Preview only show first 10 pages with watermark. For full document please download

User Guidelines And Sop - Wilfrid Laurier University

   EMBED


Share

Transcript

Date of Issue: August 2009 Revision: 9 Date of Revision: Oct 10, 2012 Olympus FV1000 Standard Operating Procedure ii TABLE OF CONTENTS DISCLAIMER ................................................................................ iv ACKNOWLEDGEMENTS ..................................................................... v 1. INTRODUCTION........................................................................ 1 1.1 Purpose of the Standard Operating Procedure ............................. 1 1.2 Emergency Contact Information .............................................. 1 1.3 User Fees and Instrument Booking ........................................... 2 1.4 Basic Theoretical Background ................................................. 2 1.4.1 Sample Preparation Suggestions......................................... 3 1.5 Instrumentation ................................................................. 5 1.5.1 Overview .................................................................... 5 1.5.2 Lasers and Filters .......................................................... 5 1.5.3 Objectives .................................................................. 5 1.5.4 Scan Head................................................................... 6 2. POTENTIAL HAZARDS ................................................................. 7 3. PERSONAL PROTECTIVE EQUIPMENT ............................................... 8 4. ACCIDENT PROCEDURES ............................................................. 8 5. WASTE DISPOSAL PROCEDURES ..................................................... 8 6. PROTOCOL ............................................................................. 9 6.1 Start-up........................................................................... 9 6.1.1 Viewing sample with transmitted light and Kolher illumination .. 10 6.1.2 Epifluorescence Imaging ................................................ 11 6.2 Preparation for Imaging ........................................................ 1 6.3 XY Image Acquisition ........................................................... 3 6.4 Z-Series or 3-D Stack Image Acquisition ..................................... 4 6.5 T-Series Image Acquisition..................................................... 4 6.6 Image Analysis ................................................................... 5 6.7 Saving and Viewing Final Images ............................................. 9 6.8 Cleaning Objectives .......................................................... 10 6.9 Shutdown ....................................................................... 10 7. TROUBLESHOOTING ................................................................ 10 8. PREVENTATIVE MAINTENANCE .................................................... 11 8.1 Monthly ......................................................................... 11 8.2 Annually or as Required ...................................................... 11 8.2.1 Replace the Mercury Burner ........................................... 11 Olympus FV1000 Standard Operating Procedure 9. iii QUICK REFERENCE GUIDE .......................................................... 12 REFERENCES .............................................................................. 13 APPENDIX 1: FLUOROCHROME PEAK EXCITATION AND EMISSION WAVELENGTHS 14 APPENDIX 2: USER LOG .................................................................. 20 APPENDIX 3: PREVENTATIVE MAINTENANCE LOG .................................... 22 Olympus FV1000 Standard Operating Procedure iv DISCLAIMER The materials contained in this document have been compiled from sources believed to be reliable and to represent the best opinions on the subject. This document is intended to serve only as a starting point for good practices and does not purport to specify minimal legal standards. No warranty, guarantee, or representation is made by Laurier as to the accuracy or sufficiency of information contained herein, and Laurier assumes no responsibility in connection therewith. Olympus FV1000 Standard Operating Procedure v ACKNOWLEDGEMENTS The following individuals of Laurier contributed to the writing, editing, and production of this manual: Gena Braun (Instrumentation Technician); Dr. Diano Marrone (Psychology); Stephanie Kibbee (Environmental/Occupational Health and Safety Office). This manual was prepared for Laurier. Any corrections, additions or comments should be brought to the attention of the Instrumentation Technician at 519-884-0710 ext. 2361. Olympus FV1000 Standard Operating Procedure 1 1. INTRODUCTION 1.1 Purpose of the Standard Operating Procedure This standard operating procedure (SOP) is NOT a substitute for training and/or reading the appropriate manuals before use. All principle investigators and supervisors must document that training has been received by students and staff who will be using the Olympus FV1000 confocal laser scanning microscope. A list of authorized users will be maintained by the Instrumentation Technician. This SOP is intended to promote consistent and safe use of the Olympus FV1000. This SOP covers the potential hazards, personal protection requirements, spill and accident procedures, waste disposal considerations, and instrument operation for the Olympus FV1000. 1.2 Emergency Contact Information The Olympus FV1000 contains several embedded class 3B lasers, and is therefore regulated under the Laurier Laser Safety Program. The Laser Safety Manual is available at the EOHS website1. The manual provides information on:  Responsibilities of the laser operator, and Laser Safety Officer  Laser registration requirements  Training requirements  Sign and labeling requirements  Eyewear requirements For more information on the specific lasers and hazards relevant to this instrument, see sections 1.5 (Instrumentation) and 2.0 (Potential Hazards) in this SOP. Emergency contacts for situations involving the Olympus FV1000 lasers are as follows: Principle Investigator: Technicians: Laser Safety Office: Dr. Diano Marrone ext. 2990 Gena Braun/Jiangxiao Sun ext 2361 or 519-500-4548 Sarah Lamb ext 3108 Immediately notify the Laser Safety Officer at extension 3108 for all laserrelated injuries and near-misses. If the contacts listed above cannot be reached, or in the case of a more general emergency, call 9-911 from any campus phone and/or Special Constable Service at ext. 3333. 1 http://www.wlu.ca/documents/42977/Laser_Safety_Manual_2010.pdf Olympus FV1000 Standard Operating Procedure 2 1.3 User Fees and Instrument Booking To recover the operating costs of the FV1000, the following user fees have been established: Internal users: Hourly: $ 20 Annually: $ 1500 (Internal users do not pay a training fee) External users: Training Fee: $ 55/hour (2-3 hours) Assisted Use: $ 40/hour Unassisted Use: $ 30/hour Following training, users may book the confocal as follows: 1. Go to instrument booking website http://www.supersaas.com/schedule/login/WLU_Instruments/ 2. Use your user name and password to log in. Log in information is provided by the Research Instrumentation Technician following training. 3. Click on the day that you wish to book. Enter the time slot in “When” and “To” boxes. 4. Make sure you select “Confocal Microscope” in the “Instruments” drop down box. 5. If you want to book the same time slot in other days, select from “Repeat” drop down box for “daily, weekly” etc. 6. Click on “Create reservation”. 7. Please make sure to cancel your reservation if you will not be using the microscope, so that others can book the instrument if they need. 1.4 Basic Theoretical Background The first commercially available laser scanning confocal microscope (LSCM) appeared on the market in 1987, and has since become a very valuable tool in biological research. LSCM is based on the same fundamentals as standard widefield microscopy and epifluorescence, with a few distinct advantages. The critical advantage arises from the use of simultaneously in focus (confocal) pinholes to direct both excitation and emission light. As a result, only a thin section of the specimen is excited and emits fluorescence that can pass through the pinhole and be detected by the photomultiplier tube (PMT) (see Figure 1-1). Exclusion of fluorescence from sample planes that are not in focus greatly reduces the background and increases the sensitivity of LSCM over other techniques. Different planes within the specimen can also be sequentially brought into focus and used to generate a 3-D image. Confocal microscopy is not recommended for all specimen types because the high laser intensity causes rapid photobleaching when compared other widefield techniques. CLSM is highly useful for examining thick specimens, Olympus FV1000 Standard Operating Procedure producing 3-D reconstructions to assess cell structure, studying the spatial distribution of labeling, imaging of live cells or other applications that require simultaneous and rapid multi-channel imaging, and studying multi-labeled specimens. Figure 1-1: Basic illustration of confocal imaging; only the signal that is confocal with the emission pinhole is detected, reducing out-of focus background signal and improving the image significantly. 1.4.1 Sample Preparation Suggestions A number of factors can affect the quality of a confocal image, including sample preparation. Special attention should be paid to the selection of the appropriate dye(s) and mounting medium. When selecting a dye or set of dyes, keep the following in mind: - Each dye should have an absorption maximum close to one of the available laser lines (405, 458, 488, 515, 543, 643). - For multi-colour labeling, each dye should have relatively narrow and well separated absorption and emission maxima to avoid bleed-through (excitation light reaching the detector). - The dye should be resistant to photobleaching. - Each dye should have high quantum efficiency, meaning a large number of photons are produced with minimal excitation power. There are a few families of dyes with specific benefits or drawbacks, so research your selected dye carefully. For example, fluorescein isothiocyanate (FITC), a very popular and common dye, is easily influenced by environment 3 Olympus FV1000 Standard Operating Procedure 4 (i.e. pH) and has a relatively broad emission spectrum, so it is not ideal for studies that require dual or triple labeling. Appendix 1 lists several dyes and their excitation/emission wavelengths. The appropriate mounting medium depends on the nature of the specimen (fixed or live), and the magnification to be used. For fixed slides, the medium should contain anti-fade reagents to minimize photobleaching. Conversely, anti-fade reagents can be toxic to living organisms and are typically not recommended for live specimens. The medium and anti-fade reagent must also be compatible with the selected dye; some anti-fade compounds can cleave the fluorescent molecules of certain dyes and cause the signal to degrade with in a few days or less. If you plan to image using the 60x or the 100x oil immersion lenses, your mounting medium must have a refractive index that is similar to oil. Consider using 50-80% glycerol or 2,2-thiodiethanol in your mounting medium to improve the image (see Staudt et al., 2007). Again, confirm that your dyes are compatible with all the components of your mounting medium before preparing the specimen. All mounting media must be completely dry before the specimen is viewed on the FV1000. All sample preparation should be done in a separate lab, NOT in the room with the FV1000. Additional information on mounting media and anti-fade reagents has been summarized by Tonny Collins (Wright Cell Imaging Facility, Toronto) at: http://www.uhnresearch.ca/facilities/wcif/PDF/Mountants.pdf Olympus FV1000 Standard Operating Procedure 5 1.5 Instrumentation 1.5.1 Overview The FV1000 inverted CLSM can be used to examine both fixed and live specimens. Light is produced by a several lasers, and passes through the appropriate filters and mirrors and a pinhole before being scanned across the image in a raster patter. High speed imaging (up to 16 frames/sec for a 256 x 256 image) and rapid spectral scanning (100 nm/sec) are carried out using galvo scanning mirrors. Wavelengths can be resolved to 2 nm on this system, and all filters are ion-sputter coated to provide improved transmission efficiency and allow imaging at lower laser intensity. Fluorescence produced by the specimen passes back through mirrors and filters and a pinhole to be detected by the appropriate photomultiplier tube (PMT). The available lasers and optics are described in more detail below. 1.5.2 Lasers and Filters The FV1000 is equipped with a multi-line argon laser, a green helium-neon (HeNe) laser, two diode lasers, and a mercury lamp. The argon laser produces excitation light at 458, 488, and 515 nm; the green He-Ne at 543 nm; the blue diode at 405 nm; and the red diode at 635 nm. The mercury lamp must be on for at least 30 minutes before it can be switched off; after turning it of, it must cool down for ~15 minutes before being turned back on. The lamp and lasers should not be switched on and off frequently as this shortens the lifetime. The light source for transmitted light is a halogen bulb, which is connected to the microscope via a fiber optic cable. Brightfield and differential interference images using transmitted light can be viewed through the oculars or collected via the computer at the same time as laser scanning. Epifluorescence illumination is conducted using the mercury lamp, and three filter cubes for DAPI, FITC, and TRITC are available. 1.5.3 Objectives The terms that follow the objective magnification indicate the corrections applied to the lens, as follows: - “Plan” indicates a flat field correction, which corrects for field curvature produced by the lens. - “S-Apo” indicates apochromatic correction which is the highest degree of correction for spherical and chromatic aberration. - “Fluor” indicates fluorite aberration correction, which is multi-colour chromatic and spherical aberration. - “NA” is the numerical aperture of the lens, which determines resolution and depth of field. - “WD” is the working distance. Olympus FV1000 Standard Operating Procedure 6 There are 3 air objectives, and two oil-immersion objectives available on the FV1000 confocal microscope: - 10x (Plan S-Apo, NA 0.4, WD 3.1 mm) - 20x (Plan S-Apo, NA 0.75, WD 0.65 mm) (DIC) - 40x (Plan Fluor, NA 0.6, WD 2.7-4.0 mm, correction collar) (DIC) - 60x oil (Plan Apo, NA 1.42, WD 0.15 mm) (DIC) - 100x oil (Plan S-Apo, NA 1.4, WD 0.12 mm) (DIC) The correction collar on the 40x objective can be adjusted to account for variation in coverslip thickness. All of the other objectives must be used with a 0.17 mm (#1) coverslip. Extra care must be taken when using oil objectives: - An oil immersion objective must be cleaned before adding any more oil to view a new slide. If it is not cleaned, excess oil can run down the side of the objective and into the microscope. - DO NOT use any of the air objectives (10x, 20x, or 40x) immediately after using an oil immersion (60x or 100x) objective. The 20x and 40x objectives in particular have a very small working distance, and can easily come into contact with any oil left on the slide. The slide must be cleaned before switching to the air objectives. 1.5.4 Scan Head The microscope scan head contains the optics required to accept, filter, and detect the laser and fluorescence signals, and to scan in a raster or bidirectional pattern across the specimen. The FV1000 scan head contains 3 internal photomultiplier tubes (PMTs) for fluorescence detection, and one external PMT to detect transmitted light. As a result, up to three fluorescent signals and a transmitted light signal can be collected simultaneously. The signal for each channel passes though a single pinhole, which can be adjusted in size from 50 to 800 m, in 5 m increments. Scanning is done using two galvo-mirrors, and can cover a standard square, a slice, a line, or a user selected region of interest. Multi-dimensional scans can be completed for Zseries (3-D) or T-Series (time series) experiments; the minimum increment for a Z-scan is 0.01 m. Olympus FV1000 Standard Operating Procedure 7 2. POTENTIAL HAZARDS The FV1000 is a class 3B laser system and uses a variety of high power lasers which present laser radiation and heat hazards. The laser radiation from this instrument can cause serious eye damage if direct or reflected laser light enters you eyes. Follow all warning labels on the equipment. Never look directly at or touch a slide while scanning. Never switch objectives while scanning. The FV1000 operates under high voltage. Never tamper with any of the connections or electrical cables, and contact the Instrumentation Technician immediately if any of the cables appear damaged. Do not expose your hand or finger to the laser beam output from the objective mount hole, objective tip or condenser lens, or your skin may be damaged. Never attempt to output the laser beam outside the system by inserting a mirror or a similar object in the light path. The laser beam may enter your eyes, and this is extremely hazardous. Make sure that the slide is laying flat on the stage. If the specimen inclines, the laser beam may reflect into your eyes, which is extremely hazardous. Do not bend or pull any of the laser fiber cables. If the laser fiber cable is damaged, the laser light may leak outside it and create a hazardous situation. Should such an event occur, immediately turn off the laser power and contact the Instrumentation Technician. The air outlet of the laser cooling fan blows out warm air and the mercury lamp housing gets hot. Do not place a flammable or non-heat-resistant object near these items. The cleaning fluids (mixture of alcohol or ether) used to clean the optics are highly flammable. Take special care in handling. The FV1000 rests on a floating table. The table dampens any vibrations to the system, and is connected to a cylinder of compressed air. Do not adjust the regulator on the compressed air cylinder or attempt to change an empty cylinder; the cylinder is maintained and replaced by the Instrumentation Technician. Olympus FV1000 Standard Operating Procedure 8 3. PERSONAL PROTECTIVE EQUIPMENT Gloves and a lab coat are encouraged for any type of lab work. Safety glasses are not required for the FV1000 because the laser path is enclosed within the live cell chamber, but never look directly at the laser light; always look through the protective pane above the microscope oculars. See the WLU Laboratory Health and Safety Manual for additional information on personal protective equipment: http://www.wlu.ca/documents/23120/Laboratory_Health_%26_ Safety_Manual__Feb_2007_Final.pdf. 4. ACCIDENT PROCEDURES All incidents must be reported to the Instrumentation Technician and if applicable, a student’s supervisor. All accidents, incidents and near misses must be reported to the Environmental/Occupational Health and Safety (EOHS) Office via the WLU Employee Accident/Incident/Occupational Disease Report form (www.wlu.ca/eohs/forms). To meet regulatory requirements, these forms must be submitted to EOHS within 24 hours of occurrence, with the exception of critical injuries, which must be reported immediately to the EOHS Office by telephone. Critical injuries include any of the following; place life in jeopardy, produce unconsciousness, result in substantial loss of blood, involve fracture of a leg or arm but not a finger or toe, involve amputation of a leg, arm, hand or foot, but not a finger or toe, consist of burns to a major portion of the body, or cause the loss of sight in an eye. Additional details regarding incident reporting can be found in the WLU Accident Incident Procedure (www.wlu.ca/eohs). The WLU Laboratory Health and Safety Manual provides detailed instructions for dealing with major and minor spills. Before using ANY hazardous materials, make sure you understand the proper clean-up procedure. All sample preparation and disposal should be done in a separate lab, NOT in the room with the FV1000. 5. WASTE DISPOSAL PROCEDURES If any hazardous chemicals are used for sample analysis or preparation, they must be disposed of properly, as outlined in the WLU Laboratory Health and Safety Manual. Olympus FV1000 Standard Operating Procedure 9 6. PROTOCOL Anyone using the confocal microscope must receive hands on training. This document is a summary of the procedure and is only intended to help you remember the various steps. 6.1 Start-up You only need to turn on the components that you plan to use. Turning lasers and the mercury lamp on and off and running them affects their lifespan, so if you are sure you won’t be using a given lamp or laser, don’t turn it on. You must turn on all the lasers you might potentially need at the start, because you should not turn lasers on once the computer is on. The 405+635 diode laser power supply also runs the laser combiner, so this component must always be turned on, regardless of the wavelengths you are interested in viewing. If you just need the computer to look at data or transfer it to a CD/DVD, turn only the computer and monitors on. If you will be using the microscope to collect z-stacks or time series images, it is recommended to turn the system on, including the temperature controller, up to two hours before imaging. This will allow all of the components to reach thermal equilibrium and reduces z-drift during imaging. ALL mounting media or other substances used to secure coverslips (i.e. nail polish) must be COMPLETELY DRY before the slide is viewed on the confocal. Mounting media may otherwise leak onto the objective and is very difficult to remove. 1. TURN ON THE FAN. 2. Make sure the window shade is down all the way. 3. Sign-in in the user log book and fill in mercury lamp hours at start, and lasers to be used. 4. Turn on the microscope components according to the numbers on labels: 1. U-RFL-T: Mercury burner power supply – required for epifluorescence observation. 2. Power bar. This will turn on the following: a. FV10-MCPSU: Power supply for the diode lasers (405 and 643 nm). b. Prior ProScan II: Stage controller. c. IX2-UCB: Microscope controller power supply (on the windowsill). d. Scan head power supply. 3. Only turn on if required for your dye: Melles Griot argon (458, 488, and 515 nm ) laser power supply. Turn on the power switch and then the key. 5. Turn on the computer and monitors. 6. Log on to the computer: a. Login: Administrator b. Password: fluoview. Olympus FV1000 Standard Operating Procedure 10 7. Double click on the Olympus FV10-ASW 1.7 software icon and log on using your username and password. 8. Wait for the software and microscope to initialize. 9. Click on Device, then Microscope Controller. 6.1.1 Viewing sample with transmitted light and Kolher illumination Before viewing specimen with transmitted light (brightfield) the microscope condenser should be properly aligned to provide even and bright illumination (commonly called Kolher). If you do not need to view or collect brightfield or differential interference contract (DIC) images, Kolher illumination is not necessary, and you can skip to section 6.1.2. Refer to figures 6-1 and 6-2 for hardware and software diagrams. 1. In the Acquisition Setting window, select the 10x objective (drop down list, #6 in Figure 6-2). 2. Manually move the objective down by turning the knob clockwise. 3. Mount your slide on the microscope, coverslip down. 4. Make sure the polarizer is in the optical path (above the stage – manual slider, #4 in Figure 6-1)). 5. Align the DIC/Wollaston prism in the light path as well (below the stage) by sliding it in until in comes to a stop (#7 in Figure 6-1). 6. Select “DICT” from the Microscope Controller window, and the DIC cube will be rotated into place automatically (#15 in Figure 6-2). 7. Use the software to select Transmitted Light mode (turn on Trans Lamp, upper button by #8 in Figure 6-2). 8. Focus on the specimen using the microscope oculars. a. Adjust the brightness using the buttons on the front of the microscope. b. The focus mechanism can be set to fine using either the F/C button below the right focus knob, or in the Micoscope Controller window. i. Close your left eye and focus on the specimen using the fine focus knob. ii. Close your right eye and focus on the specimen using the diopter ring on the left ocular. iii. Open both eyes and confirm that the focus is comfortable. c. Turn the shear knob on the DIC prism (just under the stage, #7 in Figure 6-1) to adjust the contrast. d. Adjust the DIC prism focal plane by pulling out or pushing in the small L-shaped knob just beside the shear knob. (Push it in or pull it out until the best image is observed). 9. Completely open the condenser aperture diaphram (just above the stage, #6 in Figure 6-1) and close the field diaphragm (top diaphram – outside of the live cell case, #1 in Figure 6-1) enough so that you can see the diaphragm edges in the eyepieces. Olympus FV1000 Standard Operating Procedure 11 10. Focus the condensor knob (#5 in Fugre 6-1) so that the edges of the diaphragm AND the speciman appear very sharp (do not use the fine/course focus for the objectives). 11. If nessicary, center the condensor using the centering screws (it helps to open the field diaphragm so that the edges of the diaphragm almost fill the field of view, #3 in Figure 6-1). 12. When the condensor is focused and centered, open the field diaphragm so that it completely fills the eye pieces (#1 in Figure 6-1). 13. If you need DIC images only, proceed to 6.2, otherwise proceed to section 6.1.2 for epifluorescence imaging. 6.1.2 Epifluorescence Imaging Examining your specimen using epifluorescence before collecting a confocal image allows focusing and objective/magnification selection without the photobleaching risks associated with high intensity laser light. Use epifluoresence to find the region of interest and focus at the desired magnification as follows before proceeding to laser scanning. Refer to figures 6-1 and 6-2 for hardware and software diagrams. 1. Close the manual shutter on the mercury lamp. 2. If you do not plan to acquire DIC images in addition to confocal images, pull the Wollaston prism out using the shear knob (below the objective turret) until it comes to a stop (not completely) (#7 in Figure 6-1). 3. Click on the Epi Lamp button at the top left corner of the Acquisition Setting window (#8 in Figure 6-2). 4. Choose the appropriate filter cube (mirror) for your speciman in the Microscope Controller window (DAPI, FITC, or TRITC) (#15 in Figure 6-2). 5. Open the shutter on the mercury lamp to illuminate your sample, and use the joystick and fine focus to locate your region of interest. a. Note: If you don’t see any light, check to make sure that the shutter on the mercury lamp is open. The mercury lamp brightness can also be adjusted using this shutter. 6. When you have found the region of interest and focused with the 10x objective, move sequentially up to the high powered objectives: switch to the 20x, focus, then switch to 40x if desired, and focus again. (Only use the 60x and 100x oil objectives if you have been trained by the Instrumentation Technician). 7. When you have found the region of interest and focused with the desired objective, close the shutter on the mercury lamp to avoid bleaching your sample. Olympus FV1000 Standard Operating Procedure Figure 6-1: Inverted microscope components (live cell chamber not shown) 12 Olympus FV1000 Standard Operating Procedure 1 Figure 6-2: Software windows and basic description of controls for the FV-ASW software Olympus FV1000 Standard Operating Procedure 1 6.2 Preparation for Imaging 1. Turn off brightfield or epifluorescence illumination by clicking the relevant button at the top left of the Acquisition Setting window (#8 in Figure 6-2). The microscope then automatically sets up for laser scanning (LSM). 2. Change the focus mechanism to fine (using the green buttons beside the focus knob or using the drop down list - #16 in Figure 6-2). 3. Click on the Dye List button on the left in the Image Acquisition Control window (#9, Figure 6-2). a. Double click on the dyes that you have in your sample to add them to the Selected Dyes box. Up to three dyes can be selected in addition to transmitted light. b. If the dye you are using is not in the list, it can be added by clicking on Tools  Maintenance  User Settings. Click on the Dye tab, and then click the “add” button on the far right side of the window. Type in the name of the dye and then enter the excitation and emission wavelengths near the bottom left corner of the screen. Select the laser that most closely corresponds to the excitation wavelength. Press “save and close”. c. Click Apply. The appropriate filters will automatically move into place. d. Close the Dye List window. 4. The fourth detector chanel, TD1, is for DIC or brightfield image collection only. If you plan to collect a DIC image as well, slide the Wollastin prism back into the light path (#7, Figure 6-1), then click the box beside TD1 in the Acquisition Control window. You can select any laser to correspond with the TD1 channel except the 405. 5. Set up your scanning parameters in the Acquisition Setting window: a. Mode (#1, Figure 6-2): one way raster scanning (normal) or bidirectional (high speed. Bidirecitonal is best for live cell imaging but can cause slight image abberation. This image can be corrected using the Image Adjust button, next to the scanning speed slider). b. The scanning area must be normal (rectangle) to start. i. After an image has been collected, different regions of interest (ROIs) can be selected using the buttons in this section: clip scan (diamond button; select rectangle, ellipse, or polygon), line (line button; straight or curved line), or point (point button). More than one area can be scanned at a time using the ROI manager button in the 2D Control Panel window, and laser parameters can be set individually for each area. c. Image acquisition speed (#2, Figure 6-2): Select fast scanning speed for image acquisition initially (a slower speed will give a better signal, but also causes higher photobleaching). Make sure “AutoHV” is not selected at this point. Olympus FV1000 Standard Operating Procedure 2 d. Size (#3, Figure 6-2): intially set at 512 x 512 pixels, and adjusted later to optimize resolution. e. Area (#4, Figure 6-2): leave at 1:1 initally (no zoom). f. Laser (#5, Figure 6-2): lasers are selected automatically for fluorescence according to the dye that you select. Do not adjust the laser powers at this point. g. Microscope (#6, Figure 6-2): make sure the correct objective is selected from the drop down menu. 6. In the Image Acquisition Control window: a. If you are using more than one dye, check the Sequential box to minimize bleedthrough. i. Line sequential scan minimizines the time difference between images for each channel by scanning only a line in each channel at a time. This is typically the best choice. ii. Frame sequential scan takes one full image on a given channel before scanning the next channel. b. Click on XY repeat to initiate sample scanning (#10, Figure 6-2). Alternatively, the Focus x2 or Focus x4 buttons can be used to obtain a very rough scan that skips lines, doubles or quadrupoles the scanning speed, and minimizes photobleaching. Remember that while the image is being scanned, the specimen is prone to photobleaching – avoid scanning for any longer than neccesary. c. The focus between what is seen with epi (though the oculars) and what is seen on the screen is slightly different. You wil need to move the objective up VERY CAREFULLY (while in FINE FOCUS, turn the knob very slowely counter clockwise) to bring the image into focus. Only a very small amount of adjustment should be required. If the image does not come in to focus easily, switch back to a lower magnification and/or switch back to epiflourescence mode to focus again. d. Adjust the brightness of the image: i. If the signal is low in all channels, slow the scan speed down while scanning until signal can be observed in all channels. ii. Press Ctrl+H to show the high and low limits of detection. If the signal is too high and saturating the detector, it will show up as red. Any blue on the image indicates zero signal detection. iii. Adjust HV on each of the channels (#12, Figure 6-2) independantly until just a few pixels of red can be seen (avoid setting HV higher than 700 as this will increase background instrument noise). (Note: The HV for DIC in the TD1 channel is usually much lower – try starting around 200). If the HV is quite high and the signal is still too dim, try adjusting the C.A. (confocal aperture); an increased CA will increase brightness, but also decrease resolution, and increase cross-section thickness. The scan rate can also be Olympus FV1000 Standard Operating Procedure 3 decreased futher, or the laser power can be adjusted as a last resort. Typical starting laser powers as as follows: 405, 458, 488, and 635: 0-5% 515: 10-20% 543: 45-55% iv. Avoid adjusting the gain or offset if possible (#13 and #14, Figure 6-2), however; if HV is high and the image is still too dim, the gain can be increased. Increasing the offset will turn the background darker. v. Check for bleedthrough by switching off (unchecking) all but one laser then make sure that there is only signal in the relevant channel. Do this for each laser in turn. (If you are already doing sequential scanning, this is unnessicary). vi. Kalman integration can be used to collect several images and average them to decrease S/N. This slows scanning, can dim an image, and increases photobleaching. vii. Note: for very weak signals, the photon-counting mode provides better sensitivity. e. Press Ctrl+H to return to regular viewing mode. f. If desired, optimize the resolution (Nyquist) by pressing “i” in the Image Acquisition Control window. To achieve optimal resolution the pixel size should be aprox ½ of the optical resolution. Adjust the size of the image (number of pixels) and the zoom or change objectives to optimize resolution. i. When setting the pixel size of your image, keep in mind your final format; journal, poster, etc. An image that will be blown up for a poster will need a high pixel count to avoid becoming pixelated. An image for a journal should have the dots-per-inch required by that journal, which can be determined based on image size and pixel number (i.e. the Journal of Microscopy requires 300 dpi). g. The fine focus can be adjusted again slightly to obtain the brightest image. 7. When adjustements for brightness and senstivity are complete, press the Stop button. 8. The imge is now optimized; a. to collect a single XY scan, proceed to section 6.3. b. to collect a 3-D image, proceed to section 6.4. c. to collect a time series image, proceed to section 6.5. 6.3 XY Image Acquisition 1. If you are satisfied with the image, click the XY button to do a single XY scan (#11, Figure 6-2), and then press Stop. 2. When image acquisition is complete, the image will appear in the 2D View window. This window can be used to manipulate the 2D image in a number Olympus FV1000 Standard Operating Procedure 4 of ways; see Section 6.6 for instructions on image analysis, manipulation, and saving your files. 6.4 Z-Series or 3-D Stack Image Acquisition 1. Optimize your image as described in Section 6.2. 2. Right below the XY button (#11, Figure 6-2) in the Image Acquisition Control window, select the Depth button. XYZ should now be bolded in the button. 3. Scan rapidly using XY Repeat or Focus x2 or Focus x4. 4. In the Image Acquisition window (z-stack controls indicated by #7 on Figure 6-2): a. Use the down button (or fine focus knob) to move the focus to the bottom (or just past the bottom) of your sample. Press the Set button below Start to indicate the start position. b. Use the up button (or fine focus knob) to move to the top of the sample, or to the last XY image that you want to take. Press the Set button below End. c. Click the “go” button next to Center to go the middle of your Zseries. Adjust the brightness as needed by changing the HV or laser powers. d. Stop the XY Scanning. e. Set your step size as desired, or press “Op” to the right of the Step Size box. This will automatically set the slice thickness to Nyquist/2, which garuntees that you will oversample the Z-series and capture all possible detail. Keep in mind that more sampling leads to longer scanning and increased photobleaching. The “S” at the top of the Acquisition Setting window will indicate the total time to acquire the stack. If you require a very small step size you can also increase the scan speed or decrease the depth of the stack to decrease scanning time. 5. Start the scan using the XYZ scan button. 6. Click on Series Done if you are satisfied with the stack collected, or Append Images to add additional frames. 7. If you are satisfied with the image, proceed to section 6.6 for image analysis. 6.5 T-Series Image Acquisition 1. Optimize your image as described in Section 6.2. 2. Right below the XY button in the Image Acquisition Control window, select the Time button. XYt should now be bolded in the button. 3. Set the rest interval (in the format hh:mm:ss.ms) this is the time between one acquisition start and the next) and the scanning time. Note: you can do regular LSM or visual observation during rest time (time between acquisition sets). 4. Re-start XY scanning, and if the image is acceptable, press XY(t). 5. When image acquisition is complete, press the Series Done button, or Append next to add additional frames to the image. Olympus FV1000 Standard Operating Procedure 5 6. If you are satisfied with the image, proceed to section 6.6 for image analysis. 6.6 Image Analysis When image acquisition is complete, the image will appear in the 2D View window. This window, along with the 2-D control panel, can be used to manipulate the 2D image in a number of ways, and the various tools are described in Table 6-1 and 6-2. The image can also be saved and viewed in different ways (Table 6-3), and processed (Table 6-4). 6.6.1 Add a scale bar to your image: 1. In 2D view window, click on the Pencil button, and then click on the ruler at the bottom. 2. In the area of interest, click and drag the ruler to show a scale bar. 3. Proceed to Section 6.7 to save the image. Olympus FV1000 Standard Operating Procedure 6 Table 6-1: 2D-View Window Tool button LUT (look up table) Single/Panel/Tile Zoom, 1:1, Fit to Window Slider Active Overlay Numbered buttons down the side of the window (1, 2, etc) Pencil button Animation arrows 3-D button Intensity profile button Measurement button Histogram button Series Analysis Line Series Colocalization Function Use the LUT to change the gamma, intensity, contrast, and pseudocolour of any channel to enhance dim images or highlight certain features with a different colour. Change the view of the sample, or look at more than one image at once. Channels can be viewed separately or overlaid using the Tile button. These buttons adjust the size of the image on the screen. Use to scan through the frames collected for this speciman (primarily for Z- or T-series images). Add text to the image describing Z position, time, or wavelength. Use to select the desired channels displayed in the image. Allows access to region of interest (ROI) buttons and other drawing tools (i.e. text, scale bar, grid). After a ROI is selected, imaging can be conducted on that area specifically to avoid unnecessary bleaching of entire speciman in light path. Used to play an animation or scroll through Z- or T-series. Allows viewing of a Z-series from any angle, selection of cross sections, rotation along a given axis and creation of animation (using the More arrow). Displays the intensity profile for a selected ROI. This profile can be used to determine the size of a given structure, for example. Measure the size, area, etc., of selected ROI. Also accessible from the Analysis Menu. Displays an intensity profile (intensity vs. frequency) for the selected ROI. Also accessible from the Analysis Menu. Displays intensity vales for a given ROI and each channel for each frame in a series (i.e. can use it to determine which frame best displays a given feature/ROI). Also accessible from the Analysis Menu. Illustrates the change in intensity of a line as a series (Z or T) progresses. Graph displays extent of colocalization, and statistics page calculations Pearson's coefficient, overlap and colocalization indexes. Olympus FV1000 Standard Operating Procedure 7 Table 6-2: 2D Control Panel Window Tool button Digital Zoom ROI Format and Manager buttons Load Scan data button Stepping box Tile box Intensity profiles Function Adjust digital zoom. Change the text and line formatting for labeling ROIs, and view information on ROIs selected. If accessible, the acquisition conditions for the image can be read. Allows stepping through Z, time, or animation series frames. Determines the organization of the tiles from a set of images. Click to display the vertical or horizontal intensity distribution along a line. Table 6-3: Main Fluoview Window Tool button Properties ( “i” ) Report, Thumbnail, Thumbnail + property Various buttons to organize how windows are displayed Darkroom colour button Function Display image information in Data Manager Window. Change how files are displayed in the Explorer window. Assists in organizing the screen-view when many windows are open. Dims the monitor display to minimize stray light. Olympus FV1000 Standard Operating Procedure 8 Table 6-4: Processing Menu Menu Item Filter Setting… and Filter Threshold Image Calculation Correcting Pixel Gaps Correcting Z Gaps Ratio/Concentration Edit Experiment Function Mathematical filters can be used to improve image clarity or emphasize certain characteristics. The filters available are: sharpen, average, DIC, sobel, median, shading, lapacian. Sharpen: highlights the edges of images, improves blurred images, but also increases noise. Sobel: Emphasizes contours. Lapacian: emphasizes intensity changes. Press “Preview” at bottom of Filter window first, then select various filters to observe result. Select the Single or Series button, if doing a single image, enter the desired frame in the Image Position box. If you want to save the filtered image, press New Image. This feature is used to make the image binary (two colour). The C and Z indicate the channel and frame to display. The thresholds can be adjusted by clicking on the channel in the table below threshold, and manually entering new values, or by moving the bars on the graph Used to subtract/add/divide images or constants. This window can be used to correct for image shift between channels. Image position can be used to select the desire frame. The white box in the Preview area can be dragged to view other areas of the image. To correct for Z position shift between channels. Moving the yellow bar changes the Z-slice image. The Ratio tab is used to look at changes in intensity between two channels. Use the drop down menu to select the desired operation, and the equation for that operation will be displayed in the Equation box. Equations include baseline/background subtraction, ratio of channels, FRET image comparison (photo bleached image required). The Concentration tab is used to analyze a change in ion intensity over time. This feature requires knowledge of intensity values with and without ion, and the dissociation constant (in nM). See the Help menu for an explanation of F values. Combine single channels, append or extract series, from two different images to create a new image or series Olympus FV1000 Standard Operating Procedure 9 Additional Tools: Virtual channel: Useful if fluorescent signals from two reagents overlap. Delay image acquisition: After a set time has passed, acquisition can be started or terminated using XYt imaging. Image noise reduction: Can be done using Kalman filtering (takes several scans and averages them). Live plot window: plots change in intensity in ROI over time. First outline ROI in Live View window by clicking on the “pen” tool. Then select Live Plot from Live menu. Live Tiling: allows viewing of the specimen as time elapses while the image is being acquired (button in Live View window has the image of a wrench). 6.7 Saving and Viewing Final Images 1. When you are finished collecting and analyzing your image, select File  Save As and save it to your folder as a .oif or .oib file (automatically saved in D:\FV10-ASW\users\yourusername\Image). Make sure “Include all ROI” (or selected ROIs) are checked before you save the image so that scale bars and other additions to the image are included. a. You can also right click on the image to export as a TIFF, JPEG, or other image format. b. An OIF format saves the images as TIFFs in a folder, and creates a file with the sample information. Both the file and the TIFFs are required to open the image. If you save this data to a memory key or CD, you need to copy both the .oif files and the associated sub-folder containing the images. c. An OIB format saves all of the images in a set of files. This format is recommended if analyzing the images in Metamorph or a similar program. 2. Manipulations can be done using the fluoview software on the confocal computer, or another program or fluoview software on a different computer (recommended). a. A free version of the Fluoview software can be downloaded from:. https://support.olympus.co.jp/cf_secure/en/lisg/bio/download/ ga/fv10_asw/ Our serial number is 7M10. b. ImageJ (http://rsb.info.nih.gov/ij/) and VOXX (http://www.nephrology.iupui.edu/imaging/voxx/) software can also be used to view .oif files. Note: DATA IS CLEARED FROM THE CONFOCAL COMPUTER PERIODICALLY, so SAVE YOUR DATA ELSEWHERE (disk, another computer) for safekeeping. Olympus FV1000 Standard Operating Procedure 10 6.8 Cleaning Objectives NEVER USE KIMWIPES OR OTHER TISSUE PAPER TO CLEAN OBJECTIVES. USE ONLY LENS PAPER, supplied in SR314. If you notice oil on the non-oil objectives (10x, 20x and 40x), DO NOT try to clean them. Inform the Instrumentation Technician as soon as possible and note the problem in the log book. 1. Using clean lens paper gently blot off the oil from the lens. Do NOT drag the paper across the lens, just dab off the oil. The front lens of the objective is very delicate and must be protected from scratching. 2. Wipe any oil off of the objective barrel. 3. Put a drop of ethanol on a clean piece of lens paper, and gently draw it across the lens (DO NOT RUB). 6.9 Shutdown 1. Switch back to the 10x objective, then manually lower the objective by turning the knob clockwise. 2. Remove your slide. 3. If you used the oil objectives, make sure the oil has been removed as per section 6.8. 4. Make sure you have saved your data and then exit the Fluoview software. Close the FL (mercury lamp) shutter as recommended by the Caution message. 5. If the argon laser (#3) it turned on, turn it’s key to the upright position, but DO NOT TURN OFF THE POWER SWITCH YET. This allows the fan to continue to blow until the laser has cooled down. 6. Turn off power bar (#2) 7. Turn off the Mercury power supply (#1). 8. When the fan has stopped running on the argon laser, the power can be turned off (#3). 9. Transfer your data to a CD or memory stick if desired (recommended). 10. Shutdown the computer. 11. Cover the microscope. Make sure the cover DOES NOT touch the mercury lamp, which is hot and may melt the cover. 12. Clean up anything you have left in the room. 13. Fill in the rest of the log book. 7. TROUBLESHOOTING For troubleshooting tips, see the Help menu in the Fluoview software. Olympus FV1000 Standard Operating Procedure 8. PREVENTATIVE MAINTENANCE Users are not to perform maintenance. Unless noted otherwise, these procedures are carried out by the Instrumentation Technician. 8.1 Monthly - If the system has not been used for a month or more, it will be run for ~ 8 hours to ensure long-term laser stability Objective lenses will be checked and cleaned if required Check the 5% CO2 and humidity canister for the Weather Station chamber Check the CO2 for the anti-vibration table 8.2 Annually or as Required 8.2.1 Replace the Mercury Burner The mercury burner should be replaced when the lamp hours reach 350 hours or when the epifluorescence images appear unstable due to lamp failure. 11 Olympus FV1000 Standard Operating Procedure 12 9. QUICK REFERENCE GUIDE 1. 2. 3. 4. 5. 6. 7. 8. Remove the cover from the microscope. Sign in in the user log book and fill in the relevant details. Turn on the microscope components according to the numbers on lables: Log on to the computer and the software (FV10-ASW 1.7) Wait for the software and microscope to initialize. Cafully place your slide on the stage. Set up Kohler. View the image using epifluorescence and the appropriate filter cube to fine-tune the focus. 8. When you have found the region of interest and focused with the desired objective, close the shutter on the mercury lamp to avoid bleaching your sample. 9. Turn off brightfield or epifluorescence illumination. The microscope then automatically sets up for laser scanning (LSM). 10. In the Image Acquisition Control window, select desired dyes. 9. Set up your image parameters in the Acquisition Setting window: a. One way scanning (normal) or bidirectional. b. Select fast scanning speed for initial image acquisition initially c. Set the size to 512 x 512 pixels d. Leave zoom at 1:1 initally. e. Set the laser powers should be set as low as possible to minimize bleaching (except the 543 laser, which should be at at least 50%). f. Select the correct objective from the drop down menu. 10. In the Image Acquisition Control window: a. If you are using more than one dye, check the Sequential box to minimize bleedthrough. b. Click on XY repeat or Focus x2 or Focus x4 to initiate scanning c. Use the fine focus or the up and down button in the Acquisition Setting window to align the microscope with the cross section to be observed. d. Adjust HV on the active channels until the image can be observed (do not set higher than 700). e. Adjust the brightness of the image using HV, CA, scan speed, or laser powers, and check for satration (Ctrl + H). f. If desired, optimize the resolution (Nyquist) by pressing “i” in the Image Acquisition Control window. 11. When adjustements for brightness and senstivity are complete, press the Stop button. 12. The image is now optimized. 13. Collect a single XY scan, or create a Z- or T-series. 14. When image acquisition is complete, the image will appear in the 2D View window. 15. Adjust the image as nessicary, then save as an .oif or .oib file. Transfer to a CD or memory key for long term storage on your personal computer. Olympus FV1000 Standard Operating Procedure REFERENCES Laboratory Health and Safety Manual. 2007. Wilfrid Laurier University Environmental/Occupational Health and Safety Office. Olympus website: http://www.olympusfluoview.com. Pawley, JB (Ed.). 2006. Handbook of Biological Confocal Microscopy, 3rd Edition. Springer Science + Business Media: Singapore. Staudt, T, Lang, MC, Medda, R, Engelhardt, J, & Hell, S.W. 2007. 2,2'thiodiethanol: a new water soluble mounting medium for high resolution optical microscopy. Microsc Res Tech 70: 1-9. Texas A&M University. Microscopy and Imaging Center. Olympus FV1000. http://microscopy.tamu.edu/instruments/light-microscopy/olympusfv1000-confocal-microscope.html. Accessed May 26, 2008. 13 14 Olympus FV1000 Standard Operating Procedure APPENDIX 1: FLUOROCHROME PEAK EXCITATION AND EMISSION WAVELENGTHS The following list of fluorochromes and wavelengths is from the Olympus website (http://www.olympusmicro.com/primer/techniques/fluorescence/ fluorotable2.html) Fluorochrome Excitation Wavelength Emission Wavelength Acid Fuchsin 540 630 Acridine Orange (Bound to DNA) 502 526 455-600 560-680 Acridine Yellow 470 550 Acriflavin Acridine Red 436 520 AFA (Acriflavin Feulgen SITSA) 355-425 460 Alizarin Complexon 530-560 580 Alizarin Red 530-560 580 Allophycocyanin 650 661 ACMA 430 474 AMCA-S, AMC 345 445 Aminoactinomycin D 555 655 7-Aminoactinomycin D-AAD 546 647 Aminocoumarin 350 445 Anthroyl Stearate 361-381 446 Astrazon Brilliant Red 4G 500 585 Astrazon Orange R 470 540 Astrazon Red 6B 520 595 Astrazon Yellow 7 GLL 450 480 Atabrine 436 490 Auramine 460 550 Aurophosphine 450-490 515 Aurophosphine G 450 580 BAO 9-(Bisaminophenyloxadiazole) 365 395 BCECF 505 530 Berberine Sulphate 430 550 Bisbenzamide 360 600-610 BOBO-1, BO-PRO-1 462 481 Blancophor FFG Solution 390 470 Blancophor SV 370 435 15 Olympus FV1000 Standard Operating Procedure Bodipy Fl 503 512 Bodipy TMR 542 574 Bodipy TR 589 617 BOPRO 1 462 481 Brilliant Sulphoflavin FF 430 520 Calcein 494 517 Calcien Blue 370 435 Calcium Green 505 532 Calcium Orange 549 576 Calcofluor RW Solution 370 440 Calcofluor White 440 500-520 Calcofluor White -ABT Solution 380 475 Calcofluor White- Std Solution 365 435 548(low pH) 576(high pH) 587(low pH) 635(high pH) 6-Carboxyrhodamine 6G 525 555 Cascade Blue 400 425 Catecholamine 410 470 450-490 515 504(low pH) 514(high pH) 587(low pH) 540(high pH) Coriphosphine O 460 575 Coumarin-Phalloidin 387 470 CY3.18 554 568 CY5.18 649 666 CY7 710 805 DANS (1-DimethylAminoNaphthaline-5-Sulphonic Acid) 340 525 340-380 430 Dansyl NH-CH3 in water 340 578 DAPI 350 470 DiA 456 590 Diamino Phenyl Oxydiazole (DAO) 280 460 5-(and 6-)carboxy SNARF-1 indicator Chinacrine CL-NERF DANSA (DiaminoNaphthylSulphonic Acid) Di-8-ANEPPS 488 605 310-370 520 DiI [DiIC18(3)] 549 565 DiO [DiOC18(3)] 484 501 Diphenyl Brilliant Flavine 7GFF 430 520 DM-NERF 497(low pH) 510(high pH) 527(low pH) 536(high pH) Dopamine 340 490-520 Dimethylamino-5- Sulphonic Acid 16 Olympus FV1000 Standard Operating Procedure ELF-97 alcohol 345 530 Eosin 525 545 Erythrosin ITC 530 558 Ethidium Bromide 510 595 Euchrysin 430 540 FIF (Formaldehyde Induced Fluorescence) 405 435 375-530 612 Fluorescein 494 518 Fluorescein Isothiocyanate (FITC) 490 525 Fluo 3 485 503 FM1-43 479 Flazo Orange 598 Fura-2 363(low [Ca ]) 335(high [Ca2+]) 512(low [Ca2+]) 505(high [Ca2+]) Fura Red 472(low [Ca2+]) 436(high [Ca2+]) 657(low [Ca2+]) 637(high [Ca2+]) Genacryl Brilliant Red B 520 590 Genacryl Brilliant Yellow 10GF 430 485 Genacryl Pink 3G 470 583 Genacryl Yellow 5GF 430 475 Gloxalic Acid 405 460 Granular Blue 355 425 530-560 580 Hoechst 33258, 33342 (Bound to DNA) 352 461 3-Hydroxypyrene-5,8,10-TriSulfonic Acid 403 513 7-Hydroxy-4methylcoumarin 360 455 380-415 520-530 Indo-1 350 405-482 Intrawhite Cf Liquid 360 430 Leucophor PAF 370 430 Leucophor SF 380 465 Leucophor WS 395 465 Lissamine Rhodamine B200 (RD200) 575 595 Lucifer Yellow CH 425 528 Lucifer Yellow VS 430 535 Haematoporphyrin 5-HydroxyTryptamine (5-HT) 2+ 17 Olympus FV1000 Standard Operating Procedure LysoSensor Blue DND-192, DND-167 374 425 LysoSensor Green DND-153, DND-189 442 505 384(low pH) 329(high pH) 540(low pH) 440(high pH) LysoTracker Green 504 511 LysoTracker Yellow 534 551 LysoTracker Red 577 592 Magdala Red 524 600 Magnesium Green 506 531 Magnesium Orange 550 575 Maxilon Brilliant Flavin 10 GFF 450 495 Maxilon Brilliant Flavin 8 GFF 460 495 Mitotracker Green FM 490 516 Mitotracker Orange CMTMRos 551 576 MPS (Methyl Green Pyronine Stilbene) 364 395 Mithramycin 450 570 NBD 465 535 LysoSensor Yellow/Blue NBD Amine 450 530 Nile Red 515-530 525-605 Nitrobenzoxadidole 460-470 510-650 340 490-520 Noradrenaline Nuclear Fast Red 289-530 580 Nuclear Yellow 365 495 Nylosan Brilliant Flavin E8G 460 510 Oregon Green 488 fluorophore 496 524 Oregon Green 500 fluorophore 503 522 Oregon Green 514 fluorophore 511 530 Pararosaniline (Feulgen) 570 625 Phorwite AR Solution 360 430 Phorwite BKL 370 430 Phorwite Rev 380 430 Phorwite RPA 375 430 Phosphine 3R 465 565 Phosphine R 480-565 578 Pontochrome Blue Black 535-553 605 POPO-1, PO-PRO-1 434 456 Primuline 410 550 Procion Yellow 470 600 18 Olympus FV1000 Standard Operating Procedure Propidium Iodide Pyronine 536 617 410 540 540-590 560-650 Pyrozal Brilliant Flavin 7GF 365 495 Quinacrine Mustard 423 503 R-phycoerythrin 565 575 Rhodamine 110 496 520 Rhodamine 123 511 534 Rhodamine 5 GLD 470 565 Rhodamine 6G 526 555 Pyronine B Rhodamine B 540 625 523-557 595 Rhodamine B Extra 550 605 Rhodamine BB 540 580 Rhodamine BG 540 572 Rhodamine Green fluorophore 502 527 Rhodamine Red 570 590 Rhodamine WT 530 555 Rhodol Green fluorophore 499 525 Rose Bengal 540 550-600 Serotonin 365 520-540 Sevron Brilliant Red 2B 520 595 Sevron Brilliant Red 4G 500 583 Sevron Brilliant Red B 530 590 Sevron Orange 440 530 Sevron Yellow L 430 490 SITS (Primuline) 395-425 450 SITS (Stilbene Isothiosulphonic Acid) 365 460 Sodium Green 507 535 Stilbene 335 440 Snarf 1 563 639 Sulpho Rhodamine B Can C 520 595 Sulpho Rhodamine G Extra 470 570 SYTOX Green nucleic acid stain 504 523 SYTO Green fluorescent nucleic acid stains 494 ± 6 515 ± 7 SYTO Green fluorescent nucleic acid stains 515 ± 7 543 ± 13 SYTO 17 red fluorescent nucleic acid stain 621 634 Tetracycline 390 560 TRITC (Tetramethyl Rhodamine 557 576 Rhodamine B 200 19 Olympus FV1000 Standard Operating Procedure Isothiocyanate) Texas Red 596 615 Thiazine Red R 510 580 Thioflavin S 430 550 Thioflavin TCN 350 460 Thioflavin 5 430 550 370-385 477-484 Thiozol Orange 453 480 Tinopol CBS 390 430 TOTO 1, TO-RRO-1 514 533 TOTO 3, TO-PRO-3 642 661 True Blue 365 420-430 Ultralite 656 678 Uranine B 420 520 Uvitex SFC 365 435 X-Rhodamine 580 605 Xylene Orange 546 580 XRITC 582 601 YOYO-1, YOYO-PRO-1 491 509 YOYO-3, YOYO-PRO-3 612 631 Thiolyte For additional lists of fluorescent dyes and their properties, see: Interactive database of fluorescence spectra: http://www.mcb.arizona.edu/ipc/fret/default.htm Olympus fluorchrome data tables: http://www.olympusmicro.com/primer/techniques/fluorescence/fluorotable2. html Molecular Probes dye spectra viewer: http://probes.invitrogen.com/resources/spectraviewer/ Olympus FV1000 Standard Operating Procedure APPENDIX 2: USER LOG 20 21 Olympus FV1000 Standard Operating Procedure DATE & TIME IN NAME & EXTENSION SUPERVISOR SAMPLE AND DYES USED MERCURY LAMP HOURS AT START HE-NE LASER USED (Y/N) OBJECTIVES USED (AIR: 10, 20, 40, OIL: 60 OR 100) MERCURY LAMP HOURS AT END TIME OUT PROBLEMS / COMMENTS Olympus FV1000 Standard Operating Procedure APPENDIX 3: PREVENTATIVE MAINTENANCE LOG 22 23 Olympus FV1000 Standard Operating Procedure DATE NAME EXT # TYPE OF MAINTENANCE FREQUENCY OF MAINTENANCE (I.E. WEEKLY) PROBLEMS / COMMENTS