Technical Information
Oxymax W COS61 Dissolved oxygen measurement Optical sensor acc. to the fluorescence quenching principle
Application The continuous measurement of the dissolved oxygen concentration is very important in many areas of water management: • Sewage treatment plants: Oxygen measurement and regulation in the activated sludge basin for a highly efficient biological cleaning process • Water monitoring: Oxygen measurement in rivers, lakes or seas as an indicator of the water quality • Water treatment: Oxygen measurement for status monitoring of drinking water for example (oxygen enrichment, corrosion protection etc.) • Fish farming: Oxygen measurement and regulation for optimum living and growth conditions Your benefits • Optical technology: – Minimum maintenance – Maximum availability • Compatible with tried-and-tested COS31 with COM2x3W: – Easy measuring point changeover to optical technology • Compatible with COS41 with COM2x3D with conversion kit • Sensor with digital data transmission: – Storage of calibration data in sensor – Insensitive to electromagnetic interferences thanks to digital communication to the transmitter • Maximum measurement accuracy: – Longterm-stable measurement – Long maintenance intervals – Intelligent sensor self monitoring • No flow needed - measurement possible in still water
TI387C/07/en/04.07 51517134
Oxymax W COS61
Function and system design Measuring principle
• Sensor design: – Oxygen-sensitive molecules (markers) are integrated in an optically active layer (fluorescence layer). – The surface of the fluorescence layer is in contact with the medium. – The sensor optics are directed at the underside of the fluorescence layer. • There is an equilibrium between the oxygen partial pressure in the medium and that in the fluorescence layer: – If the sensor is immersed in the medium, the equilibrium is established very quickly. • Measuring process: – The sensor optics send green light pulses to the fluorescence layer. – The markers "answer" (fluoresce) with red light pulses. – The duration and intensity of the response signals is directly dependent on the oxygen contents and the partial pressure. – If the medium is free from oxygen, the response signals are long and very intense. – Oxygen molecules "quench" the marker molecules. As a result, the response signals are shorter and less intense. • Measurement result: – The sensor returns a signal that is in proportion to the oxygen concentration in the medium. – The fluid temperature and air pressure are already calculated in the sensor.
Sensor monitoring
The optical signals are continuously monitored and analysed for plausibility. If inconsistencies occur, an error message is output via the transmitter. Aging of the sensor cap is detected. The transmitter first displays a warning for predictive maintenance and later generates an error message. In addition, the following fault conditions are detected in conjunction with the sensor check system of Liquisys M COM2x3: • Implausibly high or low measured values • Disturbed regulation due to incorrect measured values
Measuring system
The complete measuring system consists of the following at least: • Oxygen sensor • Transmitter, e.g. Liquisys M COM223/253 • Special measuring cable if extension is necessary • Assembly, e.g. COA250 flow assembly, CYA611 immersion assembly or COA451 retractable assembly Optional: • CYH101 universal immersion assembly holder for immersion operation • VS junction box (for cable extension) • Chemoclean automatic spray cleaning system 1
4
2
3 a0007362
Fig. 1: 1 2 3 4
2
Measuring system (example) CYA611 immersion assembly VS junction box (optional) Liquisys M COM253 transmitter COS61 oxygen sensor
Endress+Hauser
Oxymax W COS61
Input Measured variable
Dissolved oxygen [mg/l, ppm, % SAT or hPa] Temperature [° C, °F)
Measuring range
with Liquisys M COM 223/253: 0 to 20 mg/l (0 to 20 ppm) 0 to 200 % SAT 0 to 400 hPa
Wiring Electrical connection
Direct connection to the transmitter Connect the sensor directly to the transmitter (COM253-WX/WS) by using the special measuring cable with SXP plug.
a0004118
SXP plug
Direct connection to the transmitter (panel mounting device) • Remove the SXP connector (transmitter side!) from the cable. • Refer to the following table for the cable assignment and the assigned terminals for Liquisys M COM223-WX/WS. • Please note that the cable assignment changes depending on the sensor version (fixed cable or TOP68 connection). Terminal COM223
Endress+Hauser
Sensor with fixed cable (OMK)
Sensor with TOP68 connection (CYK71)
Core
Assignment
Core
Assignment
87
YE
+UB
YE
+UB
0
GY
0V
WH
0V
96
PK
Com. (digital)
GN
Communication (digital)
97
BU
Com. (digital)
BN
Communication (digital)
88
BN
–UB
Koax innen –UB
3
Oxymax W COS61
Connection with cable extension To lengthen the sensor connection beyond the length of the fixed cable, you require a junction box VS. Always connect the sensor cable with the SXP plug to the junction box. The cable extension to the transmitter then depends from the transmitter version, i.e. field device or panel mounted device. 1 82 / 3.23
80 / 3.15
4
YE GY PK BU BN
2
mm / inch
WH
0 96 97 88 19 18 a0004130
a0004129-en
Junction box VS to a field device 1 2
GN
3
2
87
Junction box VS to a panel mounted device 2 3 4
SXP plug to field device SXP plug from sensor
SXP plug from sensor Measuring cable (OMK) to the transmitter Connection department of the transmitter
Performance characteristics Response time
t90: 60 s
Reference operating conditions
Reference temperature: Reference pressure: Reference application:
Maximum measured error
±2 % of measuring range end
Repeatability
±0.5 % of measuring range end
Life time of the sensor cap
>2 years (under reference operating conditions, protect against direct sun light)
4
25 °C (77 °F) 1013 hPa (15 psi) Municipal wastewater treatment
Endress+Hauser
Oxymax W COS61
Installation Angle of installation
The sensor can be installed up to the horizontal in an assembly, support or a suitable process connection. Other angles and overhead installation are not recommended. Reason: possible sediment formation and resulting falsification of measured value.
!
Note! The optimum installation angle is 45° (e.g. with assembly CYA611-0E).
A a0004087
Angle of installation A
!
Permissible installation positions: 0 ... 180 °
Note! Make sure you comply with the instructions for installing sensors. You will find them in the Operating Instructions for the assembly used.
Environment Ambient temperature range
–20 to +60 °C (0 to 140 °F)
Storage temperature
–20 to +70 °C (0 to 160 °F) at 95% relative humidity, non condensing
Ingress protection
• Fixed cable versions: IP 68 (10 m (33 ft) water column at 25 °C (77 °F) in 30 days) • Top 68 plug-in head versions: IP 68 (1 m (3.3 ft) water column at 50 °C (122 °F) in 7 days)
Process Process temperature
–5 to 50 °C (20 to 120 °F)
Process pressure
max. 10 bar (145 psi)
Endress+Hauser
5
Oxymax W COS61
Mechanical construction Design, dimensions
a0004083-en
Fixed cable version
Weight
with cable length 7 m (23 ft): 0.7 kg (1.5 lbs.) with cable length 15 m (49 ft): 1.1 kg (2.4 lbs.) with TOP68 plug-in connection: 0.3 kg (0.66 lbs.)
Material
Sensor shaft: Cap with fluorescence layer: Fluorescence layer:
Process connection
G1
Maximum cable length
max. 100 m / 328 ft (including cable extension)
Temperature compensation
internal
6
a0004084-en
TOP68 version
stainless steel 1.4571 (AISI 316Ti) POM Silicone
Endress+Hauser
Oxymax W COS61
Ordering information Product structure
Certificate A Ex free version Cable length 0 Cable length: 1.5 m (4.9 ft) 1 Cable length: 7 m (23 ft) 2 Cable length: 15 m (49 ft) 8 Without cable (for TOP 68 version) 9 Special version acc. to customers specification Sensor head F G1, fixed cable with SXP plug S G1, TOP68 plug Accessories 0 without accessories COS61-
Scope of delivery
complete order code
The following items are included in the delivery: • Oxygen sensor with transport protection cap for membrane protection • Operating Instructions (on CD only) • Brief Operating Instructions (paper version)
Certificates and approvals Electromagnetic compatibilty
! Assemblies (selection)
Interference emission and interference immunity complies with EN 61326: 1997 / A1: 1998
Accessories Note! In the following sections, you find the accessories available at the time of issue of this documentation. For information on accessories that are not listed here, please contact your responsible service. Retractable assembly Cleanfit COA451 • manually driven retractable assembly, stainless steel, with ball valve, for oxygen sensors; • ordering acc. to product structure (Technical Information TI368C/07/en) Immersion assembly Dipfit W CYA611 • For sensor immersion in basins, open channels and tanks, PVC • Ordering acc. to product structure, see Technical Information TI166C/07/en Flow assembly COA250 • for sensor installation in pipe lines, PVC • ordering acc. to product structure (Technical Information TI111C/07/en)
Zero solution
• 3 units to produce 3 x 1 liter oxygen-free solution • order no. 50001041
Measuring cable
COK31 special measuring cable • for sensors COS31, COS61 and COS71 with TOP68 plug-in head • Order numbers: – Cable length 1.5 m (4.9 ft): 51506820 – Cable length 7 m (23 ft): 51506821 – Cable length 15 m (49 ft): 51506822
Endress+Hauser
7
Measuring cable OMK • for use as extension cable between junction box VS and transmitter, not terminated • sold by the metre - order no. 50004124 CYK71 measuring cable • Non-terminated cable for the connection of sensors (e.g. conductivity sensors) or the extension of sensor cables • Sold by the meter, order numbers: – non-Ex version, black: 50085333 – Ex version, blue: 51506616 Junction box
VS junction box • With plug-in socket and 7-pole plug • For cable extension from sensor (COS71, COS61, COS31, COS3 with SXP connector) to transmitter, IP 65; • Order no. 50001054
Transmitter
Liquisys M COM 223/253 • Transmitter for oxygen measurement • field or panel-mounted housing • Hart® or Profibus available • Ordering acc. to product structure, see Technical Information (TI 199C/07/en)
Instruments International Endress+Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel.+41 61 715 81 00 Fax+41 61 715 25 00 www.endress.com
[email protected]