Emc Test Technologies Sas Via Marecchiese, 273 - 47900 - Rimini (RN) P.Iva: 03768440400 Tel. +39 0541 728562 Fax. +39 0541 1792293 E.mail:
[email protected]
GTEM EMC 250-INOX Frequency Range:
100kHz to 20GHz
Max. Power-Input:
250 W
Input Connector Type:
N (Female)
Nominal Impedance:
50 Ω
Typical VSWR:
1:1.2
Power Line Filter:
2x 230V AC/DC 20A
RF Coax Feed-Thru(s):
2x SMA + 1x N (Females)
Other Feed-Thru(s): 1x Fiber optical cable Installed options:
Dimensions:
115(L) x 60(W) x 42(H)
PLF-20S; FOPT-1; FT-N; FT-SMA
Septum Height:
250 mm
PDTP
Door Size:
30(W)x20(H) cm
Uniform Area Size ±3dB:
15(L)x15(W)x10(H) cm
Reference Standards:
EUT Max Size:
20(L)x20(W)x15(H) cm
EN 61000-4-3 - EN 61000-4-20
Weight:
25 Kg
Introduction The GTEM cell is a TEM waveguide with the upper frequency limit extended to the GHz range. It is under consideration as an alternative measurement facility for both radiated emission and immunity measurements. It is included in the recently published standard IEC 61000-4-20 “Emission and Immunity Testing in Transverse Electromagnetic (TEM) Waveguides”. Theory of operation GTEM-cells (Giga-hertz Transversal Electro-Magnetic cells) are waveguide structures intended for electromagnetic compatibility measurements, as well as biomedical applications. The electromagnetic field distribution inside the cell is in TEM mode. With TEM mode propagation, there is no component of electric and magnetic field in the direction of propagation of electromagnetic wave. Therefore the field components are strictly perpendicular. Assuming the field distribution ideal TEM below the cut-off frequency of the cell (before the introduction of higher order modes), the electromagnetic field distribution can be considered static. Key Features
Engineered and completely manufactured in Italy.
Ruggedized fully INOX steel construction
Unique compact design.
Optimized for EMI and EMC.
Strong fields achieved with low input power
Broadband up to 20Ghz
High effective shielding
Fully customizable feed thrus and filters
Excellent quality at Low cost
* Specifications are subject to change without prior notice
w w w . e m c t e s t . i t
Pag. 1 of 3
Emc Test Technologies Sas Via Marecchiese, 273 - 47900 - Rimini (RN) P.Iva: 03768440400 Tel. +39 0541 728562 Fax. +39 0541 1792293 E.mail:
[email protected]
Applications
EMI and EMS devices
Radiation and susceptibility test
Specifically designed for telecom application
Biomedical and dosimetrical applications
Isotropic sensors calibration
Receiver sensitivity test
Available options INOX
INOX rugged construction (Default on GTEM EMC 250 Only)
EIA 7/8
EIA 7/8'' Input Connector (max. 3GHz)
EIA 7/16
EIA 7/16'' Input Connector (max. 3GHz)
ESI-250
Electrical safety interlock
FOPT-1
Channel for fibre optic leads (1 fiber pair)
FOPT-3
Channel for fibre optic leads (3 fiber pairs)
FPT-DB9
Filtered 9-poles pass-thru RS-232(DB9)
FPT-DB25
Filtered 25-poles pass-thru RS-232(DB25)
FPT-DC10
Filtered 10A 1000V DC banana socket pass-thru
FT-BNC
RF feed-thru BNC-BNC Female panel mount type connector
FT-N
RF feed-thru N-N Female panel mount type connector
FT-BNC/SMA
RF feed-thru BNC-SMA Female panel mount type connector
FT-RJ45
Shielded RJ45 (RJ11) feed-thru female-female connector
FT-SMA
RF feed-thru SMA-SMA Female panel mount type connector
HCOMB-10
Honey comb 10x10 cm air intake/outtake
HPTER-250
Hi Power terminations 500W continuous power up to 3GHz
ILED-250
Indoor LED lighting 50W shielded lamp
IP-CAM
Shielded RJ45 IP camera system
PDTP
Technical panel pre-drilled for options with EMC gaskets and inox bolts
PLF-20S
Single Phase AC Power Line Filter 2x 20A Phase+N+Ground
PLF-40S
Single Phase AC Power Line Filter 2x 40A Phase+N+Ground
PLF-60S
Single Phase AC Power Line Filter 2x 60A Phase+N+Ground
PLF-20T
Tri Phase AC Power Line Filter 4x 20A 3Ph+N+Ground
PLF-32T
Tri Phase AC Power Line Filter 4x 32A 3Ph+N+Ground
PLF-64T
Tri Phase AC Power Line Filter 4x 64A 3Ph+N+Ground
DCF-120
Single pole Filtered AC/DC Feedthru 120A 1700VDC/250VAC
PWS
Additional Power Socket for EUT
SAE
Door for tests acc. to SAE J1752/3 – Integrated Cir
SHW-250
Shielded Window in door for GTEM EMC 250 GTEM Vertical positioning – Plastic table over absorbers – 90° turned door
VPOS
turn of door position, plastic table over pyramids Only for GTEM 250/ 500
WLIP-CAM
Shielded IP camera system
* Specifications are subject to change without prior notice
w w w . e m c t e s t . i t
Pag. 2 of 3
Emc Test Technologies Sas Via Marecchiese, 273 - 47900 - Rimini (RN) P.Iva: 03768440400 Tel. +39 0541 728562 Fax. +39 0541 1792293 E.mail:
[email protected]
Detailed views
Locking System
Technical Panels
Shielded Window
Inside view
Multipolar Filter
Power required / Electric field Vs. frequency
Electric Field [V/m] 4 3,5 3
Power (W)
2,5 10V/m 30V/m
2
50V/m 1,5 1 0,5 0 1
10
100
1000
Freq (MHz)
* Specifications are subject to change without prior notice