Transcript
Engineering Specification SPEC, APPVL INST 9739MVD IECEx EB-20019335 Revision: AA Number of Pages: 6 Comments:
Originator:
RCS 6/29/11
Approved:
RCS 6/29/11
Rev
ECN
AA
1042090
Description Release to Approvals
Approval
Date
RCS
6/29/11
THIS DOCUMENT IS CONFIDENTIAL, CONTAINS PROPRIETARY INFORMATION, AND SHOULD NOT BE DISTRIBUTED, COPIED OR OTHERWISE REPRODUCED WITHOUT THE EXPRESS WRITTEN CONSENT OF MICRO MOTION, INC.
Micro Motion, Inc.
Page 1 of 6
SPEC, APPVL INST 9739MVD IECEx
Subject:
Equipment type
Manufactured and submitted for
EB-20019335 Revision AA
Transmitter type 9739MVD******** Micro Motion, Inc.
examination Address
Boulder, Co. 80301, USA
Standard basis
IEC60079-0:2007 IEC60079-1:2007 IEC60079-11:2006
Code for type of protection
Ex d [ib] IIB/IIC T6 Gb [Ex ib Gb] IIB/IIC
EC Type Examination Certificate
IECEx BVS 11.0063 X
General requirements Flameproof enclosure ´d´ Intrinsic safety ´i´
THIS DOCUMENT IS CONFIDENTIAL, CONTAINS PROPRIETARY INFORMATION, AND SHOULD NOT BE DISTRIBUTED, COPIED OR OTHERWISE REPRODUCED WITHOUT THE EXPRESS WRITTEN CONSENT OF MICRO MOTION, INC.
Micro Motion, Inc.
Page 2 of 6
SPEC, APPVL INST 9739MVD IECEx
1)
EB-20019335 Revision AA
Subject and Type Transmitter type 9739MVD******** Instead of the * in the complete denomination, letters and numerals will be inserted which characterize the following variations: 9 7 3 9 MV D 6 * * * * * * * Letters for factory options Language Letter for conduit connections Approval 3 = IECEx Intrinsically safe sensor outputs I = IECEx Intrinsically safe sensor outputs flameproof transmitter Display 1 = Dual Line Display 3 = No Display Power 6 = 12-30 VDC and 85-250 VAC (auto ranging)
2)
Description The transmitter is, in combination with a sensor, used for measurement of mass flow and data transmission. The electrical circuitry of the transmitters is mounted inside a flameproof metal enclosure type 9739MVD6*I*****. The 9739MVD6*3***** is not a flameproof enclosure.
THIS DOCUMENT IS CONFIDENTIAL, CONTAINS PROPRIETARY INFORMATION, AND SHOULD NOT BE DISTRIBUTED, COPIED OR OTHERWISE REPRODUCED WITHOUT THE EXPRESS WRITTEN CONSENT OF MICRO MOTION, INC.
Micro Motion, Inc.
Page 3 of 6
SPEC, APPVL INST 9739MVD IECEx
3)
Parameters
3.1
Mains circuit (see document EB-20019903 for terminals) voltage max. voltage
3.2
EB-20019335 Revision AA
Um
AC/DC AC/DC
12 -250 V 250
V
For Transmitter type 9739MVD6******* Intrinsically safe circuits type of protection Ex ib IIC / Ex ib IIB The circuits designed for connecting sensors are classified initially in Group IIC. However, when certain sensors are connected, they can also be assigned to Group IIB.
3.3.1 Drive circuit (see document EB-20019903 for terminals) voltage current power Internal resistance
Uo Io Po Ri
DC
10,5 1,04 2,11 10,12
V A W Ω
type of protection Ex ib IIC max. external inductance max. external capacitance max. inductance/resistance ratio
Lo Co Lo/Ro
33 2,41 13,05
μH μF μH/Ω
type of protection Ex ib IIB max. external inductance max. external capacitance max. inductance/resistance ratio
Lo Co Lo/Ro
131 16,8 52,6
μH μF μH/Ω
The maximum external inductance L (sensor coil) can be calculated with the following term: L = 2 x E x ( Ri + Ro / 1,5 x Uo)2 Whereby E= 40µJ for group IIC and E = 160µJ for group IIB will be inserted and Ro is the total resistance (coil resistance + series resistance).
THIS DOCUMENT IS CONFIDENTIAL, CONTAINS PROPRIETARY INFORMATION, AND SHOULD NOT BE DISTRIBUTED, COPIED OR OTHERWISE REPRODUCED WITHOUT THE EXPRESS WRITTEN CONSENT OF MICRO MOTION, INC.
Micro Motion, Inc.
Page 4 of 6
SPEC, APPVL INST 9739MVD IECEx
EB-20019335 Revision AA
3.3.2 Pick-off circuits (see document EB-20019903 for terminals) voltage Uo current Io power Po
17,3 18,05 30
V mA mW
type of protection Ex ib IIC max. external inductance max. external capacitance
Lo Co
109 353
mH nF
type of protection Ex ib IIB max. external inductance max. external capacitance
Lo Co
436 2,06
mH μF
17,3 21 91
V mA mW
3.3.3 Temperature circuit (see document EB-20019903 for terminals) voltage Umax curent Imax power Pmax
3.4
DC
DC
type of protection Ex ib IIC max. external inductance max. external capacitance
Lo Co
80,4 353
mH nF
type of protection Ex ib IIB max. external inductance max. external capacitance
Lo Co
322 2,06
mH μF
Ambient temperature range: 9739MVD6*******
Ta Ta
-30 °C to +55 °C or -40 °C to +55 °C (routine test required, only for 9739MVD6*I*****)
THIS DOCUMENT IS CONFIDENTIAL, CONTAINS PROPRIETARY INFORMATION, AND SHOULD NOT BE DISTRIBUTED, COPIED OR OTHERWISE REPRODUCED WITHOUT THE EXPRESS WRITTEN CONSENT OF MICRO MOTION, INC.
Micro Motion, Inc.
Page 5 of 6
SPEC, APPVL INST 9739MVD IECEx
4)
EB-20019335 Revision AA
Marking -30 °C ≤ Ta ≤ +55 °C for 9739MVD or -40 °C ≤ Ta ≤ +55 °C for 9739MVD (routine test required, only for 9739MVD**I*****) - type 9739MVD**I*****
- type of protection
9739MVD**3*****
[Ex ib Gb] IIC
Ex d[ib] IIC T6 Gb
5)
Special conditions for safe use / Installation instructions for 9739MVD**I*****.
5.1
For the application of the transmitter in an ambient temperature of less than - 20 °C suitable cable and cable entries or conduit entries certified for this condition shall be used.
5.2
If certified conduit entries are used for the connection of the transmitter enclosure, the associated stopping boxes shall be installed immediately at the enclosure.
5.3
The transmitter shall only be installed within the intended hazardous area if metal cable entries (with ¾”–14 NPT threading) are used which are classified as Ex d IIC and are for enclosures with >2dm3 and are certified by an authorized test station.
5.4
Entry holes which are not being used must be sealed with blanking plugs and which are classified as Ex d IIC and are certified by an authorized test station.
5.5
For installation outside the hazardous area, it is allowed to use cable entry fittings that are not flameproof.
5.6
The non-intrinsically safe end of the transmitter must only be connected to devices where there are no voltages higher than 250V.
6)
Special conditions for safe use / Installation instructions for 9739MVD*(1,3)*****.
6.1
The transmitter must be installed outside the hazardous area in such a way that it meets at least the safety requirement IP20 according to IEC 60529.
6.2
The non-intrinsically safe end of the transmitter must only be connected to devices where there are no voltages higher than 250V.
THIS DOCUMENT IS CONFIDENTIAL, CONTAINS PROPRIETARY INFORMATION, AND SHOULD NOT BE DISTRIBUTED, COPIED OR OTHERWISE REPRODUCED WITHOUT THE EXPRESS WRITTEN CONSENT OF MICRO MOTION, INC.
Micro Motion, Inc.
Page 6 of 6