Preview only show first 10 pages with watermark. For full document please download

Thermometre No Contact 2090

   EMBED


Share

Transcript

2090 EN 3 DISPLAY AND ICONS Thermometre No Contact FUNCTION 1 INTRODUCTION & CLASSIFICATION   Battery Level This thermometer is classified as a Class II Medical Device, sorted as internally- powered equipment and type B application device. Water proof rating is: IPX0. It is prohibited to use this thermometer in flammable anesthetic gas or gas mixture of air and oxygen or nitrous oxide. This is a continuous operation equipment. ICON   Beeper LCD   display Measurement mode Probe Setting buttons Reading display Memory Measurement button Measurement mode Menu   Measurement method Non-contact Effective distance 5cm~8cm Resolution Working condition Storage condition Power supply Power consumption Invisible Battery power is sufficient. Visible Flashes Beeper is on. Beeper is off. Body mode Object mode °C °F   Temperature value Temperature value of last measurement DETAILS Body/object To switch measurement mode between body and object Activates menu and save settings Increases parameter or select relevant parameter Reduces parameter or select relevant parameter Settings Human Body 35.5°C~42.9°C (95.9°F~109.2°F) Object 0°C~118°C (32°F~244°F) Human body Object ±0.2°C/0.4°F ±1.0°C/1.8°F 0.1°C/°F 10°C~40°C (50°F~104°F) RH≤80% Non-condensing -25°C~55°C (-13°F~131°F) RH≤90% Non-condensing d.c. 3V (2x AA alkaline batteries) When off When measuring ≤10µW ≤30µW Power level indicator Indication for low power level Memory Automatic recall of last reading Display LCD back-lit Reading Scale Celsius or Fahrenheit Automatic shut-off in 30 seconds Dimensions 93mm x 153mm x 41mm Net weight 125g Standards EN60601-1 EN 12470-5 ASTM 1965-98 Menu + - 5 SETTING User can change reading scale between Celsius or Fahrenheit, switch on or switch off beeper, set alarm value, set deviation value and change measurement mode between human body or object. Please refer to below chart. MENU F-1 F-2 F-3 F-4 SAVE FUNCTION Reading scale setting Beeper on/off Alarm value setting Deviation value setting - + °C °F DEFAULT REMARKS °C Off On On Decrease Increase 38°C 0.1°C 0.1°C Decrease Increase °C 0.1°C 0.1°C Save and automatically shut off When thermometer is on, it displays current measurement mode ( fig.8.1). Switch the “Body/Object” button to change measurement mode ( fig.8.2). Reading scale setting: F-1 Activates setting menu by pressing “MENU” button. F-1 will be displayed. Press “+” to select Fahrenheit ºF as temp unit (icon “°F” is flashing, fig.8.3); or press “-” to select Celsius ºC as temp unit (icon “°C” is flashing, fig.8.4). Effective range 37°C~42°C Effective range 0°C~3°C   Measurement mode setting Alarm value setting: F-3 Press MENU again to save the setup and F-3 will be displayed (fig.8.8). Defaulted value “38.0°C” will be displayed (fig.8.9). Press “+” or “-” to increase or decrease alarm value (fig.8.10). Value will be increased or decreased by 0.1°C step by step.   [fig 8.6]   [fig 8.8]     [fig 8.7]   [fig 8.9] [fig 8.10]   [fig 8.12]   Press MENU to save the setup and “SAVE” will be displayed (fig.8.14). All setup will be saved and thermometer will automatically shut off (fig.8.15). After activating setup menu, users may select specific [fig 8.14] functions(F1/F2/F3/F4) directly for specific setting. New setup will not be saved if the thermometer is shut off unexpectedly during the procedure of setting. After activating setup menu, the thermometer is unable to measure temperature. There will be no response when pressing measurement button. ! Notice Temperature reading under human body mode is obtained from dynamic compensation of environmental temperature and forehead surface temperature. Object temperature mode is to test surface temperature of an object. The temperature obtained from the forehead under this mode is merely temperature of the skin surface but not core body temperature. Deviation setting is able to adjust measurement value from 0°C to 3.0°C according to testing distance, surrounding temperature, skin difference, etc. Defaulted value is 0.8°C. For example: if the tested body temperature is 36.2°C while the temperature is supposed to be 37.0°C, “F-4” to set up the value and adjust to 1.6°C, then the measurement value will be 37.0°C. 6.1 Body temperature [fig 8.2]   [fig 8.3]   6 MEASUREMENT   [fig 8.1]   Press MENU again to save the setup and F-4 will be displayed (fig.8.11). Defaulted deviation value is 0.8°C (fig.8.12). When defaulted deviation value is flashing, press “+” or “-” to increase or decrease deviation value (fig.8.13). [fig 8.11] Deviation value will be increased or decreased by 0.1°C step by step. Save and exit (automatic shut off) BUTTON ® Press “MENU” again to save the setup and F-2   will be displayed (fig.8.5). Beeper is defaulted as “OPEN”. Press “+” to switch on beeper and “OPEN” will be displayed (fig.8.6). Press “-” to switch off beeper and “CLOS” will be displayed (fig.8.7). [fig 8.5] Deviation value setting: F-4 Celsius reading Fahrenheit reading   2 TECHNICAL PARAMETERS Accuracy Flashes The battery is exhausted and thermometer can not function properly. Please replace battery immediately. 4 FUNCTION DEFINITION OF BUTTONS Battery lid Range Visible The battery is low, but the thermometer is functioning properly. Please replace battery as soon as possible. Body temp Object temp Reading scale Beeper setting: F-2 DETAILS [fig 8.4] Press measurement button to turn on thermometer and it displays boot screen (fig.9.1). After POST and two beeps, it will display value of last reading and be ready for measurement (fig.9.2). Make sure the thermometer is under body mode.     Keep distance at 5cm to 8cm from upper eyebrows to the probe (fig.9.3). Press measurement button and when it gives a “beep” measurement is finished and value will be displayed (fig.9.4). If measurement [fig 9.1] value is exceeding alarm value (Defaulted value is [fig 9.2] 38°C), it gives 3 successive beeps as an indication.     After measurement, if the thermometer is idle for 30 seconds, it will display “OFF” (fig.9.5) and gives a “beep” and shut off automatically. [fig 9.3] [fig 9.4] ! Notice Keep the sensor and probe cavity clean before and after use. To ensure the accuracy of measurement, it is recommended to start measurement ten minutes after carrying the thermometer to a new environment. 8 MAINTENANCE & TIPS   [fig 9.5] Wait for 10 minutes to measure body temperature after measuring extremely high temperature or extremely low temperature. Wait for 5 minutes to start a measurement when the measuring target (object or human) is from an environment with enormous difference in temperature. Breeze, water, sweating, cosmetic on the forehead may affect measurement. Do not measure body temperature 30 minutes after exercise, bath or meals. 6.2 Object temperature Press measurement button to turn on thermometer (fig.9.6).     Make sure the thermometer is under object mode. Keep vertical distance at 5cm to 8cm from object to measurement probe. Press measurement button and when it gives a “beep” measurement is finished and value will be displayed (fig.9.7). [fig 9.6] [fig 9.7] After measurement, if the thermometer is idle for 30 seconds, it will display “OFF” (fig.9.8) and gives a beep and shuts off automatically. ! Notice   The value under this mode is object surface temperature instead of core temperature. The defaulted infrared emissivity is 0.95. The reading will be deviated from the real temperature according to different emissivity. For example, the reading on stainless steel will be obviously [fig 9.8] lower than real temperature. BE CAUTIOUS FOR SCALDING. ! Notice     [fig 9.10] [fig 9.11]   [fig 9.12] [fig 9.13] When surrounding temperature is lower than 10.0ºC or higher than 40.0ºC, it displays “Err” (fig.9.13). Measurement is not possible and hence accuracy is not assured. 7 BATTERY REPLACEMENT Open the battery lid and take out the exhausted battery. Put in 2x AA alkaline batteries and shut the battery lid. After new battery is installed, thermometer will give 2 successive beeps. If no beeps are heard, check if the positive and negative pole are correct (see fig.10.1). ! Notice Read this manual book thoroughly before use. Make sure that the batteries are well installed. It is not recommended to put the thermometer in any liquid or expose to strong sunlight or extremely low temperature. Strong impact to the product will cause damage. Do not dismantle this thermometer by yourself. Keep the thermometer away from children’s reach. Do not use the thermometer in cases of strong magnetic intereference. The measurement results can fluctuate due to improper measurement procedures. Please practice adequate measurement in order to improve your skill. The measurement results cannot supersede a doctor’s diagnosis. Special maintenance is not necessary for this thermometer. Please contact distributor or manufacturer in case of malfunction. 9 TROUBLESHOOTING DESCRIPTION LCD display “Lo” or “Hi”   [fig 10.1] Take out battery in case the thermometer is not used for long period. Do not discard the battery in a fire. Dispose of your batteries according to local regulations. SOLUTION Breeze, water, sweating, cosmetic on the forehead may affect measurement. Check deviation value setting. Defaulted value is 0.8°C. If the environmental temperature changes enormously or if the thermometer is used directly from high temperature to very low temperature, the measurement difference will happen. The thermometer should be kept in a relative stable environment for 10 minutes to obtain heat balance before starting a new measurement. Ensure measurement distance is 5cm to 8cm. No or improper display Take out and reassemble battery. Check if the thermometer is under menu setup. In procedure of menu setting, thermometer is unable to measure and therefore no response. Take out battery and install battery again. No beeper Check if beeper is switched off. Shut off right after switching on Check battery level or take out and install the battery again. No response when pressing measurement button 6.3 Exceeding measurement range Body mode: When measurement value is   lower than 35.5ºC, it displays “Lo” (fig.9.9) and gives 4 successive beeps. When measurement value is higher than 42.9 ºC, it displays Hi (fig.9.10) and gives 4 successive beeps. [fig 9.9] Object mode: When measurement value   is lower than 0ºC, it displays “Lo” (fig.9.11) and gives 4 successive beeps. When measurement value is higher than 118.0ºC, it displays “Hi” (fig.9.12) and gives 4 successive beeps. Make sure the sensor and probe cavity are clean otherwise accuracy will be affected. Cleaning method for probe: Use a cotton stick or soft cloth with water or alcohol to wipe the casing. Use a cotton stick or soft cloth with alcohol to wipe the sensor surface or probe cavity gently. Do not use thermometer before alcohol is vaporized. 10 QUALITY COMMITMENT AND AFTER-SALE SERVICE This product benefits of a warranty of 24 months on material and manufacturing defects, starting from the date of purchase (see sales slip). The warranty of 24 months does not include damages caused by the usual use of parties identified as “consumable” (e.g., batteries, brush heads or parts subject to usury). The legal guarantee of 24 months is void if: The product has undergone aesthetic damage due to improper use not in accordance with instructions in the manual. This product has been modified and / or tampered with. The cause of the failure was due to poor maintenance of the individual components and / or accessories and / or supplies (e.g. oxidation and / or scaling due to the retention of water or other liquids, sediment blocking the sensor, leak of corrosive liquid from batteries.) The following is excluded from the legal guarantee of 24 months: Costs related to replacements and / or repair of parts subject to wear or costs for ordinary maintenance of the product. The costs and risks involved in transporting the product to and from the store where you purchased or otherwise authorized collection centre to receive the products under warranty. Damage caused by or resulting from improper installation or improper use not in accordance with the directions in the instruction manual. Damage due to natural disasters, accidental events or adverse conditions not compatible with the product. Defects that have a negligible effect on product performance. The manufacturer, distributor and all the parties involved in the sale do not assume any liability for losses and economic damage from any malfunction of the product. In accordance with current regulations the manufacturer, distributor and all the parties involved in the sale are not responding in any case for damages, including direct, indirect ones, loss of net income, loss of savings and additional damage and other details consequences going beyond the damage caused by the breach of warranty, contract, strict liability, wrongdoing or due to other causes, resulting from the use or inability to use the product and / or paper and electronic documents, including the lack of service. For further information on the help service visit the website www.nuvitababy. com