Transcript
X-Ray imaging and Optical Equipments For Synchrotrons
High resolution X-Ray imaging microscopes For large format camera sensors
Monochromatic configuration (Low/medium X-Ray flux & energy)
Vertical configuration (with PCO 4000 camera)
Horizontal configuration (with Frelon Camera)
Microscope
3 objectives motorised nosepiece Motorised Focus Motorised camera rotation Single or triple motorized scintillator supports Adapters for different camera formats & mounts
Lightproof scintillator supports
Standard or specific supports Tiltable scintillator support Multiple motorised scintillator support
Vertical configuration (With Andor iKon camera)
References : ESRF, SLS, RWTH, BESSY, ANKA, DESY, DIAMOND, LBLN (USA) 2
High resolution X-Ray imaging microscopes For small format camera sensors
Monochromatic configuration (Low/medium X-Ray flux & energy)
Microscope with Andor Neo camera
Microscope with PCO 2000 Camera
High resolution X-Ray imaging microscopes Single objective Single objective microscope
Motorised focus One objective (range 1.25X to 40X) Scintillator support : tiltable / not tiltable Same scintillator large & miniature mounts as other microscopes Adapters for large format cameras
Miniature scintillator mount
Large scintillator mount / tiltable
References : ESRF, SLS, ANKA, IMT, DESY, Diamond, LSU-CAMD (USA) , SSRF (China), LNLS (Brazil), PAL (Korea), Indus (India), Australian Synchrotron 3
High resolution X-Ray imaging microscopes White Beam configuration (High X-Ray flux & energy)
WB Microscope with single or multiple eyepieces supports
Twin objectives WB microscope
White beam microscopes
2X-5X-7.5X-10X long working distance objectives Single or twin objective configurations Single or multiple eyepieces supports (For large format cameras) Small format cameras direct coupling Motorised focus Motorised camera rotation Low backscattering structure materials
Tiltable scintillator support
Small camera format - direct coupling & Camera motorised rotation
References : ESRF, SLS, ANKA, Australian Synchrotron, LNLS (Brazil), Indus(India) SLRI (Thailand) 4
High resolution X-Ray imaging microscopes Modular system
LOW / MEDIUM / HIGH ENERGY BEAMS Low energy (monochromatic beam) module Three objectives motorised nosepiece Objectives in the beam axis : High resolution short working distance objectives Objectives set from 1,25X/0,04 to 40X/0,95 Motorised focus Filters slot and filters wheels Scintillator support (manual/motorized)
High energy (White beam / Pink beam) module One long working distance objective Objective off beam axis (avoids optics browning by radiations) Low scattering image mirror Motorised focus Magnification change performed by eyepieces turret Scintillator support
5
High resolution X-Ray imaging microscopes Optical data for small / medium format cameras (Direct camera mounting) Monochromatic microsopes Objective Glass thickness comp. (mm)
1.25 X
2 X
White beam microscopes
4 X
10 X
20 X
20 X (Lwd)
2 X
5 X
7.5 X
10 X
-
-
-
0.17
0.17
0-2
-
-
-
-
NA
0.04
0.08
0.16
0.40
0.75
0.45
0.05
0.14
0.21
0.28
Working distance (mm)
5.10
6.20
13.0
3.10
0.65
6.6 - 7.8
34
34
35
33.5
Resolution limit (µm)
8.4
4.2
2.1
0.84
0.45
0.75
6.7
2.4
1.6
1.2
12.12 12.12 5.9
7.58 7.58 3.7
3.79 3.79 1.9
1.52 1.52 0.74
0.76 0.76 0.37
0.76 0.76 0.37
8.42 8.42 4.11
3.37 3.37 1.64
2.25 2.25 1.10
1.68 1.68 0.82
246
66
17
2.6
0.74
1.8
151
20
8.8
5.0
4.16 3.51 1.6
1.66 1.40 0.65
0.83 0.70 0.33
0.83 0.70 0.33
9.24 7.80 3.61
3.70 3.12 1.44
2.47 2.08 0.96
1.85 1.56 0.72
16
2.5
0.71
1.7
146
19
8.5
4.8
0.67 0.67 0.33
7.40 7.40 3.61
2.96 2.96 1.44
1.97 1.97 0.96
1.48 1.48 0.72
CCD PCO 2000 / PSL HR4M (2048x2048 pixels 7,4x7,4µm) Hor. Object field (mm) Vert. Object field (mm) Object Vert.&Hor. pixel size (µm) +/- Depth of focus (µm) ( Optics + camera )
sCMOS PCO Edge 5.5 / Andor Neo / Andor Zyla 5.5 (2560x2160 pixels 6,5x6,5µm) Hor. Object field (mm) 13.31 8.32 Vert. Object field (mm) 11.23 7.02 Object Vert.&Hor. pixel size (µm) 5.2 3.3 +/- Depth of focus (µm) ( Optics + camera )
237
63
sCMOS PCO Edge 4.2 / Andor Zyla 4.2 / PSL Genicam / Hamamatsu Orca Flash 4 (2048x2048 pixels 6.5x6.5µm) Hor. Object field (mm) 10.65 6.66 3.33 1.33 0.67 Vert. Object field (mm) 10.65 6.66 3.33 1.33 0.67 Object H&V pixel size (µm) 5.2 3.3 1.6 0.65 0.33 +/- Depth of focus (µm) ( Optics + camera ) CCD iKon-M 912 (512x512 pixels 24x24µm) Hor. Object field (mm) Vert. Object field (mm) Object H&V pixel size (µm) +/- Depth of focus (µm) ( Optics + camera ) CCD iKon-M 934 (1024x1024 pixels 13x13µm) Hor. Object field (mm) Vert. Object field (mm) Object H&V pixel size (µm) +/- Depth of focus (µm) ( Optics + camera )
237
63
16
2.5
0.71
1.7
146
19
8.5
4.8
9.83 9.83 19.20
6.14 6.14 12.00
3.07 3.07 6.00
1.23 1.23 2.40
0.61 0.61 1.20
0.61 0.61 1.20
6.83 6.83 13.33
2.73 2.73 5.33
1.82 1.82 3.56
1.37 1.37 2.67
412
118
29
4.7
1.29
2.7
243
33
14.7
8.3
10.65 10.65 10.40
6.66 6.66 6.50
3.33 3.33 3.25
1.33 1.33 1.30
0.67 0.67 0.65
0.67 0.67 0.65
7.40 7.40 7.22
2.96 2.96 2.89
1.97 1.97 1.93
1.48 1.48 1.44
302
84
21
3.3
0.92
2.1
182
24
10.8
6.1
6
High resolution X-Ray imaging microscopes Optical data for large format cameras (Camera mounting with image enlarging eyepieces) Monochromatic microsopes Objective Glass thickness comp. (mm)
1.25 X
2 X
White beam microscopes
4 X
10 X
20 X
20 X (Lwd)
2 X
5 X
7.5 X
10 X
-
-
-
0.17
0.17
0-2
-
-
-
-
NA
0.04
0.08
0.16
0.40
0.75
0.45
0.05
0.14
0.21
0.28
Working distance (mm)
5.10
6.20
13.0
3.10
0.65
6.6 - 7.8
34
34
35
33.5
Resolution limit (µm)
8.4
4.2
2.1
0.84
0.45
0.75
6.7
2.4
1.6
1.2
7.21 4.81 1.80
3.61 2.40 0.90
1.44 0.96 0.36
0.72 0.48 0.18
0.72 0.48 0.18
8.02 5.34 2.00
3.21 2.14 0.80
2.14 1.43 0.53
1.60 1.07 0.40
208
54
14
2.2
0.61
1.6
130
17
7.5
4.2
8.87 8.87 4.40
5.54 5.54 2.75
2.77 2.77 1.38
1.11 1.11 0.55
0.55 0.55 0.28
0.55 0.55 0.28
6.16 6.16 3.06
2.46 2.46 1.22
1.64 1.64 0.81
1.23 1.23 0.61
227
60
15
2.4
0.67
1.7
141
18
8.2
4.6
11.06 11.06 5.40
6.91 6.91 3.38
3.46 3.46 1.69
1.38 1.38 0.68
0.69 0.69 0.34
0.69 0.69 0.34
7.68 7.68 3.75
3.07 3.07 1.50
2.05 2.05 1.00
1.54 1.54 0.75
239
64
16
2.6
0.71
1.7
148
19
8.6
4.8
8.85 8.85 4.32
5.53 5.53 2.70
2.76 2.76 1.35
1.11 1.11 0.54
0.55 0.55 0.27
0.55 0.55 0.27
6.14 6.14 3.00
2.46 2.46 1.20
1.64 1.64 0.80
1.23 1.23 0.60
226
60
15
2.4
0.67
1.7
140
18
8.1
4.6
5.73 5.73 2.80
2.87 2.87 1.40
1.15 1.15 0.56
0.57 0.57 0.28
0.57 0.57 0.28
6.37 6.37 3.11
2.55 2.55 1.24
1.70 1.70 0.83
1.27 1.27 0.62
60
15
2.4
0.68
1.7
141
18
8.2
4.6
CCD PCO 4000 / PSL VHR - 2,5X Eyepiece (4008x2672 pixels 9x9µm) Hor. Object field (mm) 11.54 Vert. Object field (mm) 7.70 Object H&V pixel size (µm) 2.88 +/- Depth of focus (µm) ( Optics + camera ) CMOS PCO Dimax - 2X eyepiece (2016x2016 pixels 11x11µm) Hor. Object field (mm) Vert. Object field (mm) Object H&V pixel size (µm) +/- Depth of focus (µm) ( Optics + camera ) CCD iKon-L 936 - 2X eyepiece (2048x2048 pixels 13.5x13.5µm) Hor. Object field (mm) Vert. Object field (mm) Object H&V pixel size (µm) +/- Depth of focus (µm) ( Optics + camera ) CCD iKon-L 936 - 2.5X eyepiece (2048x2048 pixels 13.5x13.5µm) Hor. Object field (mm) Vert. Object field (mm) Object H&V pixel size (µm) +/- Depth of focus (µm) ( Optics + camera )
CCD Frelon (TH7899M) - 2.5X eyepiece (2048x2048 pixels 14x14µm) Hor. Object field (mm) 9.18 Vert. Object field (mm) 9.18 Object H&V pixel size (µm) 4.48 +/- Depth of focus (µm) ( Optics + camera )
228
7
Low magnification X-Ray imaging Macroscopes
Macroscope For large format cameras Up to 42 mm field of view Motorised focus
Zoom macroscope (With Photonic Science camera) 18 m to 3 mm field of view Motorised zoom & focus
Low magnification macro objectives
High NA 5X macro objective
High resolution custom optics
High NA 4X macro objective
Tandem macro objectives
References : ESRF, ELETTRA (Italy), SLS, LSU-CAMD (USA) 8
Mirror objective for high energy beams imaging 10X / 0,4 Mirror objective High resolution on axis optic 43 mm long working distance Object field up to 2 mm diameter Sensor up to 20 mm diagonal Beam stop interface on front mirror Internal lead shield protection
T Mount adapter (Suitable for F-Mount and C-mount cameras)
Beam stop interface
Reference : ESRF
Video microscopes applications
Motorised video microscope Multiple objective nospiece Integrated slip ring connector
Vacuum operation Long working distance optics
Drilled bending mirrors (for X-Ray beam path)
(Sensor mounting on objective attachment)
References : ESRF – SOLEIL - DESY 9
Spectroscopy Micro spectro photometer
3 * 90°- orientated objectives supports 3 axis adjustable support Long working distance objectives Polarizer on objectives Filter slider on objectives Long working distance videomicroscope XY stages 2 axis-motorised sample holder Motorised rotation Goniometric head interface
Mirror objectives for wide spectrum spectroscopy
4X/0.29 objective 50 mm working distance
4X/0.22 objective 35.5 mm working distance
Objectives with optic fibres matig sleeves _ rotatable polariser/analyser _ filters
References : ESRF – IBS – EMBL – MIT (USA) – Brookhaven Labs (USA)
10
45° observation cameras Specific optics enabling a 45° tilted observation plane Standard model : 65 mm working distance Vacuum tight model : 98 mm working distance
Standard
Vacuum tight
Optic fibres and illumination systems Specific optical fibres design and manufacture Led illumination
Vaccum tight optical fibres couplings
Miniature transmitted & polarised light illuminator ( 2 mm diameter light guide)
References : ESRF - CEA - SOLEIL - BESSY 11
Other activities MADE-TO MEASURE APPLIED TO : NUCLEAR MICRO-ELECTRONICS
OPTICAL COMPONENTS MACHINE VISION
SERVICES TRAINING SESSIONS CALIBRATION MAINTENANCE EQUIPMENTS MODIFICATION
Tel : (33) 04 78 47 78 74 FAX : (33) 04 78 83 61 74
[email protected]
www.optiquepeter.com Allée Romaine Z.A. du Charpenay 69210 Lentilly - France 2014 12