File is formatted to print at (70in. x 36in.) To request printed template, contact Midmark Corporation at 1-800-643-6275.
PLACE THIS EDGE AGAINST WALL (CONSIDERING FINISHED WALL THICKNESS)
WATER ROUND
AIR
END ROUND
ACCESSORY OPTION
ACCESSORY
VACUUM
OPTION
PANEL
PLUMBING ROUGH-IN PERIMETER
OPTION
17.00"
END
END
END
PANEL
OPTION
42.00"
General Notes NOTE: HAVE A LICENSED PLUMBER INSTALL ALL PLUMBING COMPONENTS. LOCAL PLUMBING
CODES
SHOULD
BE
FOLLOWED
FOR ALL CONNECTIONS. MIDMARK CORP. CANNOT BE HELD RESPONSIBLE FOR INSTALLATIONS WHICH DO NOT CONFORM TO LOCAL CODES.
FLOOR PLUMBING LAYOUT OSC042D
42" Side Delivery Casework 003-1982-00
Rev. A1
1. Outside edges of template indicate overall cabinet silhouette. Position
template on wall in desired location per room layout drawing 2. Stubs for water, air, and vacuum
can be placed anywhere within the bold line perimeter. Recommended
locations have been indicated.
Plumbing Specifications 1. WATER: 3/4" diameter hole required for 1/2" NPT pipe stub. Stub to extend 1" min. / 2-1/2" max. from finished wall
42
17-3/4
[1067]
[451]
surface. Install manual shut-off valve
(supplied). Flush all debris from line before installing valve. Water pressure NOT to exceed 50 psi. 2. AIR: 3/4" diameter hole required for 1/2" NPT pipe stub. Stub to extend
1" min. / 2-1/2" max. from finished wall
30-3/8
surface. Install manual shut-off valve
[772]
(supplied). Flush all debris from line before installing valve. Air pressure to be 80 psi (min.) to 100 psi (max.). 3. VACUUM: Plumbing to the casework should be specified by central vacuum supplier. Drill hole in wall as required. Extend pipe stub 1" min. / 2-1/2" max. from finished wall surface. Terminate
plumbing with a 5/8" OD tubing connector.
FRONT VIEW
END VIEW
Midmark Corporation, P.O. Box 286, 60 vista Drive
Versailles, Ohio 45380-0286, www.midmark.com 1-800-MIDMARK (1-800-643-6275), FAX 1-877-725-6495