Preview only show first 10 pages with watermark. For full document please download

Zeiss Transmitted Light, Wide-field Fluorescence And Deconvolution

   EMBED


Share

Transcript

Louise Cole Apr 2013  Zeiss Transmitted light, wide‐field Fluorescence and Deconvolution  microscope    Start‐up   Uncover microscope. Do not put dust cover on the floor.   Plug in the correct camera. The default camera is the AxioCam  HRm (monochrome camera) for fluorescence imaging (left hand  side of microscope).  The AxioCam HRc (colour  camera) is suitable for  transmitted light imaging (top  of microscope).         Turn on the microscope (green switch on the  power supply box; see right image, bottom unit)  and round silver switch on the left hand side of the  microscope (see below). The TFT display on the  microscope will come on.           If require ApoTome.2, switch on ApoTome. 2  power supply (middle unit). NB. Important: If want  to use “Smart Set‐up” in Zen Blue software. Switch  on the ApoTome.2 power supply now, even if not using ApoTome.2.   Turn on the UV lamp (if required for fluorescence imaging)(top unit). Note HBO lamp hours  (changed at 300h). Wait for the LAMP green light to come on. If it doesn’t, contact the  Microscopy Officer.      1 Louise Cole Apr 2013   Turn on the computer (Login Name: User, password:  decon).     Start the Zen Blue software (it takes a few minutes to  load).      ZEN ‐ ZEISS Efficient Navigation – is the single user interface that you will see on all new  imaging systems from Carl Zeiss.     To set up the microscope for fluorescence imaging      Load microscope slide onto stage. Note XY movement of stage is MANUAL     Use the touch screen on the RHS of the microscope  to select objectives and filters. Always start with a  low objective ie. X5 and move up and down in order  of objectives. NB. The nosepiece that carries the  objectives is MOTORIZED so must NOT be moved  manually.    Alternatively you can use the  buttons on the focus wheel  (filters on left, objectives on   the right wheel).      Note also the RL (reflected light) and TL (transmitted light) controls on the far right of the TFT  (above right).  Looking down the microscope:    Push both pins in all the way to view your sample  through the eyepieces (far right image).    If the top pin is pushed half way in, half the light  goes to the eyepieces and half goes to the colour  camera. Check that the pin on the LHS is also out.        2 Louise Cole Apr 2013     Open UV light shutter via RL (reflected light) illumination button on right of the TFT display or  the RL button on the right hand side of microscope (see below). Note you can use the  attenuator (lower image on right) to cut down UV light reaching specimen if sample very bright  or if it is bleaching rapidly. Position 1 = no attenuation. Find specimen and make sure it is in  focus. NB. Close shutter when not imaging to reduce sample bleaching.               Image acquisition   Send the image to the correct camera on the microscope. Setup as follows;                Monochrome camera setup    Bottom pin all the way out.  Colour camera setup  Bottom pin in and top pin all the way out.   Top pin halfway = half to camera and half to the eyepieces.        You are now ready to acquire FL images using the Zen software          3 Louise Cole Apr 2013     Overview of the Zen screen layout:    On the Left Tool Area the user finds the tools for image acquisition and image processing,  easily accessible via the Main tabs.     All functions needed to control the microscope can be found on the Locate tab. To acquire  images use the Acquisition tools. Arranged from top to bottom these tools follow the logic of  the experimental workflow.     The area for viewing and interacting with images is centered in the middle of the Main  Application window: the Center Screen Area. Each image (once acquired) can be displayed  and/or analyzed with many view options available through "view tabs" which can be found on  the left side of the image. According to the chosen view tab, the required view controls appear  in View control blocks below each image.     File management and data handling tools are found in the Right Tool Area.              4 Louise Cole Apr 2013     Select light/dark from design dropdown list in  workspace configuration     To zoom in or out of the workspace move the  slider to left or right          Left Hand Tool Area           The Left Tool Area contains all tools for system operation, image acquisition and image  processing.  The major functions are organized in the four Main tool tabs: Set exposure, Live, Continuous  and Snap  Locate: camera inspection in transmitted and epifluorescence light    Locate Tab  Click on Locate tab in the Left tool area. The currently set light path of the microscope is  displayed.   Note: In most FL cases, you can proceed straight to using the Acquisition Tab    5 Louise Cole Apr 2013    Depending on the level of motorization, the following parts can be controlled (Interactive  control):     Shutter Reflected Light: This turns the rear port mirror to guide the light to the specimen or to  block it from the specimen.    Reflector: Push and click reflector cube can be selected via graphical pop‐up menu.     Objective: Objectives can be selected via graphical pop‐up menu.    Condenser: Numerical aperture of the condenser is set via input box or slider. Turret position  (filter) selected from graphical pop‐up menu (only for motorized condensers).    Field Stop Filter: Opening of luminous‐field diaphragm (transmitted light) can be set via input  box or slider.    Transmitted Light Lamp: Transmission values for attenuation filter (transmitted) are set via  input box or slider for the front or rear filter wheel in accordance with the available filter steps.    Transmitted light is switched on / off via ON button in the Transmitted Light frame, setting of  light intensity can be varied via input box or slider. 3200 K color temperature for photo  documentation can be switched on via 3200 K button in the Transmitted Light frame.    Shutter Transmitted Light: This function opens or closes the internal shutter for transmitted  light.    Under Camera Tab     Under mode, select appropriate  resolution. Standard frame size is  1300 x1040 pixels (2.75 MB) for  fluorescence imaging. NB. You can  crop image but at the expense of  resolution. NB. Always check the  scalings. Use slow live speed. This  can be done under acquisition tab  also.            6 Louise Cole Apr 2013    Acquisition    You can load a predefined acquisition configuration from the Experiment Manager field by  selecting a saved configuration in the pull down menu.    To delete a configuration press the Delete button.    Smart Set‐up     Use Smart Setup for rapid selection of single or multi‐channel set‐up. The tool Smart Setup is  an intuitive, user‐friendly interface which can be used for almost all standard applications. It  configures all system hardware for a chosen set of dyes (as well as TL setup)  Note: Smart Setup offers you support when configuring multichannel acquisition experiments.  Select the fluorescent dyes and contrast techniques that you want to include in your  experiment from a large dye database. Smart Setup takes the configuration of your microscope  hardware and the properties of the selected dyes into account. Based on this information it  makes one or more suggestions for acquisition. You can adopt these into your experiment as  required and make further changes to them there. Smart Setup also configures the motorized  components of your system for the acquisition of multichannel images. It does not, however,  influence any camera parameters (e.g. exposure time or resolution) and also does not change  any parameters of other acquisition dimensions (e.g. Z‐stack, time series).             Simply choose the dye(s) used from the list dialog and assign signal display colour(s)              7 Louise Cole Apr 2013         Select Add, select Probe, Add and OK when finished    Save method under Experiment Manager. Once asterisk disappears, method has been saved.    Defining Channels: The Channel tool displays every Channel that is currently open in the Light  Path tool. The Tracks name and the used Channels for that track are indicated. Each Channel  that is highlighted will be displayed in the Channel Configuration display area, which allows for  individual settings of each Channel. A Channel will only be active when the Snap button is  pressed, if its check box is ticked. Changes in the selection of the Channels in the Channel tool  will automatically update the selection in the Light Path tool and vice versa.     Action buttons     Action buttons provide control for how the image is  acquired.     They have the following functions, regardless of the imaging mode:    Set exposure optimises the camera exposure. Set exposure can be disrupted by pressing the  Stop button, to which the set exposure button changes upon activation          8 Louise Cole Apr 2013    Live uses the maximum speed for a given image frame (and zoom setting) for all active  channels. To disrupt Live acquisition, the Stop button has to be pressed into which the Live  button changes after its activation.    Click on Live button. The live mode will be activated. You  will recognise the live mode by the green light signal and  by the hatched tab in the document bar.     In the Centre Screen Area you will see the camera live  image. By default the live image shows a cross hair helping  to navigate on the sample.      Select Set exposure button. Alternatively set the exposure  manually using the slider   Select linear display mappings (reset) on  the display window and find the optimal  focus on the monitor. (NB. You can select  Best Fit and Min/Max to alter your display  settings if required.) Use the dynamic  range indicator in the Dimensions tab to  highlight overexposed (too bright) regions  (indicated by red pixels) and underexposed (too  dark) areas (in blue).   If you want to adjust the gamma; on the Display  tab click the 0.45 button. The display curve will be  adapted to a gamma value of 0.45. This will set  the optimum colour presentation.   NB. The gamma value makes it possible to correct the display of images on computer screens  which do not allow the linear display of gray value curves. By changing the gamma value you  can emphasize certain intensity ranges within your image when it is displayed on the screen. A  value <1 emphasizes the ranges of medium pixel intensity (medium gray values), while a value  >1 emphasizes the dark and bright pixel intensities and therefore increases the contrast. The  recommended settings are 0.8 for fluorescence images, 1.2 for phase contrast or DIC images  and 0.45 for true color images. Please bear in mind that a "correct" gamma value setting  depends on numerous parameters, such as screen settings, ambient brightness, etc., and a  universal setting cannot therefore be given.       Click on Snap to take the image  9 Louise Cole Apr 2013      Viewing images      The Center Screen area can be set up to hold 1, 2 or 3 Image Containers.    Clicking the Expose Mode button in the top right corner of the Image Display opens a field of  thumbnails in the image display with one large thumbnail for each open image. Whenever you  lose track of your open images, open the Expose Mode and click once in the image you are  looking for and it will be immediately brought to the foreground of the Image Display.         2D View: This function allows to:  ‐ display a single image in frame mode,   ‐ display multiple channel images in superimposed mode.    10 Louise Cole Apr 2013      Split View: This function allows to display the individual channels of a multi channel image as  well as the superimposed image. NB. Cf. Use of splitter – see below image    Gallery View: This function allows to:  − display images (Z‐Stack, time series, combination of both) side by side in a tiled fashion,  − add data relevant to the displayed images (Z‐Stack slice distance, time of acquisition)  − extract a subset of images from the original stack and store the result as a new image  (controls for this function are in the Processing tab function Copy → Subset).    Ortho View: This function allows to  − display a Z‐Stack of images in an orthogonal view  − measure distances in three dimensions    The Information View is a view type that shows a summary information sheet of all relevant  image acquisition parameters. Parts of it can be edited – e.g. the Notes field.                                The Dimensions View Options  control block holds the tools to  modify the image display with  respect to the (multiple) dimensions  of the image data set.  For z‐stacks: The sections (slices) in  the dimensions can be scrolled with  sliders (Z‐Position) and also can be  directly addressed with setting  numbers in the spin‐boxes next to the sliders. The small button on the right hand side of the  spin‐boxes (Z‐Position) starts and stops an animation in the respective dimension. This basic  11 Louise Cole Apr 2013  animation cannot be customized. For all other player‐functionality refer to the Player View  Block.                                             Zoom In the Dimensions View Options control block, there are the Zoom and Tools function  lines related to changing the zoom factor of a displayed image. Any changes on these elements  affect the displayed image immediately. The zoom function can be performed online.  In the Display View Option control block, the intensity minimum can set to the Black value and  the intensity maximum to the White value of the image using the Min/Max button.   Clicking on the Best Fit button sets the intensity minimum to Black and percentile value and  the intensity maximum to White and percentile value.  Activating the Auto check box sets the intensity minimum and maximum values automatically.  With the spin‐boxes you can set the lower and upper percentile values.  The Black, the Gamma and the White values of the displayed image can be adjusted  individually.  With the Ch buttons, the effect of the settings can be restricted to an individual channel or be  applied to all channels. By default, the settings apply to all channels simultaneously. The  parameters can be changed using the spin‐boxes – or directly by typing in numbers in the  number‐field.   With the Reset button, the original settings are easily re‐set.  The settings in this View control block apply only to the Image Display and are NOT changed in  the original data set. To write the changed settings to the image data set, use the Apply  Display Curve in the Processing tab.  Graphics: This function allows to  − select from a set of drawing functions such as rectangles and arrows,  − add a scale bar to the image, as well as text annotations,  − use a set of interactive measurement functions for length, angle, area and size,  − add a text box that displays the coordinate    Functional Description: The  graphics function uses a plane  separate from the image plane (the  graphics plane) and therefore does  not change the content of the  image(s). Any changes done with  this function are effective  immediately. The overlay graphics  can be stored in the images or  separately in a file from where  they can be re‐loaded.    12 Louise Cole Apr 2013    Saving Images  See right tool area for Data  Management and Storage                          Open Images: In the Right Tool Area the  Open Images section is displaying the  images which are loaded in the system  memory. Double clicking brings the  image to the front. Files can be saved  and deleted/closed on this section.  Multiple images can be selected by  holding down the CTRL key on the  keyboard and clicking on the list entries.  The whole section can be hidden by  clicking on the black triangle at the left  side of the panel. Once hidden, a click  on the right boarder of ZEN lets the  Documents panel slide back into the  application.  All files can be displayed in three ways  by using the buttons underneath the  section.  1. Text View  2. Thumbnail View  3. Big Preview  The Re‐use function is available in the  ZEN Image Browser: Hardware  parameters can be loaded without  opening the files by clicking the Re‐use  button.     NB. Star ratings ***            13 Louise Cole Apr 2013  Multidimensional Acquisition Settings  The settings for multidimensional acquisition are accomplished within the tool groups eg. Z‐stack,  time series. If available the settings are displayed beside the function, e.g. how many images are  taken for the time series. Tool bars within the tool groups only become available if the box of a  multidimensional function is checked; otherwise the tool bar is not displayed. The graphical display  field above the Start Experiment button shows a representation of the experiment type with its  dimensions is show. It also lists the estimated size of the final multidimensional image.      Taking a Z‐stack       The Z‐Stack tool tab is only available if the Z‐stack checkbox is ticked in the Multi‐dimensional  acquisition field.     NB. Select All Channels per Slice if collecting multi‐ channel images    The way in which the z‐stack is acquired can be selected by pressing either the First/Last or  Center mode buttons. The modes differ in the way the first and last images of the stack are  selected.      Press the First/Last mode in the definition panel.  Press the Live action button and focus onto the  specimen.    Adjust the focus to the top of the specimen or  where you would like the z‐stack to start. Select  Set first. Focus to the bottom of the specimen or  where you should like the z‐stack to stop. Select  Set last. These focus positions are then indicated  next to Set First and Set Last buttons. The Range  display box shows the total height of the stack in  micrometers.     The Slices input box displays the number of slices of the stack. This number can be edited by  typing or using the arrow keys. Changing the value for Slices automatically adjusts the Interval  in a way that the total Range for the stack is maintained.        14 Louise Cole Apr 2013   The Interval input box displays the z‐distance (interval) between slices in μm. This number can  be edited by typing or using the arrow keys. Changing the value for the Interval automatically  adjusts the Number of Slices while keeping the total Range for the stack within the possibilities  of fitting the calculated number of slices into the Range. The total Range may therefore vary to  higher and lower numbers around the initial value of the Range.     Pressing the Optimal button sets Interval to match the Nyquist criteria (2 fold over sampling)  and accordingly adjusts the Slices to keep the Range within the limits defined by Set First and  Set Last. The optimal interval is displayed in the button in μm.     Within the Keep display field (only available in Show all mode) you have the option to switch  between Keep Interval and Keep Slice. If Keep Interval is checked, the number of slices is  adjusted to a new Range defined pressing Set First or Set Last. Likewise, when Keep Slice is  checked, the value for the Interval between the slices is adjusted to a new Range defined by  pressing Set First or Set Last.     When Start Experiment is pressed the system will perform a z‐stack     Start Auto Z‐ configuration to define top and  bottom of specimen for easy z‐stack selection            15 Louise Cole Apr 2013  Exporting images    Export images with the Export dialog in the File menu,  the acquired data can be directly exported to various file  types  •    •     •     •     •     •      Select the image to be exported.  Click on the Export menu Item in the File menu of  the menu bar. This opens the Export window or  Processing Tab.  Under Method, select the format to which the  image is to be exported to eg. Single File Export  Select the file type which the image is to be  exported under Parameters.  Choose a compression level. For some file formats  lossless compression or various other compression  levels are available. The degree of losses for the  image quality is listed according to the type of  compression.  Next select export destination and file name. Click Apply to complete the export process.  When stacks or time series are exported, each frame is stored as an individual image.  Under Options and Saving: The Auto Save panel  enables you to instruct the program to save the  results automatically and continuously during the  measurement process. All data recorded until a, for  example, a software failure, will be available this  way. This function is recommended when performing  long measurements. If the Auto Save check box is  ticked you can specify the directory in the Directory  selection box, to which the measurement should be  saved.       NB. .raw data files are still saved in the directory  specified under Save raw data during measurement.  However, if the Auto Save option is active,  temporary raw data files will be saved permanently  and not only temporarily and will be available if a  software failure occurs.         16 Louise Cole Apr 2013  Batch processing and export    This is useful for copying all parameters for export from one image to another. See Copy  parameters and paste parameters.    Make sure appropriate image(s) output storage path is selected in Browse     Various file formats are supported for export in pulldown menu for Filetype. Under Method  parameters, you can also select burn‐in annotations and create folders.       When ready to batch export, simply select Run Batch    17 Louise Cole Apr 2013  Processing  Maximum Intensity Projection    The Maximum Intensity Projection function  generates a maximum intensity projection image  along the z‐, time‐ or channel dimension of a  multidimensional image data set.    •   To open the Maximum Intensity Projection  Processing tool, select it from the list in the  Processing main tab.  •  Click on the triangle buttons to show or hide  the Method Parameters and Preview panels  of the function.  •  Select the Input Image with the Select button  and choose the dimension along which you  want to generate a maximum intensity  projection from the Coordinate pull down  menu (x, time, channel) in the Settings panel.  •   Clicking the Apply button creates the  maximum intensity projection in a new image  document, the name which is specified in the  Output image display of the Select field.    The maximum intensity projection creates an  output image whose pixels contain the maximum  value over all images in the stack at the particular  pixel location.               18 Louise Cole Apr 2013  Deconvolution software     Deconvolution (removal of out of focus blur) is  done using a mathematical algorithm. It gives  an image with less noise, better definition and  higher resolution.   A Z‐stack of your sample taken using the  optimal distance and with known appropriate  scalings is required.    Select Z‐stack image that you require to  deconvolve   Select Deconvolution in Processing Tab   Choose appropriate algorithm (see below) to  use eg. Inverse Filter seems to work well in  most instances. The deconvolution process can  take time if there are a number of channels and  z‐slices. Once process complete. Select Apply.  Observe and save the result.                19 Louise Cole Apr 2013    Deconvolution (defaults)  This method allows you to use 4 different algorithms for deconvolution.   Simple, very fast (Nearest Neighbor) button: Executes the fast Nearest Neighbor method  using default parameters. To use it, click on this button and then on Apply.   Better, fast (Regularized Inverse Filter) button: Executes the Regularized Inverse Filter  algorithm for image enhancement. To use it, click on this button and then on Apply.   Good, medium speed (Fast Iterative) button: Executes the Fast Iterative restoration method.  To use it, click on this button and then on Apply.   Excellent, slow (Constrained Iterative) button: Executes the Constrained Iterative quantitative  restoration method. To use it, click on this button and then on Apply.            20 Louise Cole Apr 2013  Glossary   Nearest Neighbor: The Nearest Neighbor method uses the simplest and fastest algorithm  (Castleman, K.R., Digital Image Processing, Prentice‐Hall, 1979). Its function is based on  subtraction of the out‐of‐focus information in each plane of a stack, taking the neighboring  sections above and below the corrected Z‐plane into account. This method is applied  sequentially to each plane of the entire 3D stack. It allows you to enhance contrast quickly,  even if image stacks have not been put together optimally.  Regularized Inverse Filter: The inverse filter is a genuine 3D method and generally achieves  better results than the Nearest Neighbor algorithm. It essentially involves dividing the Fourier  transformation ("FT") of the volume by the FT of the PSF, which can be performed very quickly.  In the real space this corresponds to deconvolution. In addition, a statistical method ("General  Cross Validation) is applied, which determines the noise component of the image and  automatically sets the restoration strength to the optimum level in line with this. This process  is also known as regularization. The method is very well suited to the processing of several  image stacks in order to preselect images for the application of the iterative "high‐end"  method. Z‐stacks must, however, have been acquired at the correct (Nyquist) distance. The  additional acquisition of Z‐planes above and below the structure of interest is recommended.  Fast Iterative: The "Fast Iterative" method is an iterative restoration method that uses only  one iteration per convolution step (see Meinel, E. S.: Origins of linear and nonlinear recursive  restoration algorithms. J.Opt.Soc.Am, 3 (6), 1986, 787‐799). No regularization is used in this  case. Due to the fast processing and convergence after just a few iterations, this method is  suitable in particular for the processing of larger time‐lapse images. The results of the method  can quickly lead to good results and remove most of the out‐of‐focus light. They do not,  however, create quantitative brightness conditions in the image. If undersampled images are  present, artifact formation may also result.  Constrained Iterative: The best image quality is achieved using the iterative maximum  likelihood algorithm (see Schaefer et al.: "Generalized approach for accelerated maximum  likelihood based image restoration applied to three‐dimensional fluorescence microscopy", J.  of Microscopy, Vol. 204, Pt 2, November 2001, pp. 99ff.). This algorithm is able to calculate  light from various focal planes back to its place of origin. Consequently, with this method it is  possible to derive the 3D structure from fluorescence images with the correct brightness  distribution and to visualize optical sections. It is also possible for missing information to be  partially restored from neighboring voxels. The spatial resolution can be increased without  artifacts up to a theoretical limit (one voxel). It is essential for Z‐stacks to have been acquired  in accordance with Nyquist. Acquiring sufficient planes above and below the structure of  interest is also imperative for achieving good results. As this is a complex mathematical  method, the calculation can take longer, depending on the image size and the PC being used.   21 Louise Cole Apr 2013  Transmitted light set‐up         Select TL brightfield set‐up in Smart Setup.     Make sure that you have set up properly for Kohler illumination with the objective that you  require for imaging.     Check condenser aperture diaphragm setting.      Under Locate tab     Select colour camera            White Balance: Only active if a colour camera has been selected.  ‐ Auto button: Sets the white balance automatically.  ‐ Pick button: Select a point in the image as a reference point for the white balance.  ‐ 3200K button: Sets the color temperature to 3200K.  ‐ 5500K button: Sets the color temperature to 5500K.    Show Channels checkbox: Only visible if the Show All mode is activated.  Activated: Displays 3 sliders to allow each color channel to be set precisely.  22 Louise Cole Apr 2013           Color Offset slider: Here you can set the color temperature.    Saturation slider: Here you can set the saturation.    Color Mode button  ‐ RGB: Sets the color mode to RGB.  ‐ B/W: Sets the color mode to B/W.    Live Speed dropdown list: Here you can select the speed of the live mode. You can choose  between Slow, Medium and Fast.    Resolution dropdown list: Here you can select the camera resolution.    Binning dropdown list: Here you can select the binning mode for the camera.    Acquisition ROI section: In this section you can define a particular area that you want to be  acquired.      After clicking on the Refresh Overview button, in the window on the left you will see a Preview  of your sample under the camera. The Pixel Size below the preview window indicates the size  of a pixel depending on the scaling set. The blue frame indicates the selected area. The entire  preview window is highlighted by default. To change the area, move the mouse pointer over  the blue frame. The mouse pointer will appear as a double‐headed arrow. If you hold down the  mouse button and move the mouse, you can change the size of the frame and in this way  highlight a certain area on the sample. To move the area, position the mouse pointer inside the  frame. The mouse pointer will appear as a four‐headed arrow. If you hold down the mouse  button and move the mouse, the frame can be moved.    Maximize button: Maximizes the frame to the maximum size in the preview window.    Center button: Positions a frame that has been changed precisely at the center of the image.    Size input field: Here you can enter the size of the frame that you want. Depending on the  camera type available, select from the predefined sizes in the dropdown list (1388x1040,  1024x1024, 512x512, 256x256, 128x128). Alternatively you also have the option of entering  the Size in X and Y, and the top left Start position. You can change the values by increments  using the arrow buttons next to the input windows or you can enter the desired values directly  using the keyboard. To do this, you must activate the number already present using the cursor.  Complete the entry by using the Tab or Enter key.    Offset input field: Here you can enter the start position for the frame, measured from the top  left corner.    Refresh Overview button: Adopts the image from the camera into the preview window.           Post‐Processing section    23 Louise Cole Apr 2013       Depending on which camera you are using, some functions are not visible. The most important  functions are explained in the section below.      Black Reference checkbox: Only active if you have defined a black reference via the Define  button.     Activated: Applies the measured black reference to the image.  Deactivated: The measured black reference is not used. The reference image is retained.    Define button: Automatically defines the black reference. To do this a reference image is  acquired, from which the black reference is measured. The light path to the camera must be  closed and the lamp must be switched on for the measurement. The measurement lasts for  several seconds. The Black Reference checkbox is then activated automatically.    Shading Correction     Activated: Applies the measured shading correction to the image.  Deactivated: The measured shading correction is not used. The reference image is retained.    Define button: Automatically defines the shading correction. To do this you will acquire a  reference image for the shading correction. An empty image without structures is required to  measure the shading image. Move the slide to an empty position on the sample with no  artifacts in the image. The Shading Correction checkbox is activated automatically after the  definition.    Enable Noise Filter checkbox: Activated: The Limit slider and the spin box/input field are active.  Enter the limit for the noise filter using the slider or spin box/input field.    Enable Unsharp Mask checkbox: Activated: Emphasizes edges in the image more clearly.            24 Louise Cole Apr 2013  Shutting down the microscope     Check microscope bookings. If there is another user coming within 1 hour then leave the  system on. If not, shut down completely. If you finish your session earlier than planned please  send an email to [email protected] to say the microscope is free OR call the next  user. If working out of normal working hours (ie. Mon‐Fri 9 am – 5 pm), please do not leave the  system on for the next user unless you are sure they are coming!   Note UV lamp hours. Turn off the Hg lamp and log your usage on the sheet provided. Report  any errors on sheet and to Dr Louise Cole directly (in person or by email:  [email protected])   Save all images and exit the software.    Remove slide from stage. If used an oil objective, please wipe excess oil from objective using  lens tissue ONLY   Leave microscope with a low objective in place and clean for other users. Switch off  microscope by the silver switch and at the power supply. Turn off ApoTome.2 if turned on.   Unplug both cameras. Shut down the computer.    Cover the microscope – rubber covers for eyepieces and transmitted light as well as the large  blue cover for microscope.                   Leave area clean and tidy. Turn any lights off in room as you leave.       25