Preview only show first 10 pages with watermark. For full document please download

Transmitter H200xph

   EMBED


Share

Transcript

Transmitter H200X pH User Manual Latest Product Information: www.hamiltoncompany.com 084800 Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender). Subject to change Return of products under warranty Please contact our Service Team before returning a defective device. Request a Return Goods Authorization number (RGA). This number assures proper tracking of your device. Ship the cleaned device to the address you have been given. If the device has been in contact with process fluids, it must be decontaminated/disinfected before shipment. In that case, please attach a corresponding certificate, for the health and safety of our service personnel. Disposal Please observe the applicable local or national regulations concerning the disposal of “waste electrical and electronic equipment”. HAMILTON Bonaduz AG Via Crusch 8 CH-7402 Bonaduz Switzerland Tel: +41 81 660 60 60 Fax: +41 81 660 60 70 Email: [email protected] Web: www.hamiltoncompany.com HAMILTON Company 4970 Energy Way Reno, Nevada 89502 USA Tel: +1 775 858 3000 Toll Free: 800 648 5950 Fax: +1 775 856 7259 Email: [email protected] Web: www.hamiltoncompany.com Hamilton Bonaduz AG Shanghai Office Room 502, No.50 Boxia Road Shanghai Pudong Software Park 201203 Shanghai China Phone: +86 21 61646567 Fax: +86 21 61063732 eMail: [email protected] 2 Table of Contents Safety Information ......................................................................... 7 Intended Use...........................................................................................................9 Registered Trademarks ........................................................................................9 Conformity with FDA 21 CFR Part 11 .......................................... 11 Overview of the H200X pH ..............................................................................15 Package Contents ......................................................................... 16 Mounting Plan ......................................................................................................17 Pipe Mounting, Panel Mounting ................................................. 18 Installation and Connection ........................................................ 20 Installation Instructions ....................................................................................20 Connection to Supply Units ............................................................................20 Terminal Assignments .................................................................. 20 Connecting the VP Cable.............................................................. 22 Wiring assignment for HAMILTON VP single coaxial cable‚ VP 6.0’ .............23 Wiring assignment for HAMILTON VP double coaxial cable‚ VP 8.0 DC’.....23 Wiring Examples ........................................................................... 24 User Interface and Display ........................................................... 28 Operation: Keypad ....................................................................... 30 Safety Functions ........................................................................... 31 Sensocheck, Sensoface Sensor Monitoring...............................................31 Hold Mode .............................................................................................................32 Alarm........................................................................................................................33 Configuration ................................................................................ 34 Menu Structure of Configuration ..................................................................35 Overview of Configuration Steps ..................................................................36 Individual Settings ..............................................................................................37 3 Table of Contents Current output: Electrode type, measuring function ............................38 Current output: Output current range, current start, current end ...40 Assignment of measured values: Current start and current end ......41 Current output: Time averaging filter ..........................................................42 Current output: Output current during Error and HOLD .....................44 Temperature compensation: Temperature unit and probe .................46 Temperature compensation: Temp detection for meas/cal, TC process medium ..................................... 48 Calibration mode.................................................................................................50 Alarm settings ......................................................................................................52 Passcodes According to FDA 21 CFR Part 11 .............................. 54 Calibration .................................................................................... 56 Automatic calibration with Calimatic (BUF -xx-) Automatic or manual temperature detection ..........................................58 Manual calibration: Automatic or manual temperature detection ..........................................60 Data entry of premeasured sensors .............................................................62 Product calibration ............................................................................................64 ORP calibration ....................................................................................................66 Temperature dependence of commonly used reference systems ...66 Temp Probe Adjustment .............................................................. 68 Measurement ................................................................................ 68 Diagnostics Functions .................................................................. 69 Cleaning ........................................................................................ 70 Operating States ........................................................................... 71 Error Messages (Error Codes) ...................................................... 72 4 Calibration Error Messages .......................................................... 74 Sensoface ...............................................................................................................76 Sensocheck ............................................................................................................76 Product Line and Accessories ...................................................... 79 Specifications ................................................................................ 80 Buffer Tables ................................................................................. 85 Glossary ......................................................................................... 94 Index .............................................................................................. 97 5 6 Safety Information Safety information Be sure to read and observe the following instructions! The device has been manufactured using state of the art technology and it complies with applicable safety regulations. When operating the device, certain conditions may nevertheless lead to danger for the operator or damage to the device. Caution! Commissioning must be carried out by trained experts. Whenever it is likely that protection has been impaired, the device shall be made inoperative and secured against unintended operation. The protection is likely to be impaired if, for example: • the device shows visible damage • the device fails to perform the intended function • after prolonged storage at temperatures above 70 °C • after severe transport stresses Before recommissioning the device, a professional routine test must be performed. This test should be carried out at the manufacturer's factory. Caution! Before commissioning it must be proved that the transmitter may be connected with other equipment. 7 Safety Information Safety precautions for installation • Observe the provisions of EN 60079-10 / EN 60079-14 during commissioning. • H200X pH may be installed in the following locations: ATEX, Zone 1 with measurement in Zone 0. Connection to Supply Units • H200X pH: must be connected to an explosion-proof power supply unit (for electrical data refer to annex of EC-Type-Examination Certificate). Before commissioning it must be proved that the intrinsic safety is maintained when connecting the device to other equipment, such as power supply and cable. Terminals: suitable for single or stranded wires up to 2.5 mm2 (AWG 14) Cleaning in a hazardous location In hazardous locations the device must be cleaned only with a damp cloth to prevent electrostatic charging. 8 Intended Use The Transmitter H200X is used for pH/mV, ORP and temperature measurement in industry, environment, food processing and sewage treatment. The sturdy molded enclosure can be fixed into a control panel or mounted on a wall or at a post. The protective hood provides additional protection against direct weather exposure and mechanical damage. The device can easily be replaced and it accepts commercially available sensors with a nominal zero point at pH 7. • H200X pH may be installed in the following locations: ATEX, Zone 1 with measurement in Zone 0. Registered Trademarks The following names are registered trademarks of Knick Elektronische Messgeräte GmbH & Co. KG. For practical reasons they are shown without trademark symbol in this manual. Sensocheck® Sensoface® Calimatic® GainCheck® HART® is a registered trademark of the HART Communication Foundation. 9 EC Declaration of Conformity 10 Conformity with FDA 21 CFR Part 11 In their directive “Title 21 Code of Federal Regulations, 21 CFR Part 11, Electronic Records; Electronic Signatures“ the American health agency FDA (Food and Drug Administration) regulates the production and processing of electronic documents for pharmaceutical development and production. This results in requirements for transmitters used for corresponding applications. The following features ensure that the transmitters of the H200X series meet the demands of FDA 21 CFR Part 11: Electronic Signature Access to the device functions is regulated and limited by individually adjustable codes – “Passcodes” (for Passcode Editor see Page 54, overview of factory settings on back of manual). This prevents unauthorized modification of device settings or manipulation of the measurement results. Appropriate use of these passcodes makes them suitable as electronic signature. Audit Trail Every (manual) change of device settings can be automatically documented. Each change is tagged with a “Configuration Change Flag”, which can be interrogated and documented using HART communication. Altered device settings or parameters can also be retrieved and documented using HART communication. 11 EC-Type-Examination Certificate 12 13 14 Overview of the H200X pH Loop current Ex ia IIC 1 Glass electrode 4 ... 20 mA Ex ib IIC pH / mV input 2 14 – Output 15 + Output mA/ HART Reference electrode 4 Auxiliary electrode 5 Not connected 6 RTD 7 RTD 8 PA (equipot. bonding) 16 Do not connect 17 Do not connect 18 Do not connect 19 Temp input 15 Assembly Package Contents Check the shipment for transport damage and completeness. The package should contain: • Front unit • Rear unit • Bag containing small parts • User manual • Specific test report 1 11 10 2 3 9 8 7 6 1 2 3 4 5 6 7 5 Jumper (2 x) Washer (1 x), for conduit mounting: Place washer between enclosure and nut Cable tie (3 x) Hinge pin (1 x), insertable from either side Enclosure screw (4 x) Sealing insert (1 x) Rubber reducer (1 x) Fig.: Assembling the enclosure 16 4 8 9 10 11 Cable gland (3 x) Filler plug (3 x) Hexagon nut (5 x) Sealing plug (2 x), for sealing in case of wall mounting Mounting Plan 105 144 15 144 27 42 84 21 43 1 80 2 1 2 72 32 6,2 3 3 4 Cable gland (3 x) Knockouts for cable gland or conduit 1/2”, dia. = 21.5 mm (2 x) Conduit hardware not included! Knockouts for pipe mounting (4 knockouts ) Knockouts for wall mounting (2 knockouts ) 4 Fig.: Mounting plan 17 Pipe Mounting, Panel Mounting 40 132 60 1 2 3 4 5 1 2 3 4 5 P/N 243084 protective hood (if required) Hose clamp with worm gear drive to DIN 3017 (2 x) Pipe-mount plate (1 x) For vertical or horizontal posts or pipes Self-tapping screw (4 x) Fig.: P/N 243082 pipe-mount kit 165 132 173 1 Fig.: P/N 243084 protective hood for wall and pipe mounting 18 1 max. 25 78 1 2 3 4 5 27 Screw (4 x) Gasket (1 x) Control panel Span piece (4 x) Threaded sleeve (4 x) 1 5 4 Panel cutout 138 x 138 mm (DIN 43700) 2 1...22 3 Fig.: P/N 243083 panel-mount kit 19 Installation and Connection Installation Instructions • Installation must be carried out by trained experts in accordance with this user manual and as per applicable local and national codes. • Be sure to observe the technical specifications and input ratings during installation. • Be sure not to notch the conductor when stripping the insulation. • All parameters must be set by a system administrator prior to commissioning. Connection to Supply Units • H200X pH must be connected to an explosion-proof power supply unit (for electrical data refer to annex of EC-Type-Examination Certificate). Terminals: suitable for single or stranded wires up to 2.5 mm2 (AWG 14) + supply/ output HART ® 4 to 20 mA + do not connect TÜV 99 ATEX 1446 2211X pH II 2(1) G Ex ib [ia Ga] IIC T6 Gb D-14163 Berlin Electrical data see EC-Type Examination Certificate 8 7 6 5 4 2 1 input shield observe grounding 0044 conditions BONDING BETWEEN CONDUIT IS NOT AUTOMATIC AND MUST BE PROVIDED AS PART OF THE INSTALLATION.SEE INSTALLATION INSTRUCTIONS Tamb - 20 to + 55 °C 20 Made in Germany 84733/0000000/1243 F-2211XPH-HAM 14 meas. el. 15 ref. el. 16 n.c. 17 aux. el. 18 RTD 19 RTD Terminal Assignments 7 mm 7 mm 4 13 mm 280 mm 5 1 2 6 3 1 Recommended stripping lengths for multi-core cables 2 Recommended stripping length for coaxial cable 3 Pulling out the terminal blocks using a screwdriver (see also 7) 4 Cable position in transmitter 5 Connecting lines for loop current 6 Cover for sensor and temperature probe terminals 7 Areas for placing the screwdriver to pull out the terminals 8 Connection of handheld terminal 7 8 Fig.: Information on installation, rear side of device 21 Connecting the VP Cable Connecting the sensor to the VP cable VP6 SC sensor cable 355108 1m 355109 3m 355110 5m 355111 10 m 355112 20 m Please note: Cable lengths > 20 m can worsen the response during pH measurement. Be sure to observe the sensor instruction manual. 22 Wiring assignment for HAMILTON VP single coaxial cable‚ VP 6.0’ Color coding in the cable VP pin pH / ORP Coax core black/transparent A pH / ORP Coax shield red B Reference Gray wire C For Pt100: shorted with pin E Blue wire D Solution ground (1) White wire E Pt 100(0) Green wire F Pt 100(0) Outer shield green/yellow Housing Shielding of connector head (2) Wiring assignment for HAMILTON VP double coaxial cable‚ VP 8.0 DC’ Color coding in the cable VP pin pH / ORP Coax core black/transparent A pH / ORP Coax shield black B Reference Coax core red/transparent C For Pt100: shorted with pin E Coax shield red D Solution ground (1) White wire E Pt 100(0) Green wire F Pt 100(0) Yellow wire G - Brown wire H - Outer shield green/yellow Housing Shielding of connector head (2) (1) if provided (2) In the electrode the housing of the VP socket is not connected with any other potential-carrying component of the sensor. 23 pH Wiring Examples Example 1: pH measurement with monitoring of glass electrode Sensor connections Transmitter H200X pH All common pH sensors can be connected (with glass electrode and separate or combined reference electrode, with or without temperature detector, with nominal zero at pH 7). Connecting cable for S7 / S8 screw cap, e.g. 5 m P/N 355 066 Temp. probe 24 pH sensor without temperature detector, with S7 / S8 screw cap Example 2: pH measurement with monitoring of glass electrode, without Liquid Earth Sensor connections Transmitter H200X pH Coax core black/transparent Coax shield red Green White Yellow/Green All common pH sensors with VP connector and nominal zero at pH 7 can be connected. e.g. VP cable VP 6.0 e.g. 1 m P/N 355 108 Easyferm Plus VP pH combination sensor with integrated temperature detector 25 pH Wiring Examples Example 3: pH measurement with monitoring of glass electrode, sensors with Liquid Earth Sensor connections Transmitter H200X pH Coax core black/transparent Coax shield red Blue Green White Yellow/Green All common pH sensors with VP connector and nominal zero at pH 7 can be connected. e.g. VP cable VP 6.0 e.g. 1 m P/N 355 108 Polilyte Plus VP pH combination sensor with Liquid Earth and integrated temperature detector 26 ORP Wiring Example Example 4: ORP (redox) measurement Sensor connections Transmitter H200X pH All common redox sensors can be connected (with separate or combined reference electrode, with or without temperature detector). Connecting cable for S7 / S8 screw cap, e.g. 5 m P/N 355 066 Polilyte RX redox sensor without temperature detector, with S7 / S8 screw cap Temp. probe 27 User Interface and Display User Interface 1 2 3 4 1 2 Display Mode indicators (no keys), from left to right: - Measuring mode - Calibration mode - Alarm - Not in use - Configuration mode 28 3 4 Alarm LED Keypad Display 1 2 3 4 5 6 7 8 9 10 20 11 19 12 18 13 17 16 15 1 Passcode entry 2 Not in use 3 Temperature 4 Current output 5 Limit values 6 Alarm 7 Sensocheck 8 Calibration 9 Interval/response time 10 Not in use 11 Unit symbols 12 Proceed using enter 13 Bar for identifying the device status, above mode indicators, from left to right: - Measuring mode - Calibration mode - Alarm - Not in use - Configuration mode 14 14 Secondary display 15 Manual temperature specification 16 Hold mode active 17 Waiting time running 18 Sensor data 19 Main display 20 Sensoface 29 Operation: Keypad cal conf   enter Cal Info: Display of asymmetry potential (zero) and slope enter Error Info: Display of last error message conf Start GainCheck device self-test + cal 30 enter Start, end calibration Start, end configuration • Select digit position (selected position blinks) • Menu navigation • Edit digit • Menu navigation • Calibration: Continue in program sequence • Configuration: Confirm entries, next configuration step • Measuring mode: Display output current Safety Functions Sensocheck, Sensoface Sensor Monitoring Sensocheck continuously monitors the sensor and its wiring. Sensocheck can be switched off (Configuration). Sensoface provides information on the sensor condition. The asymmetry potential (zero), slope and response time during calibration are evaluated. The three Sensoface indicators provide the user with information on wear and required maintenance of the sensor. GainCheck Device Self-Test A display test is carried out, the software version is displayed, and the memory and measured value transfer are checked. Start GainCheck device self-test:  +  Automatic Device Self-Test The automatic device self-test checks the memory and measured-value transfer. It runs automatically in the background at fixed intervals. 31 Safety Functions Hold Mode Display: The Hold mode is a safety state during configuration and calibration. Loop current is frozen (Last) or set to a fixed value (Fix). If the calibration or configuration mode is exited, the device remains in the Hold mode for safety reasons. This prevents undesirable reactions of the connected peripherals due to incorrect configuration or calibration. The measured value and “HOLD” are displayed alternately. The transmitter only returns to measuring mode after enter is pressed and 20 seconds have passed. 32 Outputs Loop current The loop current is controlled by the process variable selected in the configuration. The current start and end can be set to represent any desired value. To check connected peripherals (e.g. limit switches, controllers), the loop current can be manually specified (see page 70). HART communication H200X pH can be remote-controlled via HART communication. It can be configured using a handheld terminal or from the control room. Measured values, messages and device identification can be downloaded at any time. This allows easy integration also in fully automatic process cycles. If you require “Transmitter-Specific Command Specifications”, please contact Hamilton. Alarm The alarm delay is configurable. Error messages can also be signaled by a 22 mA loop current (see Configuration, page 53). The alarm LED on the front panel can be configured as follows: HOLD off: Alarm: LED blinks HOLD on: Alarm: LED on. HOLD: LED blinks 33 Configuration In the Configuration mode you set the device parameters. Activation Activate by pressing conf Enter passcode “1200“ Use  and  to edit a parameter. Press enter to confirm/continue. (Exit by pressing conf enter.) Hold During configuration the device remains in the Hold mode for reasons of safety. The loop current is frozen (at its last value or at a preset fixed value, depending on the configuration), Sensoface is off, “Configuration” mode indicator is on. HOLD icon Input errors The configuration parameters are checked during the input. In the case of an incorrect input, ”Err” is displayed for approx. 3 sec. The incorrect parameters cannot be stored. Input must be repeated. Exit Exit by pressing conf. The measured value and Hold are displayed alternately, “enter” blinks. Press enter key to exit the Hold mode. The display shows the measured value. The output current remains frozen for another 20 sec (HOLD icon on, “hourglass” blinks). 34 Menu Structure of Configuration The configuration steps are assigned to different menu groups: • Current output (code: o1.) • Temperature compensation (code: tc.) • Calibration mode (code: CA.) • Alarm settings (code: AL.) You can jump between the individual menu groups by using the arrow keys. Each menu group contains menu items for setting the parameters. Example: “o1.” is displayed with all menu items of the “Current output” menu group. Pressing enter opens a menu item. Use the arrow keys to edit a value. Press enter to confirm/save the settings. Return to measurement: Press conf. Select menu group Menu group Code Display Current output o1. Select menu item Menu item 1 Menu item 2 Menu item ...    Temperature compensation tc. Calibration mode CA. Alarm settings AL. Previous menu group:  35 Overview of Configuration Steps Code Selection / Default Menu out1 Current output (Factory setting bold print) o1.UnIT Select process variable pH / ORP o1.EL Select electrode type Enter current start (4 mA) GLAS EL (FEt EL - without function) o1.4mA o1.20mA Enter current end (20 mA) - 2.00 ... 16.00 PH (14.00 PH) - 1999 ... 1999 mV (0014 mV) o1.FtME Time averaging filter 0000 ... 0120 SEC (0000 SEC) o1.FAIL 22 mA signal in the case of error ON / OFF o1.HoLD Signal response during HOLD Last / Fix o1.FIX Enter fixed value - 2.00 ... 16.00 PH (00.00 PH) - 1999 ... 1999 mV (0000 mV) 000.0 ... 022.0 mA (021.0 mA) tc Temperature compensation tc.UnIT Select temperature unit °C / °F tc.rTD Select temperature probe Pt100/Pt1000/NTC30 tc.MEAS Temp detection during meas Auto/man (man: -20.0 .... 200 °C) (025.0 °C) (man: -004 ... 0392 °F) (0025 °F) tc.CAL Temp detection during cal Auto/man (man: -20.0 .... 200 °C) (025.0 °C) (man: -004 ... 0392 °F) (0025 °F) tc.LIN Enter TC process medium -19.00 ... 19.99 %/K (00.00 %/K) CAL Calibration mode CA.SOL Select calibration mode BUF / MAN / DAT (-08-BUF) CA.tiME Enter cal timer interval 0000 ... 9999 h (0000 h) ALrt Alarm settings AL.SnSO Select Sensocheck ON / OFF (OFF) AL.dlY Enter alarm delay AL.LED LED in HOLD mode 0000 ... 0600 SEC (0010 SEC) ON / OFF (OFF) 36 Individual Settings Code. Parameter Setting o1.UnIT Unit pH / ORP ____________ o1.EL Electrode type GLAS EL______ ____________ o1.4mA Current start ____________ ____________ o1.20mA Current end ____________ ____________ o1.FtME Filter time ____________ ____________ o1.FAIL 22mA signal ____________ ____________ ____________ o1.HoLD Hold response ____________ ____________ o1.FIX Fixed current ____________ ____________ tc.UnIT Unit °C/°F ____________ ____________ tc.rTD Temp probe ____________ ____________ tc.MEAS Measuring temp ____________ ____________ tc.CAL Calibration temp ____________ ____________ tc.LIN TC process medium ____________ ____________ CA.SOL Calibration solution ____________ ____________ CA.tiME Calibration interval ____________ ____________ AL.SnSO Sensocheck ____________ ____________ AL.dLY Alarm delay ____________ ____________ AL.LED LED HOLD ____________ ____________ 37 Configuration Current output: Electrode type, measuring function conf 1 1. Press conf key. 2. Enter passcode 1200 3. Select Current output menu group using arrow keys. All items of this menu group are indicated by the “o1.” code. 4. Press enter to select menu, edit using arrow keys (see page 39). Confirm (and proceed) by pressing enter. 5. Exit: Press conf, then enter. 2 enter Current output: 3 Select process variable enter 4 enter Select electrode type Enter current start Enter current end Set output filter 22 mA for error Signal response during HOLD 5 38 conf enter Code Display Action o1. Select configuration (Press conf) After correct input, a welcome text (CONF) is displayed for approx. 3 sec Choices Enter passcode “1200“ (Use to select the position, use to edit the number. When the display reads “1200”, press enter to confirm.) Device is in the HOLD mode (HOLD icon is active). Select process variable pH/ORP Select using arrow key  Press enter to proceed. pH/ORP Only with pH selected: Select electrode type: • Glass electrode Select using arrow key  Press enter to proceed. Glas Note: Characters represented in gray are blinking and can be edited. 39 Configuration Current output: Output current range, current start, current end conf 1 1. Press conf key. 2. Enter passcode 1200 3. Select Current output menu group using arrow keys. All items of this menu group are indicated by the “o1.” code. 4. Press enter to select menu, edit using arrow keys (see page 41). Confirm (and proceed) by pressing enter. 5. Exit: Press conf, then enter. 2 enter Current output: 3 Select process variable enter 4 enter Select electrode type Enter current start Enter current end Set output filter 22 mA for error Signal response during HOLD 5 40 conf enter Code Display Action Choices o1. Current start Enter lower end of scale, depending on the process variable selected (pH or ORP) Select using , edit number using , press enter to proceed. pH -2 ... 16 Current end Enter upper end of scale, depending on the process variable selected (pH or ORP) Select using , edit number using , press enter to proceed. pH -2 ... 16 (-1999 mV ... +1999 mV (-1999 mV ... +1999 mV) Assignment of measured values: Current start and current end Example 1: Range pH 0 ... 14 Example 2: Range pH 5 ... 7. Higher resolution in range of interest [pH] 14 [pH] 7 Output current 0 Output current 5 4 20 [mA] 4 20 [mA] 41 Configuration Current output: Time averaging filter conf 1 1. Press conf key. 2. Enter passcode 1200 3. Select Current output menu group using arrow keys. All items of this menu group are indicated by the “o1.” code. 4. Press enter to select menu, edit using arrow keys (see page 43). Confirm (and proceed) by pressing enter. 5. Exit: Press conf, then enter. 2 enter Current output: 3 Select process variable enter 4 enter Select electrode type Enter current start Enter current end Set output filter 22 mA for error Signal response during HOLD 5 42 conf enter Code Display o1. Action Choices Time averaging filter Default setting: 0 sec (inactive). To specify a time interval: Select using , edit number using , press enter to proceed. 0 sec 0...120 sec Time averaging filter To smoothen the current output, a low-pass filter with adjustable filter time constant can be switched on. When there is a jump at the input (100 %), the output level is at 63 % after the time interval has been reached. The time interval can be set from 0 to 120 sec. If the time interval is set to 0 sec, the current output follows the input. Note: The filter acts on the current output, not on the display! Time interval 0 ... 120 sec 43 Configuration Current output: Output current during Error and HOLD conf 1 1. Press conf key. 2. Enter passcode 1200 3. Select Current output menu group using arrow keys. All items of this menu group are indicated by the “o1.” code. 4. Press enter to select menu, edit using arrow keys (see page 45). Confirm (and proceed) by pressing enter. 5. Exit: Press conf, then enter. 2 enter Current output: 3 Select process variable enter 4 enter Select electrode type Enter current start Enter current end Set output filter 22 mA for error Signal response during HOLD 5 44 conf enter Code Display o1. Action Choices 22 mA signal for error message Select using arrow key  Press enter to proceed. OFF (ON) LAST Output signal during HOLD (FIX) LAST: During HOLD the last measured value is maintained at the output. FIX: During HOLD a value (to be entered) is maintained at the output. Select using arrow key  Press enter to proceed. Only with FIX selected: Enter current which is to flow at the output during HOLD Select position using  key and edit number using  key. Press enter to proceed. 21.0 mA (00.0 ... 22.0 mA) Output signal during HOLD: Output current [mA] Output signal for HOLD FIX setting = 21.0 mA Output signal for HOLD LAST setting 21 4 HOLD active HOLD active 45 Configuration Temperature compensation: Temperature unit and probe conf 1 2 enter Current output: 1. Press conf key. 2. Enter passcode 1200. 3. Select Temperature compensation menu group using arrow keys. All items of this menu group are indicated by the “tc.” code. 4. Press enter to select menu, edit using arrow keys (see page 47). Confirm (and proceed) by pressing enter. 5. Exit: Press conf, then enter. enter Temperature compensation: 3 4 enter Select °C/°F enter Select temperature probe TC during measurement TC during calibration Enter TC process medium 5 46 conf enter Code Display tc. Action Choices Specify temperature unit Select using arrow key  Press enter to proceed. °C Select temperature probe Select using arrow key  Press enter to proceed. Pt1000 (°F) (Pt100, NTC30) 47 Configuration Temperature compensation: Temp detection for meas/cal, TC process medium conf 1 1. Press conf key. 2. Enter passcode 1200. 3. Select Temperature compensation menu group using arrow keys. All items of this menu group are indicated by the “tc.” code. 4. Press enter to select menu, edit using arrow keys (see page 49). Confirm (and proceed) by pressing enter. 5. Exit: Press conf, then enter. 2 enter Current output: enter Temperature compensation 3 4 enter Select °C/°F enter Select temp probe TC during measurement TC during calibration TC process medium 5 48 conf enter Code Display tc. Action Choices Select temp detection during measurement (Auto/MAN) AUTO: Temp detection with temperature probe MAN: Manual temperature input Select using  key, press enter to proceed AUT Only with manual temp detection selected (MAN) Enter temperature. Select position using  key, edit number using key. Press enter to proceed. 25 °C Select temp detection during calibration (Auto/MAN) Select using  key, press enter to proceed AUT Only with manual temp detection selected (MAN) Enter temperature. Select position using  key, edit number using  key. Press enter to proceed. 25 °C (MAN) (xxx.x °C) (MAN) (xxx.x °C) For pH measurement only: 00.00 %/K Enter temperature compensation of (xx.xx %/K) process medium Select position using  key, edit number using key. Press enter to proceed. 49 Configuration Calibration mode conf 1 1. Press conf key. 2. Enter passcode 1200. 3. Select Calibration mode menu group using arrow keys. All items of this menu group are indicated by the “CA.” code. 4. Press enter to select menu, edit using arrow keys (see page 51). Confirm (and proceed) by pressing enter. 5. Exit: Press conf, then enter. 2 enter Current output: enter Temp compensation enter Calibration mode: 3 4 enter Calibration mode enter Cal timer interval 5 50 conf enter Code Display Action CA. For pH measurement only: Select calibration mode BUF: Calibration with Calimatic automatic buffer selection. To do so, you must select your buffer set: -00- BUF: -01- BUF: -02- BUF: -03- BUF: -04- BUF: -05- BUF: -06- BUF: -07- BUF: -08- BUF: Choices -01-BUF (-01-BUF/ -02-BUF/ -03-BUF/ -04-BUF/ -05-BUF/ -06-BUF/ Knick -07-BUF/ Mettler-Toledo Merck Titrisol, Riedel Fixanal -08-BUF/ Ciba (94) MAN/ NIST technical buffers DAT) NIST standard buffers HACH buffers WTW technical buffers Hamilton Duracal MAN: Calibration with manual buffer entry DAT: Entry of asymmetry potential and slope of premeasured sensors Select using key, press enter to proceed. Enter calibration interval: Entry of time interval within which the device is to be calibrated. With a time interval of 0000 hrs the calibration timer is not active. Select using , edit number using , press enter to proceed. 0000 h (0000 ... 9999 h) 51 Configuration Alarm settings 1 1. Press conf key. 2. Enter passcode 1200. 3. Select Alarm settings menu group using arrow keys. All items of this menu group are indicated by the “AL.” code. 4. Press enter to select menu, edit using arrow keys (see page 53). Confirm (and proceed) by pressing enter. 5. Exit: Press conf, then enter. 2 enter Current output: enter Temp compensation enter Calibration mode: enter Alarm settings: 3 4 enter Select Sensocheck enter Delay LED in Hold mode 5 52 conf enter Code Display AL. Action Choices Select Sensocheck (continuous monitoring of glass and reference electrode) Select using  key, press enter to proceed. ON / OFF Alarm delay Acts on LED indication and 22 mA signal. Select usingkey, edit number usingkey. Press enter to proceed. 0010 s (xxxx s) LED in Hold mode Select using  key, press enter to proceed. ON / OFF LED state: Setting Alarm HOLD ON on blinks OFF blinks off 53 Passcodes According to FDA 21 CFR Part 11 Access to the device functions can be protected with adjustable passcodes if required. If such a protection is not required, you should use the preset passcodes. To open the passcode editor: Press conf key and enter Administrator passcode (1989). Display 54 Action Remark 1. Press conf key 2. Enter Administrator passcode (1989): Welcome text is displayed This text is displayed for approx. 3 sec “Cal info” Edit: Arrow keys Proceed: enter Cancel: conf Default setting: 0000 “Zero adjustment” Edit: Arrow keys Proceed: enter Cancel: conf Default setting: 1001 “Calibration” Edit: Arrow keys Proceed: enter Cancel: conf Default setting: 1100 “Product calibration” Edit: Arrow keys Proceed: enter Cancel: conf Default setting: 1105 “Temp probe adjustment” Edit: Arrow keys Proceed: enter Cancel: conf Default setting: 1015 Display Action Remark “Error info” Edit: Arrow keys Proceed: enter Cancel: conf Default setting: 0000 “Configuration” Edit: Arrow keys Proceed: enter Cancel: conf Default setting: 1200 “Sensor monitor” Edit: Arrow keys Proceed: enter Cancel: conf Default setting: 2222 “Current source” Edit: Arrow keys Proceed: enter Cancel: conf Default setting: 5555 “Administrator passcode” Edit: Arrow keys Proceed: enter Cancel: conf Caution! If you have lost the Administrator passcode, the Passcode Editor cannot be called up! Please consult our technical support! • Discard new administrator passcode Proceed: enter Cancel: conf Default setting: 1989 • Take over new administrator passcode Proceed: enter Cancel: conf 55 Calibration Calibration adjusts the device to the sensor. Activation Activate by pressing cal Enter passcode: 1100 Select using , edit number using , press enter to proceed. (Exit by pressing cal enter.) Hold During calibration the device remains in the Hold mode for reasons of safety. The loop current is frozen (at its last value or at a preset fixed value, depending on the configuration), Sensoface is off, “Calibration” mode indicator is on. HOLD icon Input errors The calibration parameters are checked during the input. In the case of an incorrect input, ”Err” is displayed for approx. 3 sec. The incorrect parameters cannot be stored. Input must be repeated. Exit Exit by pressing cal. The measured value and Hold are displayed alternately, “enter” blinks. Press enter key to end the Hold mode. The measured value is displayed. The output current remains frozen for another 20 sec (HOLD icon on, “hourglass” blinks). 56 pH calibration Calibration is used to adapt the device to the individual sensor characteristics, namely asymmetry potential and slope. Calibration can be performed with Calimatic automatic buffer recognition, with manual buffer input, by entering premeasured electrode data, or by sampling the product. Caution • All calibration procedures must be performed by trained personnel. Incorrectly set parameters may go unnoticed, but change the measuring properties. • The response time of the sensor and temperature probe is considerably reduced when the sensor is first moved about in the buffer solution and then held still. • The device can only operate properly when the buffer solutions used correspond to the configured set. Other buffer solutions, even those with the same nominal values, may demonstrate a different temperature response. This leads to measurement errors. When using sensors with a zero point other than pH 7, the nominal zero point must be adjusted each time a new sensor is connected. This is important if you want to obtain reliable Sensoface messages. The Sensoface messages issued during all further calibrations are based on this basic calibration. 57 Automatic calibration with Calimatic (BUF -xx-) Automatic or manual temperature detection The device can only operate properly when the buffer solutions used correspond to the configured set. Other buffer solutions, even those with the same nominal values, may demonstrate a different temperature response. This leads to measurement errors. Display Action Remark Press cal key, enter passcode 1100. Select position using , edit number using , press enter to proceed. If an invalid passcode is entered, the device returns to measuring mode. Remove the sensor and temperature probe, clean them and immerse them in the first buffer solution (in any order). When “Manual temp detection” has been configured, enter temperature value in the secondary display using the arrow keys. Press enter to start. Device in Hold mode, measured value frozen. Sensoface inactive. Buffer recognition While the “hourglass” icon is blinking, the sensor and temperature probe remain in the first buffer solution. Buffer recognition terminated, the nominal buffer value is displayed. 58 The response time of the sensor and temperature probe is considerably reduced when the sensor is first moved about in the buffer solution and then held still. Display Action Remark Stability check: The measured mV value is displayed. To abort stability check: Press cal (accuracy reduced) Calibration with the first buffer is terminated. Remove the sensor and temp probe from the first buffer solution and rinse them thoroughly. • One-point calibration: Exit by pressing cal. Slope [%] and asymmetry potential [mV] of the sensor are displayed. Press enter to proceed. For one-point calibration only: • Two-point calibration: Immerse sensor and temperature probe in the second buffer solution. Press enter to start. The calibration process runs again as for the first buffer. Retract sensor and temp probe out of second buffer, rinse off, re-install. Repeat calibration: cal Exit calibration: enter Slope and asymmetry potential of sensor (based on 25 °C) are displayed. pH value and Hold are displayed alternately. Press enter to proceed. Hold is deactivated after 20 sec. Confirmation prompt. 59 Manual calibration Automatic or manual temperature detection For calibration with manual buffer specification, you must enter the pH value of the buffer solution used in the device for the proper temperature. This presetting enables calibration with any desired buffer solution. The MAN calibration mode and the type of temperature detection are selected in the configuration mode. Display Action Remark Press cal key, enter code 1100 Select position using , edit number using , press enter to proceed. If an invalid passcode is entered, the device returns to measuring mode. Remove the sensor and temperature probe, clean them and immerse them in the first buffer solution (in any order). When “Manual temp detection” has been configured, enter temperature value in the secondary display using the arrow keys. Press enter to start. Device in Hold mode, measured value frozen. Sensoface inactive. Enter the pH value of your buffer solution for the proper temperature. While the “hourglass” icon is blinking, the sensor and temperature probe remain in the buffer solution. 60 The response time of the sensor and temperature probe is considerably reduced when the sensor is first moved about in the buffer solution and then held still. Display Action Remark Stability check: The measured mV value is displayed. To abort stability check: Press cal (accuracy reduced) Calibration with the first buffer is terminated. Remove the sensor and temp probe from the first buffer solution and rinse them thoroughly. • One-point calibration: Exit by pressing cal. Slope [%] and asymmetry potential [mV] of the sensor are displayed. Press enter to proceed. For one-point calibration only: • Two-point calibration: Immerse sensor and temperature probe in the second buffer solution. Enter the pH value of the second buffer solution. Press enter to start. The calibration process runs again as for the first buffer. Retract sensor and temp probe out of second buffer, rinse off, re-install. Repeat calibration: cal Exit calibration: enter The slope and asymmetry potential of the sensor (based on 25 °C) are displayed. pH value and Hold are displayed alternately. Press enter to proceed. Hold is deactivated after 20 sec. Confirmation prompt. 61 Data entry of premeasured sensors You can directly enter the values for slope and asymmetry potential of a sensor. The values must be known, e.g. determined beforehand in the laboratory. The DAT calibration mode must have been preset during configuration. Display Action Remark Press cal key, enter code 1100 Select using , enter number using , press enter to proceed. If an invalid passcode is entered, the device returns to measuring mode. Ready for calibration Press enter to start. Device in Hold mode, measured value frozen. Sensoface inactive. Enter asymmetry potential [mV]. Select usingkey, edit number usingkey. Press enter to proceed. Enter slope [%]. Select usingkey, edit number usingkey. Press enter to proceed. The device displays the new slope and asymmetry potential (at 25 °C). Press enter to proceed. pH value and Hold are displayed alternately. Press enter to proceed. Hold is deactivated after 20 sec. 62 Confirmation prompt. Converting slope [%] to slope [mV/pH] at 25 °C: % 78 80 82 84 86 88 90 92 94 96 98 100 102 mV/pH 46.2 47.4 48.5 49.7 50.9 52.1 53.3 54.5 55.6 56.8 58.0 59.2 60.4 Converting asymmetry potential to sensor zero point: VAS [mV] ZERO = 7 – S [mV / pH] ZERO VAS S Sensor zero Asymmetry potential Slope 63 Product Calibration Calibration by sampling During product calibration the sensor remains in the process. The measurement process is only interrupted briefly. Procedure: During sampling the currently measured value is stored in the device. The device immediately returns to measuring mode. The calibration mode indicator blinks and reminds you that calibration has not been terminated. The sample is measured in the lab or directly on the site using a portable meter. To ensure an exact calibration, the sample temperature must correspond to the measured process temperature. The measured sample value is then entered in the device. From the difference between the stored measured value and entered sample value, the device calculates the new asymmetry potential (one-point calibration). If the sample is invalid, you can take over the value stored during sampling. In that case, the old calibration values are stored. Afterwards, you can start a new product calibration. Display 64 Action Remark Product calibration step 1: Press cal key, enter code 1105 (Press key to select position, enter number using key, confirm by pressing enter) If an invalid passcode is entered, the device returns to measuring mode. Take sample and save value. Press enter to proceed. Now the sample can be measured in the lab. Display Action Measuring mode: From the blinking CAL mode indicator, you see that product calibration has not been terminated. Product calibration step 2: When the sample value has been determined, open the product calibration once more (cal, code 1105). Remark Until the sample value is determined and can be entered, the device is in measuring mode. Display (approx. 3 sec) Enter lab value. The new asymmetry potential is calculated. Display of slope and new asymNew calibration: metry potential (based on 25°C). Press cal. Exit calibration by pressing enter. The measured value is shown in the main display alternately with “Hold”; “enter” blinks. Exit by pressing enter. After end of calibration, the outputs remain in Hold mode for approx. 20 sec. 65 ORP Calibration ORP calibration mode is automatically preset when ORP measurement is configured. The potential of a redox sensor is calibrated using a redox (ORP) buffer solution. In the course of that, the difference between the measured potential and the potential of the calibration solution is determined according to the following equation. During measurement the Transmittter adds this difference to the measured potential. mVORP = mVmeas + ΔmV mVORP = displayed ORP mVmeas = direct sensor potential ΔmV = delta value, determined during calibration The sensor potential can also be related to another reference system – e.g. the standard hydrogen electrode. In that case the temperature-corrected potential (see table) of the reference electrode used must be entered during calibration. During measurement, this value is then added to the ORP measured. Please make sure that measurement and calibration temperature are the same since the temperature behavior of the reference electrode is not automatically taken into account. Temperature dependence of commonly used reference systems Temperature [°C] 0 10 20 25 30 40 50 60 70 80 66 Ag/AgCl/KCl 1 mol/l [ΔmV] Ag/AgCl/KCl 3 mol/l 249 244 240 236 233 227 221 214 207 200 224 217 211 207 203 196 188 180 172 163 Thalamid [ΔmV] [ΔmV] Mercury sulfate [ΔmV] -559 -564 -569 -571 -574 -580 -585 -592 -598 -605 672 664 655 651 647 639 631 623 613 603 Display Action Remark Select calibration (Press cal.) Enter passcode 1100. Select using , edit number using , press enter to proceed. If an invalid passcode is entered, the device returns to measuring mode. Remove the sensor and temperature probe, clean them and immerse them in the redox buffer. Display (approx. 3 sec) Device is in the Hold mode. Enter setpoint value for redox buffer (Secondary display: sensor potential displayed for approx. 6 sec) Select using , edit number using , press enter to proceed. After approx. 6 sec the secondary display shows the measured temperature. Display of sensor data (delta value) Press enter to proceed. Rinse sensor and temperature probe and reinstall them. “Zero” and “enter” icons are blinking, Sensoface is active. The measured ORP value [mV] is shown in the main display alternately with “Hold”, “enter” blinks. Exit by pressing enter. After end of calibration, the outputs remain in Hold mode for approx. 20 sec. 67 Temp Probe Adjustment Display Action Remark Select calibration (Press cal, enter passcode 1015.) Select usingkey, edit number usingkey. Press enter to proceed. Wrong settings change the measurement properties! If an invalid passcode is entered, the device returns to measuring mode. Measure the temperature of the process medium using an external thermometer. Device is in the Hold mode. Enter measured temperature value. Select usingkey, edit number usingkey. Press enter to proceed. Press enter to exit adjustment. HOLD will be deactivated after 20 sec. Default: Value of secondary display. Measurement Display Smiley only with Sensocheck activated 68 Remark In the measuring mode the main display shows the configured process variable (pH or ORP [mV]) and the secondary display the temperature. During calibration you can return to measuring mode by pressing the cal key, during configuration by pressing conf. (Waiting time for signal stabilization approx. 20 sec). Diagnostics Functions Display Remark Display of output current Press enter key while in measuring mode. The main display shows the measured value, the secondary display the output current. After 5 sec the device returns to measuring mode. Smiley only with Sensocheck activated Display of calibration data (Cal Info) Press cal while in measuring mode and enter code 0000. The slope is shown in the main display, the asymmetry potential in the secondary display. After 20 sec the device returns to measuring mode (immediate return at pressing enter). Display of sensor potential (Sensor monitor) Press conf while in measuring mode and enter code 2222. The (uncompensated) sensor potential is shown in the main display, the measuring temperature in the secondary display. Press enter to return to measurement. Smiley only with Sensocheck activated Display of last error message (Error info) Press conf while in measuring mode and enter code 0000. The last error message is displayed for approx. 20 sec. After that, the message will be deleted. (immediate return to measurement at pressing enter). 69 Diagnostics Functions These functions are used for testing the connected peripherals. Display Action / Remarks Specify output current Press cal key, enter code 5555. The output current indicated in the main display can be modified. Select usingkey, edit number usingkey. Press enter to proceed. The actually measured current is shown in the secondary display. The device is in Hold mode. Press conf to return to measurement (Hold remains active for another 20 sec). Cleaning Use only a moistened, lint-free cloth to remove dust, dirt and spots from the external surfaces of the device. Use a mild household cleaner if necessary. 70 LED Out 1 Operating state Time out Operating States Measuring Cal Info (cal) 0000 20 s Error Info (conf ) 0000 20 s Calibration (cal) 1100 Temp adjustment (cal) 1015 Product cal 1 (cal) 1105 Product cal 2 (cal) 1105 Configuration (conf ) 1200 20 min Sensor monitor (conf ) 2222 20 min Current source 1 (conf ) 5555 20 min Explanation: Active as configured (Last/Fix or Last/Off ) LED blinks during HOLD (configurable) 71 Error Messages (Error Codes) Error ERR 01 Display Problem Possible causes Measured pH sensor value • Sensor defective blinks • Too little electrolyte in sensor Red LED and Out 1 (22 mA) x • Sensor not connected • Break in sensor cable • Wrong sensor connected • Measured pH value < –2 or > 16 • Measured ORP value < –1999 mV or > 1999 mV ERR 02 Measured ORP sensor • Sensor defective value • Sensor not connected blinks x • Break in sensor cable • Wrong sensor connected • Sensor potential < –1500 mV • Sensor potential > 1500 mV ERR 98 “Conf“ blinks System error Configuration or calibration data defective; completely reconfigure and recalibrate the device. Memory error in device program x ERR 99 “FAIL” blinks Factory settings EEPROM or RAM defective This error message only occurs in the case of a total defect. The device must be repaired and recalibrated at the factory. x 72 Error Problem Icon (blinks) Possible causes Red LED and out 1 (22 mA) ERR 03 Temperature probe Open or short circuit Temperature range exceeded x ERR 11 Current output Current below 3.8 mA x ERR 12 Current output Current above 20.5 mA x ERR 13 Current output • Current span too small / too large x ERR 33 Sensocheck: Glass electrode Reference electrode x ERR 34 • Zero error, Sensoface active, see Pg 76 • Slope error, Sensoface active, see Pg 76 • Response time exceeded, Sensoface active, see Pg 76 • Calibration interval expired, Sensoface active, see Pg 76 73 Calibration Error Messages Icon blinks: Problem Possible causes Asymmetry potential out of range (±60 mV) • Sensor worn out • Buffer solutions unusable or contaminated • Buffer does not belong to configured buffer set • Temperature probe not immersed in buffer solution (for automatic temperature compensation) • Wrong buffer temperature set (for manual temperature specification) • Nominal sensor zero point ≠ pH 7 Sensor slope out of range (80...103 %) • Sensor worn out • Buffer solutions unusable or contaminated • Buffer does not belong to configured buffer set • Temperature probe not immersed in buffer solution (for automatic temperature compensation) • Wrong buffer temperature set (for manual temperature specification) • Sensor used has different nominal slope Problems during recognition of the buffer solution • Same or similar buffer solution was used for both calibration steps • Buffer solution used does not belong to buffer set currently configured in the device 74 Icon blinks: Problem Possible causes Problems during recognition of the buffer solution (continued) • During manual calibration the buffer solutions were not used in the specified order • Buffer solutions unusable or contaminated • Wrong buffer temperature set (for manual temperature specification) • Sensor defective • Sensor not connected • Sensor cable defective Calibration was canceled after approx. 2 minutes because the sensor drift was too large. • Sensor defective • Sensor dirty • No electrolyte in the sensor • Sensor cable insufficiently shielded or defective • Strong electric fields influence the measurement • Major temperature fluctuation of the buffer solution • No buffer solution or extremely diluted 75 Sensoface (Sensocheck must have been activated during configuration.) The little smiley in the display (Sensoface) alerts to sensor problems (defective sensor, defective cable, maintenance required). The permitted calibration ranges and the conditions for a friendly, neutral or sad Sensoface are summarized in the following table. Additional icons refer to the error cause. Sensocheck Continuously monitors the sensor and leads for short circuits or open circuits. Critical values make the Sensoface “sad” and the corresponding icon blinks: The Sensocheck message is also output as error message Err 33. The red LED is lit, the output current is set to 22 mA (when configured correspondingly). Sensocheck can be switched off during configuration (then Sensoface is also disabled). Exception: After a calibration, a smiley is always displayed for confirmation. Note The worsening of a Sensoface criterion leads to the devaluation of the Sensoface indicator (Smiley becomes “sadder”). An improvement of the Sensoface indicator can only take place after calibration or removal of the sensor defect. 76 Display Problem Asymmetry potential and slope Status Asymmetry potential (zero) and slope of the sensor are still okay. The sensor should be replaced soon. Asymmetry potential and slope of the sensor have reached values which no longer ensure proper calibration. Replace the sensor. Calibration timer Over 80 % of the calibration interval has already past. The calibration interval has been exceeded. Sensor defect Check the sensor and its connections (see also Error Messages Err 33 and Err 34, page 73). 77 78 Appendix Product Line and Accessories Devices Transmitter H200X pH Mounting Accessories Pipe-mount kit Panel-mount kit Protective hood Part No. 243085-01 243082 243083 243084 Power supply units If you require power supply units for power supply and hazardous-area separation, please contact Hamilton. 79 Specifications pH/mV input Input for pH or ORP sensors Measuring range Display range -1500 ... +1500 mV pH value -2.00 ... 16.00 ORP -1999 ... +1999 mV Glass electrode input1) Input resistance Input current Reference electrode input1) Input resistance Input current Measurement error 1,2,3) pH value mV value pH sensor standardization *) Operating modes BUF MAN DAT PRD 80 > 0.5 x 1012 ohms < 2 x 10-12 A > 1 x 1010 ohms < 1 x 10-10 A < 0.02 < 1 mV TC: 0.002 pH/K TC: 0.1 mV/K pH calibration Calibration with Calimatic automatic buffer recognition: Buffer sets -00- / -01- Knick / Mettler-Toledo 2.00/4.01/7.00/9.21 -02Merck/Riedel de Haen 2.00/4.00/7.00/9.00/12.00 -03Ciba (94) 2.06/4.00/7.00/10.00 -04NIST technical 1.68/4.00/7.00/10.01/12.46 -05NIST standard 1.679/4.006/6.865/9.180 -06HACH 4.00/7.00/10.18 -07WTW techn. buffers 2.00/4.01/7.00/10.00 -08Hamilton 4.01/7.00/10.01 Manual calibration with input of individual buffer values Data entry of premeasured sensors Product calibration Zero offset Max. calibration range ±200 mV Asymmetry potential: ±60 mV Slope: 80 ... 103 % (47.5 ... 61 mV/pH) ORP sensor standardization *) Max. calibration range ORP calibration -700 ... +700 ΔmV Cal timer 0000 ... 9999 h Sensocheck Automatic monitoring of glass and reference electrode (can be disabled) Sensoface Provides information on the sensor condition: Evaluation of zero/slope, response, calibration interval, Sensocheck Sensor monitor Direct display of measured values from sensor for validation (resistance/temperature) Temperature input *) Pt100/Pt1000/NTC 30 kΩ 2-wire connection, adjustable Pt 100/Pt 1000 - 20.0 ... + 200.0 °C - 4 ... + 392 °F) NTC 30 kΩ - 20.0 ... + 150.0 °C (- 4 ... + 302 °F) Measuring range Adjustment range Resolution Measurement error 1,2,3) 10 K 0.1 °C / 1 °F < 0.5 K (< 1K for Pt100; < 1K for NTC >100°C) Temperature compensation of process medium Linear –19.99 ... +19.99 %/K (reference temp 25 °C) 81 Specifications Loop current supply voltage Measured variable *) Characteristic Overrange *) Output filter *) Measurement error 1) Start/end of scale Permissible span Current source function 4 ... 20 mA floating 12 ... 30 V pH, ORP Linear 22 mA in the case of error messages Low-pass (PT1 filter), filter time constant 0 ... 120 s < 0.3 % current value + 0.05 mA Configurable within the measuring range for pH or mV pH 2.00 ... 18.00 / 200 ... 3000 mV 3.8 mA ... 22.00 mA HART communication Digital communication by FSK modulation of loop current, reading of device identification, measured values, status, and messages, reading and writing of parameters, start of product calibration, signaling of configuration changes according to FDA 21 CFR Part 11 Display Main display Secondary display Sensoface LC display, 7-segment with icons Character height 17 mm, unit symbols 10 mm Character height 10 mm, unit symbols 7 mm 3 status indicators (friendly, neutral, sad face) Mode indication Alarm indication 5 mode indicators 'meas', 'cal', 'alarm', 'cleaning', 'config' 18 further icons for configuration and messages Red LED with alarm and HOLD, user-defined Keypad 5 keys: [cal] [conf ] [] [] [enter] 82 Service functions Current source Device self-test Display test Last Error Sensor monitor Pass codes Loop current specifiable 3.8 ... 22.00 mA Automatic memory testing (RAM, ROM, EEPROM) Display of all segments Display of last error occurred Display of direct, uncorrected sensor signal (resistance / temperature) Modifiable according to FDA 21 CFR Part 11 “Electronic Signatures” Data retention Parameters and calibration data > 10 years (EEPROM) EMC Emitted interference: Immunity to interference: EN 61326 Class B (residential area) Industry Explosion protection H200X: II 2(1)G Ex ib [ia Ga] IIC T6 Gb Nominal operating conditions Ambient temperature Transport/Storage temp Relative humidity Supply voltage –20 ... +55 °C –20 ... +70 °C 10 ... 80% not condensing 12 ... 30 V 83 Specifications Enclosure Color Mounting Dimensions Ingress protection Cable glands Weight Molded enclosure made of PBT (polybutylene terephtalate) Bluish gray RAL 7031 • Wall mounting • Pipe mounting: Ø 40 ... 60 mm, 30 ... 45 mm • Panel mounting, cutout to DIN 43 700, sealed against panel H 144 mm, W 144 mm, D 105 mm IP 65 / NEMA 4X 3 knockouts for cable glands M20x1.5, 2 knockouts for 1/2“ NPT or rigid metallic conduit Approx. 1 kg *) user-defined 1) to IEC 746 Part 1, at nominal operating conditions 2) ± 1 count 3) plus sensor error 84 Buffer Tables -00-01°C 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 Knick technical buffers (correspond to Mettler-Toledo technical buffers) pH 2.03 2.02 2.01 2.00 2.00 2.00 1.99 1.99 1.98 1.98 1.98 1.98 1.98 1.99 1.99 2.00 2.00 2.00 2.00 2.00 4.01 4.01 4.00 4.00 4.00 4.01 4.01 4.02 4.03 4.04 4.06 4.08 4.10 4.13 4.16 4.19 4.22 4.26 4.30 4.35 7.12 7.09 7.06 7.04 7.02 7.00 6.99 6.98 6.97 6.97 6.97 6.98 6.98 6.99 7.00 7.02 7.04 7.06 7.09 7.12 9.52 9.45 9.38 9.32 9.26 9.21 9.16 9.11 9.06 9.03 8.99 8.96 8.93 8.90 8.88 8.85 8.83 8.81 8.79 8.77 85 Buffer Tables -02°C 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 86 Calimat (Merck Titrisols, Riedel Fixanals) pH 2.01 2.01 2.01 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.01 2.01 2.01 2.01 2.01 2.01 4.05 4.04 4.02 4.01 4.00 4.01 4.01 4.01 4.01 4.01 4.00 4.00 4.00 4.00 4.00 4.00 4.00 4.00 4.00 4.00 7.13 7.07 7.05 7.02 7.00 6.98 6.98 6.96 6.95 6.95 6.95 6.95 6.96 6.96 6.96 6.96 6.97 6.98 7.00 7.02 9.24 9.16 9.11 9.05 9.00 8.95 8.91 8.88 8.85 8.82 8.79 8.76 8.73 8.72 8.70 8.68 8.66 8.65 8.64 8.64 12.58 12.41 12.26 12.10 12.00 11.88 11.72 11.67 11.54 11.44 11.33 11.19 11.04 10.97 10.90 10.80 10.70 10.59 10.48 10.37 -03°C 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 Ciba (94) buffers Nominal values: 2.06, 4.00, 7.00, 10.00 pH 2.04 2.09 2.07 2.08 2.09 2.08 2.06 2.06 2.07 2.06 2.06 2.05 2.08 2.07 2.07 2.04 2.02 2.03 2.04 2.05 * * * * 4.00 4.02 4.00 4.00 4.01 4.02 4.00 4.01 4.02 4.03 4.04 4.05 4.10 4.10 4.11 4.13 4.15 4.17 4.20 4.22 * * * * 7.10 7.08 7.05 7.02 6.98 6.98 6.96 6.95 6.94 6.93 6.93 6.91 6.93 6.92 6.92 6.92 6.93 6.95 6.97 6.99 * * * * 10.30 10.21 10.14 10.06 9.99 9.95 9.89 9.85 9.81 9.77 9.73 9.68 9.66 9.61 9.57 9.54 9.52 9.47 9.43 9.38 * * * * * extrapolated 87 Buffer Tables -04°C 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 Technical buffers to NIST pH 1.67 1.67 1.67 1.67 1.675 1.68 1.68 1.69 1.69 1.70 1.705 1.715 1.72 1.73 1.74 1.75 1.765 1.78 1.79 1.805 * extrapolated 88 4.00 4.00 4.00 4.00 4.00 4.005 4.015 4.025 4.03 4.045 4.06 4.075 4.085 4.10 4.13 4.14 4.16 4.18 4.21 4.23 7.115 7.085 7.06 7.04 7.015 7.00 6.985 6.98 6.975 6.975 6.97 6.97 6.97 6.98 6.99 7.01 7.03 7.05 7.08 7.11 10.32 10.25 10.18 10.12 10.06 10.01 9.97 9.93 9.89 9.86 9.83 9.83 9.83 9.83 9.83 9.83 9.83 9.83 9.83 9.83 * * * * * * * * * 13.42 13.21 13.01 12.80 12.64 12.46 12.30 12.13 11.99 11.84 11.71 11.57 11.45 11.45 11.45 11.45 11.45 11.45 11.45 11.45 * * * * * * * -05°C 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 NIST standard buffers pH 1.666 1.668 1.670 1.672 1.675 1.679 1.683 1.688 1.694 1.700 1.707 1.715 1.723 1.733 1.743 1.755 1.766 1.779 1.792 1.806 4.010 4.004 4.000 3.999 4.001 4.006 4.012 4.021 4.031 4.043 4.057 4.071 4.087 4.109 4.126 4.145 4.164 4.185 4.205 4.227 6.984 6.951 6.923 6.900 6.881 6.865 6.853 6.844 6.838 6.834 6.833 6.834 6.836 6.841 6.845 6.852 6.859 6.868 6.877 6.886 9.464 9.395 9.332 9.276 9.225 9.180 9.139 9.102 9.068 9.038 9.011 8.985 8.962 8.942 8.921 8.903 8.885 8.868 8.850 8.833 89 Buffer Tables -06°C 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 HACH buffers Nominal values: 4.00, 7.00, 10.18 pH 4.00 4.00 4.00 4.00 4.00 4.01 4.01 4.02 4.03 4.05 4.06 4.07 4.09 4.09 4.09 4.09 4.09 4.09 4.09 4.09 * * * * * * * 7.14 7.10 7.04 7.04 7.02 7.00 6.99 6.98 6.98 6.98 6.98 6.98 6.99 6.99 6.99 6.99 6.99 6.99 6.99 6.99 * * * * * * * 10.30 10.23 10.11 10.11 10.05 10.00 9.96 9.92 9.88 9.85 9.82 9.79 9.76 9.76 9.76 9.76 9.76 9.76 9.76 9.76 * * * * * * * * Values complemented Buffer values up to 60 °C as specified by Bergmann & Beving Process AB. 90 -07°C 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 WTW buffers pH 2.03 2.02 2.01 2.00 2.00 2.00 1.99 1.99 1.98 1.98 1.98 1.98 1.98 1.99 2.00 2.00 2.00 2.00 2.00 2.00 4.01 4.01 4.00 4.00 4.00 4.01 4.01 4.02 4.03 4.04 4.06 4.08 4.10 4.13 4.16 4.19 4.22 4.26 4.30 4.35 7.12 7.09 7.06 7.04 7.02 7.00 6.99 6.98 6.97 6.97 6.97 6.98 6.98 6.99 7.00 7.02 7.04 7.06 7.09 7.12 10.65 10.52 10.39 10.26 10.13 10.00 9.87 9.74 9.61 9.48 9.35 91 Buffer Tables -08°C 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 Hamilton Duracal buffers pH 4.01 * 4.01 4.00 4.00 4.00 4.01 4.01 4.02 4.03 4.04 4.06 4.08 * 4.10 * 4.13 * 4.16 * 4.19 * 4.22 * 4.26 ** * 4.30 4.35 * extrapolated 92 7.12 7.09 7.06 7.04 7.02 7.00 6.99 6.98 6.97 6.97 6.97 6.98 6.98 6.99 7.00 7.02 7.04 7.06 7.09 7.09 * * * * * * * ** * 10.19 * 10.19 10.15 10.11 10.06 10.01 9.97 9.92 9.86 9.83 9.79 9.77 * 9.75 * 9.74 * 9.73 * 9.73 * 9.73 * 9.74 ** * 9.75 9.75 93 Glossary Asymmetry potential The voltage which a pH sensor provides at a pH of 7. The asymmetry potential is different for each sensor and changes with age and wear. Buffer set Contains selected buffer solutions which can be used for automatic calibration with the Calimatic. The buffer set must be selected prior to the first calibration. Buffer solution Solution with an exactly defined pH value for calibrating a pH meter. Calibration Adjustment of the pH meter to the current sensor characteristics. The asymmetry potential and slope are adjusted. Either a one- or two-point calibration can be carried out. With one-point calibration only the asymmetry potential (zero point) is adjusted. Calimatic Automatic buffer recognition. Before the first calibration, the buffer set used must be activated once. The patented Calimatic then automatically recognizes the buffer solutions used during calibration. Combination sensor Combination of glass and reference electrode in one body. GainCheck Device self-test which runs automatically in the background at fixed intervals. The memory and measured-value transmission are checked. You can also start the GainCheck manually. Then a display test is also conducted and the software version displayed. 94 One-point calibration Calibration with which only the asymmetry potential (zero point) is taken into account. The previous slope value is retained. Only one buffer solution is required for a one-point calibration. Passcode Preset four-digit number to select certain functions. pH sensor A pH sensor consists of a glass and a reference electrode. If they are combined in one body, they are referred to as combination sensor. Response time Time from the start of a calibration step to the stabilization of the sensor potential. Sensocheck Sensocheck continuously monitors the glass and reference electrodes. The resulting information is indicated by the Sensoface smileys. Sensocheck can be switched off. Sensoface Provides information on the sensor condition. The zero point, slope, and response time are evaluated. In addition, the Sensocheck information is indicated. Sensor slope Is indicated in % of the theoretical slope (59.2 mV/ pH at 25 °C). The sensor slope is different for each sensor and changes with age and wear. Sensor zero point See asymmetry potential 95 Glossary Two-point calibration Calibration with which the sensor asymmetry potential (zero point) and slope are determined. Two buffer solutions are required for two-point calibration. Zero point See asymmetry potential 96 Index 22.0 mA for error 45 243082 18 243083 19 243084 18 A Accessories 79 Alarm 33 Alarm delay 53 Alarm LED 28 Alarm settings 52 Assembly 16 Assigning measured values, current start/end 41 Asymmetry potential 63 Automatic calibration 58 B BUF 51 Buffer tables 85 C Cable, color coding 23 Cable glands 16 Calibration 56 Calibration by sampling 64 Calibration error messages 74 Calibration mode 50 Calimatic automatic calibration 58 Cal timer interval 50 Cleaning 70 Cleaning in hazardous areas 8 Color coding in the cable 23 Conduit 1/2“ 17 Configuration 34 Configuration steps, overview 36 Conformity with FDA 21 CFR Part 11 11 Connecting the VP cable 22 Connection examples 24 97 Index Connection to supply units 20 Converting asymmetry potential to sensor zero point 63 Current end 40 Current output, configuring 38 Current start 40 D Damage of device 7 DAT 51 Data entry of premeasured sensors 62 Diagnostics functions 69 Display 29 Disposal 2 Drilling plan (wall mounting) 17 E EC Declaration of Conformity 10 EC-Type-Examination Certificate 12 Electrode type selection 38 Error and HOLD, output current 44 Error codes 72 Error messages 72 F Filler plug 16 Filter time constant 43 G Glass electrode 39 Glossary 94 H HAMILTON VP double coaxial cable‚ VP 8.0 DC’ 23 HAMILTON VP single coaxial cable‚ VP 6.0’ 23 HART communication 33 Hinge pin 16 HOLD mode 32 HOLD mode, LED 53 98 I Individual settings (for copy) 37 Installation and connection 20 Intended use 9 J Jumper 16 K Keypad 30 L LED in Hold mode 53 Loop current 33 M MAN 51 Manual calibration 60 Measurement 68 Measuring function 38 Mounting accessories 79 Mounting plan 17 N NTC30 47 O Operating states 71 ORP calibration 66 ORP measurement 39 Wiring example 27 Output current during Error and HOLD 44 Output current range 40 Output filter 43 Overview of the H200X pH 15 99 Index P Package contents 16 Panel mounting 18 Panel-mount kit, P/N 243083 19 Parameter set for copy 37 Passcode editor 54 Passcodes according to FDA 21 CFR Part 11 54 Pipe mounting 18 Process variable pH, ORP 39 Product calibration 64 Product line and accessories 79 Protective hood for wall and pipe mounting, P/N 243084 18 Pt100 47 R Redox calibration 66 Redox measurement 39 Wiring example 27 Registered trademarks 9 Return of products 2 S Safety functions 31 Safety information 7 Sealing in case of wall mounting 16 Sealing insert 16 Sensocheck 76 Sensoface 76 Sensor connection, examples 24 Slope [%] to [mV/pH] 63 Specifications 80 Supply units, connection 8 100 T Table of contents 3 TC process medium 48 Technical data 80 Technical terms 94 Temperature compensation 46 Temperature dependence of commonly used reference systems 66 Temperature detection during meas/cal 48 Temperature probe adjustment 68 Temperature unit and probe 46 Terminal assignments 20 Time averaging filter 42 U User interface 28 V Visible damage 7 VP cable 22 W Warranty 2 Wiring examples 24 ORP measurement 27 pH measurement 24 VP connection 22 101 102 TA-194.100-HAME05 20121108 Software version: 4.x