Model LBP30 Smart Pulsewave Blood Pressure Monitor
Model LBP30
Smart Pulsewave Blood Pressure Monitor INSTRUCTION MANUAL
Shenzhen RSP Medical Technology Co., Ltd
Model LBP30 Smart Pulsewave Blood Pressure Monitor
Dear Customer, Thank you for purchasing the LBP30 Smart Pulsewave Blood Pressure Monitor. In order to use the device correctly and efficiently, please read this Instruction Manual before use. In addition, please take good care of the manual after your reading to ensure you can find it easily when necessary.
Visual number: V1.1.0
This product is suitable for people who are older than 12 years of age in hospital, clinic and social medical organizations etc.
Model LBP30 Smart Pulsewave Blood Pressure Monitor
CONTENTS SAFETY PRECAUTIONS..........................................................................................4 COMPONENTS OF THE PRODUCT.......................................................................... 6 RECHARGE THE BATTERY....................................................................................... 7 APPLYING THE CUFF...................................................................................................8 POSTURES INSTRUCTIONS........................................................................................8 SETTING UNITS.............................................................................................................. 9 TAKING A MEASUREMENT.....................................................................................10 CARE AND MAINTENANCE..................................................................................... 12 ERROR INDICATORS..................................................................................................12 TROUBLESHOOTING TIPS.......................................................................................12 PRODUCT SPECIFICATIONS................................................................................... 13 APPENDIX A: PACKING LIST.................................................................................. 14 APPENDIX B: EMC....................................................................................................... 14
Model LBP30 Smart Pulsewave Blood Pressure Monitor
SAFETY PRECAUTIONS The purpose of the symbols used in the Instruction as safety identifying is to enable you use the product safe and properly, and to prevent harm to you and others or damage to property. SYMBOLS AND ABBREVIATIONS Identifiers WARNING
CAUTION
Indications It indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. It indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
Attention! Consult the manual!
BF TYPE Internal battery WARNING 1. Contact your physician for specific information about your blood pressure. Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician or licensed healthcare provider. 2. This product is only suitable for measuring blood pressure and pulse rate of adult, which values are for reference. 3. Pregnant women, pre-eclamptic, mental disorder or arrhythmia patients should use this device under practitioner’s guidance. 4.
When common arrhythmias (such as atrial premature ventricular premature ventricular premature and atrial fibrillation) appear, it cannot ensure that the blood pressure monitor will achieve the performance it claims to. It is advised to use the product under a doctor’s guidance.
Model LBP30 Smart Pulsewave Blood Pressure Monitor
CAUTION 1. Be sure to use and purchase a dedicated double-bladder cuff, or you cannot get accurate results. 2. Do not use a cellular phone near the device, or it may cause electromagnetic interference, and thus result in an operational failure. 3. Relax yourself and breathe naturally before measuring, please keep quiet and keep correct posture in measuring process. Incorrect posture (for example, incorrect sitting posture, the height of measuring position is inconsistent with the height of the heart) or non-quiet status (for example, limbs shaking, head and body swing, talking and being nervous) may cause measuring results not correct. 4. Do not attempt to maintain or modify the device by yourself. Please contact the manufacturer or authorized distributor if maintenance service needed. 5. Repeat measuring the same person with an interval of at least 1 minute, because too frequent measurements may cause injury to you due to blood flow interference and get incorrect measurements. 6. Double-bladder cuff should not be overly persistent, and the pressure should keep below the value of 300mmHg, or it may cause arm blood be unable to timely reflow. 7. Do not use the cuff when arm is wet or sweaty, please use it after wiping up. 8. Do not subject the monitor or the AC adapter to strong shocks, such as dropping on the floor, otherwise it may damage the unit . 9. Do not soak the unit or cuff in water. 10. Operate the device only as intended. Do not use the device for any other purpose. 11. Do not store or use this device outside the specified temperature or humidity range, or it may not achieve the claimed performance. 12. If the monitor and battery achieved longevity, do not arbitrarily discard them, they should be processed according to the local environmental protection regulations in order to avoid environmental pollution. 13. The cuff and Li batteries are classified as consumables. 14. The cuff still meets safety and performance requirements after cycle use for 10000 times; Li batteries capacity will decrease after charge-discharge for 300 times. NOTES: 1. When in idle state (the 2nd process of measuring blood pressure), press the button “
” for about 5 seconds, the blood pressure monitor will enter in “P mode” for
measuring. In this state, users have two methods of exit: a. Press the button “
” for
about 5 seconds; b. Wait for 15 to 20 seconds to automatically exit (no-operation), users please don’t do this operation if unnecessary. 2. Calibration should be made every three years. It should be sent back to Shenzhen RSP Medical Technology Co., Ltd or its authorized distribution for calibration.
Model LBP30 Smart Pulsewave Blood Pressure Monitor
Otherwise the device may reach less than the claimed accuracy.
COMPONENTS OF THE PRODUCT
Main unit:
Display:
Diastolic blood pressure
Unit
Pulse rate Systolic blood pressure Measurement data
Model LBP30 Smart Pulsewave Blood Pressure Monitor
DualCuff: (Applicable arm circumference: 22cm-32cm)
Connect to upper bladder Connect to lower bladder
The arrow directs down.
RECHARGE THE BATTERY
1. When the screen displays “LOP”, it means low battery, please recharge it in time;
2. Press the button “
” to turn on the device, “- 0 -” flashing indicates the battery is
low, in this state can complete the measurement, please recharge in time after the measurement.
3. In charging process, the charging indicator light is orange; when the battery has been fully charged, the charging light indicator turns green.
CAUTION
1. Please use the standard power adapter (5V-1A) or other power supply equipment
(5V-1A) to recharge products. 2. In the process of charging, please do not use the product; please do not plug in the
power adapter with wet hands.
Model LBP30 Smart Pulsewave Blood Pressure Monitor 3. If the product was not operated for 3 months, please store the device with fully
charged battery.
APPLYING THE CUFF
1. When measuring, take off the thick garments and keep the upper arm bared. Thin clothing does not affect the measurement if it does not take pressure to the arm.
2. Put your arm through the cuff loop with the monitor installed on the cuff be at the middle of the arm. After placing the cuff correctly, straining the tail of the cuff to fasten the cuff.
NOTES: 1. Do not put your thick clothes into cuffs, and the bottom of the cuff should be approximately 1cm above the elbow. 2. It will get an incorrect result when the cuff is applied too tightly or too loosely, it is advisable to put in a finger, that is to say, a finger can be put in freely between the cuff and arm.
POSTURES INSTRUCTIONS
1. Sit in a chair with your feet flat on the floor.
Model LBP30 Smart Pulsewave Blood Pressure Monitor
2. Place your arm on the table so that the cuff is at the same level as your heart.
3. Be relaxed and breathe normally before measurement.
NOTE: Your arm cuff and heart should be on the same level; otherwise, measurement accuracy will be affected.
SETTING UNITS
In shutdown status, press and hold the button “
” for over 5 seconds to enter unit
setting mode. When “mmHg” displays on the screen, it means the unit is “mmHg”; if “kPa” displays on the screen, it means the unit is “kPa”.
Model LBP30 Smart Pulsewave Blood Pressure Monitor
NOTE: If you need to reset the units, you must turn the monitor off, then operate the unit according to the above steps.
TAKING A MEASUREMENT
This monitor has two different units optional: mmHg and kPa. The default unit is “mmHg”. The following interpretation take “mmHg” for example. Users should try to relax the body before measurement, and then sit for 2-3 minutes. Users are advised to be measured at the same time daily. 1. Apply the DualCuff correctly (refer to the section “APPLYING THE DUAL CUFF”) and set right measuring postures (refer to the section “POSTURES INSTRUCTIONS”). Do not move your hands in the process of measurement.
2. Press the button “
” to start the monitor, and then it shows “-0-”, thus enter
standby mode.
3. Press the button “
” to take a measurement.
4. When the screen shows the error code of “EE” during inflation process, Please consult the “ERROR INDICATORS”.
5. After the bladder inflated successfully, the monitor begins to test the pulse wave.
Model LBP30 Smart Pulsewave Blood Pressure Monitor
6. After inflation stopped, the bladder will deflate with a constant speed, the monitor enters the measuring state and the value on the display goes down.
7. After the measurement finished, the DualCuff deflates quickly, and your blood pressure and pulse rate will show on the display. First systolic blood pressure and diastolic pressure are shown, two seconds later, pulse rate is shown. The blood pressure and pulse rate will be shown on the display circularly.
8. Press the button “
” to turn off the monitor. If you want to quit the
measurement in the process of measurement, please press the button “
”.
NOTES: 1.
The monitor will automatically turn off after 60 seconds without operation.
2.
Do not repeat the measurements in a short time, or it may cause the arm blood can’t timely return which may result in incorrect results. A new measurement can be started after at least 1 minute once the previous measurement finished.
3.
If in the measurement process the body movement occurs, an error message will appear (refer to the section “ERROR INDICATORS”) , please re-measure and keep quiet until the measurement finished,
Model LBP30 Smart Pulsewave Blood Pressure Monitor
4.
If in the measurement process a system error occurs for some reasons which result in measurement failure or the cuff is excessively inflated, please press the button “
” to turn off the device and restart it.
CARE AND MAINTENANCE
1. If the monitor gets dirty, use soft cloth dipped little water or a mild detergent to slightly wipe it. Do not use gasoline, thinners or other solvents. 2. Do not wet the DualCuff or let the liquid enter the device. 3. Do not crash or fall down the monitor or AC adapter. 4. Keep the monitor and accessories in packing box when not in use. 5. Do not subject the monitor or the AC adapter to extreme high or low temperatures, humidity or direct sunlight. 6. Do not forcefully bend the DualCuff or airway tubes. Do not fold tightly. 7. Do not start measurement when the DualCuff is not applied, or else, the DualCuff may be damaged. 8. Changes or modifications not approved by RSP Medical will void the user warranty. Do not disassemble or attempt to repair the unit or components. 9. Remove the batteries if the unit will not be used for three months or longer.
ERROR INDICATORS
Error code EE 01
Reasons The pressure of cuff reaches 270mmHg
EE 02
Air is leaking from the bladder
Others
Unknown errors
Measures Turn off the monitor and remove it quickly Return to the manufacturer or authorized distributor for maintenance Contact RSP Medical
TROUBLESHOOTING TIPS
The following chart lists common faults you may come across when you use blood pressure monitor. Please contact our after-sales service department in case problems still cannot be solved.
Model LBP30 Smart Pulsewave Blood Pressure Monitor Number
Phenomenon of fault No display appears on the unit
Possible reason
2
Unable to measure or measurement value is too high
3
Measurement values appear too high or too low
Arm cuff may not be at the same level as the heart The measurement values will change at times when you measure under tension
1
No power
Solution Use the AC adaptor to supply power Please apply the arm cuff correctly and keep quiet Take a deep breath before measurement to relax yourself and keep quiet
PRODUCT SPECIFICATIONS
Name: Smart Pulsewave Blood Pressure Monitor Model: LBP30 Measurement Range: Pressure: 0-270(mmHg) (0-36[kPa]) Pulse Rate: 40/min-180/min Accuracy: Pressure: ±3mmHg (±0.4kPa) Pulse Rate: ±5% Power Supply: DC3.7v (Li batteries) Operating Temperature / Relative Humidity: 5℃~40℃, 15%RH~80%RH Storage and Transportation Temperature / Relative Humidity: -20℃~+55℃, ≤93%RH Air Pressure: 80kPa~106kPa Storage and Transportation Air Pressure: 50kPa~106kPa Main Unit Dimension (not including the cuff): 165mm(L)×116mm(W)×67mm(H) Shock Protection: Internal power source; BF TYPE Main Unit Weight (including the cuff): Approximately 287g Applicable Arm Circumference: 22cm~32cm
Model LBP30 Smart Pulsewave Blood Pressure Monitor
APPENDIX A: PACKING LIST
When the user opens blood pressure monitor’s packaging, please check the following packing list. If objects are not complete or have other questions, please contact RSP Medical or authorized distributors.
Number
Name
Quantity
1
Main unit integrated DualCuff
1
2
Instruction manual (including the product certification and warranty card)
1
3
USB cable
1
APPENDIX B: EMC
Please install and use this instrument according to the EMC information provided in this Instruction Manual. The portable and mobile RF communications equipment can affect this instrument‘s normal operation. Please use the accessories sold by our company, the inappropriate one may result in increased emission or decreased immunity of this instrument. The instrument should not be used adjacent or stacked with other equipment and if adjacent or stacked use is necessary, please verify its normal operation in the configuration in which it will be used.
Model LBP30 Smart Pulsewave Blood Pressure Monitor Table 1: 1
Guidance and manufacturer’s declaration – electromagnetic emission
2
The Smart Pulsewave Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of Smart Pulsewave Blood Pressure Monitor should assure that it is used in such an environment.
3
Emissions test
Compliance
Electromagnetic environment - guidance
Group 1
The Smart Pulsewave Blood Pressure Monitor uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment
RF emissions 4 EN 55011 RF emissions 5
Class B EN 55011 Harmonic emissions
6
Class A EN 61000-3-2
7
Voltage fluctuations / flicker emissions
The Smart Pulsewave Blood Pressure monitor is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Complies
EN 61000-3-3
Table 2: Guidance and manufacturer’s declaration – electromagnetic immunity The Smart Pulsewave Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Smart Pulsewave Blood Pressure Monitor should assure that it is used in such an environment. Immunity test Electrostatic discharge (ESD) EN 61000-4-2
EN 60601 test level
Compliance level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
Electromagnetic environment guidance Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Model LBP30 Smart Pulsewave Blood Pressure Monitor
Electrostatic transient / burst
± 2 kV for power supply lines
± 2 kV for power supply lines
EN 61000-4-4
± 1 kV for input/output ± 1 kV for input/output lines lines
Surge
± 1 kV differential mode
EN 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines EN 61000-4-11
Power frequency (50/60 Hz) magnetic field
± 2 kV common mode
± 1 kV differential mode ± 2 kV common mode
< 5 % UT (>95 % dip in UT ) for 0.5 cycle
< 5 % UT (>95 % dip in UT ) for 0.5 cycle
40 % UT (60 % dip in UT ) for 5 cycles
40 % UT (60 % dip in UT ) for 5 cycles
70 % UT (30 % dip in UT ) for 25 cycles
70 % UT (30 % dip in UT) for 25 cycles
< 5 % UT (>95 % dip in UT ) for 5 sec
< 5 % UT (>95 % dip in UT ) for 5 sec
3 A/m
3 A/m
EN 61000-4-8 NOTE
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Smart Pulsewave Blood Pressure Monitor requires continued operation during power mains interruptions, it is recommended that the Smart Pulsewave Blood Pressure Monitor be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
UT is the a. c. mains voltage prior to application of the test level. Table 3:
Guidance and manufacturer’s declaration – electromagnetic immunity The Smart Pulsewave Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Smart Pulsewave Blood Pressure Monitor should assure that it is used in such an environment. Immunity test
EN 60601 test level Compliance level Electromagnetic environment - guidance
Model LBP30 Smart Pulsewave Blood Pressure Monitor Portable and mobile RF communications equipment should be used no closer to any part of the Smart Pulsewave Blood Pressure Monitor, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance
Conducted RF
3 Vrms 3V
EN 61000-4-6
d [
3.5 ] P V1
d [
3.5 ] P E1
80 MHz to 800 MHz
7 ] P E1
800 MHz to 2.5 GHz
150 kHz to 80 MHz
d [
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).b Radiated RF
3 V/m 3 V/m
EN 61000-4-3
80 MHz to 2.5 GHz
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey.a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from structures, objects and people. a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Smart Pulsewave Blood Pressure Monitor is used exceeds the applicable RF compliance level above, the Smart Pulsewave Blood Pressure Monitor should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Smart Pulsewave Blood Pressure Monitor.
Model LBP30 Smart Pulsewave Blood Pressure Monitor b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m. Table 4:
Recommended separation distances between portable and mobile RF communications equipment and the Smart Pulsewave Blood Pressure Monitor The Smart Pulsewave Blood Pressure Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Smart Pulsewave Blood Pressure Monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Smart Pulsewave Blood Pressure Monitor as recommended below, according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter m Rated maximum output of transmitter
150 kHz to 80 MHz
d [
3.5 ] P V1
80 MHz to 800 MHz
d [
3.5 ] P E1
800 MHz to 2.5 GHz
d [
7 ] P E1
W 0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Model LBP30 Smart Pulsewave Blood Pressure Monitor
Product Name: Smart Pulsewave Blood Pressure Monitor Model: LBP30 Shenzhen RSP Medical Technology Co. Ltd 2203, Taibang Technical Building, Gaoxin S. 4th Rd, Nanshan District, Shenzhen, 518000, China.
Website: www.lepuhomecare.com Tel: +86 -755-86952278 Email: