/extended) to be used. No analysis of Sync messages or Remote frames is made. The four CAN nodes are handled with equal priority. A node parameterized as sender, works as an autonomous gateway and does no longer require connection to a PC. SDOs are not supported; the MX471 can thus not be considered as a CANopen slave. The individual nodes’ statuses are directly displayed at the respective connection point.
CAUTION
If a CAN node uses an incorrect baud rate for transmission over the bus, the entire data transmission on the bus may fail.
QuantumX
HBM
136
8
Accessories
Accessories
System accessories Article
Description
Order No.
Module carrier QuantumX (standard)
Module carrier for maximum 9 modules of the QuantumX family; General information: − Wall or control cabinet installation (19”); − Connection of external modules via FireWire possible; − Power supply: 24 V DC; − Power consumption: Max 5 A (150 W); Note: Only modules with degree of protection IP20 can be inserted.
1-BPX001
Housing connection element for QuantumX modules
Mechanical connection elements for QuantumX modules (IP20 / IP65); Set consisting of 2 housing clips, including assembly material for fast connection of 2 modules.
1-CASECLIP
Module accessories Voltage supply IP20
IP65
AC−DC plug−in power supply unit; Input: 100-240 V AC ("10%); 1.5 m cable with international plug set Output: 24 V DC, Max. 1.25 A; 2 m cable with plug for IP20 modules
Connector − Voltage supply QuantumX (module in protection class IP65)
Order No.: 1-NTX001
Order No.: 1-CON−P1001
3 m cable for voltage supply of a QuantumX module; Plug for IP20 modules on one side and exposed wires at the other end. Note when using multiple modules: The supply voltage can be looped through FireWire connections (max. 1.5 A) Order No.: 1-KAB271-3 FireWire FireWire cable (module to module)
1−CASEFIT
HBM
IP20
IP65
FireWire connection cable between QuantumX modules in IP20 design (length: 0.2m/2m/5m); Fitted both ends with appropriate plugs. Note: The cable can be used to optionally supply connected QuantumX modules with voltage (max. 1.5 A, from source to last acceptor).
FireWire connection cable between QuantumX modules in IP65 design (length: 0.2m/2m/5m); Fitted both ends with appropriate plugs. Note: The cable can be used to optionally supply connected QuantumX modules with voltage (max. 1.5 A, from source to last acceptor).
Order No.: 1-KAB269-0.2 Order No.: 1−KAB269−2 Order No.: 1−KAB269−5
Order No.: 1-KAB272-0.2 Order No.:: 1−KAB272−2 Order No.: 1−KAB272−5
Fitting panel for mounting of QuantumX modules using case clips (1−CASECLIP), lashing strap or cable tie. Basic fastening by 4 screws
Order No.: 1-CASEFIT
QuantumX
Accessories
137
1−SCM−FW
SCM−FW FireWire Extender Package consists of 2 In−line elements to extend the FireWire connection up to 40 m; Necessary parts: 2 x 1−KAB269−x and Industrial Ethernet cable (M12, CAT5e/6, max. 30 m). KAB270−3 connection is not possible!
Order No.: 1-SCM−FW
1−CASEPROT
Case protection functioning as stacking aid for MX amplifiers, including material for fast mounting.
Order No.: 1-CASEPROT
IP20
IP65
FireWire connection cable between hub and QuantumX modules in IP20 design (length: 3 m); Fitted both ends with appropriate plugs. Note: The cable can be used to optionally supply connected QuantumX modules via the hub with voltage (max. 1.5 A, from source to last acceptor).
FireWire connection cable between hub and QuantumX modules in IP65 design (length: 3 m); Fitted both ends with appropriate plugs. Note: The cable can be used to optionally supply connected QuantumX modules or hub with voltage (max. 1.5 A, from source to last acceptor).
Order No.: 1-KAB275-3
Order No.: 1-KAB276-3
FireWire Hub to module connection cable 3 m
PC to module connection cable 3 m
FireWire connection cable between PC and QuantumX module (length: 3 m); Fitted both ends with appropriate plugs. Note: The cable cannot be used to supply QuantumX modules with voltage. Order No.: 1−KAB270-3
FireWire PC-Card
FireWire PC−Card with FireWire 1394b interface for connection of QuantumX amplifiers to notebook or PC (via PCCARD adapter) Order No.: 1-IF−001
Ethernet Ethernet crossover cable
Ethernet patch cable between PC and QuantumX module in IP65 design (length: 5 m); Fitted both ends with appropriate plugs.
Order No.: 1-KAB239-2
Order No.: 1-KAB273-5
Transducer side General information Article
Description
Order No.
D−Sub−HD 15-pin plug set with TEDS chip
Plug kit D−Sub−HD 15-pin (male) with TEDS chip for storing a sensor data sheet; housing Metallized plastic with knurled screws. Note: The TEDS chip is blank.
1-SUBHD15-MALE
Port Saver D−Sub−HD 15 pin
Four fully-wired male to female port savers protecting the wear and tear of D−Sub−HD 15 pin ports for frequent plugging and unplugging of transducers. Extends contact durability by min. 500. This adaptor attaches securely with screws.
1−SUBHD15−SAVE
SCM−SG120/350
SCM−SG350 Strain Bridge Module (120 Ohm or 350 Ohm) Signal conditioning module for QuantumX full bridge input (D−Sub−HD15 pin). Integrated 350 Ohm bridge completion resistor, shunt calibration, TEDS, D−Sub−HD-device connection. Soldering point for transducer cable.
1−SCM−SG120 1−SCM−SG350
Adapter D−Sub−HD 15-pin to D−Sub 15-pin.
Adapter D−Sub HD 15-pin to D−Sub 15-pin for connecting transducers with pre−assembled D−Sub plugs on MX840 (length approx. 0.3 m) Note: Pre-assembled for full bridge (6-wire).
1-KAB416
QuantumX
HBM
138
Accessories
MX840/MX840A accessories Article
Description
Order No.
Cold junction for thermocouple on MX840, MX840A, MX440
Electronics for temperature compensation for measurements with thermocouples on MX840, MX840A, MX460, comprising: − PT1000 cold junction − Including 1-wire TEDS chip for transducer identification Note: Mounting in D−Sub−HD 15-pin transducer plug.
1-THERMO-MXBOARD
Adapter D−Sub−HD15 to D−Sub9 (CAN) on MX840, MX840A
Adapter for connecting CAN devices to MX840, MX840A, MX460. D−Sub−HD 15-pin (plug) to D−Sub 9-pin (socket); (length: approx. 30 cm).
1-KAB418
Article
Description
Order No.
BNC adapter for IEPE sensors with BNC
Adapter for connecting IEPE sensors with BNC connection cable to MX410 (BNC socket to 15-pin D−Sub−HD; length: approx. 7 cm).
1-IEPE-MX410
Article
Description
Order No.
Bag with 10 mini thermocouple plugs, incl. RFID
Packet, comprising 10 x mini thermocouple plugs with integrated RFID chip for measuring point detection for the MX1609 thermocouple measuring amplifier of the QuantumX product family; Type K: NiCr−NiAl, RFID integrated, green, male.
1-THERMO-MINI
Bag with 10 RFIDs
Bundle, comprising 10 x RFID-chip for self−assembly on mini thermocouple plugs for the MX1609 thermocouple measuring amplifier of the QuantumX product family
1-THERMO-RFID
MX410 accessories
MX1609 and 1609P accessories
Accessories for connecting the Smart module: Product
Description
Order No.:
Smart module
External 24 V signal conditioning module, supplying IEPE with constant current (BNC socket) and feeding standardized " 10 V voltage signal.
1−EICP−B−2
Connection cable
Cable between Smart module and SubHD plug
1−SAC−EXT−MF−x−2 (x = length in meter)
Male device connector
QuantumX connector
1−SubHD15−MALE
HBM
QuantumX
Accessories
139
8.1
System accessories
8.1.1
BPX001 backplane
8.1.2
Housing connection elements
Housing clip
Covers for housings with protection class IP65
Housing clip
QuantumX
HBM
140
Accessories
8.2
Voltage supply
8.2.1
Power pack NTX001
Europe mains cable
3m NTX001
Mains
Modules UK mains cable
USA mains cable
Australia mains cable
Order No.:
HBM
1-NTX001
QuantumX
Accessories
8.2.2
141
Supply cable
3m
30 Approx. 10−15 mm
Twisted and tinned Plug ODU Medi-Snap S11M08-P04MJGO−5280
Cable LIYY 2x0.5 mm2
black brown
0V +
Power supply
Order No.: 1-KAB271-3 (Length 3 m)
8.3
FireWire
8.3.1
FireWire cable (module to module; IP20)
0.2 m 2.0 m 5.0m
Plug ODU SX1LOC−P08MFG0−0001
Order No.:
1-KAB269-2 (Length 2 m) 1-KAB269-0.2 (Length 0.2 m) 1-KAB269-5 (Length 5 m)
QuantumX
HBM
142
8.3.2
Accessories
FireWire cable (module to module; IP65)
L1=0.2 m; 2.0 m; 5.0 m
Order No.:
1-KAB272-2 (Length 2 m) 1-KAB272-0.2 (Length 0.2 m) 1-KAB272-5 (Length 5 m)
8.3.3
Connection cable (PC to module; IP20)
3m
FireWire B-plug
PC adapter socket
Order No.:
HBM
1-KAB270-3 (Length 3 m)
QuantumX
Accessories
143
8.4
General information
8.4.1
Plug kit with TEDS chip Plug kit D−Sub−HD 15-pin (male) with TEDS chip for storing a sensor data sheet. Order No.:
8.4.2
1-SUBHD15-MALE
Port saver SubHD 15pol.
Front
Back
When frequently connecting and disconnecting transducers we recommend that you use port savers for protecting the transducer sockets of the QuantumX module. The port saver is easy to screw into place and can be replaced after several hundred mating cycles. No expensive module repair is necessary any more.
QuantumX
HBM
144
Accessories
8.4.3
Adapter D−Sub−HD 15-pin to D−Sub 15-pin.
D-SUB HD plug
bk bu gy gn wh rd
D-SUB socket
Wire bridge
Order No.:
1-KAB416
CAUTION This cable is only for transducers with full bridge and 6-wire circuits! If other transducers are connected, the universal amplifier can be damaged or even destroyed.
HBM
QuantumX
Accessories
145
8.5
MX840 accessories
8.5.1
Cold junction for thermocouples
Electronics for temperature compensation for measurements with thermocouples. Board for installation in a 15-pin D−Sub−HD plug. View from above
View from below
12
12,5
6
10 1
Position of board in plug
5
PINs to be soldered TEDS chip
Cold junction
All dimensions in mm Height with components: 3 mm
Order No:
QuantumX
1-THERMO-MXBOARD
HBM
146
8.5.2
Accessories
Adapter cable (CAN)
bk bu gy gn wh rd
D-SUB HD plug
Hsg.
Order No.:
8.6
D-SUB socket
Hsg.
1-KAB418 (Length approx. 30 cm)
Accessories MX410
Order no.: 1−IEPE−MX410 The adapter from D−Sub−HD plug male to BNC plug female is to connect current fed piezo electric transducers (IEPE = Integrated Electronics Piezo Electric) with BNC connector cable to MX410 universal amplifier. Mounting: connector 1...4, screwed HBM
QuantumX
Accessories
147
8.7
MX1609 and 1609P accessories
8.7.1
Thermo-connector with integrated RFID chip
RFID
21
Package unit: 10 mini connectors for thermocouples type K Order No.:
8.7.2
1-THERMO-MINI
RFID chip for sticking onto thermo-connector
4,9
12,1 3 All dimensions in mm
Package unit: 10 RFID chips Order No.:
QuantumX
1-THERMO-RFID
HBM
148
9
Accessories
Support Headquarters world-wide Europe Hottinger Baldwin Messtechnik GmbH: Im Tiefen See 45, 64293 Darmstadt, Germany Tel. +49 6151 8030, Fax +49 6151 8039100 E−mail:
[email protected] www.hbm.com North and South America HBM, Inc., 19 Bartlett Street, Marlborough, MA 01752, USA Tel. +1-800-578-4260 / +1-508-624-4500, Fax +1-508-485-7480 E−mail:
[email protected] Asia Hottinger Baldwin Measurement (Suzhou) Co., Ltd. 106 Heng Shan Road, Suzhou 215009, Jiangsu, VR China Tel. (+86) 512 68247776, Fax (+86) 512 68259343 E−mail:
[email protected] Up to date addresses of representatives can also be found on the Internet under: www.hbm.com/Contact/Worldwide Contacts
HBM
QuantumX
Support
QuantumX
149
HBM
Im Tiefen See 45 S 64293 Darmstadt S Germany Tel. +49 6151 803−0 S Fax: +49 6151 803−9100 Email:
[email protected] S www.hbm.com
I3031−5.0 en
Hottinger Baldwin Messtechnik GmbH
7−2002.3031
E Hottinger Baldwin Messtechnik GmbH. All rights reserved. All details describe our products in general form only. They are not to be understood as express warranty and do not constitute liability whatsoever.